- Circuito aberto voltage: 4V a 12V
- Resistencia de illamento (vol. CC)tage):
- Proba voltage [V]
- Rango [M]
- Resolution [M]
- Precisión
- Corriente de proba máxima: 5.81A (at 265V); 10.10A (at 457V)
- Proba voltagrangos e: 100-265V (Line-Neutral) / 100-460V (Line-Line); 50/60Hz ±5%
- Protección de seguridade: Liña Z (Liña-Liña, Liña-Neutro, Liña-PE)
- Materiais de illamento: PVC, Rubber butyl, EPR, XLPE
- First fault current (IT systems):
- Rango (mA)
- Resolución (mA)
- Límite de contacto voltage (ULIM): 25 V, 50 V
Electrical Safety Testing
- Connect the safety tester according to the specified voltagrango.
- Select the appropriate insulation resistance test voltage.
- Measure the insulation resistance and ensure it falls within the specified accuracy range.
- Check the maximum test current and ensure it is within the safe limits for testing.
- Perform first fault current tests as required.
Probas de RCD
- Select the RCD type and rated tripping currents for testing.
- Set the Line-PE and Line-N voltage segundo as especificacións.
- Perform RCD Molded type tripping time range tests based on the specific RCD type and current settings.
"'
Circuito aberto voltage: Short circuit current: Nominal test current: Safety protection:
nominal test voltage 0% +10% <6.0mA at 500V test voltage
>1mA if load= 1k*Vnom (Vnom=50V, 100V, 250V, 500V, 1000V) the display shows an error message for input voltage > 10 V
Liña Z (Liña-Liña, Liña-Neutro, Liña-PE)
Rango []
Resolución []
Precisión
0.00 199.9 m (*) 200 1999 m (*)
0.1 m (*) 1 m (*)
(5.0 rdg + 1m) (*)
0.01 9.99 10.0 199.9
0.01 0.1
(5.0 rdg + 3dgt)
(*) By means of IMP57 optional accessory
Máxima corrente de proba
5.81A (at 265V); 10.10A (at 457V)
Proba voltage ranges: Protection type: Insulation materials:
100265V (Line-Neutral) / 100460V (Line-Line); 50/60Hz 5% MCB (B, C, D, K), Fuse (gG, aM) PVC, Rubber butyl, EPR, XLPE
Primeira corrente de falla (sistemas informáticos)
Rango (mA)
Resolución (mA)
0.1 0.9
0.1
1 999
1
Límite de contacto voltage (ULIM)
25 V, 50 V
Accuracy (5.0 rdg + 1dgt) (5.0 rdg + 3dgt)
HT ITALIA SRL Via della Boaria, 40 48018 Faenza (RA) Italy T +39 0546 621002 | F +39 0546 621144 M export@ht-instruments.com | ht-instruments.com
ONDE ESTAMOS
MACROTEST3
Professional installation safety tester
Rel 2.02 of 26/05/25 Pag 2 of 6
Proba de RCD (tipo de caixa moldeada)
RCD type:
AC ( ), A/F ( ), B/B+( ) General (G), Selective (S) and Delayed ( )
Rated tripping currents (IN)::
6mA, 10mA, 30mA, 100mA, 300mA, 500mA, 650mA, 1000mA
Liña-PE, Liña-N vol.tage:
100V 265V RCD type AC and A/F, 190V 265V RCD type B/B+
Frecuencia:
50/60Hz 5%
RCD tripping current (Molded case type RCD General)
Tipo RCD
IN
Range IN [mA] Resolution [mA]
AC, A/F, B/B+
6 mA, 10 mA
AC, A/F, B/B+ 30mA IN 300mA (0.2 1.1) IN
0.1 EN
AC, A/F 500mA IN 650mA
Accuracy IN – 0%, +10%IN
– 0%, +5%IN
RCD Molded type tripping time range [ms] (TT/TN system)
x 1 / 2
x 1
x 2
x 5
AUTO
GS
GS
GS
GS
GS
6 mA 10 mA 30 mA
m
AC 999 999 999 999 999 999 160 210 A/F 999 999 999 999 999 999 160 210 B/B+ 999 999 999 999 999 999 AC 999 999 999 999 999 999 160 210 A/F 999 999 999 999 999 999 160 210 B/B+ 999 999 999 999 999 999
AC 999 999 999 999 999 999 160 210 A/F 999 999 999 999 999 999 160 210
50 150 50 150
50 150 50 150
50 150 50 150
B/B+ 999 999 999 999 999 999 AC 999 999 999 999 999 999 160 210 100mA A/F 999 999 999 999 999 999 160 210
50 150 50 150
B/B+ 999 999 999 999 999 999
AC 999 999 999 999 999 999 160 210 300mA A/F 999 999 999 999 999 999 160 210
B/B+ 999 999 999 999 999 999
AC 999 999 999 999 999 999 160 210
500mA 650mA
A / F
999 999 999 999 999 999 160 210
B/B+
50 150 50 150
50 150
AC 999 999 999 999 999 999 160 210 1000mA A/F 999 999 999 999 999 999
B/B+
Resolution: 1ms, Accuracy: (2.0rdg + 2dgt)
RCD Molded type tripping time range [ms] (IT system) (*)
GS
310 310 310 310 310 310
310 310 310 310 310 310 310 310 310 310 310
AUTO+ G S
x 1 / 2
x 1
x 2
x 5
AUTO
GS
GS
GS
GS
GS
6mA AC 999 999 999 999 999 999 160 210
10mA A/F 999 999 999 999 999 999 160 210
30mA B/B+ 999 999
999 999
100mA 300mA
AC A/F B/B+
999 999 999
999 999 999
999 999
999 999 999
999 999 999
999 999
160 160
210 210
500mA 650mA
AC A/F B/B+
999 999
999 999
999 999
999 999
999 999
999 999
160 160
210 210
50 150 50 150
50 150 50 150
50 150
AC 999 999 1000mA A/F 999 999
B/B+
999 999 999 999
160 210
Resolution: 1ms, Accuracy: (2.0rdg + 2dgt) (*) Selection RCD type A/F and type B/B+ available only for Norway country
GS
310 310 310 310 310
AUTO+
GS
HT ITALIA SRL Via della Boaria, 40 48018 Faenza (RA) Italy T +39 0546 621002 | F +39 0546 621144 M export@ht-instruments.com | ht-instruments.com
ONDE ESTAMOS
MACROTEST3
Professional installation safety tester
Rel 2.02 of 26/05/25 Pag 3 of 6
Proba en interruptores diferenciais con retardo de fuga a terra (RCDs) (co accesorio opcional RCDX10)
RCD type: Rated tripping currents (IN):: Line-PE, Line-N voltage: Frecuencia:
AC ( ), A/F ( ), B/B+( ) General (G), Selective (S) and Delayed ( ) 0.3A 10A (type AC, A/F) ; 0.3A 3.0A (type B/B+) 100V 265V RCD type AC and A/F, 190V 265V RCD type B/B+ 50/60Hz 5%
Comprobador de retardo de fuga á terra, corrente de disparo de RCDs (RCD Xeral)
Tipo RCD
IN
Range IN [mA] Resolution [mA]
Accuracy IN
AC, A/F, B/B+ AC, A/F
300mA IN 1A 1.1A IN 10A
(0.3 1.1) IN
0.1 EN
– 0%, +5%IN
Rango de tempo de disparo dos RCD do comprobador de retardo de fuga á terra [ms] (sistema TT/TN)
x 1 / 2
x 1
x 2
x 5
AUTO
0.3A
1.0A
1.1A
3.0A
3.1A
6.5A
6.6A
10.0A
GS
GS
GS
AC 999 999 999 999 999 999 200 250
A/F 999 999 999 999 999 999 200 250
B/B+ 999 999 999 999 999 999
AC 999 999 999 999 999 999 200 250 A/F 999 999 999 999 999 999 200 250 B/B+ 999 999 999 999 999 999
AC 999 999 999 999 999 999 200 250 A/F 999 999 999 999 999 999 200 250 B/B+
AC 999 999 999 999 999 999 200 250 A/F 999 999 999 999 999 999 B/B+
G S 50 150 50 150
50 150 50 150
50 150 50 150
GS
G S 310 310 310 310 310
310 310
Resolution: 1ms, Accuracy: (2.0rdg + 2dgt)
Earth leakage delay tester RCDs trip out time range [ms] (IT system) (*)
x 1 / 2
x 1
x 2
x 5
GS
GS
GS
GS
0.3A
3.0A
AC A/F B/B+
999 999 999 999 999 999 200 250 999 999 999 999 999 999 200 250
50 150 50 150
3.1A
6.5A
AC A/F B/B+
999 999 999 999 999 999 200 250 999 999 999 999 999 999 200 250
50 150 50 150
6.6A
10.0A
AC A/F B/B+
999 999 999 999 999 999 200 250 999 999 999 999 999 999 200 250
Resolution: 1ms, Accuracy: (2.0rdg + 2dgt) (*) Selection RCD type A/F and type B/B+ available only for Norway country
AUTO G S
G S 310
310
NoTripTest Non-trip earth loop impedance
Proba voltage:
100265V (Line-PE), 50/60Hz 5%
NoTripTest Systems with Neutral wire
Rango []
Resolución []
Precisión (*)
0.01 9.99
0.01
(5 reading + N/10)
10.0 199.9
0.1
(5 reading + N)
200 1999
1
(5 reading + 3N)
(*) If IN <30mA, test current = IN/2 and N[]=30/IN; if IN 30mA, test current <15mA and N=1
NoTripTest Systems without Neutral wire
Rango []
Resolución []
Precisión (*)
1 1999
1
-0%, +(5.0 lettura +N)
(*) if IN <30mA, test current = IN/2 and N[]=(10×30)/IN ; If IN 30mA, test current IN/2 and N[]=(3×30)/IN
HT ITALIA SRL Via della Boaria, 40 48018 Faenza (RA) Italy T +39 0546 621002 | F +39 0546 621144 M export@ht-instruments.com | ht-instruments.com
ONDE ESTAMOS
MACROTEST3
Professional installation safety tester
Rel 2.02 of 26/05/25 Pag 4 of 6
Contacto voltage (RCD e proba de cero disparo)
Rango [V]
Resolución [V]
0 Utlim
0.1
Contacto voltage (EARTH test TT system)
Rango [V]
Resolución [V]
0 99.9
0.1
Contacto voltage (EARTH test TN system)
Rango [V]
Resolución [V]
0 99.9
0.1
100 999
1
Resistencia de terra co método de 3 fíos
Rango []
Resolución []
0.01 9.99
0.01
10.0 99.9
0.1
100 999
1
1.00k 49.99k
0.01 k
Corrente de proba:
<10mA 77.5Hz, Open-circuit voltage: < 20 Vrms
(*) Engadir un 5 % á precisión se as resistencias da sonda (Rs ou Rh) son > 100 x Rmeas
Accuracy -0%, +(5.0 rdg + 3V)
Accuracy -0%, +(5.0 rdg + 3V)
Accuracy -0%, +(5.0 rdg + 3V)
Precisión (*)
(5.0 rdg + 3dgt)
Resistividade do solo co método Wenner de 4 fíos
Range [m]
Resolution [m]
0.06 9.99
0.01
10.0 99.9
0.1
100 999
1
1.00k 9.99k
0.01 k
10.0k 99.9k
0.1 k
100k 999k
1k
1.00M 3.14M
0.01 M
(*) with distance d=10m, Distance “d” range:
1 10 m
Corrente de proba:
<10mA 77.5Hz, Open-circuit voltage: < 20 Vrms
Accuracy (*) (5.0 rdg + 3dgt)
Rotación da secuencia de fases co método de 1 fío
Voltagrango PN, P-PE[V]
Rango de frecuencias
100 265
50Hz/60Hz 5%
A medición só se realiza por contacto directo con pezas metálicas activas (non na vaíña de illamento)
Voltage drop on main power lines (V%)
Intervalo (%)
Resolución (%)
0 100
0.1
Accuracy (10.0 rdg + 4dgt)
Corrente de fuga (por HT96U opcional cl)amp transductor)
Rango [mA]
Resolución [mA]
0.5 999.9
0.1
Accuracy (5.0 rdg + 2dgt)
Parámetros ambientais (función AUX)
Parámetro
Rango
Temperatura [°C]
-20°C 80°C
Temperatura [°F]
-4 °F 176 °F
Humidade relativa [%HR]
0 100%HR
Saída DC voltage
0.1 mV 1.0 V
0.001Lux 20.00 Lux (*)
Iluminancia [Lux]
0.1 Lux 2000 Lux (*)
1 Lux 20 kLux (*)
(*) A precisión da sonda lux HT53 é segundo a Clase AA
Resolution 0.1 °C 0.1 °F
0.1% UR 0.1mV
0.001 0.02 Lux 0.1 2 Lux 1 20 Lux
Accuracy (2.0rdg+2dgt)
HT ITALIA SRL Via della Boaria, 40 48018 Faenza (RA) Italy T +39 0546 621002 | F +39 0546 621144 M export@ht-instruments.com | ht-instruments.com
ONDE ESTAMOS
MACROTEST3
Professional installation safety tester
Rel 2.02 of 26/05/25 Pag 5 of 6
Medición dos principais parámetros e harmónicos (PQA)
AC TRMS Voltage
Range [V] 15.0 459.9
Allowed crest factor
Resolution [V] 0.1V
1,5 ; Frequency: 42.5 69.0 Hz
Accuracy (1.0rdg + 1dgt)
Frecuencia
Alcance [Hz]
Resolución [Hz]
42.5 69.0
0.01
Permitido voltage: 15.0 459.9V ; Allowed current: 5%FS clamp FS clamp
Accuracy (2.0rdg + 2dgt)
Corrente AC TRMS
FS clamp
Rango [A]
10A
5% FS 9.99
10A FS 200
5% FS 199.9
200A FS 3000
5% FS 2999
Range: 5 999.9 mV; Values under 5mV are zeroed
Allowed crest factor
3; Frequency: 42.5 69.0 Hz
Resolución [A] 0.01 0.1 1
Precisión
1Ph: (1.0rdg + 3 dgt) 3Ph: (2.0rdg + 5 dgt)
Active power (@ 230V in 1Ph systems, 400V in 3Ph systems, cos=1, f=50.0Hz)
FS clamp
Alcance [kW]
Resolución [kW]
Precisión
10A
0.000 9.999
0.001
10A FS 200
0.00 999.99
0.01
1Ph: (2.0rdg + 5 dgt)
200A FS 1000
0.0 999.9
0.1
3Ph: (2.5rdg + 8 dgt)
1000A FS 3000
0 9999
1
Reactive power (@ 230V in 1Ph systems, 400V in 3Ph systems, cos=0, f=50.0Hz)
FS clamp
Alcance [kVAr]
Resolución [kVAr]
Precisión
10A
0.000 9.999
0.001
10A FS 200
0.00 999.99
0.01
1Ph: (2.0rdg + 7 dgt)
200A FS 1000
0.0 999.9
0.1
3Ph: (3.0rdg + 8 dgt)
1000A FS 3000
0 9999
1
Factor de potencia (230 V en sistemas monofásicos, 1 V en sistemas trifásicos, f = 400 Hz)
Rango
Resolución
Precisión
0.70c 1.00 0.70i
0.01
(4.0rdg + 10dgt) if I 10%FS (2.0rdg + 3dgt) if I > 10%FS
cos (@ 230V in 1Ph systems, 400V in 3Ph systems, f=50.0Hz)
Rango
Resolución
Precisión
0.70c 1.00 0.70i
0.01
(4.0rdg + 10dgt) if I 10%FS (1.0rdg + 7dgt) if I > 10%FS
VoltagHarmónicos e (a 230 V en sistemas monofásicos, 1 V en sistemas trifásicos, f = 400 Hz)
Intervalo [%]
Resolución [%]
Orde
Precisión
0.1 100.0
0.1
01 25
(5.0 díx + 5 díx)
Frequency of fundamental: 42.5 69.0 Hz, DC accuracy not declared
Harmónicos de corrente (f=50 Hz)
Intervalo [%]
Resolución [%]
0.1 100.0
0.1
Orde 01 9 10 17 18 25
Accuracy (5.0rdg + 5dgt) (10.0rdg + 5dgt) (15.0rdg + 10dgt)
HT ITALIA SRL Via della Boaria, 40 48018 Faenza (RA) Italy T +39 0546 621002 | F +39 0546 621144 M export@ht-instruments.com | ht-instruments.com
ONDE ESTAMOS
MACROTEST3
Professional installation safety tester
Rel 2.02 of 26/05/25 Pag 6 of 6
2. ESPECIFICACIÓNS XERAIS
DISPLAY AND MEMORY: Features: Memory: Communication:
Touch screen, color graphic LCD, 320x240mm 999 locations, 3 marker levels Optical-USB and built-in WiFi
POWER SUPPLY: Batteries: Battery life: Auto Power OFF:
6 x 1.2V(rechargeable) type AA or 6 x 1.5V type AA > 500 test for each function after 5 min of idleness (disabled)
MECHANICAL FEATURES: Dimensions (L x W x H): Weight (included batteries):
WORKING ENVIRONMENTAL CONDITIONS: Reference temperature: Working temperature: Allowed relative humidity: Storage temperature: Storage humidity:
TEST VERIFIES REFERENCE STANDARDS: Continuity test with 200mA: Insulation resistance: Earth resistance: Fault loop impedance: RCD test: Multifunction: Prospective short circuit current: Earth resistance on TN systems:
225 x 165 x 75 mm 1.2 kg
23°C 5°C 0°C 40°C <80%RH -10°C 60°C <80%RH
IEC/EN61557-4 IEC/EN61557-2 IEC/EN61557-5 IEC/EN61557-3 IEC/EN61557-6 IEC/EN61557-10 EN60909-0 EN61936-1 + EN50522
GENERAL REFERENCE STANDARDS: Safety of measuring instruments: Product type standard: Technical documentation : Insulation: Pollution degree: Encapsulation :
Sobrevoltage category: Max height of use:
IEC/EN61010-1, IEC/EN61010-031, IEC/EN61010-2-032 IEC/EN61557-1 IEC/EN61187 double insulation 2 IP40
CAT IV 300V (to ground), max 415V between inputs 2000m
Este instrumento cumpre os requisitos de Low Voltage Directive 2014/35/EU (LVD), EMC Directive 2014/35/EU and Directive RED 2014/53/EU
Este instrumento cumpre os requisitos das directivas europeas 2011/65/UE (RoHS) e 2012/19/UE (WEEE).
HT ITALIA SRL Via della Boaria, 40 48018 Faenza (RA) Italy T +39 0546 621002 | F +39 0546 621144 M export@ht-instruments.com | ht-instruments.com
ONDE ESTAMOS
Servizos de EURO-INDEX
EURO-INDEX is fabrikant, importeur en distributeur van diverse A-merken op het gebied van test- en meetinstrumenten. Daarnaast leveren wij een groot aantal diensten om het gebruik van deze instrumenten in uw bedrijfsvoering te optimaliseren. Dit omvat uiteraard onderhoud, reparatie en kalibratie van de instrumenten, maar ook kennisdeling via de EURO-INDEX Academy en verhuur van instrumenten.
Geautoriseerd Service Centrum EURO-INDEX b.v. is van alle vertegenwoordigde merken een Geautoriseerd Service Centrum. Dit betekent dat uw instrumenten worden behandeld door technici die zijn opgeleid door de fabrikant en beschikken over de juiste gereedschappen en software. Er worden uitsluitend originele onderdelen toegepast en de garantie van uw instrument, evenals de certificering (ATEX, EN50379, etc.) blijven intact.
Kalibratielaboratorium
Ons moderne service- en kalibratielaboratorium beschikt over een
RvA accreditatie naar NEN-EN-ISO/IEC 17025. Deze accreditatie
geldt voor grootheden, zoals gespecificeerd in de scope bij accreditatienummer K105.
Meer info
Kijk voor een overzicht van al onze diensten op euro-index.nl/diensten
Mobiele Service
Naast de vaste kalibratielaboratoria in Capelle aan den IJssel
en Zaventem beschikken wij ook over laboratoria op wielen met
de naam “Mobiele Service”. Dit biedt vertrouwde
service en kwaliteit, bij u voor de deur!
Meer info
KWS®
KWS® is een uniek servicesysteem voor uw meetinstrumenten
met periodiek onderhoud en kalibratie tegen vaste, lage kosten.
Uw kalibratiecertificaten zijn digitaal beschikbaar via Mijn KWS (gratis webportaal en app).
Meer info
Verhuur van meetinstrumenten
Uitgebreid assortiment
Nauwkeurigheid aantoonbaar door actueel kalibratiecertificaat
Deskundig advies Complete levering inclusief accessoires
Meer info
EURO-INDEX Academy Trainingen (individueel en klassikaal) Cursussen en workshops Demonstratie- en instructievideo’s Whitepapers
Meer info
Servicebalie
Onderhoud, reparatie en kalibratie
Cursussen en workshops
Mobiele Service
Wijzigingen voorbehouden EURO-INDEX® NL 24005
Documentos/Recursos
![]() |
HT INSTRUMENTS MACROTESTG3 Multifunction Multi Tester [pdfInstrucións MACROTESTG3 Multifunction Multi Tester, MACROTESTG3, Multifunction Multi Tester, Multi Tester |