Lambda CS-3000 Series Curve Tracer
Specificazioni
- Model: Curve Tracer CS-3000 series (CS-3100, CS-3200, CS-3300)
- Collector Supply (HV Mode): Max.Peak voltage – 3 kV, Max. Peak current – 15 A, Max. peak power – 390 W
- Collector Supply (HC Mode): Max.PPeakvoltage – 30 V, Max. peak current – 7.5 A, Pulse width – 50 us to 400 us
- Step Generator: Step rate – 1/10 Step, Pulse width range – 50 us to 400 us
- Measurement: Collector voltage range – HV: 50 mV/div to 500 V/div, HC: 50 mV/div to 5 V/div
Istruzzioni per l'usu di u produttu
Collector Supply
For HV mode, set the max. Peak voltage and current according to the specifications. For HC mode, adjust the settings as needed.
Step Generator
Set the step rate and pulse width within the specified ranges for accurate measurements.
Misurazione
Use the appropriate collector voltage range based on the mode of operation for precise readings.
Ideal for measuring the characteristics of various semiconductors such as IGBTs, MOSFETs, transistors, and diodes etc.
- Volu piccu massimutage 3,000V (high voltage mode)
- Maximum peak current 1,000A (CS-3300 high current mode)
- All models equipped with LEAKAGE mode (1pA cursor resolution)
- USB port to save screen data and setup data
- Equipped with La LAN interface for remote control
Easy-to-understand measurement screen
CS-3000 Series Specifications
Collector supply HV mode
Mudellu | CS-3100-300V * 2 | CS-3100 / CS-3200 / CS-3300 | |||
---|---|---|---|---|---|
Mode/polarity | Full wave rectification/ + – , DC/ + – , LEAKAGE/ + – , AC | ||||
Volu piccu massimutage | 300V | 30V | 3 kV | 300V | 30V |
Massima corrente di punta (maximum peak pulse current * 1 ) |
750 mA (1.5A * 1 ) |
7.5A (15A * 1 ) |
75 mA (150mA * 1 ) |
750 mA (1.5A * 1 ) |
7.5A (15A * 1 ) |
Potenza massima di punta | 120mW / 1.2W / 120W / 390W (Cannot be set when using the maximum peak voltage of 3kV) | ||||
Horizontal axis range | 50mV~500V/div |
- The maximum instantaneous current value that can be generated (varies depending on the target to be measured and conditions), the measurement should be less than or equal to the maximum peak current value.
Collector supply HC mode
Mudellu | CS-3100 / CS-3100-300V * 2 | CS-3200 | CS-3300 | |||
---|---|---|---|---|---|---|
Mode/polarity | Not equipped with HC mode | Pulse/ + – | ||||
Volu piccu massimutage | 40V | |||||
Massima corrente di punta | 400A | 40A | 1,000A | 400A | 40A | |
Potenza massima di punta | 4 kW | 400 W | 10 kW | 4 kW | 400 W | |
Larghezza di pulse | Pulse width is variable at 50µs to 400µs (10µs resolution) | |||||
Puntu di misurazione | The measurement point can be specified (10µs resolution) | |||||
Vertical axis range | 100mA~50A/div | 100mA~100A/div | ||||
Prughjettu di prova | CS-301 | CS-302 |
Specificazioni cumuni
Loop correction | Hardware | Correction ofthe floating capacitance between the collector supply and the ground | |
Software | Pseudo-looping through software optimisation | ||
Step generator | Offset | Gamma di paràmetri | -10 ~ +10 x STEP AMPLITUDE Setting |
Risoluzione | 1% of the STEP AMPLITUDE setting | ||
Modu attuale | Gamma | 50nA to 200mA per step, 21 steps, 1-2-5 switching | |
A currente massima | 2A | ||
Vol. Massimutage | 10 V o più | ||
VoltagModu e | Gamma | 50mV to 2V per step, 6 steps, 1-2-5 switching | |
A currente massima | 500mA~(~8V), 200mA~(~15V), 10mA~(~40V) | ||
Vol. Massimutage | ± 40 V | ||
Tasso di passu | 2x 50Hz or 60Hz (1x in AC mode), pulse interval in HC mode | ||
Pulse step | Larghezza di pulse | 50µs~400µs (10µs step) When in HC mode, the pulse width is 100µs wide before and after the pulse of the collector supply |
|
Numero di passi | 0~20 Step | ||
Uscita AUX | Gamma | OFF, -40V~+40V (Variable in 100mV steps) | |
Modu di misurazione | REPEAT, STOP/SINGLE, SWEEP | ||
Asse verticale (full scale 10div) |
Collettore currente | Gamma | HV mode:1µA/div to 2A/div, 20 steps 1-2-5 switching (HC mode is not included) |
Accuratezza | 2% + 0.05 x VERT/div of readout To this, add the loop correction error for the following maximum peak voltages 0.5µA(30V), 1µA(300V), 6µA(3kV), 12µA(5kV) Specify at least 10% of the maximum peak voltage for 30V, 300V, and 3kV, and at least 30% for 5kV |
||
Emettitore currente (LEAKAGE) |
Gamma | 1nA/div~2mA/div, 20 step 1-2-5 switching (Collector supply mode is LEAKAGE) |
|
Accuratezza | 2% + 0.05 x VERT/div + 1nA or less of readout | ||
Asse horizontale (full scale 10div) |
Cullettore voltage |
Gamma | HC mode:50mV/div to 5V/div, 7 steps 1-2-5 switching (HV mode is not included) |
Accuratezza | 2% + 0.05 x HORIZ/div or less of readout | ||
Base/ emettitore voltage |
Gamma | 50mV/div~5V/div, 7 step 1-2-5 switching | |
Accuratezza | 2% + 0.05 x HORIZ/div or less of readout | ||
Mostra | Mostra | LCD TFT da 8.4 pollici | |
Numero di punti di dati | 1,000 points/trace (AC, full wave rectification) 20~1,000 points/trace (SWEEP mode) |
||
Trace display | Point interpolation display, dot display | ||
A media | OFF, 2 to 255 times | ||
Persistenza | OFF, SHORT, LONG, Infinite length | ||
Save internal waveform (REF) | 4 pannellu | ||
Cursore | DOT | Vert, Horiz, β or gm | |
fLINE | Vert, Horiz, 1/grad, intercept | ||
GRATIS | Vert, Horiz, β or gm | ||
FINESTRA | Vert, Horiz, β or gm in the WINDOW area | ||
Data save/read | Memoria | Setup:256 pcs, REF:4 screens | |
Memoria esterna | USB1.1:Setup, waveform, screen hardcopy (BMP, TIFF, PNG) | ||
Remote | Equipped with a remote 10BASE-T/100BASE-TX 1 port via LAN connection | ||
Alimentazione elettrica | AC100~240V 50/60Hz, Maximum operating power:500VA (400W), Standby:50VA (7W) | ||
External dimensions(mm) Pesu |
CS-3100 / CS-3100-300V * 2 | Approx. 424W x 220H x 555D (excluding protrusions and accessories) approx. 28kg (excluding accessories) |
|
CS-3200, 3300 | Approx. 424W x 354H x 555D (excluding protrusions and accessories) approx. 43kg (excluding accessories) |
CS-3100-300V is a made-to-order product.
Accessori standard
CS-500 Adattatore di prova | |
---|---|
![]() |
This is a jig for connecting the part to be measured, provided by the customer, and the fixture. (Standard accessories) |
FAQs
Q: What should I do if the pulse width is too narrow?
A: If the pulse width is narrow, adjust the settings to widen it within the allowable range to generate the max. peak voltage.
Q: How to calibrate the collector supply output?
A: The collector supply output can be calibrated by enabling/disabling the output and adjusting the floating capacity as per the instructions provided.