Lambda CS-3000 Series Curve Tracer

Lambda-CS-3000-Series-Curve-Tracer-PRODUCT

Specifications

  • Model: Curve Tracer CS-3000 series (CS-3100, CS-3200, CS-3300)
  • Collector Supply (HV Mode): Max.Peak voltage – 3 kV, Max. Peak current – 15 A, Max. peak power – 390 W
  • Collector Supply (HC Mode): Max.PPeakvoltage – 30 V, Max. peak current – 7.5 A, Pulse width – 50 us to 400 us
  • Step Generator: Step rate – 1/10 Step, Pulse width range – 50 us to 400 us
  • Measurement: Collector voltage range – HV: 50 mV/div to 500 V/div, HC: 50 mV/div to 5 V/div

Product Usage Instructions

Collector Supply
For HV mode, set the max. Peak voltage and current according to the specifications. For HC mode, adjust the settings as needed.

Step Generator
Set the step rate and pulse width within the specified ranges for accurate measurements.

Measurement
Use the appropriate collector voltage range based on the mode of operation for precise readings.

Ideal for measuring the characteristics of various semiconductors such as IGBTs, MOSFETs, transistors, and diodes etc.

  • Maximum peak voltage 3,000V (high voltage mode)
  • Maximum peak current 1,000A (CS-3300 high current mode)
  • All models equipped with LEAKAGE mode (1pA cursor resolution)
  • USB port to save screen data and setup data
  • Equipped with La LAN interface for remote control

Lambda-CS-3000-Series-Curve-Tracer-FIG- (1)

Easy-to-understand measurement screen

Lambda-CS-3000-Series-Curve-Tracer-FIG- (2)

CS-3000 Series Specifications

Collector supply HV mode

Model CS-3100-300V * 2 CS-3100 / CS-3200 / CS-3300
Mode/polarity Full wave rectification/ + – , DC/ + – , LEAKAGE/ + – , AC
Maximum peak voltage 300V 30V 3kV 300V 30V
Maximum peak current
(maximum peak pulse current * 1 )
750mA
(1.5A * 1 )
7.5A
(15A * 1 )
75mA
(150mA * 1 )
750mA
(1.5A * 1 )
7.5A
(15A * 1 )
Maximum peak power 120mW / 1.2W / 120W / 390W (Cannot be set when using the maximum peak voltage of 3kV)
Horizontal axis range 50mV~500V/div
  • The maximum instantaneous current value that can be generated (varies depending on the target to be measured and conditions), the measurement should be less than or equal to the maximum peak current value.
scrollable

Collector supply HC mode

Model CS-3100 / CS-3100-300V * 2 CS-3200 CS-3300
Mode/polarity Not equipped with HC mode Pulse/ + –
Maximum peak voltage 40V
Maximum peak current 400A 40A 1,000A 400A 40A
Maximum peak power 4kW 400W 10kW 4kW 400W
Pulse width Pulse width is variable at 50µs to 400µs (10µs resolution)
Measurement point The measurement point can be specified (10µs resolution)
Vertical axis range 100mA~50A/div 100mA~100A/div
Test fixture CS-301 CS-302
scrollable

Common Specifications

Loop correction Hardware Correction ofthe  floating capacitance between the collector supply and the ground
Software Pseudo-looping through software optimisation
Step generator Offset Setting range -10 ~ +10 x STEP AMPLITUDE Setting
Resolution 1% of the STEP AMPLITUDE setting
Current mode Range 50nA to 200mA per step, 21 steps, 1-2-5 switching
Maximum current 2A
Maximum voltage 10V or more
Voltage mode Range 50mV to 2V per step, 6 steps, 1-2-5 switching
Maximum current 500mA~(~8V), 200mA~(~15V), 10mA~(~40V)
Maximum voltage ±40V
Step rate 2x 50Hz or 60Hz (1x in AC mode), pulse interval in HC mode
Pulse step Pulse width 50µs~400µs (10µs step)
When in HC mode, the pulse width is 100µs wide before and after the pulse of the collector supply
Number of steps 0~20 Step
AUX output Range OFF, -40V~+40V (Variable in 100mV steps)
Measurement mode REPEAT, STOP/SINGLE, SWEEP
Vertical axis
(full scale 10div)
Collector current Range HV mode:1µA/div to 2A/div, 20 steps 1-2-5 switching (HC mode is not included)
Accuracy 2% + 0.05 x VERT/div of readout
To this, add the loop correction error for the following maximum peak voltages
0.5µA(30V), 1µA(300V), 6µA(3kV), 12µA(5kV)
Specify at least 10% of the maximum peak voltage for 30V, 300V, and 3kV, and at least 30% for 5kV
Emitter
current
(LEAKAGE)
Range 1nA/div~2mA/div, 20 step 1-2-5 switching
(Collector supply mode is LEAKAGE)
Accuracy 2% + 0.05 x VERT/div + 1nA or less of readout
Horizontal axis
(full scale 10div)
Collector
voltage
Range HC mode:50mV/div to 5V/div, 7 steps 1-2-5 switching (HV mode is not included)
Accuracy 2% + 0.05 x HORIZ/div or less of readout
Base/
emitter
voltage
Range 50mV/div~5V/div, 7 step 1-2-5 switching
Accuracy 2% + 0.05 x HORIZ/div or less of readout
Display Display 8.4-inch TFT LCD
Number of data points 1,000 points/trace (AC, full wave rectification)
20~1,000 points/trace (SWEEP mode)
Trace display Point interpolation display, dot display
Averaging OFF, 2 to 255 times
Persistence OFF, SHORT, LONG, Infinite length
Save internal waveform (REF) 4-Panel
Cursor DOT Vert, Horiz, β or gm
fLINE Vert, Horiz, 1/grad, intercept
FREE Vert, Horiz, β or gm
WINDOW Vert, Horiz, β or gm in the WINDOW area
Data save/read Memory Setup:256 pcs, REF:4 screens
External memory USB1.1:Setup, waveform, screen hardcopy (BMP, TIFF, PNG)
Remote Equipped with a remote 10BASE-T/100BASE-TX 1 port via LAN connection
Power supply AC100~240V 50/60Hz, Maximum operating power:500VA (400W), Standby:50VA (7W)
External dimensions(mm)
Weight
CS-3100 / CS-3100-300V * 2 Approx. 424W x 220H x 555D (excluding protrusions and accessories)
approx. 28kg (excluding accessories)
CS-3200, 3300 Approx. 424W x 354H x 555D (excluding protrusions and accessories)
approx. 43kg (excluding accessories)

CS-3100-300V is a made-to-order product.

Standard Accessories

Lambda-CS-3000-Series-Curve-Tracer-FIG- (3)

CS-500 Test adapter
Lambda-CS-3000-Series-Curve-Tracer-FIG- (4) This is a jig for connecting the part to be measured, provided by the customer, and the fixture. (Standard accessories)

FAQs

Q: What should I do if the pulse width is too narrow?

A: If the pulse width is narrow, adjust the settings to widen it within the allowable range to generate the max. peak voltage.

Q: How to calibrate the collector supply output?

A: The collector supply output can be calibrated by enabling/disabling the output and adjusting the floating capacity as per the instructions provided.