Kinetik Wellbeing Finger Pulse Oximeter - User Manual

Model: PO6L

Instructions to User

Thank you for purchasing the Kinetik Wellbeing Finger Pulse Oximeter. This manual is compiled in accordance with medical device directives and harmonized standards. Information is subject to change without notice.

This manual describes the product's structure, functions, specifications, transportation, installation, usage, operation, repair, maintenance, storage, and safety procedures. Please read it carefully before use.

⚠️ WARNING:

1 Safety

1.1 Instructions for Safe Operations

1.2 Warnings

1.3 Attentions

1.4 Indication for Use

The Fingertip Pulse Oximeter is a non-invasive device intended for spot-checking of arterial hemoglobin oxygen saturation (SpO2) and pulse rate in adult and pediatric patients in home and hospital environments (including clinical use in internal medicine/surgery, anesthesia, intensive care, etc.). This device is not intended for continuous monitoring.

2 Overview

Pulse oxygen saturation (SpO2) is the percentage of oxyhemoglobin (HbO2) in total hemoglobin (Hb) in the blood, representing blood oxygen concentration. It is an important bio-parameter for respiration. Kinetik Wellbeing developed this Pulse Oximeter for easier and more accurate SpO2 measurement, which also measures pulse rate simultaneously.

Features include low power consumption, convenient operation, small size, and portability. Users place a finger into the sensor, and the display shows the measured Hemoglobin Saturation value.

2.1 Classification

2.2 Features

2.3 Major Applications and Scope of Application

The Pulse Oximeter measures human Hemoglobin Saturation and pulse rate via the finger, indicating pulse intensity by a bar display. It is suitable for home use, oxygen bars, social medical organizations, and general saturation oxygen and pulse rate measurement.

The product is not suitable for continuous patient supervision.

The device is not recommended for use in cases of carbon monoxide poisoning, as it may lead to overrating.

2.4 Environment Requirements

Storage Environment:

Operating Environment:

3 Principle and Caution

3.1 Principle of Measurement

The Oximeter uses Lambert-Beer Law to establish a data process formula based on the spectrum absorption characteristics of reductive hemoglobin (Hb) and oxyhemoglobin (HbO2) in glow & near-infrared zones. The instrument employs Photoelectric Oxyhemoglobin Inspection Technology and Capacity Pulse Scanning & Recording Technology. Two beams of different wavelength lights are focused onto the fingertip through a sensor. A photosensitive element obtains the measured signal, which is then processed by electronic circuits and a microprocessor for display.

[Diagram Description: Figure 1 illustrates the operating principle, showing light sources transmitting through a finger and a sensor detecting the transmitted light.]

3.2 Caution

  1. Ensure proper finger placement (refer to Figure 5) for accurate measurement.
  2. The SpO2 sensor and photoelectric receiving tube must be correctly positioned relative to the subject's arteriole.
  3. Do not use the SpO2 sensor on a location or limb with an arterial canal, blood pressure cuff, or receiving intravenous injection.
  4. Ensure the optical path is free from obstacles like rubberized fabric.
  5. Excessive ambient light (e.g., fluorescent lamps, direct sunlight) can affect measurement results.
  6. Strenuous subject actions or extreme electrosurgical interference may also affect accuracy.
  7. Users cannot use enamel or other makeup.

3.3 Clinical Restrictions

  1. Measurement relies on arteriole pulse; substantial pulsating blood flow is required. Weak pulse (due to shock, low temperature, bleeding, or vasoconstrictive drugs) can decrease the SpO2 waveform (PLETH) and increase susceptibility to interference.
  2. Inaccurate SpO2 readings may occur in individuals with significant amounts of staining dilution drugs (e.g., methylene blue, indigo green, acid indigo blue), carbon monoxide hemoglobin (COHb), methionine (Me+Hb), thiosalicylic hemoglobin, or certain icterus conditions.
  3. Drugs like dopamine, procaine, prilocaine, lidocaine, and butacaine can cause serious errors in SpO2 measurement.
  4. While SpO2 serves as a reference for anemic anoxia and toxic anoxia, some patients with severe anemia may still report good SpO2 measurements.

4 Technical Specifications

5 Accessories

6 Installation

6.1 View of the Front Panel

[Diagram Description: Figure 2 shows the front view of the pulse oximeter, indicating the display screen and the power/function switch button.]

6.2 Battery Installation

[Diagram Description: Figure 3 illustrates the process of installing two AAA batteries into the device, showing the correct orientation for positive and negative terminals.]

  1. Refer to Figure 3 and insert the two AAA size batteries properly in the correct direction.
  2. Replace the cover.

Take care when inserting batteries to avoid damaging the device.

6.3 Mounting the Hanging Rope

[Diagram Description: Figure 4 shows how to attach the hanging rope to the device by threading one end through the loop and then through the first loop to secure it.]

  1. Put one end of the rope through the hole on the device.
  2. Put the same end of the rope through the first loop and tighten it.

7 Operating Guide

  1. Insert the two batteries correctly and replace the cover.
  2. Open the clip as shown in Figure 5.
  3. Place the patient's finger into the rubber cushions of the clip, ensuring it is in the correct position, and then clip the finger.
  4. Press the switch button once on the front panel.
  5. Keep the patient at ease and avoid shaking the finger during measurement. Measurement is not recommended if the patient is moving.
  6. Read the information directly from the screen display.
  7. In the boot-strap state, pressing the button resets the device.

Fingernails and the luminescent tube should be on the same side of the finger within the sensor.

[Diagram Description: Figure 5 shows a finger correctly inserted into the pulse oximeter clip, with the nail and sensor alignment indicated.]

8 Repairing and Maintenance

High-pressure sterilization cannot be used on the device.

Do not immerse the device in liquid. It is recommended to keep the device in a dry environment, as humidity may reduce its useful life or damage it.

9 Troubleshooting

Trouble Possible Reason Solution
The SpO2 and Pulse Rate is not displaying normally 1. The finger is not properly positioned.
2. The patient’s SpO2 is too low to be detected.
1. Place the finger properly and try again.
2. Try again; go to a hospital for a diagnosis if you are sure the device works correctly.
The SpO2 and Pulse Rate displayed is not stable 1. The finger is not placed inside deep enough.
2. The finger is shaking, or the patient is moving.
1. Place the finger properly and try again.
2. Ensure the patient is calm.
The device is not turning on 1. The batteries are drained or almost drained.
2. The batteries are not inserted properly.
3. Malfunction of the device.
1. Change batteries.
2. Reinstall batteries.
3. Please contact Kinetik Wellbeing.
The display suddenly turned off 1. The device will power off automatically when it gets no signal within 5 seconds.
2. The batteries are almost drained.
1. Normal.
2. Change batteries.

10 Key of Symbols

Symbol Description
Type BF Type BF equipment
? Refer to instruction manual/booklet
SpO2% The pulse oxygen saturation (%)
PRbpm Pulse rate (bpm)
[low-voltage indicator] The battery voltage indication is deficient (change the battery in time avoiding the inexact measure)
[signal inadequacy indicator] 1. No finger inserted
2. An indicator of signal inadequacy
[battery positive] Battery positive electrode
[battery negative] Battery cathode
[power switch] Power switch
SN Serial number
[alarm inhibit] Alarm inhibit
[WEEE symbol] WEEE (2002/96/EC)
IP22 Ingress of liquids rank
CE0197 This item is compliant with Medical Device Directive 93/42/EEC of June 14, 1993, a directive of the European Economic Community.
EUROPEAN REPRESENTATIVE

11 Function Specification

Display Information

Display Information Display Mode
The Pulse Oxygen Saturation (SpO2) Digital
Pulse Rate (BPM) Digital
Pulse Intensity (bar-graph) Digital bar-graph display

SpO2 Parameter Specification

Pulse Parameter Specification

Pulse Intensity

Battery Requirement

Battery Useful Life

Dimensions and Weight

Appendix: Electromagnetism Compatibility

Guidance and manufacturer’s declaration – electromagnetic emissions for all EQUIPMENT and SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic emission

The PO6L is intended for use in the electromagnetic environment specified below. The customer or user of the PO6L should assure that it is used in such an environment.

Emission test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 The PO6L uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emission CISPR 11 Class B The PO6L is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2 N/A
Voltage fluctuations/ flicker emissions IEC 61000-3-3 N/A

Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS

Guidance and manufacturer’s declaration – electromagnetic immunity

The PO6L is intended for use in the electromagnetic environment specified below. The customer or the user of PO6L should assure that it is used in such an environment.

Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic discharge (ESD) IEC 61000-4-2 ±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC 61000-4-4 ±2 kV for power supply lines N/A Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5 ±1 kV differential mode N/A Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions, and voltage variations on power supply input lines IEC 61000-4-11 <5% UT (>95% dip in UT) for 0.5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 sec
N/A Mains power quality should be that of a typical commercial or hospital environment. If the user of the PO6L requires continued operation during power mains interruptions, it is recommended that the PO6L be powered from an uninterruptible power supply or a battery.
Power frequency (50/60Hz) Magnetic field IEC-61000-4-8 3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.

NOTE: UT is the a.c. mains voltage prior to application of the test level.

Guidance and manufacturer’s declaration – electromagnetic immunity – for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING

Guidance and manufacturer’s declaration – electromagnetic immunity

The PO6L is intended for use in the electromagnetic environment specified below. The customer or the user of PO6L should assure that it is used in such an environment.

Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Radiated RF IEC 61000-4-3 3 V/m
80 MHz to 2.5 GHz
3 V/m Portable and mobile RF communications equipment should be used no closer to any part of the PO6L, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance:
d = √(P) × 3.5 for 80 MHz to 800 MHz
d = √(P) × 7 for 800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol: [RF interference symbol].

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the PO6L is used exceeds the applicable RF compliance level above, the PO6L should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the PO6L.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING

Recommended separation distances between portable and mobile RF communications equipment and the PO6L

The PO6L is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the PO6L can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the PO6L as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitter (W) Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.39 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

Models: PO6l, Pulse Oximeter

File Info : application/pdf, 2 Pages, 545.46KB

PDF preview unavailable. Download the PDF instead.

PO6L-manual Microsoft Word for Microsoft 365

Related Documents

Preview Kinetik JPD-500G Finger Pulse Oximeter Quick Start Guide
A concise quick start guide for the Kinetik JPD-500G Finger Pulse Oximeter, detailing setup, preparation, and measurement steps. Includes information on battery installation, user preparation, and device usage.
Preview Kinetik Wellbeing Dual-Channel TENS Pain Reliever User Manual
Comprehensive user manual for the Kinetik Wellbeing Dual-Channel TENS Pain Reliever (Model AD-2126), covering intended use, operation, maintenance, troubleshooting, and safety guidelines.
Preview Kinetik WT2 TENS Pain Reliever PennyPad Quick Start Guide
Step-by-step instructions for setting up and using the Kinetik WT2 TENS Pain Reliever PennyPad for effective pain relief.
Preview Kinetik Wellbeing Carbon Reduction Plan and Net Zero Strategy
Kinetik Wellbeing's comprehensive Carbon Reduction Plan detailing their commitment to achieving Net Zero emissions by 2040/2045. Includes baseline emissions (2022), current reporting (2024), and detailed strategies for energy, transportation, supply chain, waste management, and business travel to meet sustainability goals.
Preview Kinetik DT-K111D Digital Thermometer Quick Start Guide
A concise quick start guide for the Kinetik DT-K111D Digital Thermometer, detailing setup, disinfection, and usage instructions.
Preview Kinetik Wellbeing Smart Body Analyser Scale BTS200 Instruction Manual
Comprehensive instruction manual for the Kinetik Wellbeing Smart Body Analyser Scale BTS200. Learn about setup, usage, safety, troubleshooting, and specifications for accurate body composition analysis.
Preview Troubleshooting Blood Pressure Monitor Readings
Guidance on ensuring accurate blood pressure readings by properly fitting the cuff, positioning the arm, and resting before measurement. Includes advice on averaging multiple readings.
Preview Kinetik Wellbeing BG-710 Blood Glucose Monitoring System Quick Start Guide
A quick start guide for the Kinetik Wellbeing BG-710 Blood Glucose Monitoring System, detailing the steps for accurate blood glucose testing.