ELECTRICAL SPECIFICATIONS

  • Accuracy is indicated as ± (% readings + no. of digits*resolution) at 23°C ± 5°C, <80%RH

Voltage (RCD, LOOP, Phase sequence)

Range [V]Resolution [V]Accuracy
15 ÷ 4601±(3.0% rdg + 2dgt)

Frequency

Range [Hz]Resolution [Hz]Accuracy
47.0 ÷ 63.60.1±(0.1% rdg + 1dgt)

Continuity test on protective and equalizing conductors

Range []Resolution [Ω]Accuracy (*)
0.01 ÷ 99.990.01±(5.0%rdg + 3dgt)

(*) calibrate the cables to null their resistance

  • Test current: > 200mA DC for R≤5Ω (calibration included) ; Resolution for DC current :1mA
  • Open-circuit voltage: 4V ≤ V0 ≤ 12V

Insulation resistance (DC voltage)

Test voltage[V]Range [MΩ]Resolution [MΩ]Accuracy
 

50

0.01 ÷ 9.990.01±(2.0%rdg + 2dgt)
10.0 ÷ 49.90.1
50.0 ÷ 99.90.1±(5.0%rdg + 2dgt)
 

100

0.01 ÷ 9.990.01±(2.0%rdg + 2dgt)
10.0 ÷99.90.1
100.0 ÷ 199.90.1±(5.0%rdg + 2dgt)
 

250

0.01 ÷ 9.990.01±(2.0%rdg + 2dgt)
10.0 ÷ 99.90.1
100 ÷ 4991±(5.0%rdg + 2dgt)
 

500

0.01 ÷ 9.990.01 

±(2.0%rdg + 2dgt)

10.0 ÷ 199.90.1
200 ÷ 4991
500 ÷9991±(5.0%rdg + 2dgt)
 

1000

0.01 ÷ 9.990.01 

±(2.0%rdg + 2dgt)

10.0 ÷ 199.90.1
200 ÷ 9991
1000 ÷

 

1999

1±(5.0%rdg + 2dgt)
  • Open-circuit voltage: nominal test voltage –0% +10%
  • Short circuit current: <6.0mA at 500V test voltage
  • Nominal test current: >1mA if load= 1kΩ*Vnom (Vnom=50V, 100V, 250V, 500V, 1000V)
  • Safety protection: the display shows an error message for input voltage >10V

Z Line (Line-Line, Line-Neutral, Line-PE)

Range [Ω]Resolution [Ω]Accuracy
0.00 ¸ 199.9 mΩ (*)0.1 mΩ (*)±(5.0% rdg + 1mΩ) (*)
200 ¸ 1999 mΩ (*)1 mΩ (*)
0.01 ¸ 9.99 Ω0.01 Ω±(5.0% rdg + 3dgt)
10.0 ¸ 199.9 Ω0.1 Ω

(*) By means of the IMP57 optional accessory

  • Maximum test current: 5.81A (at 265V); 10.10A (at 457V)
  • Test voltage ranges: 100÷265V (Line-Neutral) / 100÷460V (Line-Line); 50/60Hz ± 5%
  • Protection type: MCB (B, C, D, K), Fuse (gG, aM)
  • Insulation materials: PVC, Rubber butyl, EPR, XLPE

First fault current (IT systems)

Range (mA)Resolution (mA)Accuracy
0.1 ¸ 0.90.1±(5.0% rdg + 1dgt)
1 ¸ 9991±(5.0% rdg + 3dgt)
  • Limit contact voltage (ULIM) : 25V, 50V

RCD test (Molded case type)

  • RCD type: AC (HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-2 ), A ( HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-3), B(HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-4 ) – General (G), Selective (S) and Delayed ( HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1)
  • Rated tripping currents (IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N):: 6mA, 10mA, 30mA, 100mA, 300mA, 500mA, 650mA, 1000mA
  • Line-PE, Line-N voltage: 100V÷ 265V RCD type AC and A, 190V÷ 265V RCD type B
  • Frequency: 50/60Hz ± 5%

RCD tripping current (Molded case type – RCD General)

RCD typeIHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6NRange IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N [mA]Resolution [mA]Accuracy IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N
AC, A, B6mA,10mA 

(0.2 ÷1.1) IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N

 

£ 0.1 IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N

– 0%, +10%IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N
AC, A, B30mA IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N 300mA– 0%, +5%IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N
AC, A500mA IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N 650mA

RCD molded-type tripping time range [ms] (TT/TN system)

x 1/2x 1x 2x 5AUTOHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-5AUTO+HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-5
\GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1
 

6mA

AC99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111
A99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111
B999999999999999999310
 

10mA

AC99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111
A99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111
B999999999999999999310
 

30mA

AC99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111
A99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111
B999999999999999999310
 

100mA

AC99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
A99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
B999999999999999999310
 

300mA

AC99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
A99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
B999999999999999999310
500mA

650mA

AC99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
A999999999999999999160210310
B
 

1000mA

AC999999999999999999160210
A999999999999999999
B

Resolution: 1ms, Accuracy: ±(2.0%rdg + 2dgt)

RCD Molded type tripping time range [ms] (IT system)

x 1/2x 1x 2x 5AUTOHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-5AUTO+HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-5
\GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1
6mA 10mA

30mA

AC99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111
A
B
100mA

300mA

500mA

650mA

AC99999999999999999916021050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
A
B
 

1000mA

AC999999999999999999160210
A
B
  • Resolution: 1ms, Accuracy: ± (2.0%rdg + 2dgt)

Test on earth leakage delay tester RCDs (with RCDX10 optional accessory)

  • RCD type: AC (HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-2 ), A (HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-3 ), B(HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-4 ) – General (G), Selective (S) and Delayed ( )
  • Rated tripping currents (IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N): 0.3A ÷ 10A
  • Line-PE, Line-N voltage: 100V÷ 265V RCD type AC and A, 190V÷265V RCD type B
  • Frequency: 50/60Hz ± 5%

Earth leakage delay tester RCDs tripping current (RCD General)

RCD typeIDNRange IDN [mA]Resolution [mA]Accuracy IDN
AC, A, B300mA IDN £1A(0.3 ÷ 1.1) IDN 0.1IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N– 0%, +5%IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N
AC, A1.1A IDN £10A

Earth leakage delay tester RCDs trip out time range [ms] (TT/TN system)

x 1/2x 1x 2x 5AUTOHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-5AUTO+HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-5
\GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1GSHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-1
0.3A

÷

1.0A

AC99999999999999999920025050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
A99999999999999999920025050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
B999999999999999999310
1.1A

÷

3.0A

AC99999999999999999920025050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
A99999999999999999920025050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
B999999999999999999
3.1A

÷

6.5A

AC99999999999999999920025050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
A99999999999999999920025050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
B999999999999999999
6.6A

÷

10.0A

AC999999999999999999200250
A999999999999999999
B

Resolution: 1ms, Accuracy: ±(2.0%rdg + 2dgt)

Earth leakage delay tester RCDs trip out time range [ms] (IT system)

x 1/2x 1x 2x 5AUTO 
\GS GS GS GS GS GS 
0.3A

÷

3.0A

AC99999999999999999920025050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
A
B
3.1A

÷

6.5A

AC99999999999999999920025050150HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111HT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-111310
A
B
6.6A

÷

10.0A

AC999999999999999999200250
A
B

Resolution: 1ms, Accuracy: ±(2.0rdg + 2dgt)

NoTripTest – Non-trip earth loop impedance

  • Test voltage: 100÷265V (Line-PE), 50/60Hz ± 5%

NoTripTest – Systems with Neutral wire

Range [Ω]Resolution [Ω]Accuracy (*)
0.01 ÷ 9.990.01±(5% reading + N/10)
10.0 ÷ 199.90.1±(5% reading + N)
200 ÷ 19991±(5% reading + 3N)
  • If IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N <30mA, test current = IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N/2 and N[Ω]=30/IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N Ω; if IHT-Instruments-MACROEVTEST-Professional-Installation-Safety-Tester-fig-6N ≥ 30mA, test current <15mA and N=1Ω

NoTripTest – Systems without Neutral wire

Range [Ω]Resolution [Ω]Accuracy (*)
1 ÷ 19991-0%, +(5.0% lettura +N)
Contact voltage (RCD and NoTripTest)
Range [V]Resolution [V]Accuracy
0 ÷ Utlim0.1-0%, +(5.0% rdg + 3V)
Contact voltage (EARTH test – TT system)
Range [V]Resolution [V]Accuracy
0 ÷ 99.90.1-0%, +(5.0% rdg + 3V)
Contact voltage (EARTH test – TN system)
Range [V]Resolution [V]Accuracy
0 ÷ 99.90.1-0%, +(5.0% rdg + 3V)
100 ÷ 9991
Ground resistance with 3-wire method
Range [Ω]Resolution [Ω]Accuracy (*)
0.01 ÷ 9.990.01 

±(5.0% rdg + 3dgt)

10.0 ÷ 99.90.1
100 ÷ 9991
1.00k ÷ 49.99k0.01k
  • Test current: <10mA – 77.5Hz, Open-circuit voltage: < 20Vrms
  • (*) Add 5% to the accuracy if the probe resistances (Rs or Rh) > 100 x Rmeas
Soil resistivity with 4-wire Wenner method
Range [Ωm]Resolution [Ωm]Accuracy (*)
0.06 ÷ 9.990.01 

 

±(5.0% rdg + 3dgt)

10.0 ÷ 99.90.1
100 ÷ 9991
1.00k ÷ 9.99k0.01k
10.0k ÷ 99.9k0.1k
100k ÷ 999k1k
1.00M ÷ 3.14M0.01M

(*) with distance d=10m, Distance “d” range: 1 ÷ 10m
Test current: <10mA – 77.5Hz, Open-circuit voltage: < 20Vrms

Phase sequence rotation with 1-wire method
Voltage range P-N, P-PE[V]Frequency range
100 ÷ 26550Hz/60Hz ± 5%

Measurement is only carried out by direct contact with metal live parts (not on insulation sheath)

Voltage drop on main power lines (DV%)
Range (%)Resolution (%)Accuracy
0 ÷ 1000.1±(10.0% rdg + 4dgt)
Leakage current (by HT96U optional clamp transducer)
Range [mA]Resolution [mA]Accuracy
0.5 ÷ 999.90.1±(5.0% rdg + 2dgt)
Environmental parameters (AUX function)
ParameterRangeResolutionAccuracy
Temperature [°C]-20°C ¸ 80°C0.1 °C 

 

±(2.0%rdg+2dgt)

Temperature [°F]-4°F ¸ 176°F0.1 °F
Relative humidity [%HR]0 ¸ 100%HR0.1% UR
DC output voltage0.1mV ¸ 1.0V0.1mV
 

Illuminance [Lux]

0.001Lux ¸ 20.00 Lux (*)0.001 ¸ 0.02 Lux
0.1 Lux ¸ 2000 Lux (*)0.1 ¸ 2 Lux
1 Lux ¸ 20 kLux (*)1 ¸ 20 Lux
  • (*) Accuracy of HT53 lux probe is according to Class AA

Measurement of main parameters and harmonics (PQA)

AC TRMS Voltage

Range [V]Resolution [V]Accuracy
15.0 ÷ 459.90.1V±(1.0%rdg + 1dgt)
  • Allowed crest factor ≤ 1,5 ; Frequency: 42.5 ÷ 69.0 Hz

Frequency

Range [Hz]Resolution [Hz]Accuracy
42.5 ÷ 69.00.01±(2.0%rdg + 2dgt)
  • Allowed voltage: 15.0 ÷ 459.9V ; Allowed current: 5%FS clamp ÷ FS clamp

AC TRMS Current

FS clampRange [A]Resolution [A]Accuracy
≤ 10A5% FS ÷ 9.990.011Ph: ±(1.0%rdg + 3 dgt) 3Ph: ±(2.0%rdg + 5 dgt)
10A ≤ FS ≤ 2005% FS ÷ 199.90.1
200A ≤ FS ≤ 30005% FS ÷ 29991
  • Range: 5 ÷ 999.9 mV; Values under 5mV are zeroed
  • Allowed crest factor ≤ 3; Frequency: 42.5 ÷ 69.0 Hz

Active power (@ 230V in 1Ph systems, 400V in 3Ph systems, cosθ=1, f=50.0Hz)

FS clampRange [kW]Resolution [kW]Accuracy
≤10A0.000 ÷ 9.9990.001 

1Ph: ±(2.0%rdg + 5 dgt) 3Ph: ±(2.5%rdg + 8 dgt)

10A ≤ FS ≤ 2000.00 ÷ 999.990.01
200A ≤ FS ≤ 10000.0 ÷ 999.90.1
1000A ≤ FS ≤ 30000 ÷ 99991

Reactive power (@ 230V in 1Ph systems, 400V in 3Ph systems, cosθ=0, f=50.0Hz)

FS clampRange [kVAr]Resolution [kVAr]Accuracy
≤ 10A0.000 ÷ 9.9990.001 

1Ph: ±(2.0%rdg + 7 dgt) 3Ph: ±(3.0%rdg + 8 dgt)

10A ≤ FS ≤ 2000.00 ÷ 999.990.01
200A ≤ FS ≤ 10000.0 ÷ 999.90.1
1000A ≤ FS ≤ 30000 ÷ 99991

Power factor (@ 230V in 1Ph systems, 400V in 3Ph systems, f=50.0Hz)

RangeResolutionAccuracy
0.70c ¸ 1.00 ¸ 0.70i0.01±(4.0%rdg + 10dgt) if I £ 10%FS

±(2.0%rdg + 3dgt) if I > 10%FS

cosθ (@ 230V in 1Ph systems, 400V in 3Ph systems, f=50.0Hz)

RangeResolutionAccuracy
0.70c ÷ 1.00 ÷ 0.70i0.01±(4.0%rdg + 10dgt) if I £ 10%FS

±(1.0%rdg + 7dgt) if I > 10%FS

Voltage harmonics (@ 230V in 1Ph systems, 400V in 3Ph systems, f=50.0Hz)

Range [%]Resolution [%]OrderAccuracy
0.1 ÷ 100.00.101 ÷ 25±(5.0%rdg + 5dgt)
  • Frequency of fundamental: 42.5 ÷ 69.0 Hz, DC accuracy not declared

Current harmonics (f=50Hz)

Range [%]Resolution [%]OrderAccuracy
 

0.1 ¸ 100.0

 

0.1

01 ÷ 9±(5.0%rdg + 5dgt)
10 ÷ 17±(10.0%rdg + 5dgt)
18 ÷ 25±(15.0%rdg + 10dgt)

GENERAL SPECIFICATIONS

DISPLAY AND MEMORY

  • Features: Touch screen, color graphic LCD, 320x240mm
  • Memory: 999 locations, 3 marker levels
  • Communication: Optical-USB and built-in WiFi

POWER SUPPLY

  • Batteries: 6 x 1.2V(rechargeable) type AA or 6 x 1.5V type AA
  • Battery life: > 500 test for each funtions
  • Auto Power OFF: after 5 min of idleness (disabled)

MECHANICAL FEATURES

  • Dimensions (L x W x H): 225 x 165 x 75mm
  • Weight (included batteries): 1.2kg

WORKING ENVIRONMENTAL CONDITIONS

  • Reference temperature: 23°C ± 5°C
  • Working temperature: 0°C ÷ 40°C
  • Relative humidity: <80%RH
  • Storage temperature: -10°C ÷ 60°C
  • Storage humidity: <80%RH

TEST VERIFIES REFERENCE STANDARDS

  • Continuity test with 200mA: IEC/EN61557-4
  • Insulation resistance: IEC/EN61557-2
  • Earth resistance: IEC/EN61557-5
  • Fault loop impedance: IEC/EN61557-3
  • RCD test: IEC/EN61557-6
  • Phase sequence: IEC/EN61557-7
  • Multifunction: IEC/EN61557-10
  • Prospective short circuit current: EN60909-0
  • Earth resistance on TN systems: EN61936-1 + EN50522
  • Test on EVSE devices: IEC/EN61851-1, IEC/EN60364-7-722 (with EV-TEST100)

GENERAL REFERENCE STANDARDS

  • Safety of measuring instruments: IEC/EN61010-1, IEC/EN61010-031, IEC/EN61010-2-032
  • Product type standard: IEC/EN61557-1
  • Technical documentation: IEC/EN61187
  • Insulation: double insulation
  • Pollution degree: 2
  • Encapsulation: IP40
  • Overvoltage category: CAT IV 300V∼ (to ground), max 415V between inputs
  • Max height of use: 2000m

This instrument satisfies the requirements of Low Voltage Directive 2014/35/EU (LVD) and of EMC Directive 2014/35/EU This instrument satisfies the requirements of European Directive 2011/65/EU (RoHS) and 2012/19/EU (WEEE)

Frequently Asked Questions

Q: What should I do if the tester displays an error message for input voltage?
A: If the input voltage exceeds 10V, disconnect the tester immediately to prevent damage and ensure safety.

Q: How do I select the appropriate test voltage range for insulation resistance testing?
A: Refer to the insulation materials used and choose the corresponding test voltage range as per the specifications provided.

Q: What are the recommended safety protections for conducting electrical tests?
A: Use MCB (B, C, D, K) or Fuse (gG, aM) as safety protection measures during testing to prevent electrical hazards.