Daviteq MBRTU-PODO Optical Dissolved Oxygen Sensor na may Modbus output
Panimula
Optical Dissolved Oxygen Sensor na may Modbus output MBRTU-PODO
- Tumpak at mababang maintenance optical dissolved oxygen technology (luminescent quenching ).
- RS485/Modbus signal output.
- Standard sa industriya, matibay na body housing na may 3⁄4” NPT sa harap at likod.
- Flexible cable outlet: fixed cable (0001) at detachable cable (0002).
- Integrated (probe-mounted) waterproof pressure sensor.
- Awtomatikong kabayaran sa temperatura at presyon.
- Awtomatikong kompensasyon sa kaasinan na may halaga ng conductivity/konsentrasyon ng kaasinan ng user-input.
- Maginhawang pagpapalit ng cap ng sensor na may pinagsamang pagkakalibrate.
PAGSUKAT NG NATUNAW NA OXYGEN SA TUBIG
Pagtutukoy
Saklaw | DO Saturation%: 0 hanggang 500%. DO Concentration : 0 hanggang 50 mg/L (ppm). Operating Temperatura: 0 hanggang 50°C. Temperatura sa Pag-imbak: -20 hanggang 70°C. Operating Atmospheric Pressure: 40 hanggang 115 kPa. Pinakamataas na Presyon ng Bearing: 1000 kPa. |
Oras ng Pagtugon | GAWIN: T90 ~ 40s para sa 100 hanggang 10%. Temperatura: T90 ~ 45s para sa 5 – 45oC (w/ stirring). |
Katumpakan | GAWIN: 0-100% < ± 1 %. 100-200% < ± 2 %. Temperatura: ± 0.2 °C. Presyon: ± 0.2 kPa. |
Input /output/protocol | Input: 4.5 – 36 V DC. Pagkonsumo: average na 60 mA sa 5V. Output: RS485/Modbus o UART. |
Pag-calibrate |
|
DO Compensation Factors | Temperatura: awtomatiko, buong saklaw.
Salinity: awtomatiko na may user-input (0 hanggang 55 ppt). Presyon:
|
Resolusyon | Mababang hanay (<1 mg/L): ~ 1 ppb (0.001 mg/L). Katamtamang hanay (<10 mg/L): ~ 4-8 ppb (0.004-0.008 mg/L). Mataas na saklaw (>10 mg/L): ~10 ppb (0.01 mg/L).* *Ang mas mataas na hanay, ang mas mababang resolution. |
Inaasahang Buhay ng Cap ng Sensor | Ang isang kapaki-pakinabang na buhay na hanggang 2 taon ay magagawa sa pinakamabuting kalagayan na sitwasyon. |
Iba | Hindi tinatablan ng tubig: IP68 rating na may nakapirming cable. Mga Sertipikasyon: Mga RoH, CE, C-Tick (nasa proseso). Mga Materyales: Ryton (PPS) body. Haba ng cable: 6 m (may mga opsyon). |
Mga Larawan ng Produkto
PROCESS OPTICAL DISSOLVED OXYGEN SENSOR MBRTU-PODO
MBRTU-PODO-H1 .PNG
Mga kable
Mangyaring mag-wire tulad ng ipinapakita sa ibaba:
Kawad kulay | Paglalarawan |
Pula | Power (4.5 ~ 36 V DC) |
Itim | GND |
Berde | UART_RX (para sa pag-upgrade o koneksyon sa PC ) |
Puti | UART_TX (para sa pag-upgrade o koneksyon sa PC) |
Dilaw | RS485A |
Asul | RS485B |
Tandaan: Ang dalawang UART wire ay maaaring putulin kung hindi mag-upgrade/programming probe.
Pag-calibrate at Pagsukat
DO Calibration in Options
I-reset ang pagkakalibrate
a) I-reset ang 100% pagkakalibrate.
Ang user ay sumulat ng 0x0220 = 8
b) I-reset ang 0% na pagkakalibrate.
Ang user ay sumulat ng 0x0220 = 16
c) I-reset ang pagkakalibrate ng temperatura.
Ang user ay sumulat ng 0x0220 = 32
1-point calibration
Ang 1-point calibration ay nangangahulugan ng pag-calibrate ng probe sa punto ng 100% saturation, na maaaring makuha sa pamamagitan ng isa sa mga sumusunod na paraan:
a) Sa tubig na puspos ng hangin (karaniwang pamamaraan).
Ang tubig na puspos ng hangin (para sa halample ng 500 mL) ay maaaring makuha sa pamamagitan ng tuloy-tuloy na (1) paglilinis ng tubig gamit ang hangin gamit ang air bubbler o ilang uri ng aeration mga 3 ~ 5 minuto, o (2) paghalo ng tubig sa pamamagitan ng magnetic stirrer sa ilalim ng 800 rpm sa loob ng 1 oras.
Matapos maging handa ang air-saturated na tubig, isawsaw ang sensor cap at temperature sensor ng probe sa air-saturated na tubig, at i-calibrate ang probe pagkatapos maging stable ang pagbabasa (karaniwang 1 ~ 3 minuto).
Sumulat ang user ng 0x0220 = 1 , pagkatapos ay maghintay ng 30 segundo.
Kung ang huling pagbabasa ng 0x0102 ay wala sa 100 ± 0.5%, pakisuri kung ang katatagan ng kasalukuyang kapaligiran ng pagsubok o subukang muli.
b) Sa tubig-puspos na hangin (maginhawang paraan).
Bilang kahalili, ang 1-pt na pagkakalibrate ay madaling gawin gamit ang water-saturated air, ngunit ang 0 ~ 2% na error ay maaaring sanhi depende sa iba't ibang mga operasyon. Ang mga inirekumendang pamamaraan ay ibinibigay sa ibaba:
i) ilubog ang takip ng sensor at sensor ng temperatura ng probe sa sariwang/tap na tubig 1~2 minuto.
ii) lumabas sa probe at mabilis na isawsaw ang tubig sa ibabaw ng takip ng sensor sa pamamagitan ng tissue.
iii) i-install ang dulo ng sensor sa calibration/storage bottle na may basang espongha sa loob. Iwasan ang direktang pagdikit ng takip ng sensor sa anumang tubig sa bote ng pagkakalibrate/imbak sa panahon ng hakbang ng pagkakalibrate na ito. Panatilihing ~ 2 cm ang distansya sa pagitan ng cap ng sensor at ng basang espongha.
v) hintaying mag-stabilize ang mga pagbabasa (2 ~ 4 minuto ) at pagkatapos ay isulat ang 0x0220 = 2.
2-point calibration (100% at 0% saturation point)
(i) Ilagay ang probe sa air-saturated na tubig, isulat ang 0x0220 = 1 pagkatapos mag-stabilize ang DO reading.
(ii) Matapos maging 100% ang pagbabasa ng DO, ilipat ang probe sa zero oxygen water (gamitin ang sodium sulfide na idinagdag nang labis sa isang
tubig sample).
(iii) Sumulat ng 0x0220 = 2, pagkatapos mag-stabilize ang pagbabasa ng DO (~hindi bababa sa 2 min).
- (iv) Ang user na nagbabasa ng saturation sa 0x0102 para sa 1-point calibration, 0x0104 para sa 2-point calibration.
Hindi kailangan ang 2-point cal para sa karamihan ng mga application, maliban kung kailangan ng mga user ng napakatumpak na pagsukat sa mababang konsentrasyon ng DO (<0.5 ppm). - Ang pagpapatupad ng "0% calibration" na walang "100% calibration" ay hindi pinapayagan.
Pag-calibrate ng punto para sa temperatura
i) Ang user ay sumulat ng 0x000A = ang ambient temperature x100 (Hal: Kung ang ambient temperature = 32.15, ang user ay sumulat ng 0x000A=3215).
ii) Ang temperatura ng pagbabasa ng user sa 0x000A . Kung ito ay katumbas ng iyong inilagay, ang pagkakalibrate ay tapos na. Kung hindi, pakisubukang muli ang Hakbang 1.
Modbus RTU Protocol
Istraktura ng command:
- Ang mga utos ay hindi dapat ipadala nang mas maaga kaysa sa 50mS mula sa pagkumpleto ng huling tugon.
- Kung ang inaasahang tugon mula sa alipin ay hindi nakita para sa > 25mS, magtapon ng error sa komunikasyon.
- Sinusunod ng Probe ang pamantayan ng Modbus para sa mga function na 0x03, 0x06, 0x10, 0x17
Istraktura ng Serial Transmission:
- Big-endian ang mga uri ng data maliban kung iba ang nabanggit.
- Ang bawat pagpapadala ng RS485 ay magkakaroon ng: isang panimulang bit, 8 bits ng data, walang parity bit, at dalawang stop bit;
- Default na Baud rate: 9600 (ang ilan sa mga probe ay maaaring may Baudrate na 19200);
- Default na address ng Alipin: 1
- Ang 8 data bits na ipinadala pagkatapos ng start bit ay ang pinaka makabuluhang bit muna.
- Bit Sequence
Simula ng kaunti | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Tumigil ng kaunti |
Timing
- Ang mga pag-update ng firmware ay dapat na tumakbo sa loob ng 5 segundo ng power o soft reset Probe tip LED ay magiging solid blue sa panahong ito
- Ang unang command ay hindi maaaring patakbuhin nang mas maaga kaysa sa 8 segundo mula sa power on o soft reset
- Kung walang inaasahang tugon mula sa isang ibinigay na command timeout ay magaganap pagkatapos ng 200ms
Modbus RTU protocol:
Register # | R/W | Mga Detalye | Uri | Mga Tala |
0x0003 | R | LDO (mg/L) x100 | Uint16 | |
0x0006 | R | Saturation % x100 | Uint16 | |
0x0008 | R/W | Kaasinan (ppt) x100 | Uint16 | |
0x0009 | R | Presyon (kPa) x100 | Uint16 | |
x000A | R | Temperatura (°C) x100 | Uint16 | |
0x000F | R | Rate ng Baud | Uint16 | Tandaan 1 |
0x0010 | R | Address ng Alipin | Uint16 | |
0x0011 | R | Probe ID | Uint32 | |
0x0013 | R | Sensor Cap ID | Uint32 | |
0x0015 | R | Probe Firmware Bersyon x100 | Uint16 | Tandaan 2 |
0x0016 | R | Probe Firmware Minor Revision | Uint16 | Tandaan 2 |
0x0063 | W | Rate ng Baud | Uint16 | Tandaan 1 |
0x0064 | W | Address ng Alipin | Uint16 | |
0x0100 | R | LDO (mg/L) | Lutang | |
0x0102 | R | Saturation % | Lutang | |
0x0108 | R | Presyon (kPa) | Lutang | |
0x010A | R | Temperatura (°C) | Lutang | |
0x010C | R/W | Kasalukuyang Probe Datetime | 6 byte | Tandaan 3 |
0x010F | R | Mga error bit | Uint16 | Tandaan 4 |
0x0117 | R | Kaasinan (ppt) | Lutang | |
0x0132 | R/W | Offset ng Temperatura | Lutang | |
0x0220 | R/W | Pag-calibrate ng mga Bit | Uint16 | Tandaan 5 |
0x02CF | R | Serial Number ng Membrane Cap | Uint16 | |
0x0300 | W | Soft restart | Uint16 | Tandaan 6 |
Tandaan:
- Tandaan 1: Mga value ng baud rate: 0= 300, 1= 2400, 2= 2400, 3= 4800, 4= 9600, 5= 19200, 6=38400, 7= 115200.
- Tandaan 2: Ang bersyon ng firmware ay address na 0x0015 na hinati sa 100, pagkatapos ay isang decimal pagkatapos ay address na 0x0016. Halample: kung 0x0015 = 908 at 0x0016 = 29, ang bersyon ng firmware ay v9.08.29.
- Tandaan 3: Ang Probe ay walang RTC, kung ang probe ay hindi binibigyan ng tuluy-tuloy na kapangyarihan o na-reset ang lahat ng mga halaga ay mare-reset sa 0.
Ang mga byte ng datetime ay taon, buwan, araw, araw, oras, minuto, segundo. Pinakamahalaga hanggang sa pinakamaliit.
Example: iftheuserwrites0x010C=0x010203040506, thentheDatetime will set to February 3rd, 2001 4:05:06 am. - Tandaan 4: Ang mga bit ay binibilang na hindi gaanong mahalaga sa karamihan, simula sa 1:
- Bit 1 = Error sa Pag-calibrate ng Pagsukat.
- Bit 3 = Probe Temperature na wala sa saklaw, maximum na 120 °C.
- Bit 4 = Konsentrasyon sa labas ng saklaw: minimum na 0 mg/L, maximum na 50 mg/L. o Bit 5 = Probe Pressure Sensor Error.
- Bit 6 = Pressure Sensor na wala sa saklaw: minimum na 10 kPa, maximum na 500 kPa.
Gagamit ang probe ng default na presyon = 101.3 kPa. - Bit 7 = Pressure Sensor Communication error, Probe ay gagamit ng default na presyon = 101.3 kPa.
Tandaan 5:Sumulat (0x0220) 1 Patakbuhin ang 100% pagkakalibrate. 2 Patakbuhin ang 0% pagkakalibrate. 8 I-reset ang 100% pagkakalibrate. 16 I-reset ang 0% pagkakalibrate. 32 I-reset ang pagkakalibrate ng temperatura.
- Note 6: Kung ang 1 ay nakasulat sa address na ito, ang isang soft restart ay isinasagawa, ang lahat ng iba pang mga read/write ay hindi papansinin.
Tandaan 7: kung ang probe ay may built in na pressure sensor isa itong read only na address.
Tandaan 8: Ang mga Value na ito ay mga resulta ng 2 point calibration, habang ang address ng 0x0003 at 0x0006 ay nagpapakita ng mga resulta ng 1 point calibration.
Example Transmissions
CMD: Basahin ang Data ng Probe
Raw Hex: 01 03 0003 0018 B5C0
Address | Utos | Panimulang Address | # ng mga Register | CRC |
0x01 | 0x03 | 0x0003 | 0x0018 | 0xB5C0 |
1 | Basahin | 3 | 0x18 |
Example 1 tugon mula sa probe:
Raw Hex: 01 03 30 031B 0206 0000 2726 0208 0BB8 27AA 0AAA 0000 0000 0000 0BB8 0005 0001 0001 0410 0457 0000 038 0052 0001 031 2741 0000 4 XNUMX XNUMX FADXNUMX
Example 2 tugon mula sa probe:
Raw Hex: 01 03 30 0313 0206 0000 26F3 0208 0000 27AC 0AC8 0000 0000 0000 0000 0005 0001 0001 0410 0457
0000 038C 0052 0001 031A 2748 0000 5BC0
Konsentrasyon (mg/L) | Saturation % | Kaasinan (ppt) | Presyon (kPa) | Temperatura (°C) | Konsentrasyon 2pt (mg/L) | Saturation % 2pt |
0x0313 | 0x26F3 | 0x0000 | 0x27AC | 0x0AC8 | 0x031A | 0x2748 |
7.87 mg/L | 99.71% | 0 ppt | 101.56 kPa | 27.60 °C | 7.94 mg/L | 100.56% |
CMD: Patakbuhin ang 100 % Calibration
Raw Hex: 01 10 0220 0001 02 0001 4330
Address | Utos | Panimulang Address | # ng mga Register | # ng Bytes | Halaga | CRC |
0x01 | 0x10 | 0x0220 | 0x0001 | 0x02 | 0x0001 | 0x4330 |
1 | Sumulat ng Marami | 544 | 1 | 2 | Patakbuhin ang 100% Cal |
Example 1 tugon mula sa probe:
Raw Hex: 01 10 0220 0001 01BB Tagumpay!
CMD: Patakbuhin ang 0 % Calibration
Raw Hex: 01 10 0220 0001 02 0002 0331
Address | Utos | Panimulang Address | # ng mga Register | # ng Bytes | Halaga | CRC |
0x01 | 0x10 | 0x0220 | 0x0001 | 0x02 | 0x0002 | 0x0331 |
1 | Sumulat ng Marami | 544 | 1 | 2 | Patakbuhin ang 0% Cal |
Example 1 tugon mula sa probe:
Raw Hex: 01 10 0220 0001 01BB Tagumpay!
CMD: Update Salinity = 45.00 ppt, Pressure =101.00 kPa, at Temperatura = 27.00 °C
Raw Hex: 01 10 0008 0003 06 1194 2774 0A8C 185D
Address | Utos | Panimulang Address | # ng mga Register | # ng Bytes | Halaga | CRC |
0x01 | 0x10 | 0x0008 | 0x0003 | 0x06 | 0x1194 2774 0A8C | 0x185D |
1 | Sumulat ng Marami | 719 | 1 | 2 | 45, 101, 27 |
Example 1 tugon mula sa probe:
Raw Hex: 01 10 0008 0003 01CA Tagumpay!
Address | Utos | Panimulang Address | # ng mga Register | # ng Bytes | Halaga | CRC |
0x01 | 0x10 | 0x02CF | 0x0001 | 0x02 | 0x0457 | 0xD751 |
1 | Sumulat ng Marami | 719 | 1 | 2 | 1111 |
Example 1 tugon mula sa probe:
Raw Hex: 01 10 02CF 0001 304E Tagumpay!
Mga sukat
DIMENSION DRAWING NG MBRTU-PODO (Yunit: mm)
Pagpapanatili
Kasama sa pagpapanatili ng probe ang paglilinis ng cap ng sensor, pati na rin ang wastong pagkondisyon, paghahanda, at pag-iimbak ng sistema ng pagsubok.
Kapag hindi ginagamit ang probe, lubos na inirerekomendang itabi ang probe na may naka-install na cap ng sensor nito at ang calibration/storage bottle na kasama sa orihinal na packaging, na sinulid sa probe. Ang isang beaker ng malinis na tubig o isang moist/humid capping mechanism ay maaari ding sapat kung ang calibration/storage bottle ay hindi available. Ang espongha sa loob ng bote ng pagkakalibrate/imbakan ay dapat panatilihing basa-basa para sa pinakamahusay na mga resulta.
Iwasang hawakan ng cap ng sensor ang organikong solvent, gasgas, at mapang-abusong banggaan upang palakasin at pahabain ang buhay ng paggana ng cap ng sensor. Ang espesyal na pangangalaga ay dapat gawin upang linisin ang patong ng takip, isawsaw ang probe at takip sa sariwang tubig, at pagkatapos ay i-tap ang tuyo sa ibabaw gamit ang isang tissue. Huwag punasan ang ibabaw ng patong.
Palitan ang takip ng sensor, kung ang patong ng takip ay kupas o natanggal. HUWAG hawakan ang malinaw na bintana sa dulo ng probe pagkatapos tanggalin ang lumang takip. Kung mayroong anumang mga kontaminant o nalalabi sa bintana o sa loob ng takip, maingat na alisin ang mga ito gamit ang walang pulbos na punasan. Pagkatapos ay muling i-screw ang bagong sensor cap papunta sa probe.
Mga Dokumento / Mga Mapagkukunan
![]() |
Daviteq MBRTU-PODO Optical Dissolved Oxygen Sensor na may Modbus output [pdf] Gabay sa Gumagamit MBRTU-PODO Optical Dissolved Oxygen Sensor na may Modbus output, MBRTU-PODO, Optical Dissolved Oxygen Sensor na may Modbus output, Sensor na may Modbus output, Modbus output |