GASTEC No.211H Sulphide Ion Detector Tube

Instructions for Safe Operation and Use

FOR SAFE OPERATION

Carefully read this manual before use.

⚠️ CAUTION: If you do not observe the following precautions, you may suffer injuries or damage the product.

  1. When breaking the tube ends, keep away from eyes.
  2. Do not touch the broken glass tubes, broken pieces, and reagent with bare hand(s).

ℹ️ NOTES: For maintaining performance and reliability of the test results, observe the following.

  1. Use this tube within the temperature range of 0 -50°C (32 - 122°F) in water.
  2. Use this tube between pH values of 5.0 to 12.5.
  3. This tube may be interfered with by coexisting gases. Please refer to the table "INTERFERENCES" below.
  4. The shelf life and storage condition of the tube are marked on the label of the tube box.
  5. Place the higher end plug packing of the tubes above the water surface.

APPLICATION OF THE TUBE

Use this tube for detecting Sulphide Ion in the Waste Water.

SPECIFICATION

(Because of Gastec's commitment to continued improvement, specifications are subject to change without notice.)

[Description of Detector Tube Visual: A glass detector tube labeled "211H" with a Gastec logo. It shows a detecting layer at one end with calibration marks indicating concentrations from 10 to 1000 ppm.]

ParameterValue
Measuring Range10 - 1000 ppm
Sampling Time2 minutes
Detecting Limit1 ppm
Colour ChangeWhite ➡️ Brown
Reaction PrincipleS²⁻ + Pb(CH₃COO)₂ ➡️ PbS

Coefficient of Variation: 10% (for 10 to 100 ppm), 5% (for 100 to 1000 ppm)

** Shelf Life: Please refer to the validity date printed on the tube box.

** Store the tubes in a dark and cool place.

EFFECT BY ATMOSPHERIC CONDITION

Water Temperature: No effect by the water temperature between 0 - 50°C (32-122°F).

pH Value: Use the tube in the pH value of 5.0 - 12.5.

MEASUREMENT PROCEDURE

  1. Take sample water into an approximately 100 mL capacity of dry, clean beaker.
  2. Break tips off a fresh detector tube by bending each tube end in the tube tip holder (optional).
  3. Immerse the filled end of the tube into the sample water as illustrated. Capillary action occurs and the sample water instantly rises through the reagent. If the sample contains sulphide ion, the white reagent in the tube turns to brown colour.
  4. When the sample water rises up to the upper end plug, remove the tube.
  5. Read the concentration level at the interface where the stained reagent meets the unstained reagent.
  6. If the stain exceeds the highest calibration mark (1000 ppm), dilute the sample with pure water and retest using a fresh tube. Obtain true concentration by multiplying the tube reading by the dilution ratio.

True Concentration = (V1 + V2) / V1 × Tube Reading

V1: Volume of Sample water

V2: Volume of dilution (pure water)

ℹ️ NOTES: Do not immerse the tube into sample water past the upper end plug.

INTERFERENCES

SubstanceFormulaConcentrationInterferenceChanges colour by itself to
Sulphite ionSO₃²⁻≥100 ppm-No discolouration
Chloride ionCl⁻≥15000 ppm+No discolouration
Chromate ionCrO₄²⁻≥100 ppm-Yellow
Carbonate ionCO₃²⁻≥500 ppm+No discolouration
Thiosulphate ionS₂O₃²⁻≥100 ppm+No discolouration

This table of interference substances primarily expresses the interference of each coexisting substance in the concentration range, that is equivalent to the target substances concentration. Therefore, the test result may show positive results due to other substances not listed in the table. If more information is needed, please contact us or our distributors in your territory.

INSTRUCTIONS ON DISPOSAL

The reagent of the tube uses a small amount of lead. When disposing the tube regardless of whether it has been used or not, follow the rules and regulations of your local government.

WARRANTY

If you have any questions regarding gas detection and the quality of the tubes, please feel free to contact your Gastec representatives.

Models: No.211H, Sulphide Ion Detector Tube

File Info : application/pdf, 1 Pages, 109.89KB

PDF preview unavailable. Download the PDF instead.

211H Adobe Acrobat Pro DC (32-bit) 21

Related Documents

Preview Gastec 126L Chlorobenzene Detector Tube: Instructions and Specifications
Comprehensive instructions and specifications for the Gastec 126L Chlorobenzene Detector Tube, detailing safe operation, measurement procedures, temperature/humidity/pressure corrections, and interference information.
Preview Gastec No. 92L Acetaldehyde Detector Tube: Instructions and Specifications
Comprehensive instructions and specifications for the Gastec No. 92L Acetaldehyde Detector Tube, covering safe operation, measurement procedures, temperature/humidity/pressure corrections, interferences, and disposal guidelines.
Preview Benzene Detector Tube No. 121SP - Gastec
Technical specifications and performance data for the Gastec Benzene Detector Tube No. 121SP, used for measuring benzene concentrations in the air.
Preview Gastec 30m Extension Hose No. 351A-30: Instruction Manual and Usage Guide
Detailed instruction manual for the Gastec 30m Extension Hose (No. 351A-30), covering safe operation, specifications, measurement procedures, and sampling time/correction factors for gas detection.
Preview Gastec Gas Sampling Pump GV-100S and GV-110S Instruction Manual
Comprehensive instruction manual for Gastec Gas Sampling Pumps GV-100S and GV-110S, covering safe usage, operation, maintenance, and troubleshooting for accurate gas detection.
Preview GASTEC GAP-1 Semi-fixed Gas Alarm System | Safety Monitoring Equipment
Discover the GASTEC GAP-1 semi-fixed gas alarm system. This brochure details its features, specifications, and compatibility with GASTEC wireless gas detectors (GOC-200, CM-9A-BT, GOC-100-2-BT) for effective workplace safety and gas monitoring.
Preview GASTEC GSP-501FT Automatic Air Sampling Pump - Features and Specifications
Comprehensive guide to the GASTEC GSP-501FT Automatic Air Sampling Pump, detailing its features, technical specifications, operation modes, accessories, and compatible detector tubes for air quality monitoring.
Preview Instrucciones Detalladas para el Tubo Detector Gastec 152L de Metiletilcetona
Guía completa de Gastec Corporation sobre el uso, especificaciones, procedimientos de medición, correcciones y precauciones de seguridad para el tubo detector de metiletilcetona modelo 152L.