Gastec No. 126L Chlorobenzene Detector Tube

For Safe Operation

Carefully read this manual and the instruction manual of your Gastec Gas Sampling Pump.

WARNING

CAUTION

NOTES

Application of the Tube

Use this tube for the detection of Chlorobenzene in air or in industrial areas and environmental atmospheric conditions.

Specification

(Because of Gastec's commitment to continued improvement, specifications are subject to change without notice)

Parameter Value
Measuring Range (0.5) – 10 ppm / 10 – 57 ppm
Number of Pump Strokes 3 / 1
Correction Factor 1 / 5.7
Sampling Time 1.5 minutes per pump stroke
Detecting Limit 0.2 ppm (n=3)
Colour Change Yellow -> Pale bluish purple
Reaction Principle C6H5Cl + PbO2 + H2SO4 -> HCl; HCl + Base -> Chloride
Coefficient of Variation 10% (for 0.5 to 2 ppm), 5% (for 2 to 10 ppm)
Shelf Life Refer to the validity date printed on the box of tube.
Storage Condition Store tubes in the refrigerator at 10°C (50°F) or below.

Tube Diagram Description

The detector tube consists of two layers: an 'Oxidising Agent' layer followed by a 'Detecting Layer'. The minimum scale of 0.5 ppm is indicated by a scale line on the tube, as it is not printed.

Correction for Temperature, Humidity, and Pressure

Temperature Correction

Correct for temperature using the table below:

Tube Reading (ppm) 0°C (32°F) 5°C (41°F) 10°C (50°F) 15°C (59°F) 20°C (68°F) 25°C (77°F) 30°C (86°F) 35°C (95°F) 40°C (104°F)
10 - - - 14.7 10 8.3 6.3 5.3 4.1
8 - - 11.7 8 6.7 5.1 4.2 3.3 -
6 - 13.5 8.7 6 5 3.8 3.2 2.5 -
4 12.8 7.6 5.9 4 3.3 2.5 2.1 1.6 -
2 6.8 4.6 3.2 2.9 2 1.7 1.3 1.1 0.8
1 2.7 2.1 1.5 1.4 1 0.8 0.6 0.5 0.4
0.5 1.3 1 0.8 0.7 0.5 0.4 0.3 0.3 0.2

Humidity Correction

No correction is required between 0 – 90% R.H.

Pressure Correction

To correct for pressure, use the formula below:

True Concentration (ppm) = [Tube Reading (ppm) × 1013 (hPa)] / Atmospheric Pressure (hPa)

Note: The Tube Reading value used in the formula should be after any other corrections (e.g., temperature) have been applied.

Measurement Procedure

  1. For checking pump leakage, insert a fresh sealed detector tube into the pump. Follow instructions provided with the pump operating manual.
  2. Break tips off a fresh detector tube using the tube tip breaker of the pump.
  3. Insert the tube into the pump inlet with the arrow [pump direction indicator] on the tube pointing toward the pump.
  4. Ensure the pump handle is fully in. Align the guide mark on the pump body with the guide mark on the handle.
  5. Pull the handle all the way out until it locks at one pump stroke (100 mL). Wait 1.5 minutes and confirm sampling completion. Repeat this procedure two more times.
  6. For measurements higher than 10 ppm, prepare a fresh tube and perform one pump stroke.
  7. Read the concentration level at the interface where the stained reagent meets the unstained reagent.
  8. If temperature correction is necessary, obtain the true concentration using the temperature correction table. Afterwards, multiply by the correction factor for the pump stroke if necessary.
  9. If pressure correction is necessary, use the pressure correction formula.

Interferences

Substance Interference Changes colour by itself to
Chlorine, Hydrogen chloride + Pale bluish purple
Trichloroethylene + Pale bluish purple
Tetrachloroethylene + Pale bluish purple

This table of interference gases primarily expresses the interference of each coexisting gas at a concentration equivalent to the gas concentration being measured. Therefore, the test result may be positive due to other substances not listed in the table. For more information, please contact Gastec representatives.

Dangerous and Hazardous Properties

Threshold Limit Value-Time Weighted Average by ACGIH (2021): 10 ppm

Instructions on Disposal

The reagent of the tube uses a small amount of lead. When disposing of the tube, whether used or not, follow the rules and regulations of your local government.

Warranty

If you have any questions regarding gas detection and quality of the tubes, please feel free to contact your Gastec representatives.

Models: 126L Chlorobenzene Detector Tube, 126L, Chlorobenzene Detector Tube, Detector Tube, Tube

File Info : application/pdf, 2 Pages, 470.18KB

PDF preview unavailable. Download the PDF instead.

126L

References

Related Documents

Preview Gastec No. 92L Acetaldehyde Detector Tube: Instructions and Specifications
Comprehensive instructions and specifications for the Gastec No. 92L Acetaldehyde Detector Tube, covering safe operation, measurement procedures, temperature/humidity/pressure corrections, interferences, and disposal guidelines.
Preview Benzene Detector Tube No. 121SP - Gastec
Technical specifications and performance data for the Gastec Benzene Detector Tube No. 121SP, used for measuring benzene concentrations in the air.
Preview Gastec Gas Sampling Pump GV-100S and GV-110S Instruction Manual
Comprehensive instruction manual for Gastec Gas Sampling Pumps GV-100S and GV-110S, covering safe usage, operation, maintenance, and troubleshooting for accurate gas detection.
Preview GASTEC No.211H Sulphide Ion Detector Tube Instructions
Instructions for using the Gastec No.211H Sulphide Ion Detector Tube for detecting sulphide ions in waste water, including safety precautions, measurement procedures, specifications, interferences, disposal, and warranty information.
Preview Gastec 30m Extension Hose No. 351A-30: Instruction Manual and Usage Guide
Detailed instruction manual for the Gastec 30m Extension Hose (No. 351A-30), covering safe operation, specifications, measurement procedures, and sampling time/correction factors for gas detection.
Preview GASTEC GSP-501FT Automatic Air Sampling Pump - Features and Specifications
Comprehensive guide to the GASTEC GSP-501FT Automatic Air Sampling Pump, detailing its features, technical specifications, operation modes, accessories, and compatible detector tubes for air quality monitoring.
Preview GASTEC GAP-1 Semi-fixed Gas Alarm System | Safety Monitoring Equipment
Discover the GASTEC GAP-1 semi-fixed gas alarm system. This brochure details its features, specifications, and compatibility with GASTEC wireless gas detectors (GOC-200, CM-9A-BT, GOC-100-2-BT) for effective workplace safety and gas monitoring.
Preview Gastec Smoke Tube Set No.550 Instruction Manual
This document provides comprehensive instructions for the Gastec Smoke Tube Set No.550, including safety guidelines, usage instructions, product specifications, set contents, disposal procedures, and warranty information.