GASTEC GSP-501FT Automatic Air Sampling Pump

The GASTEC GSP-501FT is a compact, lightweight, and reliable automatic air sampling pump designed for use with GASTEC detector tubes and sorbent tubes. It offers quiet operation with low noise and features a large screen that simultaneously displays suction flow rate, sampling time, and integrated volume.

Visual Description: The document displays the GSP-501FT pump in various contexts. One image shows the pump mounted on a tripod. Another provides a close-up of the pump's front panel, highlighting its LCD display and control buttons (POWER, MODE, SELECT, START/STOP). A third image illustrates a person wearing the pump attached to their belt, demonstrating its portability for personal exposure monitoring.

Key Features

  • Flow range: 50-500mL/min.
  • Volume Mode flow rate: 10-500mL/min. When set between 10-49 mL/min, intermittent operation is performed at 50mL/min.
  • Equipped with a fixed flow function that automatically regulates load changes.
  • Flow rate, sampling time, and integrated volume are displayed simultaneously on the same screen.
  • Automatic start function allows sampling to commence after a preset time.
  • Supports saving up to 5 different sampling settings via the program mode.
  • Real-time flow volume and integrated flow volume are automatically corrected to standard conditions of 20°C (68°F) or 25°C(77°F).

Display and Indicators: The pump features an easy-to-read, large LCD screen that displays the selected operating mode, flow rate, sampling time, and integrated volume. Two LED lamps blink during sampling operations, providing visible status indication from a distance.

Power: The pump operates using two AA alkaline batteries or two AA nickel-metal hydride batteries.

Specifications

Parameter Value
Model GSP-501FT
Timer Mode Air pump automatically stops at set time. Settable time: 1 minute to 30 hours.
Volume Mode Air pump automatically stops at set volume. Settable integrated volume: 0.010 to 900L.
Flow Rate Setting Timer Mode: 50-500mL/min. Volume Mode: 10-500mL/min (Note: When set to 10-49 mL/min, intermittent operation is performed at 50mL/min).
Suction Pressure 10-49mL/min: 0.0-5.0kPa; 50mL/min: 0.0-40.0kPa; 100mL/min: 0.0-37.0kPa; 200mL/min: 0.0-30.0kPa; 300mL/min: 0.0-23.0kPa; 400mL/min: 0.0-16.0kPa; 500mL/min: 0.0-10.0kPa.
Display Liquid crystal digital display (with backlight). Display range: 0-600mL/min.
Functions Constant flow rate function (built-in set flow rate holding circuit), Autostart function (autostart after set standby time), Diaphragm type air pump, Program Mode (5 sampling settings).
Accuracy (Instantaneous Flow Rate) 50-500 mL/min: ±5%.
Volume Mode (10-49 mL/min): ±(2.5 × sampling time [min]) mL.
Operating Temperature 0-40℃.
Operating Humidity 10 to 90% RH (non-condensing).
Power Source 2 AA alkaline batteries (standard accessories, commercially available) or 2 AA nickel-metal hydride batteries (commercially available).
Battery Life Approximately 20 hours (with 2 AA alkaline batteries, set flow rate of 200 mL/min, suction pressure of 2kPa or less, ambient temperature of 25°C).
Dimensions (W×D×H) 80(W)×40(D)×126(H) mm.
Weight 280g (including batteries).
Standard Accessories 2 AA alkaline batteries, detector tube adaptor, tube tip holder, dust filter (5pcs), instruction manual, warranty certificate, inspection certificate.
Compliance EU Directive: 2014/30/EU (EMC), 2011/65/EU, (EU)2015/863 (RoHS).
UK Regulation: 2016 No. 1091 (EMC), 2012 No.3032 (RoHS).
EU Standards: EN 61326-1:2013.
UK Standards: BS EN 61326-1:2013.
EU Standards: EN IEC63000:2018.
UK Standards: BS EN IEC63000:2018.

Intermittent Operation

When the instantaneous flow rate is set to 10-49 mL/min in Volume Mode, the pump performs intermittent operation by sampling at an instantaneous flow rate of 50mL/min.

Visual Description of Diagram: A diagram illustrates the intermittent operation cycle. It shows a timeline with segments labeled 'sampling' and 'pause'. The cycle consists of 12 seconds of sampling at 50mL/min, followed by a 48-second pause. This pattern repeats until the target volume is achieved.

Example of Intermittent Operation: If Volume Mode is selected and the flow rate is set to 10 mL/min with a target volume of 0.1L, the pump will sample for approximately 12 seconds at 50mL/min, then pause for about 48 seconds. This sampling-pause cycle repeats until the set volume of 0.1 L is collected, taking approximately 10 minutes.

Note: If there is a sudden change in gas concentration or if sampling is performed for a short period, the error in the sampling result may be significant.

Sorbent Tubes

Sorbent tubes are glass tubes filled with sorption agents, such as activated charcoal or silica-gel, used to adsorb toxic gases in various environments. To use them, both ends of the tube are broken off, connected to a sampling pump for sample collection. The adsorbed substance is then extracted using a solvent and analyzed via gas chromatography.

Options and Accessories

  • Gas sampling pump tripod stand (GSP-TRIPOD): Allows mounting of the GSP-501FT to use at the desired height.
  • Belt loop pump holder (GSP500FT-30): Designed for personal exposure measurement.
  • Rubber caps (DTP-2-20): Reusable rubber caps that can be attached to the broken tips of detector tubes as an extra precautionary safety measure. One package contains 20 caps.
  • Protective cover for detector tube (GSP300-14): Enables the detector tube to be securely set up for measuring and helps to avoid possible injury to the user as well as damage to the detector tube itself.
  • Sampling tube holder (730): A tube holder for personal exposure measurement.
  • Tripod mounting plate (PLATE 2): Allows two pumps to be mounted on the same tripod.

Detector Tubes for Automatic Air Sampling Pump

GASTEC offers a variety of detector tubes for automatic air sampling. The table below details some common detector tubes:

Detector Tube Chemical Formula Substance Concentration Range Sampling Volume (mL) Sampling Time (min) Initial Color Color Change Detection Limit Remarks
Ammonia NH3 Ammonia 25-50 ppm 100 1 Yellow Brown 25 ppm T: Temp Correction
Benzene C6H6 Benzene 10-150 ppm 100 1 Yellow Brown 10 ppm T: Temp Correction
Carbon Monoxide CO Carbon Monoxide 10-300 ppm 100 1 Yellow Brown 10 ppm T: Temp Correction
Carbon Disulfide CS2 Carbon Disulfide 5-100 ppm 100 1 Yellow Brown 5 ppm T: Temp Correction
Chloroform CHCl3 Chloroform 5-100 ppm 100 1 Yellow Brown 5 ppm T: Temp Correction
Ethyl Acrylate C3H6O2 Ethyl Acrylate 10-200 ppm 100 1 Yellow Brown 10 ppm T: Temp Correction
Ethyl Mercaptan C2H5SH Ethyl Mercaptan 10-200 ppm 100 1 Yellow Brown 10 ppm T: Temp Correction
Formaldehyde CH2O Formaldehyde 2-40 ppm 200 2 Yellow Pink 2 ppm T: Temp Correction
Hexane C6H14 Hexane 50-500 ppm 100 1 Yellow Brown 50 ppm T: Temp Correction
Hydrogen Cyanide HCN Hydrogen Cyanide 5-100 ppm 100 1 Yellow Brown 5 ppm T: Temp Correction
Hydrogen Sulfide H2S Hydrogen Sulfide 5-100 ppm 100 1 Yellow Brown 5 ppm T: Temp Correction
Methanol CH3OH Methanol 50-500 ppm 100 1 Yellow Brown 50 ppm T: Temp Correction
Methyl Acrylate C4H6O2 Methyl Acrylate 10-200 ppm 100 1 Yellow Brown 10 ppm T: Temp Correction
Methylene Chloride CH2Cl2 Methylene Chloride 50-500 ppm 100 1 Yellow Brown 50 ppm T: Temp Correction
Toluene C7H8 Toluene 50-500 ppm 100 1 Yellow Brown 50 ppm T: Temp Correction
Vinyl Chloride C2H3Cl Vinyl Chloride 50-500 ppm 100 1 Yellow Brown 50 ppm T: Temp Correction
Xylene C8H10 Xylene 50-500 ppm 100 1 Yellow Brown 50 ppm T: Temp Correction

Key for Remarks: T: Temp Correction, H: Humidity Correction, +: Twin Tubes, *: Refrigerated Storage, ※: Correction Factor.

Disclaimer: Specifications and appearance are subject to change without notice. Please read and understand the instruction manuals completely before operating. Actual products may differ slightly from the pictures shown.

Document Version: November, 2022.

Models: GSP-501FT, Automatic Air Sampling Pump, GSP-501FT Automatic Air Sampling Pump, Air Sampling Pump, Sampling Pump, Pump

PDF preview unavailable. Download the PDF instead.

4534 ext 14 en 1 Adobe PDF library 15.00

Related Documents

Preview Gastec Gas Sampling Pump GV-100S and GV-110S Instruction Manual
Comprehensive instruction manual for Gastec Gas Sampling Pumps GV-100S and GV-110S, covering safe usage, operation, maintenance, and troubleshooting for accurate gas detection.
Preview Gastec No. 92L Acetaldehyde Detector Tube: Instructions and Specifications
Comprehensive instructions and specifications for the Gastec No. 92L Acetaldehyde Detector Tube, covering safe operation, measurement procedures, temperature/humidity/pressure corrections, interferences, and disposal guidelines.
Preview Gastec 126L Chlorobenzene Detector Tube: Instructions and Specifications
Comprehensive instructions and specifications for the Gastec 126L Chlorobenzene Detector Tube, detailing safe operation, measurement procedures, temperature/humidity/pressure corrections, and interference information.
Preview Benzene Detector Tube No. 121SP - Gastec
Technical specifications and performance data for the Gastec Benzene Detector Tube No. 121SP, used for measuring benzene concentrations in the air.
Preview Gastec 30m Extension Hose No. 351A-30: Instruction Manual and Usage Guide
Detailed instruction manual for the Gastec 30m Extension Hose (No. 351A-30), covering safe operation, specifications, measurement procedures, and sampling time/correction factors for gas detection.
Preview GASTEC GAP-1 Semi-fixed Gas Alarm System | Safety Monitoring Equipment
Discover the GASTEC GAP-1 semi-fixed gas alarm system. This brochure details its features, specifications, and compatibility with GASTEC wireless gas detectors (GOC-200, CM-9A-BT, GOC-100-2-BT) for effective workplace safety and gas monitoring.
Preview GASTEC No.211H Sulphide Ion Detector Tube Instructions
Instructions for using the Gastec No.211H Sulphide Ion Detector Tube for detecting sulphide ions in waste water, including safety precautions, measurement procedures, specifications, interferences, disposal, and warranty information.
Preview Instrucciones Detalladas para el Tubo Detector Gastec 152L de Metiletilcetona
Guía completa de Gastec Corporation sobre el uso, especificaciones, procedimientos de medición, correcciones y precauciones de seguridad para el tubo detector de metiletilcetona modelo 152L.