Momcozy Wearable Breast Pump S9 Pro User Guide

(For OTC Type of Use)

1. Product Introduction

The Momcozy Wearable Breast Pump is intended to be used by lactating women to express and collect milk from their breasts. The Momcozy Wearable Breast Pump forms a vacuum around the nipple to express breast milk. The milk is collected in a milk collector.

The device includes stimulation mode and expression mode. In stimulation mode, the breast pump begins with a quick and short sucking pattern to get your milk to start flowing. In expression mode, the breast pump begins with a slow sucking pattern for milk expression, sucking more deeply and more slowly. To prevent milk from flowing into the vacuum system, the milk collection sets include a silicone diaphragm that physically separates the milk-contacting pathway from the vacuum system.

The device is intended for a single user.

2. Warning Instructions and Precautions

The warning signs and graphic symbols in the manual are intended to enable you to use the product safely and correctly and to prevent harm to you and others. Warning marks and graphic symbols are described as follows:

Symbol Description
⚠️ Warning A warning alerts the reader about a situation that, if not avoided, could result in death or serious injury. It may also describe potential serious adverse reactions and safety hazards.
Precautions Indicating a hazard alert that warns the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. It may also be used to alert against unsafe practices. This includes the special care necessary for the safe and effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse.
Contraindications Contraindications are conditions under which the device should not be used because the risk of use clearly outweighs any possible benefit. There may be persons for whom the device should not be used because of their health status.
Notes Indicates the need for attention, if not attention may lead to incorrect use of the product or property device damage.

Symbol Description

Symbol Description Symbol Description
☂️ Keep Dry ? Pile Limit 5 layers
? Applied part of type BF ◻️ Type of protection against electric shock: Class II
LOT Batch code ? Refer to instruction manual.
? Manufacturer information: The manufacturer name and address ? Manufacture date
☀️ Distribution packages shall not be exposed to sunlight SN Serial Number
IP22 The first number 2: Protected against solid foreign objects of 12.5mm diameter and greater. The second number 2: Protected against vertically falling water drops when enclosure titled up to 15°.
♻️ Disposal in accordance with Directive 2012/19/EU (WEEE)
? MR This device has not been tested for use in an MR environment and should not be used exposed to MR environments while patients are wearing the device. Keep it outside the MRI scanner room.

3. Indications for Use / Intended Use

[Indications for Use / Intended use]
The Momcozy Wearable Breast Pump (model: S9 Pro) is intended to express milk from lactating women in order to collect milk from their breasts. The device is intended for a single user.

[Intended users]
Can read and understand the user manual
The PATIENT is an intended OPERATOR

[Patient Population]
Lactating women

[Intended use environment]
Home Healthcare Environment

4. Contraindications

Do not use this Pump if you have an active implanted device that would be affected by a magnetic field (e.g., pacemaker; implantable cardioverter-defibrillator).

5. Safety Information

When using electrical products, especially if children are present, the following basic safety information should always be maintained.

5.1 Warning

  1. Do not use this device if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electrical interference, or death.
  2. This breast pump is not intended for use by persons (including children) with reduced physical, sensory or mental capabilities, or lack of experience and knowledge. Such persons can only use this breast pump if they are supervised by or have received proper instruction concerning use of the appliance by a person responsible for their safety.
  3. Danger: To reduce the risk of electrocution:
    • Always unplug electrical devices immediately after charging.
    • Do not use while bathing, showering, driving or swimming.
    • Do not reach for electrical product that has fallen into water, please unplug the power adapter from wall outlet immediately.
  4. The Momcozy Wearable Breast Pump is a personal care item and for hygienic reasons should never be shared between mothers or resold from one mother to another. Either of these actions will void the warranty.
  5. If you are a mother who is infected with Hepatitis B, Hepatitis C or Human Immunodeficiency Virus (HIV), pumping breast milk using the Momcozy Wearable Breast Pump will not reduce or remove the risk of transmitting the virus to your baby through your breast milk.
  6. Close supervision is necessary when product is used near children or invalids. Do not allow children or pets to play with the motor unit, packaging materials or accessories.
  7. Use product only for intended use as described in this manual.
  8. Do not use attachments or other milk collection components other than those recommended by manufacturer.
  9. Check the power adapter before charging, never use the damaged power adapter.

5.2 Precaution

Statement: If you feel excessive warmth from the Wearable breast pump while pumping, stop pumping immediately and remove pump.

Notes:
Avoid direct sunlight to the main unit.

6. Product Description

The product is mainly composed of main unit and milk collection set etc.

Fig.1 unit description

Description of breast pump unit

NO. Item Description
1 Flange All of these components make up the milk collection set.
2 Linker
3 Diaphragm
4 Main unit Pump motor
5 Silicone valve All of these components make up the milk collection set.
6 Milk collector (180ml)
7 USB Type-C cable Input end is USB 2.0 interface, the output end is Type-C interface.

Other Accessories Included:
Storage Bag
Cleaning Brush
Flange Size Ruler
Nursing Bra Extender
Quick Start Guide
17mm Flange Insert
19mm Flange Insert
21mm Flange Insert
User Manual
Silicone valve

Fig.2 control panel

Buttons Functions
1 LED Display Long press and hold to turn on/off; Short press to pause, then press any button to continue your pumping session.
2 Battery Level Indicator ? Mode selection
3 On / Off / Pause Decrease suction strength
4 Mode Selection Increase suction strength
5 Increase / Decrease ? Stimulation Mode: In this mode, the breast pump begins with a quick and short sucking pattern to get your milk to start flowing.
6 Charging Port Expression Mode: The breast pump begins with a slow sucking pattern for milk expression, sucking more deeply and more slowly.
Suction Level Display: The breast pump has 9 suction levels to choose from.
Battery Display

Description of Battery Indicator

? Low Battery The first battery indicator bar flashes.
In Charging When the pump is charged, the battery indicator lights up one by one, displaying an increasing battery percentage from 25%, 50%, 75%, to 100%, then switches off and repeats this process until the pump is fully charged.
Fully Charged Solid White

7. Operating Instructions

7.1 Charging battery

Please charge the product before using it for the first time. The breast pump is powered by an internal rechargeable lithium battery.

The products are packaged with a charging cable (USB2.0/Type-C line) but not with a power adapter. Please purchase the power adapter by yourself before charging. Make sure to buy the power adapter which matches the USB 2.0 interface, and please make sure the input power of the power adapter is AC 100~240V 50/60 Hz and the output power is DC 5V 1A.

Fig. 3

Charge the battery as below:
a. Insert the USB end of charging cable into the USB interface of power adapter;
b. Insert power adapter plug into AC socket;
c. Insert the Type-C end of charging cable into the charging port.

Fig. 4 Position of Charging port

The device will take approximately 150 minutes to fully charge, during charging, the battery indicator lights up one by one, displaying an increasing battery percentage from 25%, 50%, 75%, to 100%, then switches off and repeats this process until the pump is fully charged. In this way, when the battery is fully charged, the four lights are always on solid white.

Unplug the power adapter from the socket, disconnect the Charging cable from the power adapter and main unit after charging is completed.

Warning

Caution:

Using the wrong type of adapter or purchasing a poor-quality adapter may reduce the device's safety, and may damage your property and health. Please purchase the medical adapter that meets the IEC 60601-1 and IEC 60601-1-2 standards in professional medical markets, and make sure the output of the power adapter is DC 5V 1A.

Notes:

7.2 Cleaning and disinfection

7.2.1 General description/requirement

Providing breast milk is one of the best things you can do for your baby's health and development. Pumping your milk is one way to provide breast milk to your baby. Keeping the parts of your pump clean is critical, because germs can grow quickly in breast milk or breast milk residue that remains on pump parts. Following these steps can help prevent contamination and protect your baby from infection. If your baby was born prematurely or has other health concerns, your baby's health care providers may have more recommendations for pumping breast milk safely.

1) BEFORE EVERY USE

2) AFTER EVERY USE

3) Parts that need to be cleaned and disinfected

Clean and disinfect all parts that come into contact with breast and breast milk before you use the breast pump for the first time and after every use.

4) Before cleaning/disinfection :

  1. Please Wash your hands well with soap and water for 20 seconds.
  2. Please disassemble all parts as follows:

Fig. 5

Fig. 6

1). Separate pump motor from main body. 2). Detach flange from milk collector.

Fig. 7

Fig. 8

3). Separate linker from milk collector. 4). Disassemble diaphragm and valve from the linker.

Caution:

Never put the main unit, charging cable or the adapter (If you have purchased an adapter) in water or a dishwasher, as this causes permanent damage to these parts.

7.2.2 Cleaning and disinfecting parts contacting with the milk

Step 1: Cleaning

Please manually clean the device! The safety of using a dishwasher to clean and disinfect this product has not been verified.

Warning

Do not use antibacterial or abrasive cleaning agents to clean the breast pump parts as this may cause damage.

  1. Prepare the below items:
    - Soap and hot water
    - Drinking-quality water
    - Clean wash basin used only for infant feeding items
    - Clean brush used only for infant feeding items
    - Unused dish towel or paper towel
  2. Rinse: Rinse parts of milk collection set and any other feeding items by holding them under running water to remove the remaining milk or dirt. The water can be warm or cold, whatever you prefer.

Fig. 9

  1. Wash milk collection parts:
    Place parts of the milk collection set in a clean wash basin used only for infant feeding items. Do not wash directly in the sink because it may contain germs that could contaminate these items.
    Soak parts of the milk collection set in warm 40-60 °C (104-140 °F), soapy water for 3~5 minutes.

Fig. 10

Using a clean brush used only for infant feeding items.

  1. Rinse again: Rinse by holding items under running water, or by submerging them in fresh water in a separate basin that is used only for cleaning infant feeding items. If there is still dirt, repeat step 3 and re-rinse until there is no visible dirt.
  2. Air-dry thoroughly: Place parts of the milk collection set, wash basin, and bottle brush on a clean, unused dish towel or paper towel in an area protected from dirt and dust. Do not use a dish towel to rub or pat items dry because doing so may transfer germs to the items!

Fig. 11

  1. Clean wash basin and bottle brush: Rinse the wash basins and brush well and allow them to air-dry after each use. Wash them every few days by hand with soap and warm water. If your baby is less than 3 months old, was born prematurely, or has a weakened immune system due to illness (such as HIV) or medical treatment (such as chemotherapy for cancer), wash basin and bottle brush after every use.

After Cleaning:
For extra germ removal, disinfect the parts of milk collection set, wash basin, and bottle brush at least once daily after they have been cleaned. Items can be disinfected using boiling water with a disinfection setting. Disinfection is especially important if your baby is less than 3 months old, was born prematurely, or has a weakened immune system due to illness or medical treatment.

For the disinfection of milk contact parts, please refer to step 2.

Step 2: Disinfection

Caution:

Do not sterilize pump parts in a microwave or UV sterilizer. High temperature(> 110 °C) will shorten their service life and cause safety hazards.

Before disinfection, make sure you have a cleaned milk collection set, bottle brushes, and wash basins using the methods above. Disinfect all items in contact with milk (even the bottle brush and wash basin!) by using one of the following methods.

  1. Prepare the below items:
    A household pot
    Drinking-quality water
    Clean tongs

Caution:

During disinfection with boiling water, prevent the bottle or other parts from touching the sides of the pot. This can cause irreversible product deformation or damage.

  1. Place the disassembled milk collection set in the pot and fill the pot with enough water to cover the components to disinfect. Let the water boil for 5-10 minutes, making sure the parts do not touch the sides of the pot.

Fig. 12

  1. Remove items with clean tongs.

Fig. 13

  1. Air-dry thoroughly.

Fig. 14

Step 3: Store safely

Store dry items safely until needed. Ensure the clean pump parts, bottle brushes, and wash basins have air-dried thoroughly before storing. Items must be completely dry to help prevent germs and mold from growing. Store dry items in a clean, protected area to prevent contamination.

  1. Wash hands: Wash hands well with soap and water.
  2. Put back together: Put together the clean, dry pump parts.
  3. Store safely: Place reassembled pump parts and other feeding items, wash basin, and dry bottle brush in a clean, protected area such as inside a closed kitchen cabinet that is used only to store clean dishes.

7.2.3 Clean / Disinfect the other parts

The shell of the main unit need to be cleaned and disinfected for the first time and after each use. The other parts, do not come into contact with the human body nor the milk during pumping, and only need to be cleaned when it is dirty.

Cleaning/disinfecting the items at room temperature 5 °C~ 40 °C as below:

Prepare the below items:
- Disinfection
- 70% isopropyl alcohol wipes
- Clean soft cloth

Cleaning method:

  1. Before performing this cleaning procedure, wash your with disinfection first.
  2. Remove all residues from the main unit and Charging cable using a clean soft cloth.
  3. Take a piece of 70% isopropyl alcohol wipes to swab on the parts until no visible contaminants remain.
  4. Dry the device by placing it at room temperature for 10min to evaporate the residual cleaning reagent solution on the product.

Disinfection method:

Cleaning the shell of the main unit as the above method before disinfection.

  1. Disinfect the shell of the main unit with a soft cloth dampened with the 70% isopropanol for 3 minutes.
  2. Dry it by placing the main unit at room temperature for 10min to evaporate the residual disinfection solution.

7.3 Direction for use

7.3.1 Preparation before use

Caution:

  1. Wash hands with soap and water, and inspect and assemble a clean pump kit.
  2. Duration of pumping session:
    This product is powered by a rechargeable lithium battery, after being fully charged, it can be used for 6-7 sessions (about 180 minutes of pumping), 30 minutes per session.
    A typical pumping session lasts about 10-15 minutes per breast, but you should only pump as long as it is comfortable and productive for you.

1. Select Flange Size

This Momcozy Wearable Breast Pump only provides one 24mm silicone flange. If the flange provided does not fit, you can find various flanges and flange inserts available for purchase on Amazon. Type "momcozy" in the search bar and find the flange size and flange insert that fits you.

Fig.15 select right size

2. Stimulate the Nipple

  1. Stimulate your nipple to get a full size (do not skip this step!).
  2. Place the nipple into the smallest hole you can comfortably fit into without forcing it.
  3. Select the hole that can comfortably fit the very base of your nipple.
  4. Get your measurement based on the sizing chart.

3. Test the Flange Fit

A. To find the suitable size flange, first, you need to make sure you place the flange in the right place on your breast.

B. Please press the flange on your breast until there is no gap. Make sure your nipple is right in the middle of the flange.

Fig.16

Fig.17

Too Small
Too small: During pumping, if your nipple continues to rub against the inner wall of the flange, then this flange is too small for you.

Good Fit
Suitable fitting: During pumping, if your nipple moves smoothly without rubbing or touching the inner wall of the flange, and almost no areola is sucked inside, then this flange is right for you.

Too Large
Too large: During pumping, if a large area of areola is sucked inside, then this flange is too big for you.

4. Get the Right Bra

Fig.18

1 Made of stretchy fabric 2 No underwire or padding
3 Has a flap 4 Full coverage

Fig.19

NO GAP
When you put the pump inside your bra, the silicone flange of this pump should be firmly pressed against your skin without a gap.

Fig.20

IF IT MOVES
If you feel the pump moves during use, please tighten your bra strap to adjust the pump again, or choose another bra.

5. Recommend Posture

Caution:

Always sit or stand upright while pumping to prevent the milk from leaking. This pump is not to be used when lying down or sleeping to ensure the proper orientation is maintained during use.

POSTURE

Make sure the pump and the milk collector always stay at a vertical angle. Otherwise, the milk will spill out.

Fig.21 right posture

7.3.2 Assembling the Pump

Caution:

Make sure you have cleaned and disinfected the parts that come into contact with milk or breast.

Warning

Notes:

Wash your hands thoroughly with soap and water, then assemble the pump as follows:

Fig. 22

1. Wash your hands thoroughly with soap and water.

Fig. 23

2. Put the silicone diaphragm into the linker and apply mild pressure around the edge until the silicone diaphragm seals well with the linker.

Fig. 24

3. Connect the silicone valve with the end of the linker, and apply mild pressure to push the silicone valve to fit onto the linker securely.

Fig. 25

4. Put the assembled linker part into the milk collector, and align the "ear" notch of the linker with the inner groove of the milk collector. Press the silicone diaphragm onto the round hole of the milk collector, to make them fit well.

Fig. 26

5. Connect the flange with the main body. Make sure that the flange fits securely around the rim by pressing it down with your thumbs.

The breast pump is now ready for use.

Fig. 27

6. Insert the cylindrical protrusion located on the milk collector into the hole on the pump motor, and then align the groove on the milk collector with the notch on the pump motor.

7.3.3 Pumping instructions

Warning

Caution:

1. Alignment

a. Before pumping, please make sure the pump is pressed tightly on your breast without a gap (Fig.28).

Fig. 28

b. To align, hold the flange against your breast so you can see if your nipple is centered in the flange tunnel. You may need to bend forward.

c. After you've aligned, look through the flange to confirm your nipple is centered before pumping. The tip of your nipple should NOT touch the top, sides or bottom of the flange tunnel. If it does, try again.

First Perspective

Fig. 29

Notes:

Correct nipple alignment is key to comfort expression and performance. Nipples must be positioned correctly in flange tunnel before pumping to avoid discomfort.
See Section 7.3.1 Preparation before use-Select Flange Size.

2. Secure Pump In Bra

Secure Pump in place by bra, taking care to keep pump pressed firmly against breast. If necessary, tighten the bra to keep pump securely in place and start pumping.

Fig. 30

3. Start Pumping

Do not slide the pump directly into your bra, follow step-by-step instructions.
Disconnect from Charger before use.

A. Turn on the pump

Long press the On / Off / Pause button for 2s to turn on the pump.

B. Select the mode

Modes: Stimulation Mode & Expression Mode
When you first use the pump, the default setting is level 1 suction strength of Stimulation mode.
Memory function: The pump will remember and start with the previous suction strength of the stimulation mode when you use it next time.
Short press the Mode Selection button to switch between 2 modes.

C. Select the vacuum level

Momcozy pump has 9 vacuum pressure settings for each mode, giving you control over what feels comfortable and works most efficiently in all modes.

Once you can feel the vacuum pressure on your nipples, use the Increase/Decrease button to adjust the vacuum pressure to find your preferred setting in both modes. The LED display will be off after no operation for 20 seconds, press any button can activate the LED display again.

Tip: A higher vacuum pressure does not always mean a higher volume of milk. Putting your nipple under excess stress can reduce milk production. Always ensure you are pumping at a comfortable level. If you don't feel suction, please check the pump is assembled correctly.

Caution:

4. Pause / Start Pumping

To pause while pumping, short press the On / Off / Pause button and the mode indicator will be blinking. Press any button to continue your pumping session.

5. Turn off the Pump

Long press the On / Off / Pause button for 2 seconds to turn off the pump.
In addition, the product will automatically shut down after the 30min of pumping time.
Remark: Pause/start pumping, mode and vacuum level will not reset the pumping time.

6. Remove the Pump

Use your finger to break the seal between the flange and the breast tissue.
Lean slightly forward to remove the pump from the breast after your pumping session, then keep the flange opening side up to prevent milk from pouring over.

Fig. 31

We recommend the user stop the pumping session once the expressed milk volume of the milk collector approaches 120 ml to avoid overflow.

After finishing the milk collection, please pour the newly expressed milk into other milk collectors with caps or milk storage bags in time. Screw the cap tightly onto other milk collectors and store the milk.

Warning

Warning

7.4 Storing breast milk

7.4.1 Proper storage of breast milk

Following recommended storage and preparation techniques can maintain the safety and quality of expressed breast milk for the baby's health.

These are general guidelines for storing human milk at different temperatures. Various factors affect how long human milk can be stored safely. Such factors include milk volume, room temperature when milk is expressed, temperature fluctuations in the refrigerator and freezer, and cleanliness of the environment.

Storing breast milk after expressing:

Type of Breast Milk Countertop (77 °F (25°C) or colder (room temperature)) Refrigerator (40 °F (4°C)) Freezer (23 °F (-5°C) or colder)
Freshly Expressed or Pumped Within 2-4 Hours Within 24 Hours Within 3 months
Thawed, Previously Frozen Use within 2 hours after the baby is finished feeding

These guidelines are for healthy full-term babies only and may vary for premature or sick babies. Check with your health care provider.

Storage tips:

7.4.2 Safe thawing of breast milk

8. Storage Condition

How to store the pump

9. Specification

Model No. S9 Pro
Name Momcozy wearable breast pump
Power requirements Input: 100-240Vac, 50/60Hz;
Output: 5V 1A (Charge through provided USB 2.0/Type-C Charging cable and self-purchased power adapter)
Power supply DC 3.7 V / 1600mAh rechargeable lithium battery
Rated power 5W
Pump options Single pump
Suction modes Stimulation mode and expression mode
Vacuum range Stimulation mode: -85mmHg to -145mmHg
Expression mode: -165mmHg to -285mmHg
Cycle speed Stimulation mode: 73~127(±2) cycle/min
Expression mode: 18~56(±2) cycle/min
Pump dimension 140mm X 112mm × 73mm
Weight(with all accessories) 250g
Product use life 500 hours
Noise level ≤50dBA
Type of protection against electric shock Internally powered equipment
Degree of protection against electric shock Type BF applied part (milk collection set)
Operating conditions Temperature: 0°C to 40 °C; Humidity: 15% to 90%,
Atmospheric pressure: 70 kPa to 106 kPa
Transportation & Storage environment Temperature: -20 °C to 60 °C; Humidity: 15% to 90%,
Atmospheric pressure: 70 kPa to 106 kPa
Battery charging time 150 minutes
Battery usage Time 180 minutes

10. Maintenance and Replacement Parts

The device contains no user serviceable parts inside: Opening or tampering with this device will void the warranty. In the event the device requires repair, it should be returned to the medical equipment company or to retailer directly. Modification of any kind is prohibited.

The MANUFACTURER will make available on request circuit diagrams, component part lists, descriptions, calibration instructions, or other information that will assist SERVICE PERSONNEL to repair those parts of the device that are designated by the MANUFACTURER as repairable by the SERVICE PERSONNEL. When the product is not used for a long time, the battery will discharge slowly. In order to avoid battery damage due to low voltage for a long time, please charge the device every three months.

Warning

The replacement of lithium batteries by inadequately trained personnel could result in a hazard.

Do not clean or maintain the device while the device is in use or while charging.

11. Declaration of Conformity

Shenzhen Lutejiacheng Technology Co., Ltd. declares that the device conforms to the following standards IEC60601-1, IEC60601-1-2, IEC60601-1-11, IEC62304, ISO10993-5, ISO10993–10, ISO10993–1, ISO 14971.

12. Troubleshooting

Problems Solutions
Pump does not work If the let-down has not yet occurred, consider leaving the pump on breast for a longer period of time to observe for possible delayed let-down. Note: Initial milk let-down time is unique for each person and varies by many factors.
Check nipple alignment, pause the pump, and remove it from the breast. Realign the pump as described below.
You may not have the correct Flange size (Refer to Section 7.3 Direction for Use - Select Flange Size).
Pause the pump and remove it from breast. Realign it correctly. (Refer to Section 7.3 Direction for Use – Alignment).
You may also separate the pump from the flange to realign. Then, center your nipple in the flange tunnel and bring the pump to your breast (Refer to Section 7.3 Direction for Use – Alignment).
Milk does not flow Pause the pump and remove it from breast. Realign it correctly. (Refer to Section 7.3 Direction for Use – Alignment).
You may also separate the Pump from the Flange to realign. Then, center your nipple in the flange tunnel and bring the pump to your breast (Refer to Section 7.3 Direction for Use – Alignment).
If you're experiencing excessive discomfort, please try the following (Refer to Section 7.3 Direction for Use – Alignment):
Make sure to press the Momcozy pump firmly against breast during Stimulation phase.
Tighten the bra to make sure the Momcozy pump is held firmly against the breast.
Decrease suction strength.
Realign the Momcozy pump (see above).
You may be pumping for too long. Momcozy pump shuts off automatically after 30 minutes.
You may not have the correct flange size(Refer to Section 7.3 Direction for Use - Select Flange Size). Stop and see a medical professional or breastfeeding specialist.
Decreased (low) pump suction The Pump has 9 suction settings. Press the increase button to increase suction strength.
Check the condition of your washable parts. They should be replaced after three months of use.
If that does not work, try the following:
1. Check connections between all washable parts to ensure all are secure.
2. Visually inspect all washable parts and replace them if damaged.
3. Check that power is on.
4. Press the Pump securely against the breast.
Stop pumping Pump stops pumping automatically after 30 minutes.
Short press On / Off / Pause button will pause the pump, press again will resume.
Pump does not stop pumping Press the On / Off / Pause button.
If the issue persists, then break the seal by inserting a finger between the breast tissue and the flange. Then Press the On / Off / Pause button to turn the pump off.
Pump or Charger gets wet (immersed in water). Dry the pump off immediately. Prop Pump upright with the charger port down and suction controls up, and let it air dry overnight. Do not use the pump or charger within 24 hours.
Contact Customer Care.
Pump is not charging. Ensure the USB Type-C cable with standard Charger (not included) is fully plugged into the pump charging port.

Remark: If the above methods still do not solve the problem, please contact the manufacturer, see the last page for contact information.

13. Disposal

At the end of the product life cycle, do not throw this product into the normal household garbage, but bring it to a collection point for the recycling of electronic equipment and battery. Waste Electrical and Electronic Equipment may have potentially harmful effects on the environment. Improper disposal may lead to the accumulation of harmful toxins in the air, water, and soil, which is harmful to human health. you have obligation to dispose of the device correctly. Consult your municipal authority for information about disposal.

14. Warranty

Please contact the manufacturer in case of a claim under the warranty, the contact information please refer to last page of this user manual. If you have to send the unit, enclose a copy of your receipt with clear statement of defect description.

The warranty terms are as below:

  1. The warranty period for the device is two years from the date of purchase. In case of a warranty claim, the date of purchase has to be proven by means of the sales receipt or invoice.
  2. Repairs under warranty do not extend the warranty period either for the device or for the replacement parts.
  3. The following cases are excluded under the warranty
    • All damages which are arisen due to improper treatment, e.g. nonobservance of the user instruction.
    • All damages which are arisen due to repairs or tampering by the customer or unauthorized third parties.
    • Damages which are arisen during transport from the manufacturer to the consumer or during transport to the service center.
    • Accessories which are subject to normal wear and tear.
    • Device damage due to privately disassemble devices.
    • Liability for direct or indirect consequential losses caused by the unit is excluded even if the damage to the unit is accepted as a warranty claim.

15. EMC Information

The equipment with following ESSENTIAL PERFORMANCE is intended used in Home healthcare environment facility environment.

Essential Performance: Device- Specific Function Device-Specific Pass/Fail Criteria Detection/ Testing Method
Vacuum strength (error: ±5mmHg) Suction levels (stimulation): -85mmHg to -145mmHg
Suction levels (Expression): –165mmHg to -285mmHg
Vacuum manometer Observation
Cycle speed (cycles/min) (error: ±2cycles/min) Suction levels (stimulation): 73–127
Suction levels (Expression): 18–56
Observation
Maximum vacuum -290mmHg Vacuum manometer Observation
Software for suction pressure regulation -290mmHg Vacuum manometer Observation
Backflow prevention mechanism No leakage at maximum vacuum Observation

If Essential Performance is lost or degraded due to electromagnetic disturbances, this may result in loss of product function or injury to patients, please read below important information before to avoid possible electromagnetic disturbances.

Warning

Guidance and manufacturer's declaration – electromagnetic emission

The Momcozy Wearable Breast Pump is intended for use in the electromagnetic environment specified below. The customer of the user of the Momcozy Wearable Breast Pump should assure that it is used in such an environment.

Emission test Compliance Electromagnetic environment – guidance
Conducted and Radiated RF emissions CISPR 11 Group 1 Class B The Momcozy Wearable Breast Pump uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Conducted RF emissions CISPR 11 Group 1 Class B The Momcozy Wearable Breast Pump is suitable for use in all establishments, including domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes except for near active HF surgical equipment and the RF shielded room for magnetic resonance imaging.
Conducted RF emissions CISPR 11 Group 1 Class B

Guidance and Manufacturer's Declaration – Electromagnetic Immunity

The Momcozy Wearable Breast Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Momcozy Wearable Breast Pump should assure that it is used in such an environment.

Immunity Test IEC 60601-1-2 Test level Compliance level Electromagnetic environment- guidance
Electrostatic discharge IEC 61000-4-2 ±8kV contact;
±2kV, ±4kV, ±8kV, ±15 kV air
±8kV contact;
±2kV, ±4kV, ±8kV, ±15kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%
Immunity Test IEC 60601-1-2 Test level Compliance level Electromagnetic environment- guidance
Radiated RF EM fields IEC 61000-4-3 10V/m (Home healthcare environment), 80MHz-2.7GHz 80% AM at 1kHz 10V/m (Home healthcare environment) 80MHz-2.7GHz 80% AM at 1kHz power quality should be that of a HOME HEALTHCARE ENVIRONMENT.
Electrical fast transients/bursts IEC 61000-4-4 ±2kV AC power supply lines;
±1kV DC power/Signal lines.
100 kHz repetition frequency
±2kV AC power supply lines; power quality should be that of a HOME HEALTHCARE ENVIRONMENT.
Surges IEC 61000-4-5 ±0.5kV, ±1kV lines to lines;
±1kV, ±2kV lines to earth
±0.5kV, ±1kV lines to lines; power quality should be that of a HOME HEALTHCARE ENVIRONMENT.
Conducted disturbances induced by RF fields IEC 61000-4-6 0.15MHz - 80MHz, 6V in ISM bands between 0.15MHz and 80MHz 80% AM at 1kHz 3V
0.15MHz - 80MHz, 6V in ISM and amateur radio bands between 0.15MHz and 80MHz 80% AM at 1kHz
power quality should be that of a HOME HEALTHCARE ENVIRONMENT.
Note: The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
Rated power frequency magnetic fields IEC 61000-4-8 30A/m 50Hz or 60Hz 30A/m 50Hz Power frequency magnetic fields should be at levels characteristic of a HOME HEALTHCARE ENVIRONMENT.
Voltage dips IEC 61000-4-11 0% UT, 0.5 cycle
45°, 90°, 135°, 180°, 225°, 270°and 315°;
0% UT, 1 cycle
and
70% UT, 25/30 cycle
Single phase: at 0°
Applicable power quality should be that of a HOME HEALTHCARE ENVIRONMENT.
Voltage interruptions IEC 61000-4-11 0% UT, 250/300 cycle Applicable
NOTE: UT is the a.c. mains voltage prior to application of the test level.
E.g.: 25/30 means 25 periods at 50 Hz or 30 periods at 60 Hz.

Guidance and manufacture's declaration – electromagnetic immunity

The Momcozy Wearable Breast Pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Momcozy Wearable Breast Pump should assure that it is used in such an environment.

Immunity test IEC 60601-1-2 test level Compliance level Electromagnetic environment - guidance
Proximity fields from RF wireless communications equipment IEC 61000-4-3 See the following table Complies
Proximity magnetic fields IEC 61000-4-39 See the following table

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment

Test frequency (MHz) Band a) (MHz) Service a) Modulation Immunity Test Level (V/m)
385 380-390 TETRA 400 Pulse modulation b) 18 Hz 27
450 430-470 GMRS 460, FRS 460 FM ± 5 kHz deviation 1 kHz sine 28
710
745 704-787 LTE Band 13, 17 Pulse modulation b) 217 Hz 9
780
810
870 800-960 GSM 800/900, TETRA 800, IDEN 820, CDMA 850, LTE Band 5 Pulse modulation b) 18 Hz 28
930
1720
1845 1700-1990 GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 2, 3, 4, 25; UMTS Pulse modulation b) 217 Hz 28
1970
2450 2400-2570 Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE Band 7 Pulse modulation b) 217 Hz 28
5240
5500 5100-5800 WLAN 802.11 a/n Pulse modulation b) 217 Hz 9
5785

NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.

a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, the carrier may be pulse modulated using a 50% duty cycle squire wave signal at 18 Hz, While it does not represent actual modulation, it would be worst case.

Test specifications for enclosure port immunity to proximity magnetic files

Test frequency Modulation Immunity test level (A/M)
30 kHz a) CW 8
134.2 kHz Pulse modulation b) 2.1 kHz 65 )
13.56 MHz Pulse modulation b) 50 kHz 7.5)

a) This test is applicable only to ME EQUIPMENT and ME SYSTEMS intended for use in the home healthcare environment.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c)r.m.s, before modulation is applied

16. Travel or International Use Statement

Since the performance of the Momcozy Wearable Breast Pump may be affected by the external environment, in view of the uncertainty and instability of the travel environment, please do not use the device during travel or on the aircraft.

The Momcozy Wearable Breast Pump can be internationally used, but it must be used and stored in the environment specified in this user manual, and please make sure the input power of your power adapter is AC 100~240V 50/60 Hz and output power is DC 5V 1A, and please make sure you have a converter to convert to the proper voltage of the target country. To ensure that the Momcozy Wearable Breast Pump is not affected during carrying, please check the following items before use to ensure it can operate normally:

Check the Momcozy Wearable Breast Pump to ensure that it is free from damage and
Before each usage, check the status of your Momcozy Wearable Breast Pump to ensure it can operate normally.
If there is any abnormality, please stop using it.

17. Manufacturer Information

Manufactured for Hong Kong Lute Technology Co., Limited
Room 02, 21/F, Hip Kwan Commercial Building, 38 Pitt Street, Yau Ma Tei, Kowloon, Hongkong

E-mail: support@momcozy.com

If assistance in setting up, using, or maintaining the device when needed or to report unexpected operation or events, please contact us.

Models: B0C2ZMLSR5 Hands Free Breast Pump S9 Pro Updated, B0C2ZMLSR5, Hands Free Breast Pump S9 Pro Updated, Free Breast Pump S9 Pro Updated, Breast Pump S9 Pro Updated, Pump S9 Pro Updated, S9 Pro Updated, Pro Updated

PDF preview unavailable. Download the PDF instead.

D1I7yhtpxwL Adobe PDF library 16.03 Adobe Illustrator 26.0 (Windows)

Related Documents

Preview Momcozy S9 Pro Bomba Tira-Leite Portátil Guia de Utilização
Comprehensive user guide for the Momcozy S9 Pro portable breast pump, detailing product features, safety warnings, cleaning instructions, milk expression techniques, storage guidelines, troubleshooting, and electromagnetic compatibility information. Designed for lactating women, this device offers stimulation and expression modes for efficient breast milk collection.
Preview Momcozy S12 Pro Portable Breast Pump User Manual
Comprehensive user manual for the Momcozy S12 Pro portable breast pump. This guide provides detailed instructions on setup, operation, cleaning, maintenance, troubleshooting, and safety precautions for optimal use.
Preview Momcozy S12 Pro Wearable Breast Pump User Manual & Guide
Comprehensive user manual for the Momcozy S12 Pro wearable breast pump, covering setup, usage, cleaning, maintenance, troubleshooting, and storage.
Preview Momcozy Wearable Breast Pump: Installation, Usage, and Care Guide
Comprehensive guide for the Momcozy wearable breast pump, covering installation, tips for effective pumping, cleaning, sterilization, and maintenance. Learn how to use your Momcozy pump for optimal comfort and milk collection.
Preview Momcozy S12pro Wearable Breast Pump: Assembly, Usage, and Care Guide
Comprehensive guide for the Momcozy S12pro wearable breast pump, covering assembly instructions, tips for milk output, cleaning, sterilization, and general usage advice. Ensure optimal performance and longevity of your breast pump.
Preview Momcozy Wearable Electric Breast Pump 510(k) Notification - FDA Review
FDA review of the Momcozy Wearable Electric Breast Pump (Models S9 Pro, S12 Pro) 510(k) premarket notification, confirming substantial equivalence to legally marketed predicate devices.
Preview Momcozy Wearable Breast Pump: Assembly, Usage, and Care Guide
Comprehensive guide for the Momcozy Wearable Breast Pump, covering assembly, proper usage to avoid leaks, washing and sterilization instructions, and helpful tips for optimal performance. Includes troubleshooting information.
Preview Momcozy Wearable Breast Pump M5 User Manual
This user manual provides detailed instructions for the Momcozy Wearable Breast Pump M5, covering product introduction, safety precautions, indications for use, contraindications, operating instructions, cleaning and disinfection, troubleshooting, and warranty information.