DESCRIPTION
The PMP250 series of AC/DC switching power supplies offers 220-250 watts of continuous output power. They are enclosed in a 94V-0 rated polycarbonate case with an IEC320/C14 (Class I) or IEC320/C18 (Class II) inlet for worldwide use. All models meet EN 55011 class B emission limits and are designed for medical applications.
FEATURES
- BF class insulation
- Operating altitude up to 5000 meters
- Efficiency level VI / CoC EPS V5 tier 2 requirements
- No load power consumption less than 0.15W
- With PFC circuit
- Wide input range 80 to 264 VAC
- 100% burn-in
- Overvoltage protection
- Overcurrent protection
- Compliant with RoHS requirements
SAFETY STANDARD APPROVALS
Safety standard approvals include UL ES 60601-1, CSA C22.2 No. 60601-1 (File No. E178020), and TÜV EN 60601-1.
INPUT SPECIFICATIONS
- Input voltage: 80-264 VAC
- Power derating: Derate linearly from 100% at 90 VAC to 90% at 85VAC and 80% at 80 VAC
- Input frequency: 47-63 Hz
- Input current: 2.5 A (rms) for 115 VAC, 1.25 A (rms) for 230 VAC
- Earth leakage current: 220 μA max. @ 264 VAC, 63 Hz
- Touch current: 100 μA max. @ 264 VAC, 63 Hz
OUTPUT SPECIFICATIONS
- Output voltage /current: See rating chart.
- Maximum output power: See rating chart.
- Ripple and noise: 1% peak to peak maximum
- Over voltage protection: Provided and set at 112-140% of its nominal output voltage, latching by recycle input to reset
- Short circuit protection: Automatic recovery
- Over temperature protection: Latching by recycle input to reset
- Temperature coefficient: ±0.04% /°C maximum
- Transient response: Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 µs after a 25% step load change
ENVIRONMENTAL SPECIFICATIONS
- Operating temperature: -20°C to +60°C
- Atmospheric pressure: 540 hPa to 1060 hPa
- Storage temperature: -40°C to +85°C
- Relative humidity: 5% to 95% non-condensing
- Temperature derating: Derate from 100% at +40°C linearly to 50% at +60°C
GENERAL SPECIFICATIONS
- Switching frequency: 50-130 KHz
- Power factor: 0.98 Typical at 115 VAC
- Efficiency: 89% min. at full load
- Hold-up time: 20 ms minimum at 100 VAC
- Line regulation: ±0.5% maximum at full load
- Inrush current: 130 A @ 115 VAC or 260 A @ 230 VAC, at 25°C cold start
- Withstand voltage: 4000 VAC from input to output (2 MOPP), 1500 VAC from input to ground (1 MOPP), 500 VAC from output to ground. (For class II models, 4000VAC from input to output)
MTBF & EMC PERFORMANCE
- MTBF: 100,000 hours at full load at 25°C ambient, calculated per MIL-HDBK-217F
- EMC Performance:
- EN55011: Class B conducted, Class B radiated
- EN61000-3-2: Harmonic distortion, Class A and D
- EN61000-3-3: Line flicker
- EN60601-1-2 Compliance:
- EN61000-4-2: ESD, ±15 KV air and ±8 KV contact
- EN61000-4-3: Radiated immunity, 9-28 V/m
- EN61000-4-4: Fast transient/burst, ±2 KV
- EN61000-4-5: Surge, ±1 KV diff., ±2 KV com.
- EN61000-4-6: Conducted immunity, 10 Vrms
- EN61000-4-8: Magnetic field immunity, 30 A/m
- EN61000-4-11: Voltage dip immunity, 30% reduction for 500 ms (criteria A @ 230VAC, criteria B @ 100VAC), 60% reduction for 100 ms (criteria A @ 230VAC, criteria B @ 100VAC) and >95% reduction for 20 ms
OUTPUT VOLTAGE/CURRENT RATING CHART
Model (Class I) | Model (Class II) | V1 | Min. Current | Max. Current | Tol. | Ripple & Noise | Max. Power | Average Active Efficiency (@115/230 Vac) |
---|---|---|---|---|---|---|---|---|
PMP250-12 | PMP250F-12 | 12 V | 0 A | 18.34 A | ±5% | 120 mV | 220 W | 89/89% |
PMP250-13 | PMP250F-13 | 15 V | 0 A | 14.67 A | ±5% | 150 mV | 220 W | 89/89% |
PMP250-13-1 | PMP250F-13-1 | 18 V | 0 A | 13.89 A | ±5% | 180 mV | 250 W | 89/89% |
PMP250-13-2 | PMP250F-13-2 | 19 V | 0 A | 13.16 A | ±5% | 190 mV | 250 W | 89/89% |
PMP250-13-3 | PMP250F-13-3 | 20 V | 0 A | 12.50 A | ±5% | 200 mV | 250 W | 89/89% |
PMP250-14 | PMP250F-14 | 24 V | 0 A | 10.42 A | ±5% | 240 mV | 250 W | 90/90% |
PMP250-15 | PMP250F-15 | 27 V | 0 A | 9.26 A | ±5% | 270 mV | 250 W | 90/90% |
PMP250-16 | PMP250F-16 | 30 V | 0 A | 8.34 A | ±5% | 300 mV | 250 W | 90/90% |
PMP250-16-1 | PMP250F-16-1 | 32 V | 0 A | 7.82 A | ±5% | 320 mV | 250 W | 90/90% |
PMP250-17 | PMP250F-17 | 36 V | 0 A | 6.95 A | ±5% | 360 mV | 250 W | 90/90% |
PMP250-18 | PMP250F-18 | 48 V | 0 A | 5.21 A | ±5% | 480 mV | 250 W | 90/90% |
PMP250-19 | PMP250F-19 | 54 V | 0 A | 4.63 A | ±5% | 540 mV | 250 W | 90/90% |
MECHANICAL SPECIFICATIONS
The power supply unit is enclosed in a 94V-0 rated polycarbonate case. It features an LED indicator. The AC input is via an IEC320/C14 inlet (Class I) or IEC320/C18 inlet (Class II). The output connector is a Molex Mini-Fit receptacle (P/N: 39-01-2060 or 39-01-2080) with female terminal #5556 or equivalent, mating with Molex plug 39-01-2066 (or P/N: 39-01-2086) and male terminal #5558 or equivalent. It also mates with Molex headers #5566, #5569, or equivalent.
Dimensions are approximately 204.0 mm (8.03 inches) in length, 91.4 mm (3.60 inches) in width, and 36.0 mm (1.42 inches) in height, with a tolerance of 0.5 mm [0.02 inches]. The approximate weight is 1100 grams (2.425 lbs).
The 'Output Power Derating Curve' shows that output power is 100% at ambient temperatures up to 40°C and derates linearly to 50% at 60°C.
PIN CHARTS
PIN CHART (output 18Vdc to 54Vdc)
This chart details the pin assignments for output voltages from 18Vdc to 54Vdc. Pins are designated for +V1 (positive voltage), V1 (negative voltage), and Return.
PIN CHART (output 12Vdc to 54Vdc)
This chart details pin assignments for output voltages from 12Vdc to 54Vdc, similar to the 18Vdc chart, with variations in pin usage for positive voltage (+V1), negative voltage (V1), and return (Return).
CONTACT INFORMATION
For inquiries, contact FSP Power Solution GmbH:
- Email: info@fsp-ps.de
- Website: www.fsp-ps.de / www.fsp-group.com.ru
- Address: Jakobshöhe 16 / 41066 Mönchengladbach / Germany
- Phone: +49 (0) 2161 4952490
- Fax: +49 (0) 2161 495249 21