400 Watt Medical & ITE Power Supplies
PM400 SERIES
Description
The FSP PM400 series offers AC-DC switching power supplies in a 4 x 7 x 1.58 inch package, delivering 400W continuous power with 7 CFM forced air cooling or 300W with convection cooling. Built on a printed circuit board with U-bracket for support and heat sinking, these units are designed for medical applications and certified to IEC/EN/UL 62368-1, suitable for data networking, industrial, and telecommunication sectors.
Features
- BF Class insulation
- Operation altitude up to 5000 meters
- 100-240 VAC input with active PFC
- Less than 300 μA leakage current
- 5VDC at 100mA standby output
- EN55011 / EN55032 Class B conducted emissions
- Inhibit: TTL low to disable output
- Standard PS Off and DC OK signals
- High Efficiency 92% typical
- Compliant with RoHS requirements
Input Specifications
Input voltage: | 90-264 VAC |
Input frequency: | 47-63 Hz |
Input current: | 4.2 A (rms) @115 VAC, 60 Hz 2.1 A (rms) @ 230 VAC, 50 Hz |
Earth leakage current: | 300 μA max. @ 264 VAC, 63 Hz |
Touch current: | 100 μA max. @ 264 VAC, 63 Hz |
Output Specifications
- Output voltage/current: See rating chart.
- Maximum output power: See rating chart.
- Ripple and noise: 1% peak to peak maximum
- Remote sense: Compensation for cable losses up to 0.5 V
- Over voltage protection: Set at 115-140% of nominal output voltage, automatic recovery
- Short circuit protection: Automatic recovery
- Over temperature protection: Latching by recycle input to reset
- Temperature coefficient: All outputs ±0.04% /°C maximum
- Transient response: Maximum excursion of 4%, recovering to 1% of final value within 500 μs after a 25% step load change
- Standby power: 5 V at 100 mA maximum
- Fan power: 12 V at 250 mA maximum
Environmental Specifications
Operating temperature: | -10°C to +70°C |
Storage temperature: | -40°C to +85°C |
Relative humidity: | 5% to 95% non-condensing |
Temperature derating: | Derate from 100% at +50°C linearly to 50% at +70°C, applicable to convection and forced-air cooling conditions. |
Safety Standard Approvals
- UL ES 60601-1, CSA C22.2 No. 60601-1 (File No. E178020)
- TÜV EN 60601-1 (Bauart geprüft - Type tested, Sicherheit - Safety, Regelmäßige Produktions-überwachung - Regular production monitoring)
- UL 62368-1, CSA C22.2 No. 62368-1
- TÜV EN 62368-1
General Specifications
Switching frequency: | 85 KHz (typical) |
Efficiency: | Typical 89% @ 115 VAC, 92% @ 230 VAC |
Hold-up time: | 12 ms minimum at 110 VAC & 400 W |
Line regulation: | ±0.5% maximum at full load |
Inrush current: | 20 A @ 115 VAC, or 40 A @ 230 VAC, at 25°C cold start |
Withstand voltage: | 4000 VAC from input to output (2 MOPP) 1500 VAC from input to ground (1 MOPP) 1500 VAC from output to ground |
MTBF: | 250,000 hours at full load at 25°C ambient, calculated per MIL-HDBK-217F |
EMC Performance
- EN55011/ EN55032: Class B conducted, class A radiated
- EN61000-3-2: Harmonic distortion, class A and D
- EN61000-3-3: Line flicker
- EN60601-1-2, EN55024: General EMC immunity standards
- EN61000-4-2: ESD, ±15 KV air and ±8 KV contact
- EN61000-4-3: Radiated immunity, 9-28 V/m
- EN61000-4-4: Fast transient/burst, ±2 KV
- EN61000-4-5: Surge, ±1 KV diff., ±2 KV com
- EN61000-4-6: Conducted immunity, 10 Vrms
- EN61000-4-8: Magnetic field immunity, 30 A/m
- EN61000-4-11: Voltage dip immunity, 30% reduction for 500 ms, 100% reduction for 10 ms
Interface Signals
- PFD: TTL logic high for normal operation and TTL logic low upon loss of input power. This signal appears at least 1 ms prior to master output dropping 5% below its nominal value. This signal also provides a minimum delay of 100 ms after master output is within regulation.
- Inhibit: TTL low to turn off output
- DC OK: TTL high when output voltage >95%
- PS OFF: TTL high to turn off output
Output Power Derating Curve
A graph illustrating the relationship between ambient temperature and maximum output power percentage. The curve shows that the power supply can deliver 100% of its rated output up to an ambient temperature of 50°C. Beyond 50°C, the output power is derated linearly, reaching 50% at 70°C. The x-axis represents ambient temperature in °C, ranging from -10°C to 70°C. The y-axis represents Output Power in percentage, from 0% to 100%.
Output Voltage/Current Rating Chart
Model | V1 | Min. Current | Max. Current at convection | Max. Current at 7 CFM | Tol. | Ripple & Noise | Max. Output Power | Efficiency (typical) 115/230 Vac |
---|---|---|---|---|---|---|---|---|
PM400-12B | 12 V | 0.1 A | 25.00 A | 33.34 A | ±2% | 120 mV | 300 W /400 W | 90 /92% |
PM400-13B | 15 V | 0.1 A | 20.00 A | 26.67 A | ±2% | 150 mV | 300 W /400 W | 90 /92% |
PM400-13-1B | 18 V | 0.1 A | 16.67 A | 22.23 A | ±2% | 180 mV | 300 W /400 W | 91 /92% |
PM400-14B | 24 V | 0.1 A | 12.50 A | 16.67 A | ±2% | 240 mV | 300 W /400 W | 89 /91% |
PM400-15B | 28 V | 0.1 A | 10.72 A | 14.29 A | ±2% | 280 mV | 300 W /400 W | 89 /91% |
PM400-17B | 36 V | 0.1 A | 8.34 A | 11.12 A | ±2% | 360 mV | 300 W /400 W | 90 /92% |
PM400-18B | 48 V | 0.1 A | 6.25 A | 8.34 A | ±2% | 480 mV | 300 W /400 W | 90 /93% |
Notes:
- Change suffix "B" for U-Bracket form to "C" for enclosed form with cover and fan assembly, e.g. PM400-14C.
- 300 W without moving air or 400 W with 7 CFM forced air provided by user for "B" version, 400 W for "C" version with cover and fan assembly.
- Ripple and noise is maximum peak to peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the output.
- All models may be operated at no-load without damage. At no load, output voltage fluctuates beyond 5% due to the burst-mode operation of the control IC in them for energy saving.
Mechanical Specifications
U-bracket Form
Mechanical drawing for the U-bracket form factor of the PM400 series power supply. It shows the overall dimensions: 4.7 inches (119.40 mm) length, 5.5 inches (139.70 mm) width, and 1.15 inches (29.20 mm) height. It indicates mounting holes (A and B) for #6-32 screws and M3x0.5 screws, respectively, with specific locations and counts. Key dimensions like 0.75 inches (19.05 mm) and 0.85 inches (21.50 mm) are also marked.
Enclosed Form
Mechanical drawing for the enclosed form factor of the PM400 series power supply, including a fan. It shows similar overall dimensions to the U-bracket form: 4.7 inches (119.40 mm) length, 5.5 inches (139.70 mm) width, and 1.58 inches (40.20 mm) height. It also indicates mounting holes (A and B) for #6-32 screws and M3x0.5 screws, with specific locations and counts. The presence of a fan assembly is implied by the 'Enclosed Form' designation and the slightly increased height.
Notes:
- Dimensions shown in inches [mm]; tolerance 0.02 [0.5] maximum.
- Input connector P1 is Dinkle terminal P/N DT-35-B01W-03, with nickel plated M3 screws.
- P2, P3: M4 x 0.7 screw connectors
- Connector P4: Molex header 87833-08 or equivalent, mating with Molex housing 51110-0851 or equivalent.
- Fan connector P5: JST header S2B-ZR-3.4 or equivalent, mating with JST housing ZHR-2 or equivalent.
- Weight: 1.0 Kg (2.23 lbs.) approx. for U-bracket form, 1.14 Kgs. (2.52 lbs.) approx. for enclosed form.
- Maximum penetration depth of fixing screws is 4 mm from the outer surface of chassis.
Pin Chart
Connector P1 (AC) & P5
Connector | P1 (AC) | P5 | |||
---|---|---|---|---|---|
PIN NO. | 1 | 2 | 3 | 1 | 2 |
Polarity | Ground | Live | Neutral | +12V Fan | Common Return |
Connector P2, P3 & P4
Connector | P2 | P4 | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
PIN NO. | 1 | 2 | 3 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
Polarity | +V1 | Common Return | +12V Fan | Common Return | Inhibit | +V1 Sense | +5V Standby | -V1 Sense | DC OK | PFD | PS OFF |
Company Information
FSP Power Solution GmbH
Jakobshöhe 16 / 41066 Mönchengladbach / Germany
TEL: +49 (0) 2161 4952490 / FAX : +49 (0) 2161 495249 21
Email: info@fsp-ps.de