Logo INSTRUMEN NASIONALLogo Apex Waves

Prosedur Kalibrasi Gelombang AO kanggo NI-DAQmx

Nggawe jurang antarane pabrikan lan sistem tes warisan sampeyan.

LAYANAN KOMPREHENSIF
Kita nawakake layanan perbaikan lan kalibrasi sing kompetitif, uga dokumentasi sing gampang diakses lan sumber daya sing bisa didownload gratis.
Adol SURPLUS Panjenengan
We tuku bagean anyar, digunakake, decommissioned, lan keluwihan saka saben seri NI.
Kita ngupayakake solusi sing paling apik sing cocog karo kabutuhan individu.
INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - ICON 0 Jual Kanggo Cash INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - ICON 0 GetCredit INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - ICON 0Nampa Deal Trade-In
OBSOLETE NI HARDWARE IN STOCK & READY TO SHIP
We stock New, New Surplus, Dianyari, lan Reconditioned NI Hardware.
Njaluk Kutipan PXI-6733 Instrumen Nasional Modul Output Analog | Ombak Puncak PXI-6733

Konvensi

Konvensi ing ngisor iki katon ing manual iki:

INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Ikon Kurung sudut sing ngemot angka sing dipisahake dening elipsis nggambarake sawetara nilai sing ana gandhengane karo bit utawa jeneng sinyal-kanggo mantanamplan, P0.<0..7>.
INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Ikon 1 Simbol » nuntun sampeyan liwat item menu nested lan opsi kothak dialog menyang tumindak pungkasan. Urutane File»Page Setup»Options ngarahake sampeyan narik mudhun File menu, pilih item Page Setup, banjur pilih Pilihan saka kothak dialog pungkasan.
INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Ikon 2 Lambang iki nuduhake cathetan, sing menehi tandha informasi penting.
kandel Teks kandel nuduhake item sing kudu sampeyan pilih utawa klik ing piranti lunak, kayata item menu lan opsi kothak dialog. Teks kandel uga nuduhake jeneng arameter lan label hardware.
miring Teks miring nuduhake variabel, emphasis, referensi silang, utawa introduksi kanggo konsep kunci. Font iki uga nuduhake teks sing dadi placeholder kanggo tembung utawa nilai sing kudu diwenehake.
monospace Teks monospace nuduhake teks utawa karakter sing kudu sampeyan lebokake saka keyboard, bagean kode, program examples, lan sintaksis examples. Font iki uga digunakake kanggo jeneng disk drive, path, direktori, program, subprogram, subrutin, jeneng piranti, fungsi, operasi, variabel, filejeneng, lan ekstensi.
miring monospace Teks miring ing font iki nuduhake teks sing dadi placeholder kanggo tembung utawa nilai sing kudu diwenehake.

Pambuka

Dokumen iki ngemot instruksi kanggo kalibrasi NI 671X / 672X / 673X kanggo piranti output analog (AO) PCI / PXI / CompactPCI.
Dokumen iki ora ngrembug teknik pemrograman utawa konfigurasi compiler. Driver DAQmx Instrumen Nasional ngemot bantuan files sing duwe instruksi compiler-tartamtu lan panjelasan fungsi rinci. Sampeyan bisa nambah bantuan iki files nalika sampeyan nginstal NI-DAQmx ing komputer kalibrasi.
Piranti AO kudu dikalibrasi kanthi interval biasa kaya sing ditetepake miturut syarat akurasi pangukuran aplikasi sampeyan. National Instruments nyaranake sampeyan nindakake kalibrasi lengkap paling sethithik sapisan saben taun. Sampeyan bisa nyepetake interval iki dadi 90 dina utawa nem sasi.

Piranti lunak

Kalibrasi mbutuhake driver NI-DAQmx paling anyar. NI-DAQmx kalebu panggilan fungsi tingkat dhuwur kanggo nyederhanakake tugas nulis piranti lunak kanggo calibrate piranti. Pembalap ndhukung akeh basa pamrograman, kalebu LabVIEW, LabWindows ™/CVI ™ , icrosoft Visual C++, Microsoft Visual Basic, lan Borland C++.

Dokumentasi

Yen sampeyan nggunakake driver NI-DAQmx, dokumen ing ngisor iki minangka referensi utama kanggo nulis sarana kalibrasi:

  • Bantuan Referensi NI-DAQmx C kalebu informasi babagan fungsi ing driver.
  • DAQ Quick Start Guide kanggo NI-DAQ 7.3 utawa mengko menehi instruksi kanggo nginstal lan konfigurasi piranti NI-DAQ.
  • Bantuan NI-DAQmx kalebu informasi babagan nggawe aplikasi sing nggunakake driver NI-DAQmx.

Kanggo informasi luwih lengkap babagan piranti sing sampeyan kalibrasi, waca
Pitulung Seri Output Analog.

Peralatan Tes

Gambar 1 nuduhake peralatan tes sing dibutuhake kanggo kalibrasi piranti. DMM tartamtu, kalibrator, lan sambungan counter diterangake ing bagean Proses Kalibrasi.INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Accosorish

Gambar 1. Sambungan Kalibrasi
Nalika nindakake kalibrasi, National Instruments nyaranake sampeyan nggunakake instrumen ing ngisor iki kanggo kalibrasi piranti AO:

  • Kalibrator—Fluke 5700A. Yen instrumen kasebut ora kasedhiya, gunakake volume tliti dhuwurtage sumber sing paling 50 ppm akurat kanggo 12- lan 13-dicokot Papan lan 10 ppm kanggo Papan 16-dicokot.
  • DMM—NI 4070. Yen instrumen kasebut ora kasedhiya, gunakake DMM 5.5 digit multi-digit kanthi akurasi 40 ppm (0.004%).
  • Counter-Hewlett-Packard 53131A. Yen instrumen kasebut ora kasedhiya, gunakake counter akurat nganti 0.01%.
  • Kabel plug-in EMF tembaga termal rendah-Fluke 5440A-7002. Aja nggunakake kabel banana standar.
  • Kabel DAQ-NI nyaranake nggunakake kabel shielded, kayata SH68-68-EP karo NI 671X / 673X utawa SH68-C68-S karo NI 672X.
  • Salah sawijining aksesoris DAQ ing ngisor iki:

- SCB-68-SCB-68 punika shielded I / O pemblokiran konektor karo 68 terminal meneng kanggo sambungan sinyal gampang kanggo 68- utawa 100-pin piranti DAQ.
– CB-68LP/CB-68LPR/TBX-68—CB-68LP, CB-68LPR, lan TBX-68 minangka aksesoris terminasi murah kanthi terminal sekrup 68 kanggo gampang nyambungake sinyal I/O lapangan menyang DAQ 68-pin piranti.

Pertimbangan Test

Tindakake pedoman iki kanggo ngoptimalake sambungan lan kahanan tes sajrone kalibrasi:

  • Tansah sambungan menyang NI 671X / 672X / 673X singkat. Kabel lan kabel dawa tumindak minangka antena, njupuk swara ekstra, sing bisa mengaruhi pangukuran.
  • Gunakake kabel tembaga sing dilindhungi kanggo kabeh sambungan kabel menyang piranti.
  • Gunakake kabel pasangan bengkong kanggo ngilangi gangguan lan offset termal.
  • Njaga suhu antarane 18 lan 28 ° C. Kanggo operate modul ing suhu tartamtu ing njaba kisaran iki, calibrate piranti ing suhu kasebut.
  • Tansah kelembapan relatif ing ngisor 80%.
  • Ngidini wektu anget paling sethithik 15 menit kanggo mesthekake yen sirkuit pangukuran ing suhu operasi sing stabil.

Proses Kalibrasi

Bagean iki menehi instruksi kanggo verifikasi lan kalibrasi piranti.
Proses Kalibrasiview
Proses kalibrasi duwe papat langkah:

  1. Persiyapan Awal—Konfigurasi piranti ing NI-DAQmx.
  2. Prosedur Verifikasi AO-Verifikasi operasi piranti sing wis ana. Langkah iki ngidini sampeyan ngonfirmasi manawa piranti kasebut digunakake ing jangkoan sing ditemtokake sadurunge kalibrasi.
  3. Prosedur Penyesuaian AO-Nindakake kalibrasi eksternal sing nyetel konstanta kalibrasi piranti karo volume sing dikenal.tage sumber.
  4. Tindakake verifikasi liyane kanggo mesthekake yen piranti wis operasi ing specifications sawise adjustment.Langkah-langkah iki diterangake ing rinci ing bagean ngisor. Amarga verifikasi lengkap kabeh kisaran piranti bisa njupuk sawetara wektu, sampeyan bisa uga pengin verifikasi mung kisaran kapentingan kanggo sampeyan.

Setup wiwitan
NI-DAQmx kanthi otomatis ndeteksi kabeh piranti AO. Nanging, supaya driver bisa komunikasi karo piranti, iku kudu diatur ing NI-DAQmx.
Kanggo ngatur piranti ing NI-DAQmx, rampungake langkah-langkah ing ngisor iki:

  1. Instal piranti lunak driver NI-DAQmx.
  2. Mateni komputer sing bakal nyekel piranti, lan nginstal piranti ing slot kasedhiya.
  3. Daya ing komputer, lan bukak Measurement & Automation Explorer (MAX).
  4. Konfigurasi pengenal piranti banjur pilih Self-Test kanggo mesthekake yen piranti bisa digunakake kanthi bener.

INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Ikon 2 Cathetan Nalika piranti dikonfigurasi nganggo MAX, bakal diwenehi pengenal piranti. Saben
telpon fungsi nggunakake pengenal iki kanggo nemtokake piranti DAQ kanggo calibrate.

Prosedur Verifikasi AO

Verifikasi nemtokake manawa piranti DAQ nyukupi spesifikasi. Kanthi nindakake prosedur iki, sampeyan bisa ndeleng carane piranti wis operate liwat wektu. Sampeyan bisa nggunakake informasi iki kanggo mbantu nemtokake interval kalibrasi sing cocok kanggo aplikasi sampeyan.
Prosedur verifikasi dipérang dadi fungsi utama piranti. Sajrone proses verifikasi, gunakake tabel ing bagean AO Device Test Limit kanggo nemtokake manawa piranti sampeyan kudu diatur.
Verifikasi Output Analog
Prosedur iki mriksa kinerja output analog. Priksa pangukuran nggunakake prosedur ing ngisor iki:

  1. Sambungake DMM menyang AO 0 kaya sing ditampilake ing Tabel 1.
    Tabel 1. Nyambungake DMM menyang AO <0..7>\
    Saluran Output DMM Input Positif DMM Input Negatif
    AO 0 AO 0 (pin 22) AO GND (pin 56)
    AO 1 AO 1 (pin 21) AO GND (pin 55)
    AO 2 AO 2 (pin 57) AO GND (pin 23)

    Tabel 1. Nyambungake DMM menyang AO <0..7> (Lanjutan)

    Saluran Output DMM Input Positif DMM Input Negatif
    AO 3 AO 3 (pin 25) AO GND (pin 59)
    AO 4 AO 4 (pin 60) AO GND (pin 26)
    AO 5 AO 5 (pin 28) AO GND (pin 61)
    AO 6 AO 6 (pin 30) AO GND (pin 64)
    AO 7 AO 7 (pin 65) AO GND (pin 31)

    Tabel 2. Nyambungake DMM menyang AO <8..31> ing NI 6723

    Saluran Output DMM Input Positif DMM Input Negatif
    AO 8 AO 8 (pin 68) AO GND (pin 34)
    AO 9 AO 9 (pin 33) AO GND (pin 67)
    AO 10 AO 10 (pin 32) AO GND (pin 66)
    AO 11 AO 11 (pin 65) AO GND (pin 31)
    AO 12 AO 12 (pin 30) AO GND (pin 64)
    AO 13 AO 13 (pin 29) AO GND (pin 63)
    AO 14 AO 14 (pin 62) AO GND (pin 28)
    AO 15 AO 15 (pin 27) AO GND (pin 61)
    AO 16 AO 16 (pin 26) AO GND (pin 60)
    AO 17 AO 17 (pin 59) AO GND (pin 25)
    AO 18 AO 18 (pin 24) AO GND (pin 58)
    AO 19 AO 19 (pin 23) AO GND (pin 57)
    AO 20 AO 20 (pin 55) AO GND (pin 21)
    AO 21 AO 21 (pin 20) AO GND (pin 54)
    AO 22 AO 22 (pin 19) AO GND (pin 53)
    AO 23 AO 23 (pin 52) AO GND (pin 18)
    AO 24 AO 24 (pin 17) AO GND (pin 51)
    AO 25 AO 25 (pin 16) AO GND (pin 50)
    AO 26 AO 26 (pin 49) AO GND (pin 15)

    Tabel 2. Nyambungake DMM menyang AO <8..31> ing NI 6723 (Lanjutan)

    Saluran Output DMM Input Positif DMM Input Negatif
    AO 27 AO 27 (pin 14) AO GND (pin 48)
    AO 28 AO 28 (pin 13) AO GND (pin 47)
    AO 29 AO 29 (pin 46) AO GND (pin 12)
    AO 30 AO 30 (pin 11) AO GND (pin 45)
    AO 31 AO 31 (pin 10) AO GND (pin 44)
  2. Pilih tabel saka bagean AO Device Test Limit sing cocog karo piranti sing sampeyan verifikasi. Tabel iki nuduhake kabeh setelan sing bisa ditampa kanggo piranti kasebut. Sanajan NI nyaranake sampeyan verifikasi kabeh kisaran, sampeyan bisa uga pengin ngirit wektu kanthi mriksa mung kisaran sing digunakake ing aplikasi sampeyan.
  3. Nggawe tugas nggunakake DAQmxCreateTask.
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxCreateTask kanthi paramèter ing ngisor iki:
    Jeneng tugas: MyAOVoltageTask
    nangani tugas: &taskHandle
    LabVIEW ora mbutuhake langkah iki.
  4. Tambah AO voltage tugas nggunakake DAQmxCreateAOVoltageChan (DAQmx Nggawe Virtual Channel VI) lan ngatur saluran, AO 0. Gunakake tabel ing bagean AO Piranti Test Watesan kanggo nemtokake nilai minimal lan maksimum kanggo piranti.
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxCreateAOVoltageChan kanthi paramèter ing ngisor iki:
    nangani tugas: Handle tugas
    saluran fisik: dev1/aoO
    jenengToAssignToChannel: AOVoltageChannel
    minVal: -10.0
    maxVal: 10.0
    unit: DAQmx_Val_Volt
    customScaleName: NULL
    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram
  5. Miwiti akuisisi nggunakake DAQmxStartTask (DAQmx Start Task VI).
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxStartTask kanthi paramèter ing ngisor iki:
    nangani tugas: Handle tugas
    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram 1
  6. Tulis voltage menyang saluran AO nggunakake DAQmxWriteAnalogF64 (DAQmx Tulis VI) nggunakake meja kanggo piranti ing bagean AO Piranti Test Watesan.
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxWriteAnalogF64 kanthi paramèter ing ngisor iki:
    nangani tugas: Handle tugas nomerampsPerChan: 1miwiti otomatis: 1

    wektu entek: 10.0

    dataLayout:

    DAQmx_Val_GroupByChannel nulisArray: & data sampsPerChanWritten: &sampkurangDitulis

    dilindhungi undhang-undhang: NULL

    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram 2
  7. Mbandhingaké nilai asil ditampilake dening DMM kanggo wates ndhuwur lan ngisor ing meja. Yen nilai ana ing antarane watesan kasebut, tes kasebut dianggep wis lulus.
  8. Mbusak akuisisi nggunakake DAQmxClearTask (DAQmx Clear Task VI).
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxClearTask kanthi parameter ing ngisor iki:

    nangani tugas: Handle tugas

    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram 3
  9. Baleni langkah 4 nganti 8 nganti kabeh nilai wis diuji.
  10. Copot DMM saka AO 0, lan sambungake maneh menyang saluran sabanjure, nggawe sambungan kaya sing ditampilake ing Tabel 1.
  11. Baleni langkah 4 nganti 10 nganti sampeyan wis verifikasi kabeh saluran.
  12. Pedhot DMM saka piranti.

Sampeyan wis rampung verifikasi tingkat output analog ing piranti sampeyan.

Verifikasi Counter
Prosedur iki verifikasi kinerja counter. Piranti AO mung duwe siji timebase kanggo verifikasi, dadi mung counter 0 sing kudu dicenthang. Ora bisa nyetel timebase iki, dadi mung verifikasi sing bisa ditindakake.
Tindakake mriksa nggunakake prosedur ing ngisor iki:

  1. Sambungake input counter positif menyang CTR 0 OUT (pin 2) lan input negatif counter menyang D GND (pin 35).
  2. Nggawe tugas nggunakake DAQmxCreateTask.
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxCreateTask kanthi paramèter ing ngisor iki:
    Jeneng tugas: MyCounterOutputTask
    nangani tugas: &taskHandle
    LabVIEW ora mbutuhake langkah iki.
  3. Tambah saluran output counter kanggo tugas nggunakake DAQmxCreateCOPulseChanFreq (DAQmx Nggawe Virtual Channel VI) lan ngatur saluran.
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxCreateCOPulseChanFreq kanthi paramèter ing ngisor iki:
    nangani tugas: Handle tugas
    counter: dev1/ctr0
    jenengToAssignToChannel: CounterOutputChannel
    unit: DAQmx_Val_Hz
    idleState: DAQmx_Val_Low
    tundha wiwitan: 0.0
    freq: 5000000.0
    siklus tugas: .5
    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram 4
  4. Ngatur counter kanggo generasi gelombang kothak terus nggunakake DAQmxCfgImplicitTiming (DAQmx Wektu VI).
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxCfgImplicitTiming kanthi paramèter ing ngisor iki:
    nangani tugas: Handle tugas
    sampleMode: DAQmx_Val_ContSamps
    sampsPerChan: 10000
    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram 5
  5. Miwiti generasi gelombang persegi nggunakake DAQmxStartTask (DAQmx Start Task VI).
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxStartTask kanthi parameter ing ngisor iki:
    nangani tugas: Handle tugas
    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram 6
  6. Piranti kasebut bakal miwiti ngasilake gelombang persegi 5 MHz nalika fungsi DAQmxStartTask ngrampungake eksekusi. Mbandhingake nilai sing diwaca dening counter sampeyan karo watesan tes sing ditampilake ing meja piranti. Yen nilai kasebut ana ing antarane watesan kasebut, tes kasebut dianggep wis lulus.
  7. Mbusak generasi nggunakake DAQmxClearTask (DAQmx Clear Task VI).
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxClearTask kanthi parameter ing ngisor iki:
    nangani tugas: nangani tugas
    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram 3
  8. Copot sambungan counter saka piranti.
    Sampeyan wis verifikasi counter ing piranti.

Prosedur Imbuhan AO

Gunakake prosedur pangaturan AO kanggo nyetel konstanta kalibrasi output analog. Ing pungkasan saben prosedur kalibrasi, konstanta anyar iki disimpen ing area kalibrasi eksternal EEPROM. Nilai kasebut dilindhungi sandi, sing nyegah akses ora sengaja utawa modifikasi konstanta kalibrasi sing diatur dening laboratorium metrologi. Sandi standar yaiku NI.
Kanggo nindakake pangaturan piranti nganggo kalibrator, tindakake langkah-langkah ing ngisor iki:

  1. Sambungake kalibrator menyang piranti miturut Tabel 3.
    Tabel 3. Nyambungake Calibrator menyang Piranti
    671X / 672X / 673X Pin Kalibrator 
    AO EXT REF (pin 20) Output Dhuwur
    AO GND (pin 54) Output Kurang
  2. Setel kalibrator kanggo output a voltage saka 5 V.
  3. Bukak sesi kalibrasi ing piranti nggunakake DAQmxInitExtCal (DAQmx Initialize External Calibration VI). Sandi standar yaiku NI.
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxInitExtCal kanthi paramèter ing ngisor iki:
    Jeneng piranti: dev1
    sandi: NI
    calHandle: &calHandle
    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram 8
  4. Nindakake pangaturan kalibrasi eksternal nggunakake DAQmxESeriesCalAdjust (DAQmx Adjust AO-Series Calibration VI).
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxAOSeriesCalAdjust kanthi paramèter ing ngisor iki:
    calHandle: calHandle
    referensiVoltage:5
    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram 9
  5. Simpen pangaturan menyang EEPROM, utawa memori onboard, nggunakake DAQmxCloseExtCal (DAQmx Close External Calibration). Fungsi iki uga ngirit tanggal, wektu, lan suhu pangaturan menyang memori onboard.
    Panggilan Fungsi NI-DAQ LabVIEW Diagram Blokir
    Telpon DAQmxCloseExtCal kanthi paramèter ing ngisor iki:
    calHandle: calHandle
    tumindak: DAQmx_Val_
    Tumindak_Komit
    INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Diagram 10
  6. Copot calibrator saka piranti.

Piranti saiki wis dikalibrasi babagan sumber eksternal sampeyan.
Sawise nyetel piranti, sampeyan bisa uga pengin verifikasi operasi output analog. Kanggo nindakake iki, baleni langkah-langkah ing bagean Prosedur Verifikasi AO nggunakake watesan tes 24 jam ing bagean Watesan Tes Piranti AO.

Watesan Test Piranti AO

Tabel ing bagean iki dhaftar specifications akurasi digunakake nalika verifikasi lan nyetel NI 671X / 672X / 673X. Tabel kasebut nampilake spesifikasi kanggo interval kalibrasi 1 taun lan 24 jam. Kisaran 1 taun nampilake spesifikasi sing kudu ditindakake piranti yen wis setaun antarane kalibrasi. Nalika piranti wis dikalibrasi karo sumber eksternal, nilai sing ditampilake ing tabel 24 jam minangka spesifikasi sing bener.
Nggunakake Tabel
Dhéfinisi ing ngisor iki njlèntrèhaké cara nggunakake informasi saka tabel ing bagean iki.
Range
Range nuduhake vol maksimum allowabletage sawetara sinyal output.
Titik Tes
Titik Tes yaiku voltagnilai e sing digawe kanggo tujuan verifikasi. Nilai iki dipérang dadi rong kolom: Lokasi lan Nilai. Lokasi nuduhake ing ngendi nilai tes pas ing sawetara tes. Pos FS singkatan saka positif full-scale lan Neg FS singkatan negatif full-scale. Nilai nuduhake voltagNilai e kanggo diverifikasi lan ana ing volt.
24-Jam Range
Kolom Range 24 Jam ngemot watesan ndhuwur lan watesan ngisor kanggo nilai titik tes. Yaiku, nalika piranti ana ing interval kalibrasi 24 jam, nilai titik tes kudu ana ing antarane nilai watesan ndhuwur lan ngisor. Watesan ndhuwur lan ngisor ditulis ing volt.
Range 1 Taun
Kolom Range 1 Taun ngemot watesan ndhuwur lan watesan ngisor kanggo nilai titik tes. Yaiku, nalika piranti kasebut ana ing interval kalibrasi 1 taun, nilai titik tes kudu ana ing antarane nilai watesan ndhuwur lan ngisor. Watesan ndhuwur lan ngisor ditulis ing volt.
Counter
Ora bisa nyetel resolusi counter / timer. Mulane, nilai kasebut ora duwe periode kalibrasi 1 taun utawa 24 jam. Nanging, titik tes lan wates ndhuwur lan ngisor diwenehake kanggo tujuan verifikasi.
NI 6711/6713—Resolusi 12-Bit
Tabel 4.
NI 6711/6713 Nilai Output Analog

Range (V) Titik Tes 24-Jam Range 1-Taun Range
minimal maksimal Panggonan Nilai (V) Limit Ngisor (V) Limit Ndhuwur (V) Limit Ngisor (V) Limit Ndhuwur (V)
–10 10 0 0.0 –0.0059300 0.0059300 –0.0059300 0.0059300
–10 10 Pos FS 9.9900000 9.9822988 9.9977012 9.9818792 9.9981208
–10 10 Neng FS –9.9900000 –9.9977012 –9.9822988 –9.9981208 –9.9818792

Tabel 5. NI 6711/6713 Nilai Counter

Titik Setel (MHz) Limit ndhuwur (MHz) Limit Ngisor (MHz)
5 5.0005 4.9995

NI 6722/6723—Resolusi 13-Bit
Tabel 6. NI 6722/6723 Nilai Output Analog

Range (V) Titik Tes 24-Jam Range 1-Taun Range
minimal maksimal Panggonan Nilai (V) Limit Ngisor (V) Limit Ndhuwur (V) Limit Ngisor (V) Limit Ndhuwur (V)
–10 10 0 0.0 –0.0070095 0.0070095 –0.0070095 0.0070095
–10 10 Pos FS 9.9000000 9.8896747 9.9103253 9.8892582 9.9107418
–10 10 Neng FS –9.9000000 –9.9103253 –9.8896747 –9.9107418 –9.8892582

Tabel 7. NI 6722/6723 Nilai Counter

Titik Setel (MHz) Limit ndhuwur (MHz) Limit Ngisor (MHz)
5 5.0005 4.9995

NI 6731/6733—Resolusi 16-Bit
Tabel 8. NI 6731/6733 Nilai Output Analog

Range (V) Titik Tes 24-Jam Range 1-Taun Range
minimal maksimal Panggonan Nilai (V) Limit Ngisor (V) Limit Ndhuwur (V) Limit Ngisor (V) Limit Ndhuwur (V)
–10 10 0 0.0 –0.0010270 0.0010270 –0.0010270 0.0010270
–10 10 Pos FS 9.9900000 9.9885335 9.9914665 9.9883636 9.9916364
–10 10 Neng FS –9.9900000 –9.9914665 –9.9885335 –9.9916364 –9.9883636

Tabel 9. NI 6731/6733 Nilai Counter

Titik Setel (MHz) Limit ndhuwur (MHz) Limit Ngisor (MHz)
5 5.0005 4.9995

CVI™, LabVIEW™, Instrumen Nasional™, NI™, ni.com™, lan NI-DAQ™ minangka merek dagang saka National Instruments Corporation. Jeneng produk lan perusahaan sing kasebut ing kene minangka merek dagang utawa jeneng dagang saka perusahaan kasebut. Kanggo paten sing nyakup produk National Instruments, deleng lokasi sing cocog: Bantuan»Paten ing piranti lunak sampeyan, patents.txt file ing CD, utawa ni.com/patents.
© 2004 National Instruments Corp. Kabeh hak dilindhungi undhang-undhang.
Kabeh merek dagang, merek, lan jeneng merek minangka properti sing nduweni.

ZEVNI A04672 Berton 10 Inch H Black Wall Sconce - lambang 1-800-9156216
hama 00188313 5000mAh Slim5-HD Power Pack - Sumbi www.apexwaves.com
Maxxima MCL 710600D Sensor Gerak Lampu Langit-langit LED - Ikon 4 sales@apexwaves.comINSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI DAQmx - Ikon370938A-01
Juli04

Dokumen / Sumber Daya

INSTRUMEN NASIONAL Prosedur Kalibrasi Gelombang AO kanggo NI-DAQmx [pdf] Pandhuan pangguna
PXI-6733, PCI-6711, PCI-6713, PXI-6711, PXI-6713, DAQCard-6715, 6715, 6731, 6733, PCI-6731, PCI-6733, PXI-6731, PXI-6733, PXI-6722 6722, PXI-6722, 6723, PCI-6723, PXI-6723, Prosedur Kalibrasi Gelombang AO kanggo NI-DAQmx, Prosedur Kalibrasi Gelombang AO, Prosedur Kalibrasi kanggo NI-DAQmx, Prosedur Kalibrasi Gelombang, Prosedur Kalibrasi NI-DAQmx

Referensi

Ninggalake komentar

Alamat email sampeyan ora bakal diterbitake. Kolom sing dibutuhake ditandhani *