This manual describes the functions, configuration, mounting and operating instructions of the ITP17 universal process indicator (hereinafter referred to as the device).
Connection, setup and maintenance of the device must be performed only by fully qualified personnel after reading this user guide.

شرایط او لنډیزونه

PC – personal computer akYtec Tool Pro – configuration software
USB (Universal Serial Bus) – serial communication interface

اوورview

The device is designed to measure and indicate signals of resistance temperature detectors (RTD), thermocouples (TC), pyrometers, DC voltage and DC signals (U / I signals).

افعال

  • measuring and displaying a measured value on the digital display;
  • signalling by color-coded indication about exceeding the set thresholds of the measured value;
  • signalling when the value is in the critical zone;
  • setting of the measured value according to the on/off- law using a discrete output based  on a transistor switch;
  • indication of a break or short circuit in the “device-sensor” communication line.

مشخصات

جدول 1 مشخصات

پیرامیټر ارزښت
بریښنایی
د بریښنا رسول 24 (10..30) VDC
د بریښنا مصرف، اعظمي. 1 W
د وسایلو ټولګي III
Galvanic isolation between domain of combined power and output interface and input domain 500 وی
ننوتلي سيګنالونه
شمیره 1
Input resistance at voltage measuring, min. 100 کیلو
ننوتۍ جلدtage drop (at current measuring), max. 1.6 وی
Input signals supported وګورئ برخه 6
Sampling time, max. 1 s
Full-scale accuracy , max. RTD, U / I signals

TC, pyrometers

 ± 0,25٪

± 0,5٪

د حرارت درجه اغیزه 0,2 of full-scale accuracy limit/ 10 °C
محصول
NPN transistor, loading capacity 200 mА, 42 VDC
Length of signal line, max. 30 متر
ترتیب انٹرفیس
Connector for configuration with akYtec Tool Pro مایکرو USB
ښودل
 شاخص one 4–digit and 7– segment indicator
رنګونه 3
د کرکټر لوړوالی 14 ملي متره
میخانیکي
ابعاد 48 × 26 × 72 mm
IP code (front / rear) (IP65 / IP20)
MTBF ۸ ساعته
د اوسط خدمت ژوند ۵ کاله
وزن نږدې 150 ګرامه

د چاپیریال شرایط

The device is designed for natural convection cooling which should be taken into account when choosing the installation site.
The following environmental conditions must be observed:

  • clean, dry and controlled environment, low dust level;
  • closed non-hazardous areas, free of corrosive or flammable gases.

جدول 2 د چاپیریال شرایط

حالت جواز لري حد
محیطي حرارت -40…+60 °C
نسبي رطوبت 30…80% (غیر کنډیشن)
دانتقال او زېرمه حرارت -25… +55 C
Transportation and storage relative humidity 5…95% (غیر کنډیشن)
لوړوالی up to 2000 m ASL
EMC emission / immunity conforms to IEC 61000-6-3-2016

نوټ
When operating the device at an altitude above 1000 m above sea level, it is  necessary to take into account the reduction of the electrical insulation as well as  the reduction in the cooling effect of the air.

ننوتلي سيګنالونه

Table 3 Signals and sensors 

اشاره تفصیل اندازه کول حد*
RTD
С 50 Cu50 (α = 0,00426 °С-1) -50…+200 °С
50 С 50М (α = 0,00428 °С -1) -180…+200 °С
۲۳ مخ Pt50 (α = 0,00385 °С-1) -200…+850 °С
50P 50P (α = 0,00391 °С -1) -200…+850 °С
C100 Cu100 (α = 0,00426 °С-1) -50…+200 °С
100C 100М (α = 0,00428 °С-1) -180…+200 °С
P100 Pt100 (α = 0,00385 °С -1) -200…+850 °С
100P 100P (α = 0,00391 °С -1) -200…+850 °С
100n 100N (α = 0,00617 °С -1) -60…+180 °С
P500 Pt500 (α = 0,00385 °С -1) -200…+850 °С
500P 500P (α = 0,00391 °С -1) -200…+850 °С
C500 Cu500 (α = 0,00426 °С -1) -50…+200 °С
500C 500М (α = 0,00428 °С -1) -180…+200 °С
500n 500N (α = 0,00617 °С -1) -60…+180 °С
C 1.0 Cu1000 (α = 0,00426°С-1) -50…+200 °С
1.0 C 1000М (α = 0,00428 °С-1) -180…+200 °С
۲۳ مخ Pt1000 (α = 0,00385 °С-1) -200…+850 °С
۲۴ مخ 1000P (α = 0,00391 °С-1) -200…+850 °С
1.0 ن 1000N (α = 0,00617 °С-1) -60…+180 °С
TC
tC.L L -200…+800 °С
tP.HA К -200…+1300 °С
tC.J J -200…+1200 °С
tC.n N -200…+1300 °С
tC.t Т -200…+400 °С
tC.S S 0…+1750 °С
tC.r R 0…+1750 °С
tC.b В +200…+1800 °С
tC.A1 AA-1 0…+2500 °С
tC.A2 AA-2 0…+1800 °С
tP.A3 AA-3 0…+1800 °С
TC in accordance with DIN 43710
tC.dL L -200…+900 °С
I signals**
I 0.5 0…5 mА ۰…۱۰۰٪
I0.20 0…20 mА ۰…۱۰۰٪
I4.20 4…20 mА ۰…۱۰۰٪
U signals**
U-5.5 -50…+50 mV*** ۰…۱۰۰٪
U 0. 1 ۱۵۰…۵۳۰ وی ۰…۱۰۰٪
U0.10 ۱۵۰…۵۳۰ وی ۰…۱۰۰٪
U2.10 ۱۵۰…۵۳۰ وی ۰…۱۰۰٪
Pyrometers
PIr.1 RК-15 +400…+1500 °С
PIr.2 RК-20 +600…+2000 °С
PIr.3 RS-20 +900…+2000 °С
PIr.4 RS-25 +1200…+2500 °С

At the temperature over 999,9 and under -199,90C the value of the least significant digit equals 10C.
The values depend on the di.Lo and di.Hi parameters.
Accuracy is not standardized.

خوندیتوب

خبرداری
خطرناکه ټوکtage!
Electric shock could kill or seriously injure.

  • All work on the device must be performed by a fully qualified electrician.
  • ډاډ ترلاسه کړئ چې اصلي حجمtage د حجم سره سمون لريtage په وسیله په نښه شوی.
  • Ensure that the device is provided with its power supply line and electric  fuse.
  • The device may not be used in aggressive environments, in atmospheres in  which there are chemically active substances.
  • The output port and internal electrical elements of the device must be protected from the humidity.

خبرتیا

  • De-energize the device before working on it. Switch on the power supply  only after completing all work on the device.

پورته کول

د وسیله نصبولو لپاره:

  • Prepare the mounting cutout with Ø of 22.5 mm in the switchboard where the device should be mounted (see Fig. 2).
  • Carefully position the supplied gasket (see Fig.
  •  Place the device with the installed gasket in  the prepared mounting cutout and tighten the nut (included in the scope of delivery) to fix the device.

خبرتیا
Do not use any tools to tighten the nut. Tighten the nut only by hand.

HEMOMATIK -ITP17 -Universal-Process-Indicator (2)Removing proceeds in the reverse order. HEMOMATIK -ITP17 -Universal-Process-Indicator (3)

پیوستون

عمومي معلومات
Signal cables should be routed separately from the power supply cables as well as from the cables which are sources of high-frequency and impulse interference.

  • For high-quality clamping and reliable electrical connections, it is recommended to use:
  • copper multicore wires, diameter after tinning – 0.9 mm (17 cores, AWG 22) or 1.1 mm (21 cores, AWG 20);
  • copper wires with single-wire cores, diameter from 0.51 to 1.02 mm (AWG 24-18).
  • The ends of the wires should be stripped of insulation by 8 ± 0.5 mm (see Fig. 4) and, if  necessary, tinned.

HEMOMATIK -ITP17 -Universal-Process-Indicator (4)

 

تارونه

احتیاط
To protect the device input from the influence of industrial electromagnetic interference, the “device-sensor” communication lines should be shielded. To protect the device input circuits from possible breakdown by static electricity charges accumulated on the “device-sensor” communication lines, their wires  should be connected to the shield ground screw for 1–2 s before connecting to  the device terminal block.
Connect the “device – sensor” communication lines to the primary converter and the device input and connect the device to the power supply (see Fig. 6).

HEMOMATIK -ITP17 -Universal-Process-Indicator (5)انځر .6 د تار کولو ډیاګرام
To protect the device against microseconds’ impulse noise of the output device (open collector) on the terminals “DO” and “-“, it is recommended to use connecting lines no longer than 30 meters or install devices for protection against impulse noise on the DC line. The VD1 diode should be located as close as possible to the terminals of the relay winding. The parameters of the diode are selected in accordance with the following rules:

The forward current of the diode must be at least 1.3 R1 (1.3 of the relay coil current).

Indication and control

  • The 4– digit and 7– segment indicator on the front panel is designed to display measured values, alarms and device parameters. The segments of the digital indicator can light up in  one of the following colors (see Section 13):
  • شین؛
  • سور؛
  • ژیړ.

HEMOMATIK -ITP17 -Universal-Process-Indicator (6)There is a micro-USB connector on the bottom of the enclosure.

HEMOMATIK -ITP17 -Universal-Process-Indicator (1)

Error indication and remedy

Being powered the device starts to operate. If the indicated values do not correspond to the real measured values, check the following:

  • sensor function and communication line integrity;
  • correctness of sensor connection;
  • settings of scaling parameters (di.Lo and di.Hi).

Table 5 Error indication and remedy 

اشاره تفصیل درملنه
HHHH The measured input value is above the upper limit  

Check the sensor code and measured value compliance

ایل ایل ایل The measured input value is below the lower limit
 Hi The calculated value exceeds the maximum possible positive value that can be displayed on 4 digits of the indicator  

 

Readjust the dP.t پیرامیټر

 

Lo

The calculated value is less than the minimum possible negative value that can be displayed on 4 digits of the indicator
 |——|  

د سینسر ماتول

Check the signal line. If the signal line isn’t broken and

the connection is correct, contact akYtec service staff

Er.[ ] Cold junction sensor (CJS) failure Contact akYtec service staff
اشاره تفصیل جواز لري

ارزښتونه

فابریکه ترتیبات
in.t د ننوتلو سیګنال وګورئ برخه 6 4…20 mA
td د ډیجیټل فلټر وخت ثابت 0…10 ثانیې 0
SQrt Square root function (for U signals) په / OFF د
di.Lo Signal lower limit (for I / U signals) – ۵۰۰…۵۰۰۰ 0
di.Hi Signal upper limit (for I / U signals) – ۵۰۰…۵۰۰۰ 100
dP.T د لسیزې نقطې موقعیت -a-u-to-

-.-.- –

– – -.-
2u3u RTD connection: 2–wire or 3–wire 2-Ln

3-Ln

3-Ln
کور Offset correction of the measured input value –1999 …9999 0
 

سي این ټي

د کنټرول فعالیت:

OFF / Heating / Cooling /

Alarm within limits (П) / Alarm outside limits (U) (see انځور 10)

 

oFF/HEAt/ CooL/П/U

 

U

SP.Lo Setpoint lower limit – ۵۰۰…۵۰۰۰ 0
SP.Hi Setpoint upper limit – ۵۰۰…۵۰۰۰ 30
 

 

 

A.HYS

Hysteresis. When “Alarm within limits (П)” or “Alarm outside limits (U)” are selected, the hysteresis blocks the actuation of the output unit with minor

fluctuations at the SP.Lo and SP.Hi boundary. The parameter is not

کله چې ښکاره شي سي این ټي = oFF/HEAt/CooL

۱۶…۲۷  

 

0

di.Sh Characteristics offset –1999 …9999 0
out.E Output device state in case of sensor failure په / OFF د
d.FnC د فلش کولو فعالیت په / OFF د
Zon.1 Thresholds for changing the color of indicator zones ۱۶…۲۷ 0
Zon.2 50
Zon.3 80
Zon.4 100
Zon.5 100
CoL.1 Indicator zone color GRN/سور/YEL GRN
CoL.2 YEL
CoL.3 سور
CoL.4 سور
CoL.D Basic indication color outside color

زونونه

GRN/سور/YEL GRN
Br.R Brightness of red* ۱۶…۲۷ 100
Br.G Brightness of green* ۱۶…۲۷ 100
Br.Y Brightness of yellow* ۱۶…۲۷ 100
BL.YR Balance of red/green in yellow* ۱۶…۲۷ 100
 نوټ

* The parameter doesn’t change by resetting to factory settings.

د الارم ترتیبات

د رنګ نښه
You can set the indicator color modes depending on the input value using Zon.n and COL.n parameters. Zon. n parameters must be recorded sequentially from the lowest to the highest.HEMOMATIK -ITP17 -Universal-Process-Indicator (2)

Zon.1 = 50.0; Zon.2 = 80.0; Zon.3 = 100.0; CoL.1 = YELL; CoL.2 = rEd; CoL.d=Grn
Fig. 9 Changing of indication color

د الارم منطق
The output device can be used for control or alarm indication.
You can select the alarm logic with the Cnt parameter (see Table 6) in accordance with Fig. 10. HEMOMATIK -ITP17 -Universal-Process-Indicator (3)

د خدماتو مینو

Table 7 Service menu

اشاره تفصیل
DEV.t د وسیلې ډول
VEr.F د فرم ویئر نسخه
CJS.E Cold junction sensor on/ off
d.rSt د فابریکې تنظیماتو ته تنظیم کړئ:

اوسنی حالت: 0.

کله چې دا ټاکل کیږي 1, all device settings are reset to default values and the  device restarts

Configuration using akYtec Tool Pro
You can configure the device using akYtec Tool Pro software.
To connect the device to akYtec Tool Pro:

  1. Connect the device to a PC with the USB — micro USB cable.
  2. Start akYtec Tool Pro.
  3. وسیلې اضافه کړئ کلیک وکړئ.
  4. Select the COM port assigned to the device in the Interface drop-down menu of the  Network parameters tab. You can check the port number and name in Windows Device Manager.
  5. Select Modbus RTU in the Protocol drop-down menu.
  6. Select the necessary device in the Measuring devices category of the Devices drop- down menu.
  7. HEMOMATIK -ITP17 -Universal-Process-Indicator (4)If the device is connected for the first time, select Manually in the Connection setup tab and set the following values: HEMOMATIK -ITP17 -Universal-Process-Indicator (1)
  8. Select Find device.
  9. Enter the address of the connected device (default address — 16).
    نوټ
    The device is available under addresses from 1 up to 255.
  10. Click Search. The device with the address will be displayed in the window.
  11. Select the checkbox next to the device and click the ОК button.

To get more information about the connection and operation of the device, use the HELP menu of akYtec Tool Pro or press F1 to call up HELP in the program.

ساتنه

The safety requirements must be observed when the maintenance is carried out.

خبرداری
Cut off all power before maintenance.

The maintenance includes:

  • cleaning of the housing and terminal blocks from dust, dirt and debris
  • checking the device fastening
  • checking the wiring (connecting wires, terminal connections, absence of mechanical damages).

خبرتیا
The device should be cleaned with a dry or slightly damp cloth only. No abrasives or solvent-containing cleaners may be used.

ترانسپورت او ذخیره کول

Pack the device in such a way as to protect it reliably against impact for storage and  transportation. The original packaging provides optimum protection.  If the device is not taken immediately after delivery into operation, it must be carefully  stored at a protected location. The device should not be stored in an atmosphere with  chemically active substances.
The environmental conditions must be taken into account during transportation and storage.

خبرتیا
The device may have been damaged during transportation.
Check the device for transport damage and completeness!
Report the transport damage immediately to the shipper and akYtec GmbH!

د سپارلو ساحه

  • ITP17 universal process indicator 1 pc.
  • User guide 1 pc.
  • Set of mounting elements 1 pc.

نوټ
The manufacturer reserves the right to make additions to the scope of delivery.

تلیفون: +46 (0) 8 771 02 20
info@hemomatik.se

اسناد / سرچینې

HEMOMATIK ITP17 Universal Process Indicator [pdf] د کارونکي لارښود
ITP17, HM 2503, ITP17 Universal Process Indicator, ITP17, Universal Process Indicator, Process Indicator, Indicator

حوالې

یو نظر پریږدئ

ستاسو بریښنالیک پته به خپره نشي. اړین ساحې په نښه شوي *