Kangaroo™ Joey Enteral Feeding Pump

Normal operation "Feed Only Set” directions for use

Battery operation

Unplugging the pump from the AC line will automatically place the pump on battery power. To recharge the battery, plug the pump into an AC wall outlet. Pump will automatically begin charging.

General instructions

  1. Fill or spike feeding container prior to setting up pump.
  2. Turn pump on: Press Power in lower right-hand corner.
  3. In order to achieve proper accuracy, the fluid line in the feeding set bag must be 6 inches above the top of the feeding pump, when initiating the feeding cycle.
  4. Select Keep Settings or Clear Settings (for day to day use with the same feeding regimen the Keep Settings option should be selected).
  5. Load the Feeding Set per diagram. Do not over stretch the tubing.

Priming the pump

  1. Press Prime Pump to access the pump priming options.
  2. Press Auto Prime to automatically prime the pump set. The pump will quickly prime the feed line and stop the formula before reaching the end of the feeding line.
  3. Press the Hold To Prime Feed menu selection to manually top off the line.
  4. Press Done.

Setting the feed rate

  1. Select Adjust Feed then Feed Rate use the buttons on the left to program the pump from 1 to 400 in increments of 1ml. Select Enter when desired rate is set.

Note the Volume To Be Delivered VTBD Rate – is an Optional Feature.

Re-priming the pump after the feed bag empties

  1. A pump set bag that has been emptied will trigger the Feed Error screen. In this condition the pump set bag can be refilled to continue the feeding, but only after the pump set has been re-primed.
  2. Disconnect the feeding line from the patient.
  3. Refill the bag.
  4. Press Continue to begin the pump running.
  5. Press Hold, then press Adjust Settings, then press Prime Pump.
  6. Press Hold To Prime to prime the line. (Do Not Use Auto Prime)
  7. Press Done, then select Run.

To change rate or clear volume

  1. Select Hold.
  2. Select Clear Vol Fed to clear the volume.
  3. Select Adjust Settings to adjust all settings.
  4. Select Run to return to normal operations.

Pump sets should not be reused after 24 hours of initial usage.

Troubleshooting guide

Cardinal Health™ Kangaroo™ Joey Enteral Feeding Pump

Status LEDs:
  • Red: Error
  • Yellow: Hold/Pause
  • Green: Normal operation
Power button Programming and menu selection buttons

Listed below are some of the probable causes of alarm conditions and their corrections.

Symptom Probable cause Corrections
Hold error Pump in Holding mode for more than 10 minutes. Press Continue to return to Holding screen, or press and hold the Power button to turn off. Several options are available from the Holding screen, such as Run or Adjust Settings.
Flow error Occlusion in pump-patient line. (downstream occlusion) Check for and correct the occlusion in the line between pump and patient. Replace Pump Set if error cannot be resolved.
Feed error Empty bag or occlusion in bag- pump line. (upstream occlusion) Check for empty bag and refill, or check for and correct the occlusion in the line between the pump and feed bag. Replace Pump Set if error cannot be resolved.
Flush error Empty bag or occlusion in flush line. Check for empty bag and refill, or check for and correct the occlusion in the line between the pump and feed bag. Replace Pump Set if error cannot be resolved.
Flow error use > 24 Hrs Pump Set in use beyond recommended length of time. Informational warning message blinks on upper left of screen. It is recommended that the Pump Set be replaced.
Pump set dislodged Pump Set not properly loaded to pump. Check black ring retainer (MISTIC) and valve; reload Pump Set and restart pump to continue. Replace Pump Set if error cannot be resolved.
Battery low Battery charge too low. Immediately plug pump into A/C outlet to recharge battery and to provide power for pump.
Rotor error Pump Set tubing not properly loaded on rotor, or some other unusual rotor operating condition. Check that the tubing is loaded on the rotor and that it is not damaged, torn, etc. Reload tubing or replace Pump Set.
System error General error caused by many factors. See Manual for error list. Power down pump and attempt restart to clear error. Call customer service with screen code if error cannot be resolved.

For more information, please contact your sales representative, call 800.964.5227 or visit cardinalhealth.com/kangaroo

© 2020 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO, ESSENTIAL TO CARE and KANGAROO are trademarks of Cardinal Health and may be registered in the US and/or in other countries. All other marks are the property of their respective owners. Lit. No. 2MS20-1195642 (07/2020)

cardinalhealth.com

Models: Joey Enteral Feeding Pump, Joey, Feeding Pump, Joey Feeding Pump, Enteral Feeding Pump, Pump

File Info : application/pdf, 2 Pages, 170.90KB

PDF preview unavailable. Download the PDF instead.

kangaroo-joey-normal-operation-instructions

References

PDFium PDFium

Related Documents

Preview Kangaroo Joey Enteral Feeding Pump Troubleshooting Alerts
A guide to understanding and resolving alerts and error messages on the Kangaroo Joey Enteral Feeding Pump, including troubleshooting tips for common issues like hold errors, flow errors, feed errors, pump set dislodged, battery low, and rotor errors.
Preview Kangaroo Joey Enteral Pump: Quick Reference Guide for Loading and Rinsing
A quick reference guide detailing the correct procedures for loading, unloading, and rinsing feeding sets with the Kangaroo Joey enteral pump. Essential for healthcare professionals and caregivers.
Preview Kangaroo Joey Enteral Feeding Pump User Guide
User guide for the Kangaroo Joey Enteral Feeding Pump, covering feed and hydration mode, pump set loading, priming, programming, and alarm information.
Preview Kendall SCD 700 Series Controller Error Key and Troubleshooting Guide
A comprehensive guide to understanding and resolving error codes for the Kendall SCD 700 Series controller, including descriptions, error types, and troubleshooting steps.
Preview Cardinal Health: Supply Disruption Notification for Dover Urine Drainage Bag 4000ML
Notification from Cardinal Health regarding a supply disruption for the Dover™ Urine Drainage Bag 4000ML (8887600909) due to a raw material shortage. Provides an alternative product and contact information for assistance.
Preview Cardinal Health Announces Product Transition for Breathable Underpads
Cardinal Health is discontinuing item #988B12 Breathable Underpad and introducing item #988C72 as a suggested substitute, effective December 2025. This notice provides details on the product transition and ordering.
Preview Cardinal Health Strep A Dipstick Rapid Test Instructions for Use
Instructions for use and performance characteristics of the Cardinal Health Strep A Dipstick Rapid Test, a qualitative immunochromatographic assay for the detection of group A streptococcal antigen from throat swab specimens.
Preview Cardinal Health Premier Plus -30°C Laboratory Freezers: Installation and Operation Manual
This manual provides comprehensive installation and operation instructions for Cardinal Health Premier Plus -30°C Laboratory Freezers, including model specifications, safety precautions, setup, maintenance, and troubleshooting.