Error Key for Kendall SCD™ 700 Series Controller

When the Kendall SCD™ 700 series controller detects a fault condition, it:

Once you confirm the cause of the error and remedy the problem, you can reset the error. Some errors will remain active until you turn off the controller or the battery runs out of charge (if operating on battery power). If a Garment Mismatch error is triggered, the user may remedy the problem by pressing the corresponding A and B Button(s).

Kendall SCD™ 700 Series Controller Error Types

The Kendall SCD™ 700 series controller has four error types:

Error Code Details

Error code Error type Description Trouble shooting
Garment Mismatch Error User Resettable The garment detection procedure has detected a garment configuration (leg or foot flashing green) that does not match the user-selected configuration (leg or foot red). Press the port configuration button(s) to turn the foot selection on/off depending on what type of garment is connected to the controller. If the proper garment is selected and the problem persists, have the controller serviced by a professional.
High Pressure (Leg Sleeves) E2 User Resettable Leg sleeve pressure is greater than 47 mmHg for 10 consecutive cycles or pressure is above 65 mmHg for 5 consecutive cycles. Check for a tight leg sleeve and adjust fit appropriately. Also check for a partially occluded tube.
High Pressure (Foot Cuffs) E2 User Resettable Foot cuff pressure is greater than 135 mmHg for 10 consecutive cycles or pressure is above 160 mmHg for 5 consecutive cycles. Check for a tight foot cuff and adjust fit appropriately. Also check for a partially occluded tube.
Low Pressure (Leg Sleeves) E3 User Resettable Leg sleeve pressure is less than 43 mmHg for 10 consecutive cycles. Check for leaks in the sleeve or the tube connections.
Low Pressure (Foot Cuffs) E3 User Resettable Foot cuff pressure is less than 125 mmHg after 10 consecutive cycles. Check for leaks in the cuff or the tube connections.
Low Pressure (Leg Sleeves) E4 User Resettable Leg sleeve pressure is not between 35-55 mmHg for 12 consecutive cycles. Check for leaks in the sleeve or the tube connections.
Low Pressure (Foot Cuffs) E4 User Resettable Foot Cuff pressure is not between 110 and 150 mmHg for 12 consecutive cycles. Check for leaks in the cuff or the tube connections.
Vent Error E8 User Resettable The pressure in the garment is greater than 20 mmHg at the end of any vent period. Check tubing for kink or occlusion. Check garment application (too loose or tight). Service technician only: Check for kinked internal tubing.
Tubing Disconnect Error E12 User Resettable Pressure measured in the inflatable garment is below 10 mmHg for 10 consecutive cycles or no garments are detected during startup. Check for disconnected tube sets or garments and reconnect.
Temperature Error E9 Manual Reset Required If the internal case temperature of the controller drops below 5°C (41°F) or exceeds 55°C (131°F). High temperature: Make sure the controller is not covered by bedding and that the fan port, located near the power cord, is not obstructed. Low temperature: Allow the system to warm to room temperature.
System High Pressure Error E1 Manual Reset Required System pressure has exceeded 90 mmHg (leg sleeve) or 180 mmHg (foot cuff). Check for kinked tubes or patient interference with the garments, like pressing foot againt foot board.
Valve Feedback Error E5 Service Required If a valve electrically malfunctions, this error will be displayed. Service technician only: Verify that the valve assembly wires are properly connected and confirm solenoid actuation.
Software Error E6 Service Required Upon startup and periodically during operation the microprocessor performs diagnostic tests. If a software error is detected, this error will be triggered. Return to Cardinal Health for service.
Compressor Error E7 Service Required If the compressor electrically malfunctions, this error will be displayed. Service technician only: Verify that the compressor wires are properly connected.
Battery Error E10 Service Required Safe battery operation of the controller cannot be ensured. Service Technician only: Ensure that an unauthorized battery pack has not been used with the controller. Replace pack or return to Cardinal Health for service.
Pressure Transducer Error E13 Service Required The system could not sense a pressure rise of more than 5 mmHg during an inflation cycle or during startup. Service technician only: Check the transducer tube inside the controller to ensure it is not kinked or disconnected.
Low Battery Error Recharge Battery Less than 15 minutes of battery charge remains. The pumps and valves will continue to operate as long as there is enough power. Plug the controller into an AC power outlet.

Further Assistance

Consult the Operation and Service Manual as your warranty may require certain repairs to be performed by an authorized Cardinal Health service technician.

PDF preview unavailable. Download the PDF instead.

kendall-scd-700-series-controller-error-key-guide Adobe PDF Library 15.0

Related Documents

Preview Kangaroo Joey Enteral Feeding Pump Troubleshooting Alerts
A guide to understanding and resolving alerts and error messages on the Kangaroo Joey Enteral Feeding Pump, including troubleshooting tips for common issues like hold errors, flow errors, feed errors, pump set dislodged, battery low, and rotor errors.
Preview Kangaroo Joey Enteral Feeding Pump: Normal Operation and Troubleshooting
A guide to the normal operation and troubleshooting of the Kangaroo Joey Enteral Feeding Pump, including setup, priming, setting feed rates, and resolving common errors.
Preview Cardinal Health Announces Product Transition for Breathable Underpads
Cardinal Health is discontinuing item #988B12 Breathable Underpad and introducing item #988C72 as a suggested substitute, effective December 2025. This notice provides details on the product transition and ordering.
Preview Cardinal Health Home Healthcare Product Guide: Solutions for Daily Living and Patient Care
Discover the Cardinal Health Home Healthcare Product Guide, featuring a comprehensive catalog of medical equipment and supplies for daily living, mobility, diabetes care, and more. Find solutions from trusted brands for patients and caregivers.
Preview Kangaroo Joey Enteral Pump: Quick Reference Guide for Loading and Rinsing
A quick reference guide detailing the correct procedures for loading, unloading, and rinsing feeding sets with the Kangaroo Joey enteral pump. Essential for healthcare professionals and caregivers.
Preview Cardinal Health: Supply Disruption Notification for Dover Urine Drainage Bag 4000ML
Notification from Cardinal Health regarding a supply disruption for the Dover™ Urine Drainage Bag 4000ML (8887600909) due to a raw material shortage. Provides an alternative product and contact information for assistance.
Preview Cardinal Health Clinical Digital Thermometers
Cardinal Health offers a comprehensive range of clinical digital thermometers designed for accuracy and speed in healthcare settings. This product catalog details standard and quick-read oral/rectal thermometers, probe covers, and forehead temperature indicators, including product specifications, model numbers, and packaging options.
Preview Cardinal Health Strep A Dipstick Rapid Test Instructions for Use
Instructions for use and performance characteristics of the Cardinal Health Strep A Dipstick Rapid Test, a qualitative immunochromatographic assay for the detection of group A streptococcal antigen from throat swab specimens.