Troubleshooting tips: What the alerts mean

If there is an issue with the Kangaroo™ Joey Enteral Feeding Pump, an alert will appear on the screen along with a red or yellow light.

Hold error

The pump has been in Holding mode for more than 10 minutes.

Press >Continue to return to Holding screen or press and hold the >Power button to turn off. Several options are available from the Holding screen, such as Run and Adjust Settings.

Description: The pump is in a paused state for an extended period.

Pump Set use >24 hours

The feeding set has been in use beyond the recommended length of time.

When this warning message blinks, it's recommended that the feeding set be replaced.

Description: The feeding set has exceeded its recommended usage time.

Flow error

There is a clog in the tubing between the pump and the patient.

Check the tubing. If the tube can't be cleared, load a new feeding set, prime it, and press >Continue to restart the feeding.

Description: An obstruction is detected in the patient-side tubing.

Pump Set dislodged

The feeding set is not properly loaded to the pump.

Open the door and check the areas where the pump set is loaded on the pump, reload the feeding set, and restart the pump to continue.

Description: The feeding set is not correctly inserted into the pump.

Feed error

There is a clog in the tubing between the bag and the pump, or the bag is empty.

First, see if the bag is empty and refill. Then, check for a clog. Check the valve pocket in the pump set loading area for dampness or dirt. If the error can't be resolved, load a new feeding set, prime it, and press >Continue to restart the feeding.

Description: The feeding bag is empty or there's a clog in the tubing from the bag to the pump.

Battery low

The battery is not charged enough to operate the pump.

Immediately plug the pump into an outlet to recharge battery and power the pump.

Description: The pump's battery charge is critically low.

Rotor error

The feeding set tubing is not properly loaded on the rotor or other unusual rotor operating condition.

Check that the tubing is loaded on the rotor and that it's not damaged, torn, etc. Reload tubing or replace the feeding set, prime it, and press >Continue to restart the feeding.

Description: An issue with the pump's rotor mechanism or feeding set loading.

System error

This is a general error that can be caused by many reasons. See your manual for the full error list.

Press and hold the >Power button for 3 seconds to power down the pump. Then, press and hold the >Power button to power up the pump. This should reboot the pump and clear the alarm.

Call your home healthcare provider or Cardinal Health customer service.

Description: A general error requiring potential reboot or further investigation.

If you're still having trouble

You can find more information, including a full list of alerts and system error codes in your Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp Operation and Service Manual that came with your pump or is available at www.cardinalhealth.com/kangaroo.

Contact your home healthcare provider or Cardinal Health customer service for any further questions you may have.

Models: 2MS19 Joey Enteral Feeding Pump, 2MS19, Joey Enteral Feeding Pump, Enteral Feeding Pump, Feeding Pump, Pump

File Info : application/pdf, 2 Pages, 134.59KB

PDF preview unavailable. Download the PDF instead.

kangaroo-joey-troubleshooting-guide

References

PDFium PDFium

Related Documents

PreviewKangaroo Joey Enteral Feeding Pump: Normal Operation and Troubleshooting
A guide to the normal operation and troubleshooting of the Kangaroo Joey Enteral Feeding Pump, including setup, priming, setting feed rates, and resolving common errors.
PreviewKangaroo Joey Enteral Feeding Pump User Guide
User guide for the Kangaroo Joey Enteral Feeding Pump, covering feed and hydration mode, pump set loading, priming, programming, and alarm information.
PreviewKangaroo Joey Enteral Pump: Quick Reference Guide for Loading and Rinsing
A quick reference guide detailing the correct procedures for loading, unloading, and rinsing feeding sets with the Kangaroo Joey enteral pump. Essential for healthcare professionals and caregivers.
PreviewKendall SCD 700 Series Controller Error Key and Troubleshooting Guide
A comprehensive guide to understanding and resolving error codes for the Kendall SCD 700 Series controller, including descriptions, error types, and troubleshooting steps.
PreviewCardinal Health: Supply Disruption Notification for Dover Urine Drainage Bag 4000ML
Notification from Cardinal Health regarding a supply disruption for the Dover™ Urine Drainage Bag 4000ML (8887600909) due to a raw material shortage. Provides an alternative product and contact information for assistance.
PreviewCardinal Health Announces Product Transition for Breathable Underpads
Cardinal Health is discontinuing item #988B12 Breathable Underpad and introducing item #988C72 as a suggested substitute, effective December 2025. This notice provides details on the product transition and ordering.
PreviewCardinal Health Strep A Dipstick Rapid Test Instructions for Use
Instructions for use and performance characteristics of the Cardinal Health Strep A Dipstick Rapid Test, a qualitative immunochromatographic assay for the detection of group A streptococcal antigen from throat swab specimens.
PreviewCardinal Health Premier Plus -30°C Laboratory Freezers: Installation and Operation Manual
This manual provides comprehensive installation and operation instructions for Cardinal Health Premier Plus -30°C Laboratory Freezers, including model specifications, safety precautions, setup, maintenance, and troubleshooting.