Cardinal Health™ Premier Plus -30°C Laboratory Freezers

Installation and Operation

REF CAX1230FA, CAX1230FD
CAX2330FA, CAX2330FD
CAX3030FA, CAX3030FD
CAX5030FA, CAX5030FD

Preface

IMPORTANT Read this instruction manual. Failure to follow the instructions in this manual can result in damage to the unit, injury to operating personnel, and poor equipment performance.

CAUTION All internal adjustments and maintenance must be performed by qualified service personnel. Material in this manual is for informational purposes only. The contents and the product it describes are subject to change without notice. Cardinal Health makes no representations or warranties with respect to this manual. In no event shall Cardinal Health be held liable for any damages, direct or incidental, arising from or related to the use of this manual.

For your future reference, please have the following information readily available. It can be found on the dataplate attached to your unit.

Model Number: _________________________
Serial Number: _________________________

The following information, if available, is helpful for contacting the factory.

Date Purchased: _________________________
Purchase Order Number: _________________________
Source of Purchase: _________________________ (manufacturer or specific agent/rep organization)

1. Model

The following table shows the units covered in this operation and installation manual by model number. Refer to Section 6.1 for details on the voltage specifications for the various models.

Table 1. Applicable Models Model (*)
Cardinal Health - CH1230F* A/D
Cardinal Health - CH2330F* A/D
Cardinal Health - CH3030F* A/D
Cardinal Health - CH5030F* A/D

2. Safety Precautions

In this manual, the following symbols and conventions are used:

The following are important safety precautions that apply to this product:

EMC

EMC Registration is done on this equipment for business use only. It may cause interference when the product would be used in home.

This equipment has been tested and found to comply with the limits for a Class A digital device. Class A covers devices for usage in all establishments other than domestic and that are not directly connected to a low voltage power supply network, which supplies domestic environment.

This ISM device complies with Canadian ICES-001.

FCC

This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

3. Unpacking

At the time of delivery, be sure to inspect the unit packaging for damage before signing for the shipment. If packaging damage is present, request immediate product inspection and file a claim with the carrier.

Note: Packaging damage does not denote that unit damage exists.

If concealed damage is found (damage that is not apparent until the item has been unpacked), stop further unpacking and save all packing for carrier's inspection. Make a written request for inspection to delivering carrier. This must be done within 15 days after delivery. Then file a claim with the carrier.

Do not return goods to the manufacturer without written authorization.

4. Packing List

Inside the freezer cabinet is a bag containing:

If the unit was ordered with shelves, the bag will also include:

If the unit was ordered with a chart recorder, the bag will also include:

If specified on the order, the bag may also include:

Other items with your unit include:

5. General Recommendations

This section includes some general recommendations for your unit.

5.1 Temperature Monitoring

IMPORTANT NOTE: We recommend the use of a redundant and independent temperature monitoring system so that the device can be monitored continuously for performance commensurate with the value of the product stored. Please use the probe in glycerin bottle (if present) as the temperature reading reference point for all calibration / validation procedures.

5.2 Intended Use

The -30°C Laboratory Freezers described in this manual are high performance units for professional use. These products are intended for use as cold storage in research use and as a general purpose laboratory freezer, storing samples or inventory at operating temperatures between -35°C and -15°C.

Expected users of this equipment include but are not limited to personnel from the following areas: Professional and clinical laboratories, Pharma and Biotech facilities, Academic, Industrial, and Government facilities or those trained in laboratory protocols put in place at your facility. The units are not for use by the general public.

It is not considered a medical device and has therefore not been registered with a medical device regulatory body (e.g. FDA): that is, it has not been evaluated for the storage of samples for diagnostic use or for samples to be re-introduced to the body.

This unit is not intended for use in classified hazardous locations, nor to be used for the storage of flammable or corrosive inventory.

Storage of unsealed corrosive substances may cause the interior of the unit to corrode.

6. Operating Environmental Conditions

The freezers described in this manual are classified for use as stationary equipment in a Pollution Degree 2 and Overvoltage Category II environment.

These units are designed to operate under the following environmental conditions:

Operation outside of these conditions will affect the performance of the unit and the samples stored inside.

6.1 Unit Specifications

The last character in the model number listed on the freezer identifies the electrical specifications for your unit. The dataplate is located on the upper left side of the unit.

Table 2. Unit Specifications Rated Voltage Rated Current Frequency / Phase Power Module Plug Approximate Weight Kg (lbs) Exterior Dimensions (D x W x H)
12A 115 V 7.2 A 60 Hz/1 128 (281) 79.0 x 61.9 x 185.4 cm (31.1 x 24.0 x 73.0 in)
12D 208-230 V 3.6 A 60 Hz/1
23A 115 V 10.0 A 60 Hz/1 162 (358) 96.2 x 71.1 x 199.4 cm (37.9 x 28.0 x 78.5 in)
23D 208-230 V 7.4 A 50 Hz/60Hz/1 IEC C19 153 (337)
30A 115 V 10.0 A 60 Hz/1 168 (371) 96.2 x 86.4 x 199.4 cm (37.9 x 34.0 x 78.5 in)
30D 208-230 V 7.4 A 60 Hz/1
50A 115 V 16 A 60 Hz/1 243 (535) 96.2 x 143.5 x 199.4 cm (37.9 x 56.5 x 78.5 in)
50D 208-230 V 9.6 A 60 Hz/1

7. Installation

WARNING: Risk of Shock. Do not exceed the electrical rating printed on the dataplate located on the upper left side of the unit.

Do not move the unit using the drain pan on the back. This could cause damage to the equipment.

7.1 Location

Install the unit on a level area free from vibration with a minimum of 6 inches of space on the sides and rear and 12 inches at the top. Do not position the equipment in direct sunlight or near heating diffusers, radiators or other sources of heat.

WARNING: Risk of Injury. Do not move the unit while loaded. Move by pushing slowly at handle level or lower. Use caution on uneven surfaces.

7.1.1 Installation Instructions

The unit must be level both front to back and side to side when installed. If the unit is out of level, you may need to shim the corners or casters with thin sheets of metal. Be sure to set the brakes for units equipped with casters.

An unlevel unit may result in instability and performance issues for the doors and drawers.

7.2 Wiring

The wiring diagrams are attached on the back of the cabinet.

CAUTION: Connect the equipment to the correct power source. Incorrect voltage can result in severe damage to the equipment.

CAUTION: Risk of Shock. For personal safety and trouble-free operation, this unit must be properly grounded while in use. Failure to ground the equipment may cause personal injury or damage to the equipment. Always conform to the National Electrical Code and local codes. Do not connect the unit to overloaded power lines.

Do not position the unit in a way that impedes access to the disconnecting device or circuit breaker in the back of the unit.

Always connect the unit to a dedicated (separate) circuit. Each unit is equipped with a service cord and plug designed to connect it to a power outlet which delivers the correct voltage. Supply voltage must be within ±10% of the unit rated voltage. If cord becomes damaged, replace with a properly rated power supply cord.

Table 3. Power Cord Specifications Power Cord Specifications
12/23/30 A 3-G 12 AWG, NEMA 5-15P, 15A/125V
50 A 3-G 12 AWG, NEMA 5-20P, 20A/125V
D 3-G 14 AWG, NEMA 6-15P, 15A/250V

CAUTION: Risk of Shock. Never cut the grounding prong from the service cord plug. If the prong is removed, the warranty is invalidated.

In an emergency, the power cord is a disconnect device.

7.3 Shelves

Single door laboratory freezers come standard with 4 shelves and double door laboratory freezers come standard with 8 shelves.

Maximum shelf capacity is 45 kg (100 lbs).

For safety in shipping, the shelves are packaged and secured inside the cabinet. Insert the shelf support hangers (included inside the unit) into the built-in shelf supports (located on the inside walls of the cabinet interior) at the desired locations. Position the shelves on the flat supports (refer to Figure 1). Do not position shelving to disrupt the thermal bottle(s).

Figure 1. Shelf Support Hanger

7.4 Drawers

Single door laboratory freezers have the option to be ordered with 7 drawers and double door laboratory freezers have the option to be ordered with 14 drawers.

Drawers are shipped in a factory installed position that promotes uniform spacing between the drawers. The drawer slides are adjustable to higher and lower positions in the cabinet. You can position these slides in the vertical slots which are spaced at one-inch intervals. The drawers must be removed in order to adjust the position of the drawer slides.

For CH1230, CH2330 and CH5030, the maximum drawer capacity is 27 kg (60 lbs). For CH3030, the maximum drawer capacity is 36 kg (80 lbs).

CAUTION: Risk of Abrasion. Hidden sharp edges may be present on drawers. Use appropriate Personal Protective Equipment (such as gloves) while handling the drawers to avoid possible sharp points related injury.

7.4.1 Removing the Drawers

To remove the drawers, complete the following steps (refer to Figure 2):

  1. Pull the drawer towards you until the slides are fully extended.
  2. Lift the back of the drawer to disengage the mounting tabs from the slots on the slides.

Note: The drawers fit snugly between the slides. Push the back of the drawer from underneath to remove the drawer.

7.4.2 Reinstalling the Drawers

To reinstall the drawers, complete the following steps (refer to Figure 2):

  1. Pull both the slides of the drawer towards you until the slides are fully extended.
  2. Position the drawer between the slides and with the back facing end raised at 45 degree angle, insert the mounting clips at the front into the slots on front of the slides.
  3. Push the back of the drawer down between the slides and insert the basket tabs into the slots at the back.

Note: The drawers fit snugly between the slides. Push on the back of the drawer from the inside to insert the drawer tabs completely into the slots. Make sure both the drawer tabs are aligned with the slots on the slides before pushing the drawer down between the slides.

CAUTION: Be careful when reinstalling the drawers to avoid possible pinching.

7.4.3 Changing Drawer Position

Drawer slides have a small wire safety clip at the front pilaster which prevent the slides from falling when the drawer is removed. To change the position of the drawer slides, complete the following steps:

  1. Locate the safety clip.
  2. Slip a small screwdriver under the bottom of the wire clip and pry the clip towards the inside of the freezer.

Figure 3. Clip Removal

3. Lift up the slide at the front. The slide is free to move from the front pilaster.

Figure 4. Front of slide Removal

4. The drawer slide must be removed from the rear pilaster at approximately a 45 degree angle towards the center of the cabinet.

Figure 5. Slide Rotation

5. Pull the slide towards the front of the cabinet.

6. Determine desired location for the slide and insert the slide into the rear pilaster at a 45 degree angle towards the center of the cabinet.

7. Once the slide is seated in the rear pilaster, rotate the slide until it is parallel with the unit wall and insert slide into front pilaster.

Figure 6. Slide Installation

CAUTION: Ensure the slide is level. If the slide is not seated in a level orientation, the drawers/baskets will not function as intended. This could cause injury or product damage.

8. Replace the safety clip by hooking the clip under the front pilaster slide bracket. Then rotate the wire safety clip to position the other bend of the wire safety clip over the top of the same bracket.

Figure 7. Clip Replacement

CAUTION: Drawer slides do not require lubrication. Additional lubricant could impede movement of the drawers when lubricant is cold.

7.5 Baskets

Baskets can be ordered as an accessory for this unit.

The basket slides are adjustable to higher and lower positions in the cabinet. You can position these slides in the vertical slots which are spaced at one-inch intervals. The baskets must be removed in order to adjust the position of the slides. The procedure for removing, reinstalling and changing position of the baskets can be found in Section 7.4.1, Section 7.4.2 and Section 7.4.3.

The maximum basket capacity is 27 kg (60 lbs).

7.6 Temperature Sensor Bottles

The temperature displayed on the control panel is measured by a probe inserted in sensor bottle inside the cabinet. The sensor probe must be inserted in the glycerin bottle prior to operation.

The sensor bottle is positioned on the left wall towards the top of the unit.

To install the sensor(s) into the bottles, complete the following steps:

  1. Remove the bottle from the bracket and verify that each bottle is full of liquid.
  2. Remove the solid cap from the bottle and save for the next time the unit needs to be moved.
  3. Remove the taped lid and temperature sensor(s) from the wall.
  4. Insert the temperature probe(s) into the bottle.
  5. Secure the lid to the bottle and return the bottle to the bracket.

7.7 Door Operation

The doors on all units are designed to stay open if opened 90 degrees or more. The door spring tension cannot be adjusted.

If the self-closing door does not work properly, make sure the unit is level.

Door seal integrity is critical for unit's performance. Never route anything through the door seal. A loose fitting gasket allows moist air to be drawn into the cabinet, resulting in quicker frost buildup on the evaporator coil, longer running time, poor temperature maintenance, and increased operation cost.

CAUTION: Keep hands and body parts clear of closing doors. The moving parts create a potential pinch point.

WARNING: Risk of Fire. There are port holes in the walls of the cabinet to help facilitate routing of tubing, independent temperature sensors, and other non-electrical items. The ports are not intended to supply power to devices inside the device. Be sure to seal the holes after routing items to prevent undesired air exchange.

7.8 Remote Alarm (Optional)

All units have factory-installed local alarm contacts that can be used for remote alarm systems. The maximum distance between a freezer and a remote alarm depends on the wire gauge used. Refer to Table 4 below:

Table 4. Wire Gauges and Distance to Remote Alarm Total Wire Length (feet) Distance to Alarm 1/2 Wire Length (feet)
20 530 265
18 840 420
16 1330 665
14 2120 1060
12 3370 1685

Remote alarm terminals are located at the rear of the machine compartment.

The three terminals are: COMMON, OPEN ON FAIL (Normally Closed), and CLOSE ON FAIL (Normally Open).

REMOTE ALARM CONNECTIONS

CONNEXIONS DE L'ALARME À DISTANCE

CONTACT RATING 1 AMP MAX
CLASS 2 CIRCUIT ONLY

VALEUR NOMINALE DE CONTACT DE 1 AMP MAX.
SUR UN CIRCUIT DE CLASSE 2 UNIQUEMENT

WIRE COLOR
(FIL DE COULEUR)

FUNCTION
(FONCTION)

PURPLE (VIOLET)

COMMON (COMMUNE)

BLACK (NOIR)

OPEN ON FAIL
(EN POSITION OUVERTE)

RED/WHITE
(ROUGE/BLANC)

CLOSE ON FAIL
(EN POSITION FERMÉE)

Figure 8. Remote Alarm Diagram

To install the remote alarm, make the following connections:

  1. Connect the COMMON terminal on the cabinet switch to the COMMON wire on the alarm.
  2. 2a. To get an alarm when the switch contacts open, connect the OPEN ON FAIL terminal on the cabinet to the OPEN ON FAIL wire on the alarm.
  3. 2b. To get an alarm when the switch contacts close, connect the CLOSE ON FAIL terminal on the cabinet to the CLOSE ON FAIL wire on the alarm. The COMMON and CLOSE ON FAIL wires must be tied together in this application.
  4. 3. Plug the alarm system service cord into an electrical outlet.

The contacts will trip in the event of a power outage, high temperature alarm or low temperature alarm.

7.9 Final Checks

Before start up, be sure to complete the following steps:

  1. Make sure that the unit is free of all wood or cardboard shipping materials, both inside and outside.
  2. Check the positions of the shelves, drawers and baskets. If you want to adjust the positions, see instructions in Section 7.3, Section 7.4 and Section 7.5.
  3. Verify that the unit is connected to a dedicated circuit.

8. Startup

8.1 Initial Startup

To start up the freezer, complete the following steps:

  1. Verify that the sensor probe(s) have been installed in the bottle(s) as described in Section 7.6.
  2. Ensure the double pole circuit breaker switch located next to the power inlet is in “ON” position (i.e. “I” position).
  3. Insert the key in the switch and turn to the Power On position. The display will show the actual cabinet temperature and the compressor and evaporator should start within approximately 10 minutes.
  4. Allow the unit to reach operating temperature before loading it with any product. To stabilize the temperature profile, a 24-hour waiting period is recommended.
  5. If you desire to enable the alarms, turn the three position key switch one turn further clockwise to the Alarm On position. To avoid nuisance alarms, wait until the unit has cooled down to the desired operating temperature (-30°C default).
  6. If you have a remote alarm, hook it up at this point (refer to Section 7.8).
  7. If desired, lock the cabinet door using the key. Place duplicate key copies in a safe place.

All controls should now be fully operational, the alarm active (if enabled), and all visual indicators active.

CAUTION: Use gloves when handling samples to avoid potential frost bite.

8.2 Product Loading and Unloading Guidelines

When loading your freezer, take care to observe the following guidelines:

Figure 9. Load Limit Line

9. Operation

9.1 Control Panel

Figure 10. Freezer Control Panel

The control panel is located on the top right side of your freezer. You can use the three pushbuttons (#5, #8, and #9 in Figure 10) to change the temperature display (#1) or to adjust temperature and alarm setpoints, as given in Section 9.3. The thermometer display (#2) provides a quick visual indicator of current cabinet temperature and alarm conditions.

  1. Main temperature display - during normal operation, shows cabinet temperature in degrees Celsius, as measured by the primary sensor inside the cabinet. You can use the buttons to display other values such as setpoints and highest and lowest recorded temperatures. The number in the main display flashes when the value can be modified.
  2. Thermometer - shows cabinet temperature and alarm conditions. There are 10 horizontal bars: 9 are displayed during normal operation, the tenth (top) bar indicates a warm alarm condition. The number of bars illuminated indicates approximate cabinet temperature. With the default settings, 4 to 6 bars illuminated indicate that the cabinet is at desired setpoint. For example, suppose that the cabinet temperature setpoint is -30°C and that the cold and warm alarm setpoints are -35°C and -20°C. Then the number of bars illuminated indicates cabinet temperature as follows:
Table 5. Thermometer Display on Control Panel (Setpoint -30°C) Temperature (°C) Bars Displayed Temperature (°C)
Bulb Only -35 (cold alarm) 6 Bars -26
1 Bar -33.5 7 Bars -24.5
2 Bars -32 8 Bars -23
3 Bars -30.5 9 Bars -21.5
4 Bars -29 10 Bars -20 (warm alarm)
5 Bars -27.5

When cabinet temperature exceeds the warm alarm setpoint, the top bar of the thermometer flashes. When temperature is lower than the cold alarm setpoint, the bulb flashes. When you are in programming mode (described in Table 8) the thermometer shows the setpoint value you are changing.

  1. Power failure - illuminates when the main power supply is interrupted. In this case the audible alarm also sounds.
  2. Service mode - illuminates when the controller is in service programming mode.
  3. Increase - pushbutton used to increase setpoint values in programming mode and for various display functions.
  4. Door ajar - illuminates when the freezer door is open longer than the duration specified in the service parameters table; default is approximately 3 minutes (when the alarm is activated and the key switch is turned to the alarm position).
  5. Battery low - illuminates when the backup battery is low (refer to Section 12.6).
  6. Decrease - pushbutton used to decrease setpoint values in programming mode and for various display functions.
  7. Scan - pushbutton used to change the main display and for various other functions.
  8. Audible alarm - illuminates during warm and cold alarm conditions.
  9. Key Switch - Switch used to turn the power and alarms on and off for the unit.
  10. Alarm On - Setpoint Security - when the key switch is in the alarm on position, the alarms are on and the setpoints cannot be modified.
  11. Power On - when the key switch is in the on position, the unit power is on with no alarms active.
  12. Off - when the key switch is in the off position, the unit is off.

For full descriptions of display, programming and service functions, refer to Tables 6, 7 and 8.

9.2 Display Functions

Function Meaning Sequence Display
Normal operation Default display while freezer is running. Temperature display and control panel thermometer icon show cabinet temperature.
Coldest logged temperature Show coldest cabinet temperature since last startup or reset. Press Display shows coldest logged temperature while button is pressed.
Warmest logged temperature Show warmest cabinet temperature since last startup or reset. Press Display shows warmest logged temperature while button is pressed.
Mute Silence audible alarm for approximately 6 minutes. Press SCAN Display and thermometer show cabinet temperature, alarm icon continues to flash.
Reset Return to default display after alarm condition, clears temperature log. Press and hold and simultaneously. Excursion values are reset; temperature display shows cabinet temperature.
Alarm Test Test by simulating warm alarm. Key switch must be in alarm mode. Press , and SCAN simultaneously, hold for approximately 5 seconds First, the display will show "AtSt" to show entry of the test. Display and thermometer show simulated cabinet temperatures (warm), alarms flash and sound as appropriate. Alarms clear when test is completed.

9.3 Programming Functions

You can enter the programming mode by pressing the Scan button (SCAN) and holding for approximately 5 seconds. The display will then flash "Prg" to indicate that you have entered the programming mode. Use and buttons to modify the values and press the Scan (SCAN) button to save the value and go to the next screen.

Note: If the alarms are enabled, all parameters are read only and cannot be edited.

Screen/Button Function Summary
1 Initial Screen Unit Setpoint The temperature that the unit is set to run
2 Press Scan (SCAN) Cold Alarm Temperature The temperature that will actuate the cold alarm (if active)
3 Press Scan (SCAN) Warm Alarm Temperature The temperature that will actuate the warm alarm (if active)
4 Press Scan (SCAN) Exit Program Mode Returns to normal operating screen

If the unit is left idle for a period of approximately 30 seconds during program mode, the program mode shall be exited.

If at any point the scan button is pressed and held for a period of approximately 10 seconds, the unit shall enter the service mode.

9.4 Service Parameters

You can enter the service mode by pressing the Scan button (SCAN) and holding for approximately 10 seconds while in Programming mode (refer to Section 9.3). The display will then flash “SEr” to indicate that you have entered the service mode followed by the software checksum values flashing on the screen. The service icon () will also illuminate.

Use up () and down () buttons to modify the values and press the scan (SCAN) button to save the value and go to the next screen.

Note: If the alarms are enabled, the unit will not enter Service Mode.

Resetting any of the following parameter values could adversely affect the performance of your unit. Be sure to understand your product requirements before making any adjustment to the service parameter values. These settings very rarely need to be changed from normal operation. Call technical support if you have any questions prior to making any adjustments to service parameter values.

Screen/Button Display Function Summary
1 Initial Screen SEr -30 Temp of the unit Allows the user to see what temperature the unit is designed to run.
Varies CPU Checksum Shows the current checksum of the CPU board for validation.
Varies Relay Board Checksum Shows the current checksum of the Relay board for validation.
2 Press Scan (SCAN) typ Model Type Unit type 30 for -30 Freezer. This parameter should never be changed.
3 Press Scan (SCAN) CFg Type of unit This parameter should only be changed by trained service personnel.
4 Press Scan (SCAN) CuFt Size of the unit Setting for size of the unit.
5 Press Scan (SCAN) Pd oFSt Control Probe Pulldown Offset The difference in temperature between the control probe value and the average compartment temperature that determines when the control system changes from pull down mode to steady state operation. This can be useful to adjust if the average compartment temperature has an undershoot or overshoot relative to setpoint during pull down mode. This offset is also used on some freezer models as a response to door openings.
6 Press Scan (SCAN) Cnt oFSt Control Probe Offset The difference in temperature between the control probe value and the average compartment temperature that is used to control the cooling system to the unit setpoint during steady state operation. This can be useful to adjust if there is a discrepancy between the unit setpoint and the average cabinet temperature.
7 Press Scan (SCAN) dIS OFSt Display Probe Offset The difference in temperature between the display probe value seen on the User Interface display and the average compartment temperature at the unit setpoint during steady state operation. This can be useful to adjust the display during unit calibration procedures.
8 Press Scan (SCAN) Ctr CAI Display Control Probe with Offset (If present) Normally OFF (00). If ON (01), the Upper Bottle Temperature output on main display is replaced by the control probe temperature value with offset when exiting the service menu.
9 Press Scan (SCAN) Cnt ucl Control Upper Hysteresis The degree increase in temperature from the setpoint that will trigger the unit to begin cooling to prevent the unit from getting too warm.
10 Press Scan (SCAN) Cnt Icl Control Lower Hysteresis The degree decrease in temperature from the setpoint that will trigger the unit to stop cooling to prevent the unit from getting too cold.
11 Press Scan (SCAN) door AJAr Door Alarm Time The approximate time the door can remain open before the door ajar alarm activates in minutes.
12 Press Scan (SCAN) AdP dEF Adaptive Defrost Setting for adaptive defrost ON (01) or OFF (00).
13 Press Scan (SCAN) dEF Int Defrost Interval The time interval between defrost cycles in hours.
14 Press Scan (SCAN) dEF dur Defrost Duration The maximum amount of time the defrost cycle will run in minutes. Value will be "nA" for units with hot gas defrost.
15 Press Scan (SCAN) dEF Hi Defrost Temp Out The maximum temperature that the defrost probe can reach before stopping the defrost cycle.
16 Press Scan (SCAN) PAn Hi Drip Pan Temp Out Maximum temperature drip pan probe can reach before stopping the defrost drip time.
17 Press Scan (SCAN) qUA Quality Mode Normally OFF(00). Used for production line testing.
18 Press Scan (SCAN) bot BOT mode Normally OFF(00). If ON(01), the unit is forced to run at coldest possible temperature for approximately 24 hours. Defrosts will still occur at scheduled.
19 Press Scan (SCAN) PEr Perimeter heater Duty Cycle Controls the perimeter heater ON time.
20 Press Scan (SCAN) Cnt prb Control Probe Temperature Displays control probe temperature (Read Only).
21 Press Scan (SCAN) Anb prb Ambient Probe Temperature Displays ambient probe temperature in deck (Read Only).
22 Press Scan (SCAN) dEF Prb Defrost Probe Temperature Displays defrost probe temperature (Read Only).
23 Press Scan (SCAN) PAn Prb Defrost Drip Pan Probe Temperature Displays defrost drip pan probe temperature.
-- Enter Defrost cycle Pressing the scan button after displaying the defrost Drip Pan probe temperature while not in an alarm mode will force the unit into a defrost cycle. To avoid defrost cycle, leave unit screen idle for approximately 30 seconds.

9.5 Temperature Settings

The factory default temperature setting is -30°C for laboratory freezers. To change the factory temperature settings, refer to the instructions in Section 9.3.

The freezer described in this manual is designed for optimum performance at -30°C. It is advisable to call Technical Service before changing setpoints.

9.6 Alarms

The alarm system is designed to provide visual and audible warning signals for both power failure and rise in temperature. The alarm is equipped with a battery backup.

Default cold and warm alarm values are -35°C and -20°C. These values may be adjusted, following instructions in Section 9.3.

The alarm system is activated only when the key switch is turned to the Alarm On position. The audible warning signal sounds when there is a power failure, temperature alarm condition, or when the door is ajar for approximately 3 minutes with default door alarm setting.

The Mute function (pressing the SCAN button) allows you to turn off the audio warning without turning off the visual indicators. The alarm will ring back after approximately 6 minutes if the alarm is still active.

During a warm alarm, a flashing of the uppermost bar of the thermometer, the temperature display, and the speaker icon will occur at the same time. If there has been a warm alarm since the last alarm reset, but the temperature is not currently in an alarm state, the uppermost bar of the thermometer and the speaker icon will slowly alternate in flashing.

During a cold alarm, a flashing of the bulb of the thermometer, the temperature display, and the speaker icon will occur at the same time. If there has been a cold alarm since the last alarm reset, but the temperature is not currently in an alarm state, the bulb of the thermometer and the speaker icon will slowly alternate in flashing.

During a power failure, the power failure icon will illuminate, the thermometer will display without the bulb, and the temperature will flash approximately every 3 seconds. If there has been a power failure since the last alarm reset, the power failure icon and the speaker icon will slowly alternate in flashing.

During a low battery condition, the low battery icon will illuminate.

The alarms can be reset by pressing and holding the and simultaneously.

10. Chart Recorder (Optional)

Panel-mounted six inch recorders are available as a factory installed option. Recorder operation begins when the system is powered on.

10.1 Set Up and Operation

To prepare the recorder to function properly, complete the following steps:

  1. Open the recorder door to access the recorder.
  2. Connect the nine volt DC battery located at the recorder's upper right corner. This battery provides back-up power.
  3. Install clean chart paper (refer to Section 10.2 at the next page).
  4. Close the recorder door.

Note: The recorder may not respond until the system reaches temperatures within the recorder's range.

Figure 11. Chart Recorder

Label Description
1 Pressure Sensitive Chart Buttons
2 Imprinting Stylus
3 Reference Mark
4 Hub-Nut and Retaining Wire
5 Chart

10.2 Changing Chart Paper

To change the chart paper, complete the following steps:

  1. Locate the pressure sensitive buttons at the front, upper left of the recorder panel.
  2. Press and hold the change chart button (#3 in Figure 12) for approximately one second. The pen will move off the scale.
  3. Unscrew the center nut, remove the old chart paper, and install new chart paper. Carefully align the day and time with the reference mark on the recorder panel (a small groove on the left side of the panel, as shown in Figure 11).
  4. Replace the center nut and hand tighten. Press the change chart button again (#3) to resume temperature recording.

Note: Be careful not to pull up on the chart recorder arm while changing paper. Pulling on the arm may damage the recorder.

CAUTION: Do not use sharp or pointed objects to depress the chart buttons. This may cause permanent damage to the recorder.

Figure 12. Chart Buttons

10.3 Power Supply

The recorder normally uses AC power when the system is operating. If AC power fails, the LED indicator on the recorder flashes to alert you to a power failure. The recorder continues sensing cabinet temperature and the chart continues turning for approximately 24 hours with back-up power provided by the nine-volt battery.

The LED indicator glows continuously when main power is functioning and the battery is charged.

When the battery is low, the LED flashes to indicate that the battery needs to be changed.

10.4 Calibration and Adjustments

This recorder has been accurately calibrated at the factory and retains calibration even during power interruptions.

If required, however, adjustments can be made as follows:

  1. Run the unit continuously at the control setpoint temperature. Continue steady operation for at least two hours to provide adequate time for recorder response.
  2. Measure cabinet center temperature with a calibrated temperature monitor.
  3. Compare the recorder temperature to the solution temperature. If necessary, adjust the recorder by pressing the left and right chart buttons (#1 and #2 from Figure 12) for approximately five seconds.

Note: The stylus does not begin to move until the button is held for approximately five seconds.

11. Temperature Transmitter (Optional)

11.1 Powered Temperature Transmitter

A powered temperature transmitter is an option for all units. This temperature transmitter allows for remote temperature monitoring of your system.

To connect your remote temperature monitoring system to the powered temperature transmitter, refer to Figure 13 below.

Figure 13. Powered Temperature Transmitter

Label Description
1 Terminal Strip
2 Rear of Unit
3 Controller and/or Indicator
4 Customer Provided System

Note: The building management system load wiring must have a resistance less than 340 ohms.

11.2 Unpowered Temperature Transmitter

An unpowered temperature transmitter is an option for this unit. This temperature transmitter allows for remote temperature monitoring of your system.

To connect your remote temperature monitoring system to the unpowered temperature transmitter, refer to the diagram attached on the back of the unit.

12. Maintenance

WARNING: Risk of Shock. Risk of Shock. Disconnect equipment from main power before attempting any maintenance to equipment or its controls unless stated otherwise.

12.1 Cleaning the Cabinet Interior

To clean the cabinet interior, remove the shelves, drawers or baskets following the instructions in Section 7.3, Section 7.4 and Section 7.5. Use a solution of water and a mild detergent for cleaning. Rinse the interior storage components and wipe them dry with a soft cloth.

12.2 Cleaning the Condenser Filter

Clean the condenser filter every three months. There is one condenser filter located in the back cage of the unit that can be accessed without removing the back cage or disconnecting the power.

To clean the filter, complete the following steps:

  1. Remove the filter by pulling upwards through the slot in the back cage.
  2. Shake the filter to remove loose dust.
  3. Rinse the filter in clean water.
  4. Shake the excess water from the filter and let it dry.
  5. Reinstall filter by pushing downwards through the slot in the back cage.

CAUTION: Risk of Abrasion. Do not pull the filter downwards from the bottom. The condenser has sharp surfaces.

12.3 Cleaning the Condenser

Condensers should be cleaned at least every six months; more often if the laboratory area is dusty. In heavy traffic areas, condensers load with dirt more quickly. Failure to keep the condenser clean can result in equipment warm-up or erratic temperatures.

CAUTION: Risk of Abrasion. Never clean around the condensers with your fingers. Some surfaces are sharp.

The condenser is located in the top rear of the machine compartment. To clean the condenser, complete the following steps:

  1. Disconnect the power.
  2. Remove the back filter.
  3. Vacuum the condenser and clean up any loose dust.
  4. Replace the back filter.
  5. Reconnect power.

12.4 Automatic Defrost

The defrosting process on all -30°C freezers initiates automatically in response to a built-in timer. The defrost interval timer adapts to user conditions and will either be approximately every 6 hours or 15 hours. During this time, there may be a slight increase in display temperature.

If a door or port has been open for an extended duration allowing room air to flow into the freezer, frost may build up inside of the freezer and negatively affect the performance of the freezer. The unit defrost may not be able to overcome this and the unit must be manually defrosted by turning the freezer off.

12.5 Gasket Maintenance

Periodically check the gaskets around the door for punctures or tears. Leaks are indicated by condensation or frost which forms at the point of gasket failure. Make sure that the cabinet is level (refer to Section 7.1.1 for leveling information).

Keep the door gaskets clean and frost free by wiping gently with a soft cloth.

To check the door seal, complete the following steps:

  1. Open the door.
  2. Insert a strip of paper (a couple of inches wide) between the door gasket and the cabinet flange and close the door.
  3. Slowly pull the paper strip from the outside. You should feel some resistance.
  4. Repeat this test at 4-inch intervals. If the door does not seal properly, replace the gasket.

12.6 Alarm Battery Maintenance

Have a certified technician replace the alarm battery every twelve months at most or when the alarm is active. The part number for a replacement battery is 322533H01.

12.7 Preparation for storage

If the unit is going to be stored in an off condition, allow the unit to warm up and dry out with the door open before moving into storage.

13. Troubleshooting

WARNING: Risk of Shock. Troubleshooting procedures involve working with high voltages which can cause injury or death. Troubleshooting should only be performed by trained personnel.

This section is a guide for troubleshooting equipment problems.

Problem Cause Solution
Unit does not operate or Power Failure Indicator is on Power supply Check that the cord is securely plugged in. Plug another appliance into the outlet to see if it is live. Check that the double pole circuit breaker located next to the power inlet is in “ON” position (i.e“I” position). Try cycling the switch to OFF position (i.e"O" position) and then bring to ON position ("I"). Test the voltage and verify that it is correct for your unit (refer to Table 2). If the outlet is dead, check the circuit breaker or fuses. The unit should not be connected to a GFCI (Ground Fault Circuit Interrupter) protected outlet as it may be subject to nuisance tripping.
Temperature fluctuates Temperature Control Make sure that the control is set correctly. Refer to Section 9.3.
Condenser clogged Make sure the condenser and filter are clean. Refer to Section 12.2 and Section 12.3.
Solution bottle Make sure the solution bottles for the temperature sensors are full. The solution is a 50/50 mixture of glycerine and distilled water.
Other causes If the temperature control is set correctly, the condenser is clean, but temperature continues to fluctuate, call an authorized service representative.
Low battery icon is lit 12V backup battery needs to be replaced. Replace the battery. It is located on the top right hand side of the cabinet. Call an authorized service representative.
Condensation around door frame Incorrect Perimeter Heater Duty Cycle Increase the Perimeter Heater Duty Cycle, Refer to Section 9.4.
Gaps exist in unit port holes Ensure all port holes in the cabinet top, sides, and back are sealed properly to prevent warm airflow into the cabinet. Seal any gaps.
Unit is warm around door frame Door seal is broken Verify nothing is placed through the door seal such as a sensor. Check the door seal following instructions in Section 12.5.
Perimeter Heater is ON This is a normal function of the unit and is a result of the perimeter heater to reduce condensation.
Door is open Make sure the door is completely closed.
Door seal Check the door seal, following instructions in Section 12.5.
Unit warms up Warm product recently loaded in unit Allow ample time to recover from loading warm product.
Power supply Check for proper voltage to the unit. If there is no voltage to the unit, call an electrician.
Setpoints need to be adjusted To adjust the setpoint, refer to Section 9.3.
Extended air leak into freezer Warm air entering the freezer from extended door openings, leaking gaskets, or unsealed port holes may create excess frost and ice buildup in the freezer that the automatic defrost may not be able to remove. This will inhibit freezer cooling. The freezer needs to be manually defrosted by turning off the unit to allow all frost and ice to melt.
"E01" on display Invalid Algorithm Check to ensure the model type is set correctly in service mode. Refer to Section 9.4.
"E02" on display Control Probe Failure Check for loose probe connector. Replace control probe.
"E03" on display Defrost Probe Failure Check for loose probe connector. Replace defrost probe
"E05" on display Ambient Probe Failure Check for loose probe connector. Replace ambient probe.
"E06" on display Drip Pan Probe Failure Check for loose probe connector. Replace drip pan probe.
"Err" on display Upper Bottle Probe Failure Check for loose probe connector. Replace upper bottle probe.
"---" on display Lost Communication Call customer service.

14. End of Life Care

Be sure to follow local regulations when disposing of an old unit. Some suggestions are listed in the following:

  1. Remove items and defrost unit. Be sure to clean up any biological safety hazards.
  2. Remove the cabinet door to help prevent entrapment inside of a unit.
  3. Have a certified technician remove the refrigerant and compressor, then drain the compressor and oil from the system. Dispose of components following local regulations.

15. Warranty

During the first twenty four (24) months from shipment from the manufacturing facility, the manufacturer through its authorized Dealer or service organizations, will at its option and expense repair or replace any part found to be non-conforming in material or workmanship with the exception of the compressor which is covered for an additional 5 years (60 months) from the time of the shipment from the manufacturing facility. The manufacturer reserves the right to use replacement parts, which are used or reconditioned. Replacement or repaired parts will be warranted for only the unexpired portion of the original warranty. Except as expressly stated above, the MANUFACTURER MAKES NO OTHER WARRANTY, EXPRESSED OR IMPLIED, WITH RESPECT TO THE PRODUCTS AND EXPRESSLY DISCLAIMS ANY AND ALL WARRANTIES, INCLUDING BUT NOT LIMITED TO, WARRANTIES OF DESIGN, MERCHANT ABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

An authorized representative of the manufacturer must perform all warranty inspections. In the event of a defect covered by the warranty, we shall, as our sole obligation and exclusive remedy, provide free replacement parts to remedy the defective product.

The warranty provided hereunder shall be null and void and without further force or effect if there is any (i) repair made to the product by a party other than the manufacturer or its duly authorized service representative, (ii) misuse (including use inconsistent with written operating instructions for the product), mishandling, contamination, overheating, modification or alteration of the product by any customer or third party or (iii) use of replacement parts that are obtained from a party who is not an authorized dealer of products.

Heating elements, because of their susceptibility to overheating and contamination, must be returned to the factory and if, upon inspection, it is concluded that failure is due to factors other than excessive high temperature or contamination, the manufacturer will provide warranty replacement. As a condition to the return of any product, or any constituent part thereof, to the factory, it shall be sent prepaid and a prior written authorization from the manufacturer assigning a Return Materials Number to the product or part shall be obtained.

IN NO EVENT SHALL THE MANUFACTURER BE LIABLE TO ANY PARTY FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, OR FOR ANY DAMAGES RESULTING FROM LOSS OF USE OR PROFITS, ANTICIPATED OR OTHERWISE, ARISING OUT OF OR IN CONNECTION WITH THE SALE, USE OR PERFORMANCE OF ANY PRODUCTS, WHETHER SUCH CLAIM IS BASED ON CONTRACT, TORT (INCLUDING NEGLIGENCE), ANY THEORY OF STRICT LIABILITY OR REGULATORY ACTION.

Models: CAX1230FA, CAX1230FD, CAX2330FA, CAX2330FD, CAX3030FA, CAX3030FD, CAX5030FA, CAX5030FD, CAX1230FD 30 Degree C Laboratory Freezers, CAX1230FD, 30 Degree C Laboratory Freezers, Laboratory Freezers, Freezers

File Info : application/pdf, 36 Pages, 871.11KB

PDF preview unavailable. Download the PDF instead.

328020H01 RevF ; size=892022

References

Adobe PDF Library 15.0 Adobe InDesign 16.1 (Windows)

Related Documents

Preview Cardinal Health Refrigerators and Freezers: Premier, Pro, and Select Series Product Catalog
Explore the range of Cardinal Health laboratory refrigerators and freezers, including Premier Plus, Pro, and Select Series. Find detailed specifications, features, and model information for critical sample storage and laboratory applications.
Preview Cardinal Health Premier Plus Blood Bank Refrigerators: Installation and Operation Manual
Comprehensive guide for installing and operating Cardinal Health Premier Plus Blood Bank Refrigerators. Covers safety, setup, usage, maintenance, and troubleshooting for models like CAX1204B and CH2304B.
Preview Cardinal Health Home Healthcare Product Guide: Solutions for Daily Living and Patient Care
Discover the Cardinal Health Home Healthcare Product Guide, featuring a comprehensive catalog of medical equipment and supplies for daily living, mobility, diabetes care, and more. Find solutions from trusted brands for patients and caregivers.
Preview Cardinal Health: Supply Disruption Notification for Dover Urine Drainage Bag 4000ML
Notification from Cardinal Health regarding a supply disruption for the Dover™ Urine Drainage Bag 4000ML (8887600909) due to a raw material shortage. Provides an alternative product and contact information for assistance.
Preview Cardinal Health Clinical Digital Thermometers
Cardinal Health offers a comprehensive range of clinical digital thermometers designed for accuracy and speed in healthcare settings. This product catalog details standard and quick-read oral/rectal thermometers, probe covers, and forehead temperature indicators, including product specifications, model numbers, and packaging options.
Preview Cardinal Health Announces Product Transition for Breathable Underpads
Cardinal Health is discontinuing item #988B12 Breathable Underpad and introducing item #988C72 as a suggested substitute, effective December 2025. This notice provides details on the product transition and ordering.
Preview Cardinal Health Strep A Dipstick Rapid Test Instructions for Use
Instructions for use and performance characteristics of the Cardinal Health Strep A Dipstick Rapid Test, a qualitative immunochromatographic assay for the detection of group A streptococcal antigen from throat swab specimens.
Preview Kangaroo Joey Enteral Feeding Pump Troubleshooting Alerts
A guide to understanding and resolving alerts and error messages on the Kangaroo Joey Enteral Feeding Pump, including troubleshooting tips for common issues like hold errors, flow errors, feed errors, pump set dislodged, battery low, and rotor errors.