EMS Kontrol OXYGEN O₂ TRANSMITTER
OT-401/441/461 USER MANUAL
Manufacturer: EMS Kontrol
Website: www.emskontrol.com
WHAT IS IT?
The Oxygen Transmitter measures the oxygen value precisely and outputs it as an analog signal.
HOW DOES IT WORK?
It operates with a supply voltage between 12 V DC and 24 V DC. Measurement values are taken from the analog output terminals and transferred to the desired location.
GENERAL FEATURES
- Accurate and Precise Measuring
- Analog Output
- Durable and Handy Design
- 2 Years Sensor Operating Life
- Easy Replaceable Sensor
- Easy Assembling
- IP 67 Plastic Case (Excluding Sensor)
AREAS OF USE
- HVAC Applications
- Poultry Automation and Poultry Farms
- Cold Storage
- Incubation Rooms
- Food Storage
- Air Conditioning Cabinets
- Clean Rooms and Laboratories
RULES TO BE CONSIDERED FOR SAFETY
- Always read the user manual before using the device and its apparatus.
- Damages caused by opening, breaking, or misuse of the plastic parts of the device and its apparatus are considered out of warranty.
- Keep the device and its apparatus away from external influences such as liquids, high dust, high temperatures, etc., and protect them.
- Do not expose the device cables to any jamming or pressure.
- Disconnect the electrical power when your device is not used for a long time.
- EMS Kontrol devices and apparatus should be used by paying attention to the points in the user manual. In case of damages and malfunctions arising from external use (liquid contact, falling to the ground, etc.), ask for help from the service.
- Failures due to electrical connection errors and electrical voltage or current errors are not covered by the warranty.
Product Codes and Output Signals
Product Code | Output Signal |
---|---|
OT-401 | 0-10 V |
OT-441 | 4-20 mA |
OT-461 | Modbus RTU |
Technical Data
Product Name: | Oxygen O₂ Transmitter |
Supply Voltage: | 12-24 V DC |
Output: | 0-10 V / 4-20 mA / Modbus |
Measurement Range: | 0-25 % Vol |
Precision: | ± 0.1 % Vol |
Accuracy: | ± 1 % Vol |
Operating Temperature: | (-10°C) - (+55°C) |
Storage Temperature: | (-20°C) - (+60°C) |
Protection Class: | IP 67 (Excluding Sensor) |
INSTALLATION
- Unpack the product and open the top cover.
- Make cable connections accordingly.
Wiring Diagrams:
- 0-10 V Wiring Diagram: Connect '+' and '-' terminals for O₂ input, 'GND' and '24 VDC' for power (12-24V DC).
- 4-20 mA Wiring Diagram: Connect '+' and '-' terminals for O₂ input, 'GND' and '24 VDC' for power (12-24V DC).
- Modbus RTU Wiring Diagram: Connect 'A' and 'B' terminals for O₂ communication, 'GND' and '24 VDC' for power (12-24V DC).
- If the product will be mounted on the wall, use the screws and dowels provided in the package.
- The product starts measuring 30 seconds after being energized. It is recommended that the product remains in the environment for at least 5 minutes to obtain stable measurement values.
- It is recommended to use a shielded cable as the communication cable to prevent communication signals from being affected by external influences.
- Since the communication cable will create resistance, check the measurement values again after cable installation.
MODBUS RTU COMMUNICATION STRUCTURE
The default RS485 Baud rate for Modbus communication is 9600. Stop bit: 1, Parity: NONE. The default Slave ID is 85. The Oxygen value is stored in the holding register at address 2 (40003). Oxygen information can be obtained by reading these registers.
Register | Address | Default Value | Read/Write | Min. Value | Max. Value |
---|---|---|---|---|---|
Slave ID | 0 (40001) | 85 | R/W | 1 | 247 |
Baud rate | 1 (40002) | 3 | R/W | 0 | 8 |
O₂ Val | 2 (40003) | - | R | - | - |
Slave ID and Baud rate can be changed via Modbus communication. To change the Slave ID, write the desired ID to the register at address 0 (40001). To change the Baud rate, write the value corresponding to the desired Baud rate from the table below to the register at address 1 (40002).
Min. Value | Max. Value | Baud Rate |
---|---|---|
0 | 1200 | 1200 |
1 | 2400 | 2400 |
2 | 4800 | 4800 |
3 | 9600 | 9600 |
4 | 14400 | 14400 |
5 | 19200 | 19200 |
6 | 38400 | 38400 |
7 | 57600 | 57600 |
8 | 115200 | 115200 |
To reset changes made to the Slave ID or Baud rate to their default values, press and hold the "Calibration" button until the Status (green) LED turns off (approximately 30 seconds). After the LED turns off, release the button. This will restore the Slave ID and Baud rate to their default settings.
LED Meanings
LED | Status | Meaning |
---|---|---|
Power (Red) | On | Normal Operating |
Off | No Energy (Error) | |
Status (Green) | On | Measurement |
Off | Error - Fault | |
Flashing | Normal Measurement |
BEFORE CALIBRATION
- The product is sold calibrated. Please do not calibrate unless necessary.
- Perform calibration only when it is certain that the product is measuring incorrectly.
- Calibration is only carried out at the production facilities. Please contact EMS Kontrol for calibration.
SENSOR LIFE - IMPORTANT CRITERIA
- Sensor life is 2 years. After 2 years, sensor replacement is required. Sensor replacement can be done easily in the area of use.
- Do not expose the sensor to high volumes of natural gases, as it may cause the sensor to malfunction.
- The sensor power connection must not be repeatedly switched on and off.
DECLARATION OF CONFORMITY
Headquarters and place of production: Halkapınar Mah. 1376 Sok. Boran Plaza No:1/L Konak / IZMIR - TÜRKİYE
EMS KONTROL ELEKTRONİK VE MAKİNE SAN. TİC. A.Ş. declares that the product marked with CE, whose name and specifications are given below, covers the specified directives and provisions.
Brand: EMS KONTROL
Product Name: OT-4XX
Product Type: Oxygen Transmitter
Compatible Directives:
- Electromagnetic Compatibility Directive 2014/30/EU (EMC EN 61000-6-3: 2007 + A1: 2011, EN 61000-6-1: 2007)
- Low Voltage Directive 2014/35/EU (LVD EN 60730-2-9:2010, EN 60730-1:2011)
Additional information: This product can be used in combination with other devices, and compliance with the directives covers only the product. EMS KONTROL is not responsible for the compliance of the entire system with the directives. This declaration is not valid if the product is modified without EMS Kontrol's approval.
WARRANTY TERMS
- The warranty period of the devices and apparatus starts from the invoice date and is guaranteed for 2 years against manufacturing defects.
- Devices and apparatus are delivered to the customer in working condition from EMS Kontrol. On-site commissioning is subject to a service fee.
- The repair of devices and apparatus under warranty is carried out at EMS Kontrol's facilities. Transportation costs for warranty repairs belong to the customer. On-site services will incur transportation and accommodation expenses for service personnel, borne by the customer.
- Maintenance of devices and apparatus is performed at EMS Kontrol's facilities. Transportation fees for maintenance belong to the customer.
- In case of malfunction during the warranty period, EMS Kontrol will test the device to determine if the fault is customer-induced or manufacturer-induced and provide a report.
- If a manufacturer-induced fault is detected within the warranty period, the customer may request a replacement or repair at no cost, provided it does not exceed the product price.
- If faults are determined to be customer-induced during the warranty period, all costs belong to the customer.
- If the customer is aware of defects from the start of the warranty period or should be aware, they cannot benefit from the replacement/full repair clause (Article 6).
- Failures arising from the use of devices and apparatus contrary to the user manual are not covered by the warranty.
- Devices and apparatus are not covered by the warranty if they are physically damaged (beaten, broken, scratched) by the customer.
- Damages caused by the use of devices and apparatus of other brands and models without EMS Kontrol's approval are not covered by the warranty.
- Errors caused by rusting, oxidation, and liquid contact due to working in dusty, acidic, or humid environments are not covered by the warranty.
- Damages that may occur during transportation are not covered by the warranty. Customers may opt for transportation insurance.
- Damages caused by mains voltage or faulty electrical installation are not covered by the warranty.
- Devices and apparatus are not covered by the warranty in case of malfunctions caused by force majeure events such as fire, flood, or earthquake.
- All parts of the devices and apparatus are covered by EMS Kontrol's warranty.
- If a device malfunctions within the warranty period, the time spent on repair will be added to the warranty period. The maximum repair time is 20 working days from the notification of the malfunction. If the malfunction is not resolved within 20 business days, EMS Kontrol must provide a similar replacement product until the repair is completed.
- If a device fails at least four times within a year or six times within the warranty period, and these failures render the device unusable, or if repair exceeds the maximum time, or if repair is deemed impossible by EMS Kontrol, the customer may request a refund or a price reduction.
- Customers may file complaints and objections with consumer courts or consumer arbitration committees.
- The warranty certificate must be kept by the customer. A duplicate will not be issued. Loss of the certificate will result in repair and replacement services being charged.
WEEE Disposal Information
This device is classified as Waste Electrical and Electronic Equipment (WEEE) according to European directives 2002/96/EC. Before disposal, prevent potential negative consequences for the environment and human health. This symbol indicates that the product should not be treated as household waste and must be delivered to designated electrical and electronic waste collection points. Disposal must comply with local environmental regulations.
Manufacturer Information
Title: EMS KONTROL ELEKTRONİK VE MAKİNE SAN. VE TİC. A.Ş.
Address: Halkapınar Mah. 1376 Sokak Boran Plaza No:1/L Konak / İzmir-TÜRKİYE
Telephone: 0 (232) 431 2121
E-Mail: info@emskontrol.com
Product Details
Type: Oxygen Transmitter | Brand: EMS Kontrol |
Model: OT-4XX | Warranty Duration: 2 Years |
Maximum Repair Time: 20 Days | Banderol and Serial Number: ***_******_******_***______ |
Vendor Information
Title: _________________________
Address: _________________________
Telephone: __________ Fax: __________
E-Mail: _________________________
Invoice Date and Number: _______________
Delivery Date and Place: _______________
Signature of Authorized Person: _______________
Company Stamp: _________________________
EMS Kontrol reserves the right to make changes and improvements to the product specifications and user manual. For all changes, please visit emskontrol.com.