Rossmax NE100 Compressor Nebulizer

Model: NE100

Introduction

Thank you for purchasing the Rossmax NE100 Compressor Nebulizer. With proper care and use, your nebulizer will provide you with many years of reliable treatments. This unit operates on standard AC power. Treatments are delivered quickly, safely, and conveniently, making this unit ideal for all ages in healthcare settings like hospitals and home care. We encourage you to thoroughly read this guidebook to learn about the features of your nebulizer. Your compressor nebulizer should be used under the supervision of a licensed physician and/or a respiratory therapist. Together with your physician and/or respiratory therapist, you can feel comfortable and confident knowing that you are obtaining the most effective inhalation treatments for your respiratory condition.

NOTE: Your nebulizer is intended for use in the treatment of asthma, COPD, and other respiratory ailments in which an aerosolized medication is required during therapy. Please consult with your physician and/or pharmacist to determine if your prescription medication is approved for use with this nebulizer. For type, dose, and regime of medication, follow the instructions of your doctor or respiratory therapist.

This device fulfills the provisions of the EC directive 93/42/EEC (Medical Device Directive) and the European Standard EN 13544-1:2007+A1:2009 Respiratory therapy equipment - Part 1: Nebulizing systems and their components.

Please read this manual carefully before use and be sure to keep this manual.

Product Cautions

READ THE FOLLOWING BEFORE USING:

Operating Cautions

Storage Cautions

Cleaning Cautions

Medical Disclaimer

This manual and product are not meant to be a substitute for advice provided by your doctor or other medical professionals. Don't use the information contained herein or this product for diagnosing or treating a health problem or prescribing any medication. If you have or suspect that you have a medical problem, promptly consult your doctor.

Durable Periods

Durable periods are as follows, provided the product is used to nebulize 2ml of medication 2 times a day for 8 minutes each time at room temperature (25°C). Durable period may vary depending on usage environment.

Product Specifications

Specification Value
Power AC 230V/50Hz or AC 220V/60Hz or AC 110V/60Hz
Power Consumption < 130W
Sound Level < 60 dBA (1 meter away from NE100)
Compressor Pressure Range > 29 psi (200 kPa)
Operating Pressure Range > 15 psi (103 kPa)
Operating Flow Range > 3.5 lpm
Operating Temperature Range 10°C to 40°C (50°F to 104°F)
Operating Humidity Range 10 ~ 90% RH
Operating Atmospheric Pressure Range 700-1060 hPa
Storage Temperature Range -20°C to 60°C (-4°F to 140°F)
Storage Humidity Range 10 ~ 90% RH
Dimension (L x W x H) 204.5mm x 148mm x 108mm (8.05x5.83x4.25 inches)
Weight 1360g (without accessories)
Medication Capacity 5ml(cc)
Particle Size (MMAD) < 2.6µm
Average Nebulization Rate > 0.35 ml/min
Standard Accessories Nebulizer Kit, Air Tube, Mouthpiece, Filters (5pcs), Adult and child masks

* Subject to technical modification without prior notice.

* Performance may vary with drugs such as suspensions or high viscosity. See drug supplier's data sheet for further details.

A. Product Identification

  1. Nebulizer Kit
  2. Nozzle
  3. Angled Mouthpiece
  4. Air Filter
  5. Child Mask
  6. Adult Mask
  7. Air Tube
  8. Air Output
  9. Air Filter Slot
  10. Air Intake
  11. Power Cord
  12. Power Switch

B. Assembling Your Nebulizer Kit

Follow the cleaning instructions in this guidebook under "Cleaning technique" prior to using your nebulizer for the first time or after it has been stored for an extended period of time.

REMEMBER: Always unplug the compressor and make sure the power switch is turned to the "OFF" position before cleaning, assembling, and before or after each use.

  1. Place the compressor on a flat, stable surface within reach.
  2. Gently twist and pull straight up on the lid of the nebulizer kit to separate into two parts (medication cup and cover).
  3. The stem inside the medication cup inserts into the tube of the nozzle.
  4. Add the prescribed amount of medication to the medication cup.
  5. Reassemble the nebulizer kit by carefully twisting the medication cup and cover together. Make sure that the two parts fit securely.

WARNING: The symbol on this product means that it's an electronic product and following the European directive 2012/19/EU, the electronic products have to be disposed of at your local recycling centre for safe treatment.

C. Operating Your Nebulizer

The nebulizer kit is operable at up to a 45° angle. If the angle is greater than 45°, no aerosol will be generated.

  1. Attach one end of the air tube connector to the air output.
  2. Carefully attach the opposite end of the air tubing connector to the stem at the base of the nebulizer kit.
  3. Attach the angled mouthpiece or mask to the top of the nebulizer kit.
  4. The capacity of the medication cup is 2~5 ml.

NOTE: A 30-minute interval is recommended after each use. The compressor will automatically shut off if it becomes overheated. If/when this happens, immediately:

  1. Press the power switch to the "OFF" position.
  2. Unplug the power cord from the outlet.
  3. Allow the motor to cool for 30 minutes.

Before restarting the unit, make sure that the air vents are not obstructed.

Operating Your Nebulizer After Every Use:

  1. Unplug the unit from the power source.
  2. Allow the unit to completely cool.
  3. Carefully detach the air tubing from the nebulizer and pour out any remaining medication.
  4. Follow the cleaning procedures provided in this guidebook.

D. Cleaning Procedures

Rinsing Technique (performed after each treatment or before first use):

  1. Make sure that the power switch has been turned to the "OFF" position and the unit has been disconnected from the power source.
  2. Disconnect the air tube from the nebulizer device.
  3. Gently twist and pull up the cover of the nebulizer kit to open and separate.
  4. Rinse the nebulizer kit and components with hot tap water.
  5. Dry with clean towels or completely air dry.
  6. Reassemble the nebulizer kit.

NOTE: For the first time cleaning or after the unit has been stored for an extended period of time, thoroughly clean all components, except the air tube. The nebulizer kit is dishwasher safe.

Cleaning the Compressor:

Wipe the compressor daily using a soft cloth.

NOTE: Any other form of cleaning or cleaning agents may damage the finish of the unit.

E. Changing the Air Filter

It is important to change the air filter approximately when the air filter turns gray. It is recommended to change the air filter every 2 months.

  1. Remove the air filter cover by gently pulling forward.
  2. Discard the gray filter.
  3. Replace with a new, clean air filter.
  4. Securely re-attach the air filter cover to the unit.

NOTE: Air filters cannot be cleaned or washed. Only NE100 air filters can be used. Do not substitute alternate material such as cotton. Do not operate without an air filter.

Troubleshooting

If any abnormality occurs during use, please check and correct the following:

  1. Unit does not operate when power switch is pressed: Check the AC connection to the outlet.
  2. No misting or low rate of misting:
    • Check that there is medication in the nebulizer cup.
    • Check the main unit for any physical damage.
    • Check the position of the nozzle inside the nebulizer.
    • Make sure that the air tube and other components are properly attached.
    • Check the air filter and replace if necessary.

Safety Information

Electromagnetic Compatibility Information

1. This device needs to be installed and put into service in accordance with the information provided in the user manual.

2. WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the NE100, including cables specified by the manufacturer. Otherwise, degradation of the performance of this device could result.

If higher IMMUNITY TEST LEVELS than those specified in Table 9 are used, the minimum separation distance may be lowered. Lower minimum separation distances shall be calculated using the equation specified in 8.10.

Manufacturer's declaration-electromagnetic immunity

The NE100 is intended for use in the electromagnetic environment specified below. The customer or the user of the NE100 should assure that it is used in such an environment.

Immunity IEC 60601 test level Compliance level Electromagnetic environment guidance
Conducted RF IEC 61000-4-6 3 Vrms: 0.15 MHz – 80 MHz 6 Vrms: in ISM and amateur radio bands between 0.15 MHz and 80 MHz 80% AM at 1 kHz 3 Vrms: 0.15 MHz – 80 MHz 6 Vrms: in ISM and amateur radio bands between 0.15 MHz and 80 MHz 80% AM at 1 kHz Portable and mobile RF communications equipment should be used no closer to any part of the NE100, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance: d = 1.2√P 80MHz to 800 MHz, d = 2.3√P 800MHz to 2.7 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).
Radiated RF IEC 61000-4-3 10 V/m 80 MHz – 2.7 GHz 80% AM at 1 kHz 10 V/m 80 MHz – 2.7 GHz 80% AM at 1 kHz Interference may occur in the vicinity of equipment marked with the following symbol: ⊗

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

More information on EMC compliance of the device can be obtained from the Rossmax website: www.rossmax.com.

Accessories (Optional)

Name Model No Material Disinfecting Method: Boiling
Nebulizer bottle set (N4 Bottle + Mouthpiece) NB_AC_009_00 PP O
Nebulizer bottle (N4) NB_AC_027_00 PP O
Mouthpiece NB_AC_021_00 PP O
Adult Mask NB_AC_003_00 PVC X
Child Mask NB_AC_004_00 PVC X
Air Tube (1.5M) NB_AC_005_00 PVC X
Air Filters (Ø15x10mm) x 5pcs NB_AC_006_00 PU X

* Please follow the disinfecting methods to boil and wash the accessories.

Warranty Card

This instrument is covered by a 3-year guarantee from the purchase date. The guarantee is valid only on presentation of the guarantee card completed by the dealer confirming the purchase date or the receipt. Nebulizer Components are not included. Opening or altering the instrument invalidates the guarantee. The guarantee does not cover damage, accidents, or noncompliance with the instruction manual. Please contact your local seller/dealer or www.rossmax.com.

Customer Name: _________________________________________________________________________

Address: ________________________________________________________________________________

Telephone: ____________________________ E-mail address: ___________________________________

Product Information

Date of purchase: _________________________________________________________________________

Store where purchased: _________________________________________________________________________

Models: NEBULIZER, Product Name NE100, Model Numbers 5, NEBULIZER Respiratory Diseases, NEBULIZER, Respiratory Diseases, Diseases

File Info : application/pdf, 2 Pages, 4.04MB

PDF preview unavailable. Download the PDF instead.

OBM IB NE100N4 V9 SWCMC ver2408

References

Adobe PDF Library 10.0.1 Adobe InDesign CS6 (Macintosh)

Related Documents

Preview Rossmax NA100 Piston Nebulizer: Instruction Manual & User Guide
Comprehensive instruction manual for the Rossmax NA100 Piston Nebulizer. Learn how to operate, clean, assemble, and troubleshoot your device for effective respiratory therapy.
Preview Rossmax PF120 Peak Flow Meter: User Manual and Specifications
Comprehensive guide to the Rossmax PF120 Peak Flow Meter, including usage instructions, troubleshooting, cleaning, specifications, and warranty information for monitoring respiratory conditions like asthma.
Preview Rossmax HS200 Blood Glucose Monitoring System User Manual
Comprehensive guide to the Rossmax HS200 Blood Glucose Monitoring System, covering setup, testing, troubleshooting, specifications, and safety for accurate blood glucose management. Visit www.rossmax.com for more information.
Preview Rossmax CH155 Automatic Blood Pressure Monitor | Accurate Home Health
Discover the Rossmax CH155 Automatic Blood Pressure Monitor, designed for accurate and convenient home health monitoring. Features include Movement Detection, Real Fuzzy Technology, Hypertension Risk Indication, Irregular Heartbeat Detection, and clinically validated accuracy by ESH and BIHS.
Preview Rossmax Aspirator V3 Instruction Manual - Safe and Effective Suction Therapy
Comprehensive instruction manual for the Rossmax Aspirator V3 medical suction unit. Learn about intended use, safety precautions, setup, operation, cleaning, and troubleshooting for effective suction therapy.
Preview Rossmax HA500 Non-Contact Temple Thermometer User Manual
Comprehensive user manual for the Rossmax HA500 non-contact temple thermometer, detailing features, operation, troubleshooting, and specifications for accurate body temperature measurement. The device is designed for accurate human body temperature measurement and complies with relevant medical device directives.
Preview Rossmax HC700 Non-Contact Temple Thermometer User Manual
Comprehensive user manual for the Rossmax HC700 non-contact infrared thermometer, detailing its features, operation, troubleshooting, and specifications for accurate body and object temperature measurement.
Preview Rossmax Z1 Automatic Blood Pressure Monitor User Manual
This document provides instructions for using the Rossmax Z1 Automatic Blood Pressure Monitor, including setup, operation, troubleshooting, and technical specifications. It covers features like Real Fuzzy Technology, Cuff Wrap Detection, Movement Detection, Hypertension Risk Indication, and Irregular Heartbeat Detection.