Elitech GSP-6 Temperature and Humidity Data Logger
This document provides an overview of the Elitech GSP-6, a versatile data logger designed for precise temperature and humidity monitoring.
Product Description
The Elitech GSP-6 data logger is primarily used to record the temperature and humidity of pharmaceutical products, vaccines, medications, food, and chemical products. It is widely applicable to every link in cold chain storage and logistics, including refrigerated containers, refrigerated trucks, refrigeration bags, refrigerated cabinets, first-aid kits, cold rooms, and laboratories.
The device can automatically limit the recording interval in case of exceeding temperature and humidity limits. It features a buzzer and a flashing red indicator light to signal when limits are exceeded.
Image Description: A front view of the Elitech GSP-6 data logger, an orange device with a large LCD screen displaying temperature (e.g., 40.2°C), humidity (e.g., 20.3%), and log count (16000). Two external probes are connected to the device. The Elitech logo is visible.
Key Features
- Records up to 16,000 temperature points and 16,000 humidity points.
- Offers 3 different types of temperature and humidity measurement (selectable via ElitechLog software).
- Includes 1 external temperature sensor and 1 external combined temperature and humidity sensor.
- Wide temperature measurement range: -40 °C to 85 °C, with a maximum accuracy of ± 0.5 °C.
- Humidity measurement range: 10% to 99% RH, with a maximum accuracy of ± 3% RH.
- Multiple alarm notifications: buzzer, LED indicators, and display screen.
- Two built-in magnets for convenient attachment to metallic surfaces.
- Large LCD screen for direct viewing of key information, including temperature and humidity data, and alarm status.
Image Description: Multiple images are shown. One is a close-up of the Elitech GSP-6 device. Another shows a calibration certificate with fields like "Inspector", "Model", "Serial Number", "Expiry Date", and "Validation Points" with readings. Other images show the device displaying different readings like "60.0°C / 90.0%" and "85.1°F / 80.4% / 81". Text annotations indicate "Celsius or Fahrenheit Converted", "Temperature Measuring Range: -40°F~185°F", "Humidity Range: 10%~99%", and "Buzzer Alarm and Light Alarm".
Image Description: Images show the Elitech GSP-6 device connected to a laptop via USB. Close-up views of the device highlight the "USB Port" and the connections for the "Temperature-humidity Combined Sensor" and the "Temperature Sensor".
Parameters
- Temperature measurement range: -40 °C ~ 85 °C
- Temperature accuracy: ± 0.5 °C (-20 °C ~ 40 °C; ± 0.1 °C for other ranges)
- Humidity range: 10% ~ 99% RH
- Humidity accuracy: ± 3% RH (at 25 °C, 20% ~ 90% RH); ± 5% RH for other ranges
- Resolution: Temperature 0.1 °C, Humidity 0.1% RH
- Record capacity: 16,000 points (max.)
- Logging interval: 10 seconds to 24 hours, continuously configurable
- Data interface: USB
- Power supply: Single-use 3.6V lithium battery or USB powered
- Battery life: 2 years at room temperature with a 15-minute logging interval and buzzer alarm disabled.
- Ambient operating temperature: -30 °C ~ 70 °C. In low-temperature environments, the LCD screen may display readings slightly slower but will return to normal when the temperature recovers.
Image Description: Images display the two types of external sensors: a needle-type temperature sensor and a combined temperature/humidity sensor with a cable. It also shows the back of the data logger with two circular magnets labeled "Two Built-in Magnets" and the battery compartment open, revealing a 3.6V lithium battery.
Software and Data Management
Compatible Software (Mac / Win): When using the data logger for the first time, install the data management software. Once connected to your computer via USB, you can configure the logging interval, temperature unit (°C / °F), and alarm limits. Data can be saved, graphed, and exported in Excel / PDF format.
Calibration
Each unit's Serial Number (S/N) comes with a calibration certificate.