Elitech GSP-6 Pro Temperature and Humidity Data Logger

Overview

The GSP-6 Pro is a temperature and humidity data logger featuring an external probe. It offers a large LCD screen, audible-visual alarms, automatic interval shortening during alarms, and built-in magnets for easy mounting. This device is ideal for recording temperature and humidity during the storage and transportation of medicines, chemicals, and other goods across various stages of the cold chain, including cooler bags, cooling cabinets, medicine cabinets, refrigerators, and laboratories.

Key Components

Specifications

FeatureDetails
ModelGSP-6 Pro / GSP-6Pro Gle
Temperature Measurement Range-40°C to +85°C (-40°F to 185°F)
Temperature Accuracy±0.3°C/±0.6°F (-20°C to +40°C), ±0.5°C/±0.9°F (others)
Humidity Measurement Range0%RH to 100%RH
Humidity Accuracy±3%RH (at 25°C, 20%RH to 80%RH), ±5%RH (others)
Resolution0.1°C/°F; 0.1%RH
MemoryMaximum 100,000 points
Logging Interval10 seconds to 24 hours
Data InterfaceType-C
Start ModePress button; Use software
Stop ModePress button; Auto-stop; Use software
SoftwareElitechLog, for macOS & Windows system
Report FormatPDF/EXCEL/TXT* (TXT for Windows ONLY) via ElitechLog software
Power1.5V AA battery (LR6/Type-C) / Type-C (Non-rechargeable)
Shelf Life2 years
CertificationCE, FCC, WEEE
External ProbeTemperature-humidity probe; Glycol bottle probe (optional)
Dimensions118 x 64.6 x 19.6 mm
Weight130g

* The glycol bottle contains 8ml propylene glycol.

Operation Instructions

1. Activate Logger

Pull out the battery insulator strip.

2. Install Probe

Install the probes into the corresponding jacks on the right side of the logger.

3. Install Software

Download and install the ElitechLog software (macOS and Windows) from the official Elitech websites: Elitech US: www.elitechustore.com/pages/download or Elitech UK: www.elitechonline.co.uk/software or Elitech BR: www.elitechbrasil.com.br.

4. Configure Parameters

Connect the data logger to your computer via USB cable. Wait for the logger icon to appear on the LCD.

Using Software:

For first-time use or after battery replacement, click "Quick Reset" under the Summary menu to synchronize local time. To change parameters, navigate to the Parameter menu, enter desired values, and click "Save Parameter".

Warning: Ensure local time is configured before use to avoid time or time zone errors. The "Interval Shortened" parameter is disabled by default; enabling it will automatically shorten the logging interval to once per minute if temperature/humidity limits are exceeded.

5. Start Logging

Press Button: Press and hold the button for 5 seconds until the logging icon appears on the LCD.

Note: If the icon flashes, it indicates a configured start delay; logging will commence after the set delay.

6. Stop Logging

Press Button: Press and hold the button for 5 seconds until the stop icon appears on the LCD.

Auto Stop: The logger stops automatically when the maximum memory capacity is reached.

Use Software: Connect the logger to your computer, open the ElitechLog software, go to the Summary menu, and click the "Stop Logging" button.

Note: The default stop method is via button press. If disabled, the button stop function will be invalid, requiring software intervention.

7. Download Data

Connect the data logger to your computer via USB cable. Wait for the logger icon on the LCD.

Using ElitechLog Software: The logger typically auto-uploads data. Click "Export" to select your desired file format. If auto-upload fails, manually click "Download" and repeat the export process.

8. Reuse the Logger

First, stop the logger. Connect it to your computer and use the software to save or export data. Then, reconfigure the logger by repeating step 4. Finally, follow step 5 to restart logging.

Warning: Reconfiguration deletes all previous logging data. Save data before applying new configurations. If data is lost, try locating it in the History menu of the software.

For APP operation instructions, visit: www.elitechlog.com

APP Operation Instructions

Bluetooth Function Indicator

Flashing: Waiting for Bluetooth connection. Constant light: Bluetooth connected.

Install Software

Download the "Elitech iCold" app from Google Play or the App Store. You can register an account or use the app without registration by clicking "Bluetooth Data Logger" to access the parameter configuration page.

Modify Parameters

  1. Click the top left corner and select "Bluetooth Data Logger".
  2. Click on the device corresponding to its SN code.
  3. For first-time use, configure parameters: time zone, recording interval, and upper/lower temperature and humidity limits. Click "Save" in the upper right corner to complete configuration. Configuration is only possible when the device is not recording; otherwise, it will jump directly to the data list.
  4. After successful configuration, the device status will display "Wait for start".
  5. Long press the GSP-6 Pro Home button for 5 seconds. The device status will change from "Wait for start" to "Logging", indicating data recording has begun.

Data Viewing

  1. Click the top right corner menu to refresh the device list. Click "Data Logger" to highlight the device list and bring the selected device to the top. Click the device to read data.
  2. After data reading, click the top right corner to stop recording.
  3. Navigate to the data menu to view a detailed data list. Click "Generate Report" to create a report in the required format (Supports PDF). You can also print the data list using a compatible Bluetooth printer.

Status Indication

Button Operation Function

LCD Screen Indicators

The LCD screen can display various information:

The display of maximum and minimum values can be configured via the APP to show continuously for 1 to 15 days or for the entire logging process.

LCD Interface Examples

Common displays include:

Buttons-LCD-LED Indication

The logger provides visual and audible feedback:

Display / LED StatusBuzzer AlertsMeaning
LCD: flashes once every 15 seconds / LED: flashes onceBeeps onceConnected to PC ?
LCD: flashes twice / LED: flashes twiceBeeps twiceHigh/Low temperature/humidity limit exceeded ?
LCD: flashes 3 times / LED: flashes 3 timesBeeps 3 timesNot started
LCD: flashes 4 times / LED: flashes 4 timesBeeps 4 timesStarted ▶️
LCD: always on / LED: always on-Stopped ⏹️
LCD: not displayed / LED: not displayed-Start delay

Note: To enable the buzzer function, open ElitechLog software, go to the Parameter menu, select Buzzer, and choose "Enable".

Battery Replacement

  1. Open the battery cover and remove the old battery.
  2. Install a new dry battery (1.5V AA, LR6) into the compartment.
  3. Close the battery cover.

What's Included

Important Warnings

PDF preview unavailable. Download the PDF instead.

huong-dan-su-dung-elitech-gsp-6pro Adobe Illustrator 26.5 (Windows) Adobe PDF library 16.07

Related Documents

Preview Elitech GSP-6 Pro Temperature and Humidity Data Logger User Manual
Comprehensive user manual for the Elitech GSP-6 Pro, detailing its features, specifications, operation, software configuration, and app usage for temperature and humidity monitoring in cold chain applications.
Preview Elitech GSP-6 Temperature and Humidity Data Logger - Product Overview
Detailed overview of the Elitech GSP-6, a high-precision temperature and humidity data logger for cold chain monitoring, pharmaceuticals, vaccines, food, and chemical products. Features include extensive recording capacity, multiple alarm notifications, and USB connectivity.
Preview Elitech GSP-6Pro Bluetooth Temperature and Humidity Data Logger | Product Overview
Discover the Elitech GSP-6Pro, a high-capacity Bluetooth data logger for precise temperature and humidity monitoring. Features 100,000 data points, 1-year battery life, and easy data export via APP. Ideal for cold storage, warehouses, and more.
Preview Elitech GSP-6 Multi-use Temperature & Humidity Data Logger User Manual
Comprehensive guide to the Elitech GSP-6 data logger, covering its features, specifications, operation, data management, and troubleshooting for temperature and humidity monitoring in cold chain logistics.
Preview Elitech GSP-6 Multi-use Temperature & Humidity Data Logger
Comprehensive guide to the Elitech GSP-6, a multi-use data logger for temperature and humidity monitoring in logistics, pharmaceuticals, and food industries. Features include LCD display, external sensors, alarm functions, and data management software.
Preview Elitech RC-61/GSP-6 User Manual: Temperature and Humidity Data Logger
This user manual provides comprehensive guidance for the Elitech RC-61/GSP-6 temperature and humidity data logger, covering its features, specifications, operation, and important usage notes for cold chain monitoring applications.
Preview Elitech GSP-6/6G/8 Multi-Use Data Logger for Vaccine Monitoring
Explore the Elitech GSP-6/6G/8 series multi-use data loggers, designed for accurate temperature and humidity monitoring, ideal for vaccine storage and transport. Features include large LCD, USB connectivity, and advanced software.
Preview Elitech RC-61/GSP-6 Temperature and Humidity Data Logger User Manual
Comprehensive user manual for the Elitech RC-61/GSP-6 temperature and humidity data logger. Learn about its features, specifications, operation, status indications, battery replacement, and default parameters.