Elitech MS-800 Series Manifold Instrument User Manual
1 Product Introduction
Welcome to the Elitech MS-800 series manifold instrument. Please read this manual carefully before use to avoid improper operation that could cause harm to you or the product. The MS-800 series manifold instrument integrates pressure-temperature mode, pressure-holding mode, vacuum mode, and data logging mode. It is suitable for daily troubleshooting, maintenance, and repair of refrigeration and HVAC systems, providing users with more real and accurate data. It significantly improves the efficiency of on-site work and can completely replace existing mechanical gauge valves. Featuring a dot-matrix LCD screen for a rich interface effect and more intuitive data viewing, it also supports an APP for viewing reports and data analysis. It allows for instant updates of commonly used refrigerants and can be upgraded via remote firmware, continuously enhancing the user experience.
2 Safety Instructions
- This product is not suitable for maintenance of ammonia (ammonia-containing) refrigerant systems.
- This product contains batteries and should not be placed in high-temperature environments or thrown into fire, as there is a risk of explosion.
- Do not use this product during thunderstorms to avoid being struck by lightning, which could endanger life and damage the product.
- Strictly follow the safety precautions of the refrigeration system.
- When using this product, please wear safety goggles and protective gloves, and read the maintenance manual of the system unit carefully before connecting to the system for maintenance operations.
- If the product is found to be damaged, please contact Elitech in time. Do not disassemble the product privately, as this may further damage the product and, in severe cases, cause the battery to catch fire or even explode.
3 Environmental Protection
- Comply with local environmental policies. Refrigerants should not be directly discharged into the atmosphere and must be recovered using professional equipment.
- At the end of the product's life, please recycle it according to local regulations. Do not dispose of it arbitrarily to avoid environmental pollution.
- Take the scrapped old batteries to the specified waste battery collection point.
4 Overview
The MS-800 series manifold instrument includes the following components:
- 1. LCD screen
- 2. Control buttons
- 3. Liquid sight glass
- 4. Valve knob
- 5. Refrigerant tube holder
- 6. Temperature clamp connector
- 7. Refrigerant pipe connection port (1/4SAE)
- 8. Suspension device
- 9. Battery compartment (AA batteries*3)
- 10. Information sticker
The manual displays two models: the MS-810 series on the left and the MS-870 series on the right. Both units are digital manifold gauges featuring an orange and black design with a prominent LCD screen and multiple connection ports for refrigerant lines and temperature sensors.
5 Button Functions
Button/Action | Function |
---|---|
[Power Button] | 1. Long press: power on/off 2. Short press: turn on/off Bluetooth |
[Mode/Setting Button] | 1. Long press: Enter the setting interface 2. Short press: ① Mode switching (pressure temperature/holding pressure/vacuum) ② Setting interface to switch the main option |
[Backlight/Calibration Button] | 1. Short press: Turn on/off the backlight 2. Long press: ① Data calibration (air pressure calibration range: -14.5-100psi) ② Vacuum interface timer zero |
[Test/Exit Button] | Short press: ① Pressure-holding interface to open pressure-holding test ② Vacuum interface to open the leakage test Long press: ① Pressure-holding interface ends pressure-holding test ② Exit pressure holding data result display interface ③ Vacuum interface ends leakage test ④ Exit vacuum data result display interface |
[Refrigerant/Option Button] | Short press: ① Switch refrigerant when unlocking ② Switching sub-options in the setting screen |
Press and hold [Up Arrow] and [Down Arrow] simultaneously | Press and hold for 3 seconds to unlock selection of refrigerant state. |
6 Specification
6.1 Host parameter
Characteristics | Parameters |
---|---|
Measurement range | -14.5~800psi |
Accuracy | ±0.5%FS |
Resolution | 0.5psi |
Units | psi, kg/cm², kPa, MPa, bar, inHg |
Refrigerant type | Device default 18 types, can be updated through APP |
Selectable refrigerants | R11 R12 R13 R22 R290 R32 R134a R404A R410A R407C R408A R409A R458A R448A R449A R500 R502 R600 |
Interface | 1/4 SAE*3, 3/8SAE*1 (MS-870S) |
Power | supply 3*1.5V, AA batteries |
Battery life | not less than 150 hours (Bluetooth and backlight off) |
Display | LCD screen |
Main unit (Size) | 186*186*66mm, Weight: about 1300g |
Main unit (Operating Conditions) | Working temperature: -10~50℃/14~122°F, Storage Temperature: -20~60℃/-4~140°F |
6.2 Temperature parameter
Characteristics | Parameters |
---|---|
Measurement range | -40~150°C |
Accuracy | ±0.5°C (-40~130℃), ±1℃ (130~150°C) |
Resolution | 0.1°C |
Units | °C, °F |
Interface | PS/2 |
7 Quick Operation Guide
7.1 Measuring Pressure and Temperature
- Press the
[Power Button]
to turn on the device and enter the main interface for PT. - Hold the
[Up Arrow]
and[Down Arrow]
buttons for 3 seconds to unlock the refrigerant selection. - Refrigerant area blinks, short press
[Up Arrow]
or[Down Arrow]
to select the desired refrigerant. - Close all valves and connect the high and low-pressure sides of the product to the system being tested.
- Start the system being tested to real-time monitor the system's pressure and saturation temperature among other parameters.
Note: Ensure proper connection according to the diagram shown below.
Diagram Description: The diagrams illustrate the connection setup for both the MS-810 series and MS-870 series units to a refrigeration system. The connections involve the high-pressure side, low-pressure side, and a service port. Valves on the unit and the system are shown in specific ON/OFF states relevant to the measurement process.
7.2 Hold Pressure Measurement
- Turn on the device and press the
[Test Button]
to enter the HOLD pressure interface. - Charge the system being tested with an appropriate amount of nitrogen. When the pressure reaches the target value, close the high-pressure side valve.
- Press the
[Test Button]
to start the hold pressure test.
Note: Ensure the connection is made correctly according to the diagram shown below.
Diagram Description: Similar to the previous connection diagrams, these illustrate the setup for a hold pressure test. The MS-810 and MS-870 series units are shown connected to a refrigeration system, with valves positioned to perform the test after reaching a target pressure.
7.3 Refrigerant Charging
- Turn on the device and enter the PT interface.
- Connect the refrigerant tank to the middle interface.
- Close the high-pressure valve and open the low-pressure valve.
- Open the refrigerant line valve to charge refrigerant.
Note: Ensure the connection is made correctly as shown in the diagram below.
Diagram Description: These diagrams show the MS-810 and MS-870 series units connected for refrigerant charging. The refrigerant tank is connected to the central port, and the valves are configured to allow refrigerant flow into the system.
8 More Details
For instructions in your language, scan the QR code. Instructions are available in English, German, French, Italian, Spanish, Portuguese, Dutch, Turkish, and Chinese.
9. Family Products
Explore other products in the Elitech family. The Elitech MS-800 series manifold instruments are part of a broader range of HVAC and refrigeration tools offered by Elitech. The product image displays various related devices, including vacuum pumps and other digital gauges.