Elitech LogEt 8 Series Data Logger User Manual

Model Numbers: LogEt 8, LogEt 8 TE, LogEt 8 TH, LogEt 8 THE, LogEt 8 PTE, LogEt 8 UTE

Manufacturer: Elitech Technology, INC. / Jiangsu Jingchuang Electric Co., Ltd.

Overview

The Elitech LogEt 8 series data loggers are designed for comprehensive monitoring across various stages of storage and cold chain logistics. Applications include refrigerated containers/trucks, cooler bags, cooling cabinets, medical cabinets, freezers, and laboratories. These loggers feature an intuitive LCD screen and two-button interface, supporting multiple start/stop modes, configurable threshold settings, and two data storage options (stop when full & cyclic record). They automatically generate PDF reports, allowing users to review data without requiring additional software.

Model Selection

Model Type Measurement Accuracy
LogEt 8 Internal Temperature -30°C~70°C (-22°F~158°F) Digital Temperature Sensor ±0.3°C/0.5°F
LogEt 8 TE Internal Temperature -40°C~85°C (-40°F~185°F) Digital Temperature Sensor ±0.3°C/0.5°F
LogEt 8 TH Internal Temperature and Humidity Temperature: -30°C~70°C (-22°F~158°F)
Humidity: 0%RH~100%RH
Digital Temperature and Humidity Sensor
Temperature: ±0.3°C (-20°C~40°C); ±0.5°F (-4°F~104°F)
±0.5°C (-50°C~85°C); ±0.9°F(-58°F~185°F)
±1°C (-100°C~150°C); ±1.8°F (-148°F~302°F)
±2°C/3.6°F (others)
Humidity: ±3%RH (25°C: 20%RH~80%RH), ±5%RH (others)
LogEt 8 THE External Temperature and Humidity Temperature: -40°C~85°C (-40°F~185°F)
Humidity: 0%RH~100%RH
Digital Temperature and Humidity Sensor
Temperature: ±0.3°C (-20°C~40°C); ±0.5°F (-4°F~104°F)
±0.5°C (-50°C~85°C); ±0.9°F(-58°F~185°F)
±1°C (-100°C~150°C); ±1.8°F (-148°F~302°F)
±2°C/3.6°F (others)
Humidity: ±3%RH (25°C: 20%RH~80%RH), ±5%RH (others)
LogEt 8 PTE External Low Temp -85°C~150°C (-121°F~302°F) Platinum Resistance Temperature Sensor PT100
Temperature: ±0.5°C (-50°C~85°C); ±0.9°F(-58°F~185°F)
±1°C (-100°C~150°C); ±1.8°F (-148°F~302°F)
LogEt 8 UTE External Ultra-low Temp -196°C~150°C (-320.8°F~302°F) Platinum Resistance Temperature Sensor PT100
Temperature: ±0.5°C (-50°C~85°C); ±0.9°F(-58°F~185°F)
±1°C (-100°C~150°C); ±1.8°F (-148°F~302°F)

Specifications

Operation

  1. Install Software: Please download and install the free ElitechLog software (macOS and Windows) from www.elitechlog.com/softwares.
  2. Configure Parameters:

    First, connect the data logger to the computer's USB port and wait until the [USB connected] icon shows on the LCD. Then, configure parameters using the ElitechLog software:

    • If default parameters are acceptable: Click "Quick Reset" under the "Summary" menu to synchronize local time before usage.
    • If parameters need changing: Click the "Parameter" menu, enter your preferred values, and click "Save Parameter" to complete the configuration.

    Warning! For first-time users or after battery replacement: To avoid time or time zone errors, ensure you click "Quick Reset" or "Save Parameter" before usage to sync and configure your local time into the logger.

  3. Start Logging:
    • Press Button: Press and hold the left button for 5 seconds until the [logging] icon shows on the LCD, indicating the logger has started logging.
    • Auto Start:
      • Immediate Start: The logger starts logging immediately after being unplugged from the computer.
      • Timed Start: The logger starts a countdown after removal from the computer and will begin logging automatically at the set date/time.

    Note: If the [start delay] icon keeps flashing, it indicates the logger is configured with a start delay; logging will commence after the set delay time elapses.

  4. Mark Events: Press and hold the left button for 5 seconds to mark the current temperature and time. Up to 10 groups can be marked. After marking, the LCD will display the currently marked groups.
  5. Stop Logging:
    • Press Button: Press and hold the right button for 5 seconds until the [stop] icon shows on the LCD, indicating the logger has stopped logging.
    • Auto Stop: The logger stops automatically when the recorded data points reach the maximum memory capacity.
    • Use Software: Open the ElitechLog software, navigate to the "Summary" menu, and click the "Stop Logging" button.

    Note: Stopping via the button is the default. If disabled, this function is invalid, and you must use the ElitechLog software's "Stop Logging" button.

    Note: The "Auto Stop" function is automatically disabled if "Circular Logging" is enabled.

  6. Download Data:

    Connect the data logger to your computer's USB port and wait for the [USB connected] icon to appear on the LCD. Then, download data:

    • Without ElitechLog Software: Locate and open the "ElitechLog" removable storage device. Save the auto-generated PDF report to your computer for viewing.
    • With ElitechLog Software: After the logger auto-uploads data to the ElitechLog software, click "Export" and select your preferred file format. If auto-upload fails, manually click "Download" and repeat the process.
  7. Re-use the Logger:

    To reuse a logger, stop it first. Then, connect it to your computer and use the ElitechLog software to save or export the data. Next, reconfigure the logger by repeating the steps in "2. Configure Parameters". Finally, follow "3. Start Logging" to restart the logger for new logging sessions.

    Warning! Reconfiguration will delete all previous logging data. Save or export data before reconfiguring. If data was not saved, try locating it in the "History" menu of the ElitechLog software.

  8. Repeat Start: To restart a stopped logger, you can press and hold the left button to quickly start logging without reconfiguration.

    Note: Always back up data before restarting by repeating "6. Download Data".

Status Indication

1. Buttons

Operation Function
Press and hold the left button for 5s Start logging
Press and hold the right button for 5s Stop logging
Press and release the left button Check/Switch interfaces
Press and release the right button Back to main menu
Press and hold the left button for 5s (under logging status) Mark events

2. LCD Screen

Function Display Meaning
High limit alarm [alarm] Indicates a high-temperature or high-humidity alarm has been triggered.
Low limit alarm [alarm] Indicates a low-temperature or low-humidity alarm has been triggered.
Working status [logging] / [stop] / [start delay] Shows if the logger is logging, stopped, or in a start delay period.
Battery Level [battery] Indicates the remaining battery power.
Temperature/Humidity Value e.g., 25.5°C / 60%RH Displays the current measured temperature and/or humidity.
Date e.g., 2023-10-27 Displays the current date.
Time e.g., 10:30:00 Displays the current time.

3. LCD Interface

Meaning Display Meaning Display
1. Timed start [timed start] 15. Low humidity limit e.g., 20%RH
Delay start [delay start] 16. Maximum humidity e.g., 100%RH
2. Not start [not started] 17. Minimum humidity e.g., 0%RH
3. Started [started] 18. Average humidity e.g., 50%RH
4. Recorded points e.g., 16000 19. Current date & time e.g., 2023-10-27 10:30
5. Maximum temperature e.g., 70°C 20. Sensor error [sensor error]
6. Minimum temperature e.g., -30°C 21. PDF generation progress e.g., 50%
7. MKT value e.g., 25°C 22. No alarm [no alarm]
8. Average temperature e.g., 20°C 23. Alarmed [alarm]
9. High temperature limit 3 e.g., 60°C 24. Marked [marked]
10. High temperature limit 2 e.g., 50°C 25. Data deleted [data deleted]
11. High temperature limit 1 e.g., 40°C 26. USB communicating [USB communicating]
12. Low temperature limit 1 e.g., 2°C
13. Low temperature limit 2 e.g., -20°C
14. High humidity limit e.g., 90%RH

Battery Replacement

  1. Remove the screws, open the battery cover, and take out the old battery.
  2. Install a new battery with the negative (-) terminal towards the spring.
  3. Mount the battery cover and tighten the screws.

Warning! Do not remove the battery if the logger is logging. The LCD screen will automatically turn off after 15 seconds of inactivity (default). Press a button to turn it back on. Do not use the logger for long-distance transport if the battery is not full.

What's Included

Default Parameter Configurations

Parameter Default Parameter Default
Logging Interval 10 min High Temperature Limit H3 /
Start Mode Press Button High Temperature Limit H2 /
Start Delay Enable High Temperature Limit H1 8°C
Stop Mode - Press Button Enable Low Temperature Limit L1 2°C
Temporary PDF Enable Low Temperature Limit L2 /
Temperature unit C Calibration Temperature 0°C
Time Zone UTC+0:00 High Humidity Limit /
Repeat Start Disable Low Humidity Limit /
Circular Logging Disable Calibration Humidity /
Alarm Mode Multiple Alarms
Alarm Type Single
Alarm delay /

Warning! Please store your logger at room temperature. If you use the logger for the first time, please use ElitechLog software to synchronize system time and configure parameters. Any parameter configuration on ElitechLog software will delete all logged data inside the logger. Please save data before applying any new configurations.

Technical Requirements (for printing/manual)

Color Requirements

PDF preview unavailable. Download the PDF instead.

LogEt-8-ute Adobe PDF library 10.01

Related Documents

Preview Elitech LogEt 5 Series User Manual: Temperature and Humidity Data Loggers
Comprehensive user manual for the Elitech LogEt 5 Series temperature and humidity data loggers. Covers overview, model selection, specifications, operation, status indication, battery replacement, and default configurations.
Preview Elitech LogEt 5 Series Data Logger Specifications and User Manual
Comprehensive specifications, features, and operating instructions for the Elitech LogEt 5 series of temperature and humidity data loggers, designed for cold chain monitoring.
Preview Elitech IoT & Cold Chain Data Loggers | Comprehensive Product Catalog
Explore Elitech's extensive range of IoT and cold chain data loggers, including temperature, humidity, and multi-use devices. Discover features, technical specifications, and software solutions for reliable monitoring and compliance.
Preview Elitech LogEt 1 Series User Manual: Temperature and Humidity Data Loggers
Comprehensive user manual for the Elitech LogEt 1 series of single-use data loggers, covering LogEt 1 (temperature), LogEt 1TH (temperature-humidity), and LogEt 1 Bio (high accuracy temperature). Includes specifications, operation, status indication, and warnings.
Preview Elitech LogEt 1 Series Single-Use PDF Data Logger User Manual
User manual for the Elitech LogEt 1, LogEt 1TH, and LogEt 1Bio single-use PDF data loggers, covering appearance, technical specifications, indication instructions, and operation.
Preview Elitech LogEt 8 TH Temperature and Humidity Data Logger | Withnell Sensors
Explore the Elitech LogEt 8 TH, a high-accuracy temperature and humidity data logger for sensitive products in life sciences and cold chain logistics. Features include 16,000 recordings, PDF reports, and USB connectivity.
Preview Elitech LogEt 6 Ultra-Low Temperature Data Logger - Specifications and Manual
Comprehensive guide for the Elitech LogEt 6 ultra-low temperature data logger, covering safety, specifications, operation, software, and warranty. Ideal for cold chain monitoring and ultra-low temperature applications.
Preview Elitech LogEt 5i User Manual: Single-Use IoT Data Logger
Comprehensive user manual for the Elitech LogEt 5i single-use IoT data logger. Covers product features, technical specifications, operating instructions, safety precautions, and data management for temperature, humidity, and light monitoring in logistics and cold chain applications.