Elitech LogEt 5 Series Multilingual Specification

The Elitech LogEt 5 series data loggers are designed for comprehensive temperature and humidity monitoring across various stages of storage and cold chain logistics. They are suitable for use in refrigerated containers, trucks, cooler bags, medical cabinets, freezers, and laboratories.

These loggers feature an LCD screen and a dual-button design to prevent accidental activation. They support multiple start and stop modes, configurable alarm thresholds, and two data storage options: stop when full or cyclic record. Users can easily check data through automatically generated PDF reports without the need for additional software.

Product Overview

The LogEt 5 series offers various models with different sensor types and measuring ranges:

ModelSensor TypeMeasuring Range
LogEt 5 TInternal Temperature-30°C~70°C
LogEt 5 TEExternal Temperature-40°C~85°C
LogEt 5 THInternal Temperature and Humidity-30°C~70°C, 0%RH~100%RH
LogEt 5 THEExternal Temperature and Humidity-40°C~85°C, 0%RH~100%RH
LogEt 5 TLEExternal Cryogenic Temperature-196°C~150°C

Key components include:

  • ① USB Port
  • ② LCD Screen
  • ③ Button
  • ④ Internal Sensor
  • ⑤ External Sensor

Technical Specifications

  • Sensor Accuracy: Temperature: ±0.3°C (-20°C~40°C), ±0.5°C (-50°C~85°C), ±1°C (-100°C~150°C), ±2°C (others); Humidity: ±3%RH (25°C: 20%RH~80%RH), ±5%RH (others)
  • Resolution: Temperature: 0.1°C/0.1°F; Humidity: 0.1%RH
  • Memory: 32,000 points (max)
  • Logging Interval: 10 seconds ~ 24 hours
  • Start Mode: Press button or use software
  • Stop Mode: Press button, use software, or auto stop
  • Alarm Threshold: Configurable; Temperature: up to 3 high limits and 2 low limits; Humidity: 1 high limit and 1 low limit
  • Alarm Type: Single, cumulative
  • Alarm Delay: 10 seconds ~ 24 hours
  • Data Interface: USB port
  • Report Type: PDF data report
  • Battery: 3.0V disposable lithium battery CR2450
  • Battery Life: 2 years (storage and use)
  • Protection Level: IP65
  • External probe length: 1.2m
  • Dimensions: 97mm×43mm×12.5mm (LxWxH)

Operation Instructions

1. Install Software

Download and install the free ElitechLog software (macOS and Windows) from www.elitechlog.com/softwares.

2. Configure Parameters

Connect the data logger to the computer's USB port. Configure parameters via the ElitechLog software. For first-time use or after battery replacement, ensure local time is synchronized by clicking 'Quick Reset' or 'Save Parameter'.

3. Start Logging

Press Button: Press and hold the left button for 5 seconds. The logger starts logging.

Auto Start: Immediate Start (starts upon disconnection) or Timed Start (starts at a preset date/time).

4. Mark Events

Double-click the left button to mark the current temperature and time (up to 10 groups).

5. Stop Logging

Press Button: Press and hold the right button for 5 seconds. The logger stops logging.

Auto Stop: Stops automatically when memory is full.

Use Software: Open ElitechLog software and click 'Stop Logging'.

6. Download Data

Connect the logger to the computer via USB. A PDF report generates automatically. Data can also be exported via the ElitechLog software.

7. Re-use the Logger

After stopping, save or export data. Reconfigure parameters and restart logging.

8. Repeat Start

Press and hold the left button to quickly restart a stopped logger without reconfiguring.

Status Indication

1. Button Functions

  • Press and hold left button (5s): Start logging
  • Press and hold right button (5s): Stop logging
  • Press and release left button: Check/Switch interfaces
  • Press and release right button: Back to main menu
  • Double click left button: Mark events (under logging status)

2. LCD Screen Indicators

The LCD screen displays various information including temperature/humidity values, alarm status, working status, battery level, date, time, and logging progress.

3. LCD Interface Icons

Various icons indicate logging status, alarms, USB communication, and parameter configuration.

4. LCD-LED Indication

  • Not Started: Both LEDs flash once.
  • Started: LED flashes to indicate status (e.g., alarm, no alarm).
  • Mark Data: LED flashes twice.
  • Stop Logging: LED flashes 5 times.

Battery Replacement

(1) Turn the battery cover counter-clockwise and remove it.

(2) Insert a new battery with the '+' side up.

(3) Turn the battery cover clockwise to tighten.

What's Included

  • Temperature (Humidity) Data Logger x 1
  • User Manual x 1
  • Certificate of Verification x 1

Default Parameter Configurations

ParameterDefaultParameterDefault
Logging Interval10 minAlarm Delay/
Start ModePress ButtonHigh Temperature Limit H3/
Start Delay/High Temperature Limit H2/
Stop Mode - Press ButtonEnableHigh Temperature Limit H18°C
Temporary PDFEnableLow Temperature Limit L12°C
Temperature Unit°CLow Temperature Limit L2/
Time ZoneUTC+0:00Calibration Temperature0°C
Repeat StartDisableHigh Humidity Limit/
Circular LoggingDisableLow Humidity Limit/
Alarm ModeMultiple AlarmsCalibration Humidity/
Alarm TypeSingle
Models: LogEt 5 Series, LogEt 5 Series USB Data Logger, USB Data Logger, Data Logger

PDF preview unavailable. Download the PDF instead.

Loget 5 instructions Adobe PDF library 17.00

Related Documents

PreviewElitech LogEt 1 Series User Manual: Temperature and Humidity Data Loggers
Comprehensive user manual for the Elitech LogEt 1 series of single-use data loggers, covering LogEt 1 (temperature), LogEt 1TH (temperature-humidity), and LogEt 1 Bio (high accuracy temperature). Includes specifications, operation, status indication, and warnings.
PreviewElitech IoT & Cold Chain Data Loggers | Comprehensive Product Catalog
Explore Elitech's extensive range of IoT and cold chain data loggers, including temperature, humidity, and multi-use devices. Discover features, technical specifications, and software solutions for reliable monitoring and compliance.
PreviewElitech LogEt 1 Series Single-Use PDF Data Logger User Manual
User manual for the Elitech LogEt 1, LogEt 1TH, and LogEt 1Bio single-use PDF data loggers, covering appearance, technical specifications, indication instructions, and operation.
PreviewElitech LogEt 6 Ultra-Low Temperature Data Logger - Specifications and Manual
Comprehensive guide for the Elitech LogEt 6 ultra-low temperature data logger, covering safety, specifications, operation, software, and warranty. Ideal for cold chain monitoring and ultra-low temperature applications.
PreviewElitech LogEt 1 TH Disposable PDF Data Logger: Temperature and Humidity
A comprehensive overview of the Elitech LogEt 1 TH, a single-use PDF data logger for temperature and humidity monitoring. Features include a wide temperature range (-30°C to +70°C), humidity range (0% to 100% RH), and a capacity of 16,000 records. Ideal for pharmaceuticals, food, and other temperature-sensitive products during storage and transport.
PreviewElitech Glog 5 Data Logger: Specifications and User Guide
Explore the Elitech Glog 5, a versatile data logger designed for precise environmental monitoring. This document provides comprehensive specifications, operational instructions, and safety guidelines for temperature, humidity, light, and vibration logging applications, available in multiple languages.
PreviewElitech RC-5+ Temperature Data Logger: User Manual and Specifications
This manual provides detailed instructions for using the Elitech RC-5+ USB temperature data logger, covering setup, operation, data download, and analysis with Elitech software. Features and technical specifications are also included.
PreviewElitech RC-5 Series User Manual: Temperature Data Loggers
Comprehensive user manual for the Elitech RC-5, RC-5+, and RC-5+TE temperature data loggers. Learn about specifications, operation, battery replacement, and software installation for accurate temperature monitoring.