Elitech LogEt 1 Series Single-Use PDF Data Logger User Manual
User Manual LogEt 1 / LogEt 1TH / LogEt 1Bio
Appearance
1. USB Protective Cover
2. LCD Screen
3. Button
4. LED Indicator
5. Light Sensor
6. Humidity Sensor
7. Shelf life
Diagram Description: A visual representation of the data logger showing the location of the USB protective cover, LCD screen, button, LED indicator, light sensor, and humidity sensor. The shelf life is also indicated.
Indication Instructions
Status Indication
- Start Recording: Rec......
- Stop Recording: Stop...
- Mark Success: Mark
- Mark Failure: Mark...
LED Blinks:
- Not Started: No LED blink
- Delayed Start/Timing Start: Green LED blinks once every 10 seconds
- Started - OK: Green LED blinks once every 10 seconds
- Started - Alarm: Red LED blinks once every 10 seconds (Green LED stops blinking)
- Stopped - OK: No LED blink
- Stopped - Alarm: Red LED blinks once every 10 seconds (Green LED stops blinking)
LCD Screen Indication
1. Current Temperature & Humidity: Displays current temperature and humidity readings.
2. Recorded Points: Shows the number of data points recorded.
3. Maximum Temperature & Humidity: Displays the maximum recorded temperature and humidity.
4. Minimum Temperature & Humidity: Displays the minimum recorded temperature and humidity.
5. System Date: Month-Day
6. System Time: Hour: Minute
7. Battery indicator: Shows the remaining battery life.
8. Alarm threshold: Indicates if any alarm thresholds have been met (e.g., AH1, AH2, AH3, AL1, AL2).
Diagram Description: The LCD screen displays various information including current, min, and max temperature and humidity, recorded points, date, time, battery status, and alarm status. The alarm status is indicated by codes like AH1, AH2, etc.
Technical Specifications
Parameter | Specification |
---|---|
Temperature Range | -30°C~70°C, 0%RH~100%RH |
Temperature Accuracy | ±0.5°C (-20°C ~ +40°C), others ±1.0°C ±0.3°C (-30 °C ~ +70°C) - for LogEt 1Bio only |
Humidity Accuracy | ±3%RH (20%RH~80%RH), others ±5%RH - for LogEt 1TH only under 25°C |
Resolution | 0.1°C, 0.1%RH |
Data Storage Capacity | Max. 16,000 points |
Shelf Life / Battery | 2 years / CR2450 button cell |
Recording Interval | 12 minutes (Default, others on request) |
Recording Duration | Up to 120 days (Default, others on request) |
Start Mode | Button or software |
Stop Mode | Button, software or stop when full |
Protection Class | IP67 (Not for LogEt 1TH) |
Reprogrammable | Via Elitech software |
Single Use | Via Online Configuration Web |
Certifications | EN12830, CE, RoHS |
Validation Certificate | As Hardcopy |
Software | ElitechLog Data Management Software |
Compatible OS | Windows XP/7/10, Mac OS 10.10 or higher |
Report Generation | Automatic PDF report |
Password Protection | Software Password Protection |
Connection Interface | USB 2.0 (Standard Type A Connector) |
Alarm Configuration | Optional, Up to 5 thresholds |
Notes:
- Shelf life and battery life are based on optimal storage conditions (15°C to 23°C / 45% to 75% RH).
- Recording duration may be shortened at very low or high temperatures.
Operations
- Configuration: Connect the data logger to a computer. Configure parameters and synchronize time using Elitech Log software or the online configuration site.
- Start Recording: Press and hold the button for 5 seconds to start recording.
- Mark Events: Double-click the button quickly to mark the current time and temperature.
- Stop Recording: Press and hold the button for 5 seconds to stop recording.
- View Data: Connect the logger to a computer. Open the PDF report in the "Elitech Log" removable storage disk or use the ElitechLog software to view data.
Important Notes:
- Read the User Manual carefully before use.
- The data logger is ideal for monitoring temperature (and humidity for LogEt 1TH) during storage and transportation of pharmaceuticals and other products.
- System time may not be synchronized via the online configuration site.
- If Start Delay is set, recording begins after the delay time.
- If Start Mode is set to Immediate Start, recording begins immediately after removal from the computer.
- If Start Mode is set to Timing Start, recording begins automatically at the scheduled date and time.
- Loggers do not need to be stopped for data viewing.
- The logger stops automatically when the recording points reach the set limit.
- Store the data logger at room temperature.
- Humidity parameters are only for the LogEt 1TH model.
- The logger cannot be re-configured after it is started.
- The LCD screen turns off automatically after 15 seconds of inactivity; a single click of the button will turn it back on.
- After starting, the green LED blinks every 10 seconds. If alarms are triggered, the red LED blinks every 10 seconds, and the green LED stops.
- If the battery indicator shows only half an icon, do not use the logger for long-distance transport.
- Avoid contact with unstable chemical solvents or organic compounds, especially ketene, acetone, ethanol, isopropanol, toluene, etc.
Temperature Monitoring
CAUTION
SHIPPER:
CONTAINER NO.:
TRUCK NO.:
B/L NO.:
REF NO.:
LOGGER SERIAL NO.:
START DATE:
START TIME:
DEPARTURE PORT:
ARRIVAL PORT:
CONTENTS:
TEMPERATURE REQUIRED: °C °F
Diagram Description: A form for recording temperature monitoring details, including shipper information, container/truck numbers, bill of lading number, reference number, logger serial number, start/departure details, arrival port, contents, and required temperature.