RC-51H User Manual

Multi-use Temperature & Humidity Data Logger

Product Overview

This temperature and humidity data logger is primarily used in fields such as medicine, food, life science, flowers, breeding industry, ice chests, containers, shady cabinets, medical cabinets, refrigerators, laboratories, and greenhouses. The RC-51H is plug-and-play, allowing users to generate a data report directly without needing to install data management software. The data can still be accessed even if the battery runs out.

Structure Description

The RC-51H data logger consists of the following components:

Product Label Information: Typically includes Model (RC-51H), Version (e.g., 1.0), Serial Number (S/N), Barcode, Firmware Version, and Certification information (e.g., EN 12830-S, T, C, D, I, CE, RoHS).

Parameter Instruction

Users can reconfigure parameters using the data management software. Note that reconfiguring will clear existing data.

Operating Instructions

The data logger can be started and stopped via button press or software. For detailed operations, refer to the menu and status indicators.

Logger Status Indicators:

StatusAction of the indicatorsStatusAction of the indicators
Not startedThe red and green indicators flash 2 times simultaneously.Started-alarmThe red indicator flashes once. The red light flashes once per minute automatically.
Start delayThe red and green indicators flash once simultaneously.Stopped-normalThe green light flashes 2 times. The green light flashes once per minute automatically.
TimingThe green indicator flashes once.Stopped-alarmThe red light flashes 2 times.
Started-normalThe green indicator flashes once.

Starting and Stopping the Logger:

ActionOperationLCD IndicatorIndicator Status
Instant-onDisconnect to USB[EC]Green indicator flashes 5 times.
Timing startDisconnect to USB[EC]Green indicator flashes 5 times.
Manual startPress and hold for 5s[EC]Green indicator flashes 5 times.
Manual start (delayed)Press and hold for 5s[EACE]Green indicator flashes 5 times.
Manual stopPress and hold for 5s[EC]Red indicator flashes 5 times.
Over-Max-record-capacity (disable manual stop)Reach the Max capacity[EC]Red indicator flashes 5 times.
Over-Max-record-capacity (enable manual stop)Reach the Max capacity or press and hold the button for 5s[EC]Red indicator flashes 5 times.

Viewing Data: When the data logger is connected to a computer's USB port, a data report is automatically generated. The red and green indicators flash alternately during report creation. The LCD shows the PDF report creation progress. Once created, both indicators light up simultaneously, and the data report can be viewed. Document creation takes up to 4 minutes.

Steps to View Data:

  1. Rotate the transparent cap in the direction of the arrow [arrow-right] and remove it.
  2. Insert the data logger into the computer's USB port.
  3. View the data report (e.g., XXX.pdf).

Technical Specifications

ParameterSpecificationParameterSpecification
Recording OptionsMulti-UseCertificationsEN12830, CE, RoHS
Temperature Range-30°C to 70°CValidation Certificate HardcopyOptional on request
Humidity Range10%~95%Report GenerationAutomatic PDF report
Temperature & Humidity Accuracy±0.5°C (-20°C/+40°C); ±1.0°C (other range)
±3%RH (25°C, 20%~90%RH), ±5%RH (other range)
Temperature & Humidity Resolution0.1°C (Temperature)
0.1%RH (Humidity)
Data Storage Capacity32,000 readingsPassword ProtectionOptional on request
SoftwarePDF/ElitechLog Win or Mac (latest version)ReprogrammableWith free Elitech Win or MAC software
Connection InterfaceUSB 2.0, A-TypeAlarm ConfigurationOptional, up to 5 points, Humidity only supports upper and lower limit alarm
Shelf Life/Battery2 years¹ / ER14250 button cellDimensions131mm x 24mm x 7mm (LxD)
Recording Interval15 minutes (standard)
Startup ModeButton or software
Stop ModeButton, software or stop when full
Weight60g

¹ Depending on optimal storage conditions (±15°C to +23°C / 45% to 75% rH).

Report Description

The data logger generates an automatic PDF report upon connection to a computer. The report includes various sections detailing the logged data and device configuration.

Report Sections:

Indicator Status Descriptions in Report:

Combined Indicators and Other Status:

Note: The data shown in the report is for explanation purposes; refer to the actual document for specific information.

Menu and Status Indicator

Description of the indicator flashing status:

StatusAction of the indicatorsStatusAction of the indicators
Not startedThe red and green indicators flash 2 times simultaneously.Started-alarmThe red indicator flashes once. The red light flashes once per minute automatically.
Start delayThe red and green indicators flash once simultaneously.Stopped-normalThe green light flashes 2 times. The green light flashes once per minute automatically.
TimingThe green indicator flashes once.Stopped-alarmThe red light flashes 2 times.
Started-normalThe green indicator flashes once.

Description of the menus:

MenuDescriptionExampleMenuDescriptionExample
1Countdown of the (timing) start03:426Average humidity value4.5
2Countdown of the (delayed) start03:427Maximum temperature value9.8
3Current temperature value2.38Maximum humidity value9.8
4Points of the records320009Minimum temperature value2.7
5Average temperature value4.810Minimum humidity value2.7

Description of the combined indicators and other status:

DisplayDescriptionDisplayDescription
(group) Rate of progress (record stop time)Shows the progress of data logging or report generation.(group) Minimum valueDisplays the minimum recorded value.
(group) Document creation timeTime taken for report creation.(group) Maximum valueDisplays the maximum recorded value.
(group) Actual stop mode (different from the item C)The mode in which the logger was stopped.(group) Clear dataIndicates data has been cleared.
(group) No alarmIndicates no alarm condition is active.In USB communicationShows the logger is communicating via USB. The '2' indicator should be in a state of blinking.
(group) Already alarmedIndicates a past or current alarm condition.

Note: The data shown in the report is for explanation purposes; refer to the actual document for specific information.

What's Included

Replace Battery

Follow these steps to replace the battery:

  1. Press the bayonet in the direction of the arrow [arrow-right] and remove the battery cover.
  2. Place a new battery.
  3. Install the battery cover in the direction of the arrow [arrow-left].

PDF preview unavailable. Download the PDF instead.

rc-51h-instructions Adobe PDF library 10.01 Adobe Illustrator(R) 16.0

Related Documents

Preview Elitech RC-4HC Temperature and Humidity Data Logger Specifications
Comprehensive specifications and features for the Elitech RC-4HC PDF temperature and humidity data logger, detailing its capabilities for cold chain monitoring, industrial applications, and data export.
Preview Elitech RC-51H Multi-use Temperature & Humidity Data Logger User Manual
User manual for the Elitech RC-51H multi-use temperature and humidity data logger. Covers product overview, structure, parameter settings, operating instructions, technical specifications, indicator status, menu functions, battery replacement, and report details.
Preview Elitech RC-51H Temperature and Humidity Data Logger User Manual
Comprehensive user manual for the Elitech RC-51H multi-use temperature and humidity data logger. Covers product overview, structure, parameters, operation, technical specifications, and report generation.
Preview Elitech RC-61/GSP-6 User Manual: Temperature and Humidity Data Logger
This user manual provides comprehensive guidance for the Elitech RC-61/GSP-6 temperature and humidity data logger, covering its features, specifications, operation, and important usage notes for cold chain monitoring applications.
Preview Elitech RC-51 Multi-use Temperature Data Logger: Specifications & User Guide
Comprehensive technical specifications, usage instructions, and overview for the Elitech RC-51 multi-use temperature data logger, designed for monitoring food, pharmaceuticals, and chemicals in storage and transportation.
Preview Elitech IoT & Cold Chain Data Loggers | Comprehensive Product Catalog
Explore Elitech's extensive range of IoT and cold chain data loggers, including temperature, humidity, and multi-use devices. Discover features, technical specifications, and software solutions for reliable monitoring and compliance.
Preview Elitech RC-4 Series Temperature and Humidity Data Loggers - User Manual
Comprehensive user manual for Elitech RC-4 and RC-4HC series temperature and humidity data loggers. Covers specifications, operation, setup, and troubleshooting for cold chain monitoring.
Preview Elitech RC-51H User Manual: Temperature & Humidity Data Logger
Comprehensive user manual for the Elitech RC-51H multi-use temperature and humidity data logger. Covers product overview, structure, parameters, operating instructions, technical specifications, report generation, and battery replacement.