OMRON Automatic Blood Pressure Monitor
Model: JPN610T
Approval No.: 衛部醫器輸字第033943號
This product involves the collection, processing, and utilization of personal data, and complies with the Personal Data Protection Act.
1. Introduction
Thank you for choosing the OMRON Electronic Blood Pressure Monitor. This monitor measures blood pressure using the oscillometric method. It detects the pulse oscillations in your brachial artery and converts these oscillations into a digital value.
2. Important Safety Information
This user manual provides important information for the OMRON Electronic Blood Pressure Monitor. To ensure safe and correct use, please read and understand all instructions. If you have any questions or concerns regarding these instructions before using this monitor, please consult your local OMRON representative. For detailed information about your personal blood pressure, please consult your physician.
2.1 Warnings
WARNING indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
- Do not use this product on infants, children, or individuals who cannot express their consent.
- Do not adjust your medication based on the readings from this blood pressure monitor. Always take medication as prescribed by your doctor. Only a physician is qualified to diagnose and treat high blood pressure.
- Do not use this blood pressure monitor on an arm with an injury or a treated arm.
- Do not apply the cuff over clothing or while an intravenous infusion or blood transfusion is taking place.
- Do not use this blood pressure monitor in areas with high-frequency (HF) surgical equipment, Magnetic Resonance Imaging (MRI) or Computed Tomography (CT) scanners, as this may cause the monitor to malfunction or result in inaccurate readings.
- Do not use this blood pressure monitor in an oxygen-rich environment or near flammable gases.
- If you have any of the following conditions, consult your doctor before using this device: frequent premature beats or other arrhythmias, arteriosclerosis, poor perfusion, diabetes, pregnancy, pre-eclampsia, kidney disease. Please note that these conditions, as well as patient movement, shivering, or trembling, may affect the measurement readings.
- Never self-diagnose or self-treat based on the measurement readings. Always consult your doctor.
- To prevent suffocation, keep the air tube and AC adapter cable away from infants and children.
- This product contains small parts. If accidentally swallowed by infants or children, it may pose a choking hazard.
Data Transmission
This product transmits radio frequency (RF) in the 2.4 GHz band. Do not use this product in locations where RF is restricted, such as on an airplane or in a hospital. In RF-restricted areas, turn off the Bluetooth function on this blood pressure monitor and remove the batteries and/or AC adapter.
2.2 Cautions
CAUTION indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate personal injury or damage to property.
- If you experience skin irritation or discomfort, discontinue use of the blood pressure monitor and consult your doctor.
- Measurement with this blood pressure monitor may cause temporary interference with blood flow for individuals with: vascular access, intravascular therapy, or arteriovenous (A-V) shunt; it may also cause injury. Consult your doctor before use.
- If you have had a mastectomy, consult your doctor before using this blood pressure monitor.
- If you have severe blood flow problems or blood disorders, bruising may occur due to cuff pressure. Consult your doctor before using this blood pressure monitor.
- Do not take unnecessary measurements, as this may cause bruising due to impaired blood flow.
- Do not inflate the cuff unless it is properly wrapped around the arm.
- If the cuff does not deflate, remove the cuff.
- The unit may become warm during a malfunction. If this occurs, do not touch the unit.
- The monitor is intended for measuring blood pressure only. Do not use it for any other purpose.
- During measurement, ensure that no mobile devices or other electronic devices emitting electromagnetic fields are within 30 cm of the blood pressure monitor, as this may cause the monitor to malfunction or result in inaccurate readings.
- Do not disassemble or attempt to repair the blood pressure monitor or accessories, as this may result in incorrect measurement readings.
- Do not use the blood pressure monitor in wet conditions or in areas where water may splash onto it, as this may damage the unit.
- Do not operate the blood pressure monitor in moving vehicles (e.g., cars) or airplanes.
- Avoid dropping the blood pressure monitor or subjecting it to strong impacts or vibrations.
- Avoid placing the blood pressure monitor in areas of extreme humidity, dryness, or temperature. Refer to Section 5 of this manual.
- During measurement, monitor your arm to ensure the cuff does not cause long-term circulation impairment.
- Do not use this blood pressure monitor in settings where it will be used frequently, such as medical clinics or doctor's offices.
- Do not use this blood pressure monitor simultaneously with other medical electrical (ME) equipment, as this may cause malfunction or inaccurate readings.
- At least 30 minutes before taking a measurement, avoid bathing, drinking alcohol or caffeinated beverages, smoking, exercising, and eating.
- Rest for at least 5 minutes before measuring blood pressure.
- During measurement, remove any tight-fitting or thick clothing from your arm and remove any accessories.
- Remain still and do not talk during measurement.
- Do not use if your arm circumference is outside the cuff's applicable range.
- Before measuring blood pressure, ensure that this blood pressure monitor is at room temperature. Measuring blood pressure under extreme temperature changes may lead to inaccurate readings. If the blood pressure monitor was stored at the maximum or minimum storage temperature, it is recommended to leave it in the usage environment for approximately 2 hours to allow it to acclimatize to room temperature before use. For further information on usage environment and storage/transport temperatures, refer to Section 5 of this manual.
- Do not use this blood pressure monitor after its expiration date. Refer to Section 5 of this manual.
- Do not excessively fold or kink the cuff or air tube.
- Do not fold or kink the air tube during measurement. This may cause injury due to interrupted blood flow.
- To disconnect the air tube, pull the plastic connector at the base of the air tube, not the air tube itself.
- Only use the AC adapter, cuff, batteries, and accessories specified for this blood pressure monitor. Using incompatible AC adapters, cuffs, or batteries may cause damage and/or hazards to this blood pressure monitor.
- Only use an approved cuff for this blood pressure monitor. Using other cuffs may result in incorrect readings.
- Over-inflation to unnecessarily high pressure may cause injury to the arm wearing the cuff. Note: For more information, refer to "If Your Systolic Pressure Exceeds 210 mmHg" in Section 9 of the user manual.
Data Transmission
Do not change batteries or disconnect the AC adapter while data is being transmitted to your smart device. This may cause the blood pressure monitor to operate incorrectly and prevent your blood pressure data from being transmitted.
AC Adapter (Optional Accessory) Handling and Use
- Insert the AC adapter fully into the outlet.
- When unplugging the AC adapter from the outlet, always pull by the AC adapter body, not the cable.
- When handling the AC adapter cable:
- - Do not damage. / Do not break. / Do not modify.
- - Do not pinch. / Do not bend or pull forcefully. / Do not twist. If it is bundled, do not use.
- - Do not place heavy objects on it.
- Wipe dust from the AC adapter.
- Unplug the AC adapter when not in use.
- Unplug the AC adapter before cleaning this blood pressure monitor.
Battery Management and Use
- Insert batteries, aligning the positive (+) and negative (-) ends correctly.
- This blood pressure monitor uses only 4 "AA" alkaline or manganese batteries. Do not use other types of batteries. Do not mix new and old batteries. Do not mix batteries of different brands.
- If the blood pressure monitor will not be used for an extended period, remove the batteries.
- If battery fluid accidentally gets into your eyes, rinse immediately with plenty of water and consult your doctor.
- If battery fluid accidentally contacts your skin, wash immediately with plenty of clean, warm water. If irritation, injury, or pain persists, consult your doctor.
- Do not use expired batteries.
- Periodically check the batteries to ensure they are functioning properly.
2.3 General Precautions
- When measuring on the right arm, the air tube should be on the inner side of the elbow. Be careful not to rest your arm on the air tube.
- Blood pressure may differ between the right and left arms, and may result in different readings. Continue to use the same arm for measurements. If there is a significant difference between the readings on the right and left arms, consult your doctor to determine which arm to use for blood pressure measurement.
- When using the optional AC adapter, do not place the blood pressure monitor in a location where it is difficult to plug in or unplug the AC adapter.
- To stop a measurement, press the [START/STOP] button during measurement.
- Dispose of used batteries according to local regulations.
- The lifespan of the batteries included with the packaging may be shorter than that of new batteries.
3. Error Messages and Troubleshooting
If any of the following problems occur during measurement, check to ensure that no other electrical devices are within 30 cm of the blood pressure monitor. If the problem persists, refer to the table below.
Display/Problem | Possible Cause | Solution |
---|---|---|
E1 Display, or cuff air plug not fully inserted into the blood pressure monitor. | Cuff not wrapped correctly. Cuff air plug not fully inserted into the blood pressure monitor. | Press the [START/STOP] button again to turn off the blood pressure monitor. Firmly insert the air plug and wrap the cuff correctly, then press the [START/STOP] button. Firmly insert the air plug. |
E2 Display, or cuff not inflated properly. | Cuff not wrapped correctly. Cuff air leak. | Wrap the cuff correctly and take another measurement. Refer to Section 7(2) of the user manual. Replace the cuff with a new one. Refer to Section 13(2) of the user manual. |
E2 Display, or cuff not inflated properly. | You moved or talked during measurement, and the cuff did not inflate sufficiently. | Remain still and do not talk during measurement. If "E2" appears repeatedly, manually inflate the cuff until the systolic pressure is 30 to 40 mmHg higher than the previous reading. Refer to Section 9(2) of the user manual. |
E3 Display, or unable to measure due to systolic pressure exceeding 210 mmHg. | Systolic pressure exceeds 210 mmHg. | Do not touch the cuff and/or kink the air tube during measurement. If manually inflating the cuff, refer to Section 9 of the user manual. |
E4 Display, or movement or talking during measurement. Vibration interferes with measurement. | Movement or talking during measurement. Vibration interferes with measurement. | Remain still and do not talk during measurement. |
E5 Display, or irregular pulse not detected correctly. | Irregular pulse not detected correctly. | Wrap the cuff correctly and take another measurement. Refer to Section 7(2) of the user manual. Remain still and assume the correct sitting posture. |
Display | If the [heart symbol] symbol continues to display, we recommend consulting your doctor. | |
Er Display | Blood pressure monitor malfunction. | Press the [START/STOP] button again. If "Er" still appears, contact your local OMRON representative. |
E Display | Blood pressure monitor cannot connect to or transmit data to the smart device. | Follow the instructions displayed in the "OMRON connect" application. If "Err" still appears after checking the application, contact your local OMRON representative. |
P Flashing | Blood pressure monitor is waiting to be paired with the smart device. | Refer to Section 5(2) of the user manual to pair your blood pressure monitor with your smart device, or press the [START/STOP] button to cancel pairing and turn off the blood pressure monitor. |
P Solid | Blood pressure monitor is ready to transmit your readings to the smart device. | Open the "OMRON connect" application to transmit your readings. |
Display | Cannot transmit more than 80 readings. | Pair or transmit your readings to the "OMRON connect" application so you can save them in the application's memory, and this error symbol will disappear. |
Flashing | Date and time have not been set. | |
Display | Memory contains 100 readings waiting for transmission. | |
Display | Low battery power. | Replace all 4 batteries with new ones. Refer to Section 4(2) of the user manual. |
Display, or measurement unexpectedly shuts off during measurement. | Battery power depleted during measurement. Battery polarity not aligned correctly. | Immediately replace all 4 batteries with new ones. Refer to Section 4(2) of the user manual. Check if the batteries are installed correctly. Refer to Section 4(2) of the user manual. |
Display | No display on the blood pressure monitor screen. | Check if the batteries are installed correctly. Refer to Section 4(2) of the user manual. |
Display | Extremely high or low readings displayed. | Blood pressure fluctuates constantly; many factors affect blood pressure, including stress, time of day, and how tightly the cuff is wrapped. Review Section 2(2) of the user manual. |
Display | Any other communication problems. | Follow the instructions displayed in the smart device or visit the "Help" section in the "OMRON connect" application. If the problem persists, contact your local OMRON representative. |
Display | Any other problems. | Press the [START/STOP] button to turn off the blood pressure monitor, then turn it back on and take another measurement. If the problem persists, remove all batteries, wait 30 seconds, and then reinsert the batteries. If the problem still persists, contact your local OMRON representative. |
4. Maintenance
Follow the instructions below to prevent damage to this blood pressure monitor.
Note: Do not disassemble or attempt to repair the blood pressure monitor or accessories, as this may result in incorrect measurement readings.
4.2 Storage
- When not in use, store the blood pressure monitor in the storage bag.
- 1. Remove the cuff from the blood pressure monitor.
- Note: To disconnect the air tube, pull the plastic connector at the base of the air tube, not the air tube itself.
- 2. Gently fold the air tube into the cuff. Note: Do not excessively bend or kink the air tube.
- 3. Store the blood pressure monitor and other accessories in the storage bag.
- Store the blood pressure monitor and other accessories in a clean and safe location.
- Do not store this device and its accessories under the following conditions:
- - When the blood pressure monitor and accessories are wet.
- - In locations exposed to extreme temperatures, humidity, direct sunlight, dust, or corrosive gases (e.g., bleach).
- - In locations that may be subject to vibration or impact.
4.3 Cleaning
- Do not use detergents or volatile cleaners.
- Use a dry soft cloth or a soft cloth dampened with mild (neutral) detergent, then wipe dry with a dry cloth.
- Do not wash or immerse the blood pressure monitor, cuff, or other accessories in water.
- Do not use gasoline, thinner, or similar solvents to clean the blood pressure monitor, cuff, or other accessories.
4.4 Calibration and Servicing
The measurement accuracy of this blood pressure monitor has been thoroughly tested for long-term use.
It is generally recommended to recalibrate the unit every two years to ensure proper function and accurate measurements. Please consult your authorized OMRON dealer or OMRON customer service via the website or contact information provided in the packaging or accompanying leaflet.
5. Specifications
Description | Model | Details |
---|---|---|
Product Description | OMRON Electronic Blood Pressure Monitor | |
Model | JPN610T | |
Cuff Pressure Measurement Range | 0~299 mmHg | |
Blood Pressure Measurement Range | Systolic: 60~260 mmHg / Diastolic: 40~215 mmHg | |
Measurement Accuracy | Pressure: ±3 mmHg / Pulse: ±5% of displayed reading | |
Inflation Method | Electric pump inflation | |
Measurement Method | Oscillometric method | |
Wireless Communication | Bluetooth (BLE) | |
Operating Mode | Continuous operation | |
Deflation Method | Automatic air release valve | |
Frequency Range | 2.4GHz (2400~2483.5 MHz) / Modulation: GFSK | |
Effective Radiated Power | <20 dBm | |
IP Classification | Blood Pressure Monitor: IP20 / Optional AC Adapter: IP21 | |
Maximum Temperature of Body Surface | Below +43℃ | |
Power Source | DC6 V 4.0 W | |
Battery Life | Approx. 1000 measurements (using new alkaline batteries) | |
Service Life (Usage Lifespan) | Blood Pressure Monitor: 5 years / Cuff: 5 years / Optional AC Adapter: 5 years | |
Operating Environment Conditions | +10~+40℃ / 15~90% RH (non-condensing) / 800~1060 hPa | |
Storage/Transport Conditions | -20~+60℃ / 10~90% RH (non-condensing) | |
Weight (without batteries) | Blood Pressure Monitor: Approx. 340g / Cuff: Approx. 130g | |
Dimensions (Approximate) | Blood Pressure Monitor: 105mm (W) x 85mm (H) x 152mm (L) / Cuff: 145 mm x 466 mm (Air tube: 750 mm) | |
Arm Circumference Range for Blood Pressure Monitor Cuff | 220 to 420mm | |
Memory | Stores up to 100 readings | |
Contents | Blood pressure monitor, cuff (HEM-FL31), 4 "AA" batteries, storage bag, user manual (1 & 2), and setup guide | |
Protection Against Electric Shock | Internally powered ME equipment (when using batteries only) / Type BF applied part (cuff) / Class II ME equipment (with optional AC adapter) |
Note:
- This blood pressure monitor has undergone clinical investigation in accordance with ISO 81060-2:2013. In the clinical validation study, K5 was used to measure diastolic pressure in 85 subjects.
- This blood pressure monitor and the optional AC adapter are protected against solid foreign objects with a diameter of 12.5 mm or larger (e.g., fingers). The optional AC adapter is protected against dripping water at an angle, preventing problems during normal operation.
About Wireless Communication Interference
This product operates in the unlicensed ISM band of 2.4 GHz. If this product is used near other wireless devices operating in the same frequency band, such as microwave ovens or wireless LANs, interference may occur. If interference occurs, stop the operation of other devices or move this product away from other wireless devices before attempting to use it.
6. Guidance and Manufacturer's Statement
The pressure sensor in this product is manufactured in Japan and is a core component of OMRON blood pressure monitors.
Operation Method
Standard Accessories
JPN610T
HEM-FL31
Preparing for Measurement
30 Minutes Before Measurement:
- [No Alcohol Allowed]
- [No Smoking Allowed]
- [No Strenuous Exercise Allowed]
- [No Coffee Allowed]
- [No Eating Allowed]
5 Minutes Before Measurement: Relax and Rest
- Download the "OMRON connect" application.
- Install Batteries: [Diagram showing battery installation with '+' and '-' indicators].
- Pair with your Smart Device: [Diagram showing Bluetooth pairing process]. Follow Instructions: If the blood pressure monitor is paired with the application, the date and time will be set automatically.
- Manual Date and Time Setting: If the blood pressure monitor is paired with your smart device, the date and time will be set automatically. If you need to set them manually, set Year > Month > Day > Hour > Minute.
Setting Date and Time
[Diagram showing the device display for setting Year, Month, Day, Hour, Minute with navigation buttons labeled Back/Forward in multiple languages: Chinese (Traditional), Korean, Burmese, Hindi, Thai, Malay, Vietnamese, Indonesian].
5. Wrapping the Cuff on the Left Arm
- [Diagram showing cuff placement: Click sound indicates proper connection].
- [Diagram showing cuff position on the arm, ensuring the air tube is on the inner side of the elbow, 1-2 cm above the crease].
- [Diagram showing the cuff snugly wrapped around the arm, with the air tube positioned on the inner side of the arm].
- [Diagram showing the final cuff position, ensuring it is snug and the air tube is on the inner side].
The air tube should be positioned on the inner side of the elbow, 1-2 cm above the elbow crease.
Ensure the air tube is on the inner side of the arm and the cuff is wrapped securely so it does not slip.
If measuring on the right arm, refer to Section 2.3 of the user manual.
[Reference to Instruction Manual]
6. Correct Sitting Posture
[Diagram illustrating correct posture: seated comfortably with back support, arm at heart level, feet flat on the floor, legs uncrossed, remaining still and silent].
- Sit comfortably with stable support for your back and arm.
- Position the cuff on your arm at the same level as your heart.
- Place both feet flat on the floor, do not cross your legs, remain still, and do not talk.
7. Measurement
Press the [START/STOP] button to start measurement and save the reading. Open the application to transmit the data.
[Diagram showing the START/STOP button being pressed, followed by measurement display showing SYS, DIA, PULSE values].
If your systolic pressure exceeds 210 mmHg: After the cuff begins to inflate, press and hold the [START/STOP] button until the blood pressure monitor inflates the cuff to 30 to 40 mmHg higher than the expected systolic pressure.
Taking Measurements in Guest Mode
Guest mode can be used to take a single measurement for another user. No readings will be saved to memory.
[Diagram showing the process for guest mode measurement, possibly involving holding START/STOP for a specific duration or a specific button sequence].
8. Checking Readings
Refer to the following symbols for guidance:
- [High Reading Indicator]: Appears if "Systolic Pressure" is 135 mmHg or higher and/or "Diastolic Pressure" is 85 mmHg* or higher.
- [Irregular Pulse Indicator] (flashing): Detected during measurement. If it appears repeatedly, OMRON recommends consulting your doctor.
- [Movement Indicator] (flashing): Occurs during measurement due to body movement. Remove the cuff, wait 2-3 minutes, and try again.
- [Cuff Loose Indicator] (flashing): Cuff is not wrapped tightly. Re-wrap the cuff more securely.
*High blood pressure definition is based on the 2018 ESH/ESC guidelines.
**Irregular pulse definition is a pulse that deviates from the average pulse by 25% or more during measurement.
For error messages or other problems, please refer to Section 3 of the Instruction Manual.
9. Using Memory Function
9.1 Values Stored in Memory
[Diagram showing memory display with time, date, and measurement values (SYS, DIA, PULSE)].
9.2 Average of the Last 2 or 3 Readings Within 10 Minutes
[Diagram illustrating the calculation of the average (AVG) of recent readings].
9.3 Delete All Readings for One User
[Diagram showing the process to delete all stored readings for a specific user].
10. Other Settings
Turn Bluetooth Off/On
[Diagram showing how to toggle Bluetooth on or off. Default setting is Bluetooth enabled].
Restore to Default Settings
[Diagram showing the steps to reset the device to its factory default settings].
Optional Medical Accessories
Cuff (HEM-FL31)
Intelli Wrap™ Cuff (Fit Cuff)
Model: HEM-FL31
Type B
Arm Circumference: 22-42 cm
*Do not discard the air connector. The air connector can be used with other sold cuffs.
Other Optional Accessories
AC Adapter (HHP-AM01)
[Image of AC Adapter]
HHP-AM01 (for Taiwan)
HHP-KSM01 (for Korea)
OMRON Connect App Compatible Mobile Device Operating Systems
- iOS: Mobile devices with iOS version 10.3.3 or later (including 10.3.3).
- Android: Mobile devices with Android version 6.0.1 or later (including 6.0.1).
The OMRON Connect App can transmit and record measurement values from the device for health management.
Manufacturer Information
Distributor Name: Taiwan OMRON Healthcare Co., Ltd.
Manufacturer Name: OMRON HEALTHCARE Co., Ltd. Matsusaka Factory
Manufacturer Address: 1855-370, Kubo-cho, Matsusaka-shi, Mie, 515-8503 Japan