OMRON HEM-7142T1 Automatic Blood Pressure Monitor

Thank you for purchasing an OMRON Automatic Blood Pressure Monitor. This blood pressure monitor uses the oscillometric method of blood pressure measurement, meaning the monitor detects your blood's movement through the brachial artery and converts this movement into a digital reading. This instruction manual provides important information about your OMRON Automatic Blood Pressure Monitor. To ensure safe and proper use of this monitor, READ AND UNDERSTAND all of these instructions. If you do not understand these instructions or have any questions, contact your local OMRON representative before attempting to use this monitor. For more information about your blood pressure, consult your doctor.

Important Safety Information

Intended Use: This device is a digital monitor intended for use in measuring blood pressure and pulse rate on adult patient populations. The monitor detects the occurrence of irregular heartbeats during measurement and displays an icon with the measurement results. This device is primarily intended for household use.

Read the Important Safety Information in this instruction manual before using the monitor. Follow this instruction manual thoroughly for your safety. Keep it for future reference. For more information about your blood pressure, CONSULT YOUR DOCTOR.

Unpack the monitor and check for damage. If the monitor is damaged, DO NOT USE and consult your local OMRON representative.

Safety Precautions:

General Precautions:

Data Transmission:

Handling and Use of AC Adapter (optional accessory):

Handling and Use of Batteries:

CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient, or cause damage to equipment or other property.

1. Getting to Know Your Monitor

1.1 Contents/Product Included

Monitor, arm cuff, 4 “AA” batteries, instruction manual, setup instructions

1.2 Monitor and Arm Cuff

Front of the monitor

Left, back, and bottom of the monitor

Arm cuff

1.3 Display and Symbols

OK Symbol: Flashes when your monitor is successfully connected to your smart electronic device.

Bluetooth ON Symbol: Appears when your readings are being transferred.

Bluetooth OFF Symbol: Appears when Bluetooth is deactivated.

Systolic blood pressure reading

Diastolic blood pressure reading

Heartbeat Symbol: Flashes during measurement.

Deflating Symbol: Appears during cuff deflation.

Cuff Wrapping Guide (OK) Symbol: Appears if the arm cuff is wrapped correctly around the arm during measurement and when viewing previous readings.

Cuff Wrapping Guide (Loose) Symbol: Appears if the arm cuff is not wrapped correctly around the arm during measurement and when viewing previous readings.

Sync Symbol: Flashes/appears when your data needs to be transferred because the stored memory is nearly or completely full. Once you have paired the monitor with a smart electronic device, transfer your readings immediately before the monitor deletes old readings. The monitor’s internal memory can store up to 14 readings.

Memory Symbol: Appears when viewing readings stored in memory.

Hypertension Symbol: Appears if your systolic blood pressure is 135 mmHg or higher and/or your diastolic blood pressure is 85 mmHg or higher.

Movement Error Symbol: Appears along with your reading when your body moves during measurement. If it appears, remove the arm cuff and wait for 2 to 3 minutes, then reapply the cuff, remain still, and take the measurement again.

Irregular Heartbeat Symbol: Appears when an irregular rhythm is detected 2 or more times during measurement. An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while your monitor measures blood pressure. The irregular heartbeat symbol is displayed when the monitor detects arm movement. See chapter 7 for more information. If it continues to appear, we recommend you consult and follow your doctor’s directions.

Low Battery Symbol: Flashes when the batteries are nearly depleted.

Dead Battery Symbol: Appears when the batteries are depleted.

Pulse rate/memory count display: Pulse rate appears after measurement. When pressing the Bluetooth button, the memory count appears for approximately 1.5 seconds before the pulse rate reappears on the display.

A: Cuff Air Plug

B: Air Plug Hole

C: AC Adapter Port (for optional AC adapter)

D: Battery Compartment

E: START/STOP Button

F: Memory (Connection) Button

G: Display

H: Arm Cuff (Upper Arm Circumference: 22 - 32 cm)

I: Air Tube

J: Air Plug

K: Monitor

L: Automatic Blood Pressure Monitor

M: Short Long

N: Irregular Heartbeat Normal Heartbeat

2. Preparation

2.1 Installing Batteries

  1. Press down on the battery cover hook and pull upwards.
  2. Insert 4 “AA” batteries as indicated in the battery compartment.
  3. Close the battery cover.

Note: When the symbol flashes on the display, you are advised to replace the batteries. To replace the batteries, turn off the monitor and remove all batteries. Replace with 4 new batteries at the same time. Replacing the batteries will not erase previous readings. The supplied batteries may have a shorter life than new batteries. Disposal of used batteries must be in accordance with local regulations.

2.2 Tips for Blood Pressure Measurement

To help ensure accurate readings, follow these instructions:

WARNING:

2.3 Wearing the Arm Cuff

The following are the steps for wearing the arm cuff for your left arm. When taking a measurement on the right arm, follow the cuff wearing instructions for the right arm described in “To take a measurement on the right arm” located at the end of this subsection.

Note: Blood pressure can differ between the left and right arms, and the measured blood pressure readings may differ. OMRON recommends always using the same arm for measurement. If the readings between the two arms differ substantially, ask your doctor to determine which arm to use for measurement.

To take a measurement on the right arm: When you take a measurement on the right arm, the air tube should be on the inside of your elbow. Be careful not to rest your arm on the air tube.

WARNING:

  1. Attach the arm cuff to your monitor by inserting the air plug into the air plug hole securely until it clicks.
  2. Place your hand through the cuff loop. Pull the cuff up to your left upper arm.

Note: The bottom edge of the arm cuff should be 1 to 2 cm above the inner elbow. The air tube is on the inside of the arm and aligned with the middle finger. Ensure the air tube is positioned on the inside of your arm and tighten it so it does not move around your arm.

CAUTION:

2.4 Taking a Measurement

You can use the monitor without pairing it with a smart electronic device. To pair it with a smart electronic device, refer to subsection 4.1.

Note: To stop a measurement, press the [START/STOP] button once to deflate the arm cuff.

WARNING:

  1. Press the [START/STOP] button. All symbols appear on the display before measurement starts.
  2. Remain still and do not talk until the entire measurement process is complete. The heart symbol flashes along with the pulse rate.

Note: Symbol appears if the cuff is worn correctly. Conversely, the cuff is not worn correctly. Press the [START/STOP] button to turn off the monitor, then wear the cuff correctly.

  1. Press the [START/STOP] button to turn off the monitor.

Note: After your blood pressure and pulse rate are detected, the cuff automatically deflates. Your blood pressure and pulse rate are displayed. If the systolic or diastolic reading is high (see section 1.3), the Hypertension Symbol appears.

Note: The monitor will automatically turn off after 2 minutes. Wait 2 to 3 minutes to take another measurement. The waiting time allows the artery to reduce pressure and return to its pre-measurement state. You may need to increase the waiting time according to individual physiological characteristics.

If your systolic pressure is over 210 mmHg:

Note: Do not inflate above 299 mmHg.

WARNING: Inflating to an excessive pressure may cause bruising on the arm where the cuff is placed.

To take a measurement, you must relax and sit comfortably in a room with a comfortable temperature. Place your hand on a table.

Note: If your systolic pressure is over 210 mmHg, take the measurement manually. After the arm cuff starts inflating, press and hold the [START/STOP] button until the monitor inflates 30 to 40 mmHg higher than the actual systolic pressure.

Note: If “Err” appears, follow the instructions in the “OMRON connect” application. Be aware that OMRON will not be responsible for any loss of data and/or information in the application. “OMRON connect” is the only application we recommend using with your monitor to correctly transfer data.

3. Using Your Monitor Without a Smart Electronic Device

START INFLATING DEFLATING COMPLETED

Note: If your systolic or diastolic reading is high (see section 1.3), the Hypertension Symbol appears.

Note: If “Err” appears, follow the instructions in the “OMRON connect” application. Be aware that OMRON will not be responsible for any loss of data and/or information in the application. “OMRON connect” is the only application we recommend using with your monitor to correctly transfer data.

4. Using Your Monitor With a Smart Electronic Device

4.1 Pairing Your Monitor With a Smart Electronic Device

  1. Turn on Bluetooth on your smart electronic device.
  2. Download and install the “OMRON connect” application on your smart electronic device. If you already have the “OMRON connect” application and have created an account, open the application and add your new monitor.
  3. Open the application and follow the pairing instructions displayed on your smart electronic device.
  4. Confirm that your monitor is successfully connected. When your monitor is successfully connected with your smart electronic device, the OK Symbol will appear.
  5. Press the [START/STOP] button to turn off your monitor.

Note: Review the list of compatible smart electronic devices at www.omronconnect.com/devices/

4.2 Taking a Measurement

Refer to subsection 3.1 to start a measurement. Open the application on your smart device and follow its instructions.

Note: Ensure that your monitor is within 5 m of your smart electronic device.

4.3 Enabling/Disabling Bluetooth

Disable Bluetooth on your monitor in areas that prohibit the use of wireless equipment. In airplanes • In hospitals • Overseas Bluetooth is enabled by default.

  1. When the monitor is off, press and hold the Bluetooth button for more than 10 seconds. “oFF” flashes 3 times.
  2. Press the [START/STOP] button to turn off the monitor.

Note: To enable Bluetooth, press and hold the Bluetooth button for more than 3 seconds. “on” appears on the display. The monitor will automatically turn off 10 seconds after completion. When Bluetooth is disabled, the Bluetooth Symbol appears.

Automatic Data Transfer:

When Bluetooth is enabled, readings will be transferred to your smart electronic device after taking a measurement.

5. Using the Memory Function

5.1 Viewing Stored Readings

Your monitor automatically stores up to 14 readings. Press the Bluetooth button repeatedly to scroll through the previously stored readings in memory.

  1. Press the Bluetooth button. The memory count appears for 1.5 seconds before the pulse rate is displayed. The latest reading stored in memory will be numbered 1.
  2. After the memory count appears, the latest reading will appear with previous readings.

Note: If readings are high (see subsection 1.3), the Hypertension Symbol appears. If there are no readings stored in memory, the display on the right will be shown. If the memory is full, new readings will replace old readings, starting with the oldest reading.

6. Other Settings

6.1 Restoring Your Monitor to Default Settings

To erase all stored information on your monitor, follow the instructions below. Ensure your monitor is turned off.

  1. Press the Bluetooth button.
  2. While holding down the Bluetooth button, press and hold the [START/STOP] button for more than 4 seconds. The display flashes as shown on the right.
  3. While holding down the Bluetooth button, press and hold the [START/STOP] button for more than 4 seconds. The display stops flashing. Your monitor is restored to default settings.
  4. Press the [START/STOP] button to turn off your monitor.

Note: Restoring to default settings does not erase information in the application. The monitor will automatically turn off 10 seconds after completion.

7. Error Messages and Troubleshooting

Display/ProblemPossible CauseSolution
E1 appears or cuff does not inflate.The air plug is not fully plugged into the monitor.Plug the air plug securely into the monitor.
The arm cuff is not worn correctly.Wear the arm cuff correctly, then take the measurement again. Refer to subsection 2.3.
Air is leaking from the arm cuff.Replace the arm cuff with a new one. Refer to chapter 9.
E2 appears or measurement cannot be completed after the arm cuff inflates.You moved or talked during measurement and the arm cuff did not inflate sufficiently.Remain still and do not talk during measurement. If “E2” appears repeatedly, manually inflate the arm cuff until the pressure is 30 to 40 mmHg higher than the previous reading. Refer to subsection 3.1. Because systolic pressure is above 210 mmHg, measurement cannot be performed.
E3 appears.The arm cuff was inflated beyond the maximum allowable pressure.Do not touch the arm cuff and/or kink the air tube during measurement. If manually inflating the arm cuff, refer to subsection 3.1.
E4 appears.You moved or talked during measurement. Vibration interferes with measurement.Remain still and do not talk during measurement.
E5 appears.Pulse rate was not detected correctly. The symbol does not flash during measurement.Wear the arm cuff correctly, then take the measurement again. Remain still and sit correctly during measurement. If the symbol continues to appear, we recommend you consult a doctor.
Er appears or the monitor is turned off suddenly during measurement.Monitor malfunction.Press the [START/STOP] button again to turn off the monitor, then press it again to take a measurement. If the problem persists, remove all batteries and wait for 30 seconds, then reinsert them. If the problem still persists, contact your local OMRON representative.
Communication error appears.The monitor cannot connect to the smart electronic device or send data correctly.Follow the instructions displayed in the “OMRON connect” application. If the “Err” symbol still appears after checking the application, contact your local OMRON representative.
Other communication problems appear.The monitor is waiting to be connected to a smart electronic device.Refer to subsection 4.1 for connecting the monitor to a smart electronic device or press the [START/STOP] button to cancel pairing and turn off the monitor.
Other problems appear.More than 11 readings are not transferred.Pair or transfer your readings to “OMRON connect” so you can store them in the application memory and this error symbol disappears.
Low Battery Symbol flashes.Low batteries.It is recommended to replace the 4 batteries with new ones. Refer to subsection 2.1.
Dead Battery Symbol appears or the monitor turns off suddenly during measurement.Depleted batteries.Replace the 4 batteries with new ones immediately. Refer to subsection 2.1.
Nothing appears on the monitor display.Depleted batteries.Check the battery installation for proper installation. Refer to subsection 2.1.
Battery poles are not aligned correctly.Check the battery installation for proper installation. Refer to subsection 2.1.
Sync Symbol flashes.The monitor is not paired or connected to your smart electronic device. There are 14 readings to be transferred.Pair or transfer your readings to “OMRON connect” so you can store them in the application memory and this error symbol disappears.
Cuff is ModerateThe cuff is not worn correctly, then take the measurement again. Remain still and sit correctly during measurement. If the symbol continues to appear, we recommend you consult a doctor.Wear the cuff correctly, then take the measurement again. Refer to subsection 2.3. Remain still and sit correctly during measurement. If the symbol continues to appear, we recommend you consult a doctor.
Cuff is LooseThe cuff is not worn correctly, then take the measurement again. Refer to subsection 2.3. Remain still and sit correctly during measurement. If the symbol continues to appear, we recommend you consult a doctor.Wear the cuff correctly, then take the measurement again. Refer to subsection 2.3. Remain still and sit correctly during measurement. If the symbol continues to appear, we recommend you consult a doctor.
Irregular Heartbeat Symbol appears.The monitor detects arm movement. See chapter 7 for more information. If it continues to appear, we recommend you consult and follow your doctor’s directions.The monitor detects arm movement. See chapter 7 for more information. If it continues to appear, we recommend you consult and follow your doctor’s directions.
Hypertension Symbol appears.Systolic blood pressure is 135 mmHg or higher and/or diastolic blood pressure is 85 mmHg or higher.Consult your doctor.
Movement Error Symbol appears.Your body moves during measurement. If it appears, remove the arm cuff and wait for 2 to 3 minutes, then reapply the cuff, remain still, and take the measurement again.Your body moves during measurement. If it appears, remove the arm cuff and wait for 2 to 3 minutes, then reapply the cuff, remain still, and take the measurement again.
Bluetooth ON Symbol appears.Your readings are being transferred.Wait for the transfer to complete.
Bluetooth OFF Symbol appears.Bluetooth is deactivated.Enable Bluetooth in the monitor settings if needed.
OK Symbol appears.Your monitor is successfully connected to your smart electronic device.No action needed.
Sync Symbol appears.Your data needs to be transferred because the stored memory is nearly or completely full.Transfer your readings to the “OMRON connect” application.
Low Battery Symbol flashes.Batteries are nearly depleted.Replace the batteries.
Dead Battery Symbol appears.Batteries are depleted.Replace the batteries immediately.

8. Maintenance

8.1 Maintenance

To protect your monitor from damage, follow the instructions below:

CAUTION:

8.2 Storage

  1. Remove the cuff from the monitor.
  2. Gently fold the air tube into the arm cuff.

Note: Do not excessively bend or kink the air tube.

Store your monitor and other components in a clean and safe location.

CAUTION: Do not store the monitor and other components:

8.3 Cleaning

Do not use abrasive or volatile cleaners.

Use a dry soft cloth or a soft cloth slightly dampened with mild (neutral) detergent to clean the monitor and arm cuff. Wipe with a dry cloth.

Do not wash or immerse the monitor and arm cuff or any other components in water.

Do not use gasoline, thinners, or similar solvents to clean the monitor and arm cuff or any other components.

8.4 Calibration and Service

The accuracy of the blood pressure monitor has been carefully tested and designed for a long service life. It is generally recommended that the unit be checked every two years to ensure proper function and accuracy. Please consult your authorized OMRON dealer or OMRON Customer Service at the address provided on the packaging or enclosed information.

WARNING:

9. Optional Medical Accessories

Arm Cuff Medium

HEM-CR24

Type B

22 – 32 cm

Soft Wide Cuff

HEM-RML31

Type B

22 – 42 cm

Small Cuff

HEM-CS24

Type B

17 – 22 cm

Note: Do not discard the air plug. The air plug can be used on manual cuffs.

10. Other Optional Components

AC Adapter

For Australia and New Zealand: HHP-OH01

For India: HHP-BH01

AC Adapter

HHP-CM01

HHP-BFH01 (for Sri Lanka)

Note: Make sure not to place your monitor in a location that is difficult to plug and unplug the AC adapter.

  1. Insert the AC adapter into the AC adapter port on the back of your monitor.
  2. Plug the AC adapter into the power outlet.

To unplug the AC adapter, unplug the AC adapter from the power outlet and detach the AC adapter port from the monitor.

WARNING: ONLY use the AC adapter, arm cuff, batteries, and accessories specified for this monitor. Use of AC adapters, arm cuffs, and batteries that do not support this monitor may damage and/or endanger this monitor.

11. Specifications

Product Description: Automatic Blood Pressure Monitor

Model: HEM-7142T1

Display: Digital LCD display

Cuff Pressure Range: 0 to 299 mmHg

Measurement Range:

Blood Pressure: SIS: 60 to 260 mmHg

DIA: 40 to 215 mmHg

Pulse Rate Measurement Range: 40 to 180 beats/min

Accuracy: Pressure: ±3 mmHg

Pulse Rate: ±5% of displayed reading

Inflation: Automatic by electric pump

Deflation: Automatic pressure release valve

Measurement Method: Oscillometric method

Transmission Method: Bluetooth®

Wireless Communication: Frequency Range: 2.4 GHz (2400 – 2483.5 MHz)

Modulation: GFSK

Effective radiated power: < 20 dBm

Operating Mode: Continuous operation

IP Classification:

Monitor: IP20

Optional AC Adapter: IP22 for HHP-BH01 and HHP-BFH01, IP21 for other AC adapters

Power: DC 6V 4W

Power Source: 4 “AA” 1.5V batteries or optional AC adapter (INPUT AC 100 – 240V 50 – 60 Hz 0.12 – 0.065A)

Battery Life: Approx. 900 measurements (Using new alkaline batteries and including cuff usage. Depends on battery type and cuff usage.)

Shelf Life (Service Life):

Monitor: 5 years

Arm cuff: 5 years

Optional AC adapter: 5 years

Operating Conditions: +10 to +40 °C / 15 to 90% RH (non-condensing) / 800 to 1060 hPa

Storage/Transport Conditions: -20 to +60 °C / 10 to 90% RH (non-condensing)

Contents: Monitor, arm cuff, 4 “AA” batteries, instruction manual, setup instructions

Protection against electric shock: Internally powered ME Equipment (when using batteries only)

Class II ME Equipment (optional AC adapter)

Weight (excluding batteries):

Monitor: Approx. 250 g

Arm cuff: Approx. 110 g

Dimensions (approximate):

Monitor: 103 mm (W) x 82 mm (H) x 140 mm (D)

Arm cuff: 146 mm x 466 mm (air tube: 610 mm)

Applicable cuff circumference for the monitor: 17 to 42 cm (including arm cuff: 22 to 32 cm)

Note: Memory Stores up to 14 readings

Applied Part: Type BF (arm cuff)

Note: These specifications are subject to change without notice. This monitor has been clinically validated according to the requirements of EN ISO 81060-2:2014 and complies with EN ISO 81060-2:2014 and EN ISO 81060-2:2019+A1:2020. In the clinical validation study, K5 was used on 85 subjects for the determination of diastolic blood pressure. The IP classification is the degree of protection provided by enclosures in accordance with IEC 60529. This monitor and the AC adapter are protected against solid foreign objects of 12.5 mm diameter or more, such as fingers. The optional AC adapters (except for HHP-BH01 and HHP-BFH01) are protected against vertically falling water drops that may cause problems during normal operation. The optional AC adapters HHP-BH01 and HHP-BFH01 are protected against horizontally falling drops of water that may cause problems during normal operation. The operating mode is classified according to IEC 60601-1. Regarding wireless communication interference: This product operates on an unlicensed ISM band at 2.4 GHz. If this product is used near other wireless devices such as microwaves and wireless LANs, which operate on the same frequency band as this product, there is a possibility of interference. If interference occurs, stop operating the other devices or move this product away from the wireless devices before attempting to use it. The HEM-7142T1 manufactured by OMRON HEALTHCARE Co., Ltd. complies with the Electromagnetic Compatibility (EMC) standard EN60601-1-2:2015. Further documentation according to this EMC standard is available at: https://www.omronhealthcare-ap.com/emc-information See EMC information for HEM-7142T1 on the website. To indicate a generally increasing, potentially hazardous level of radiation, or to indicate equipment or systems, for example in medical electrical areas that include RF transmitters or that intentionally apply RF electromagnetic energy for diagnosis or treatment. This blood pressure monitor is designed in compliance with European Standard EN1060, non-invasive sphygmomanometers Part 1: General requirements and Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems. Hereby, OMRON HEALTHCARE Co., Ltd., declares that the radio equipment type HEM 7142T1 is in compliance with Directive 2014/53/EU. OMRON products are manufactured under the strict quality system of OMRON HEALTHCARE Co., Ltd., Japan. The Core Components for OMRON blood pressure monitors, namely the Pressure Sensor, are manufactured in Japan. Please report to the manufacturer and the competent authority of the Member State where you are located about any serious incident that has occurred in connection with this device. This marking shown on the product or its information, indicates that the product should not be disposed of with other household waste at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of waste and recycle it responsibly to promote the sustainable reuse of material resources. Household users should either contact the retailer where they purchased this product, or their local government office, for details of where and how they can take this item for environmentally safe recycling. Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial waste for disposal. Manufacturer OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, KYOTO, 617-0002 JAPAN Production Facility OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD. No.28 VSIP II, Street 2, Vietnam-Singapore Industrial Park II, Binh Duong Industry-Services-Urban Complex, Hoa Phu Ward, Thu Dau Mot City, Binh Duong Province, Vietnam https://www.omronhealthcare-ap.com/ EU Importer OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS www.omron-healthcare.com Asia Pacific Head Office OMRON HEALTHCARE SINGAPORE PTE LTD. 438A Alexandra Road, #05-05/08 Alexandra Technopark, Singapore 119967 www.omronhealthcare-ap.com

12. Manufacturer's Guide and Declarations

Description of Symbols

Applied Part – Type BF

Degree of protection against electric shock (leakage current)

CE Marking

Class II Equipment.

Protection against electric shock

Required ingress protection rating by IEC 60529

Serial Number

LOT Number

Medical Device

Indoor use only

Temperature Limit

Humidity Limit

Atmospheric Pressure Limit

Connector polarity indicator

IP cuff marking for device

OMRON trademark technology for blood pressure measurement

Arm circumference

Cuff position indicator for left arm

Marking on cuff to be positioned over artery

Brachial artery alignment and range indicators

Arm circumference range indicator to help select the correct cuff size

User needs to consult this instruction manual

User needs to follow this instruction manual thoroughly for safety

Direct current

Alternating current

IP XX

Important information regarding Electromagnetic Compatibility (EMC)

Correct disposal of this product (waste electrical and electronic equipment)

Date of manufacture

KC Mark

BIS Mark

Power supply efficiency level

Not made with natural rubber latex

Description of Symbols

Made in Vietnam

The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by OMRON HEALTHCARE Co., Ltd. is under license. Other trademarks and trade names are those of their respective owners. Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries and regions. App Store is a service mark of Apple Inc. Google Play and the Google Play logo are trademarks of Google LLC.

PDF preview unavailable. Download the PDF instead.

e9850f1be37640768498ee67268d2f14 Adobe PDF library 10.01

Related Documents

PreviewOMRON EVOLV Automatic Arm Blood Pressure Monitor User Manual
User manual for the OMRON EVOLV Automatic Arm Blood Pressure Monitor (HEM-7600T-E), providing instructions on setup, measurement, troubleshooting, and maintenance. Learn how to accurately measure blood pressure at home and understand the results.
PreviewOmron HEM-7142T1-1 Automatic Blood Pressure Monitor Instruction Manual
Comprehensive user manual for the Omron HEM-7142T1-1 automatic upper arm blood pressure monitor. Provides detailed instructions on setup, safe usage, measurement procedures, and maintenance.
PreviewOmron HEM-7140T1 Blood Pressure Monitor Instruction Manual
Comprehensive instruction manual for the Omron HEM-7140T1 automatic blood pressure monitor, covering setup, usage, Bluetooth connectivity, and maintenance.
PreviewOMRON M3 Comfort & X3 Comfort Automatic Blood Pressure Monitor User Manual
Comprehensive user manual for OMRON M3 Comfort (HEM-7155-E) and X3 Comfort (HEM-7155-EO) automatic upper arm blood pressure monitors. Covers setup, operation, safety precautions, troubleshooting, specifications, and warranty information.
PreviewOMRON BP7250 Upper Arm Blood Pressure Monitor Instruction Manual
Instruction manual for the OMRON BP7250 Upper Arm Blood Pressure Monitor, covering setup, usage, memory functions, troubleshooting, specifications, and warranty information.
PreviewOMRON M3 (HEM-7154-E) Automatic Upper Arm Blood Pressure Monitor Instruction Manual
Comprehensive instruction manual for the OMRON M3 (HEM-7154-E) automatic upper arm blood pressure monitor, covering setup, usage, safety, troubleshooting, and technical specifications.
PreviewOMRON M4 Intelli IT & X4 Smart Automatic Upper Arm Blood Pressure Monitor Instruction Manual
Comprehensive instruction manual for OMRON M4 Intelli IT and X4 Smart automatic upper arm blood pressure monitors, covering safe usage, features, troubleshooting, and maintenance for accurate home health monitoring. Learn how to use your OMRON blood pressure monitor effectively.
PreviewOMRON HEM-7156-A Automatic Blood Pressure Monitor Instruction Manual | User Guide
Download the OMRON HEM-7156-A Automatic Blood Pressure Monitor Instruction Manual. Get detailed guidance on setup, usage, safety, troubleshooting, and specifications for accurate home blood pressure monitoring. Multilingual support available.