Drive Medical Blue Streak Wheelchair

Model Shown: BLS18FBD-SF

Weight Capacity: 250 LBS

Product Overview

The Drive Medical Blue Streak Wheelchair is a product of quality engineering, manufactured to meet high specifications for dependable service. This manual provides essential information for safe usage and maintenance.

Key Components:

Safety and Usage Instructions

CAUTION: Stability

Wheelchairs are designed for safety and stability when the center of gravity is maintained during normal activities. Avoid activities that shift your center of gravity or reaching too far out of the chair. When reaching for objects, ensure the front casters are pointing forward to extend the wheelbase for stability.

Ramps

Assess your capabilities before using ramps. Seek guidance from a qualified professional for correct ramp negotiation techniques. Assistance is recommended for inclines greater than 10°. Avoid sudden movements when resuming movement on an incline, as this can cause instability. Do not use wheel locks to slow descent on inclines. Abrupt changes in direction on inclines can lead to instability or tipping.

Wheel Locks

Engage both wheel locks before entering or exiting the chair, leaning forward, or when in an elevator or wheelchair lift. Ensure wheel locks are adjusted to engage at least 1/8" when locked. Proper adjustment involves loosening the nut on the carriage bolt, sliding the lock shoe towards the rear wheel, and then tightening the nut and bolt.

Tires

Solid rubber tires require minimal maintenance and can be cleaned with a damp cloth. Replace if severely worn or cracked.

Front Casters

Periodically check and lubricate front casters. Lubricate by removing the fork and greasing the caster stem bearings. Reassemble by tightening nuts until there is no play, but the wheels roll freely.

Removable Arms

Arms can be removed by depressing the spring button and lifting the arms out of the front and rear sockets.

Swing-Away Foot Rests/Leg Rests

To remove: Release the lock by pushing the front rigging lever, then swing the footrest/leg rest out and sideways. Lift to remove.

Adjusting Footrest/Leg Rest Length

With the user seated, loosen the bolt on the clamp assembly. Raise or lower the footplate to the desired position for maximum thigh support. Ensure the front rigging is not too low, causing discomfort. The footrest should not be lower than 2-1/2" from the ground for clearance.

Adjusting/Elevating Leg Rest

Lift the footplate on the leg rest assembly to the desired position. Release the assembly; it will lock automatically. To lower, push the leg rest lever outward and lower the footplate.

Wheels

Inspect rear wheels monthly for tightness. If a wheel wobbles, release the lock nut and slightly tighten the axle bolt. Test by tilting the chair and spinning the adjusted wheel.

General Care

Wipe chrome parts weekly with a clean, soft cloth. Metal parts can be cleaned with automotive wax. Dry immediately if exposed to moisture. Avoid abrasive cleansers.

Opening the Chair

Tilt the chair to one side and push down on the seat rail until fully opened.

Closing/Folding the Chair

Fold foot plates to a vertical position. Grab the seat with both hands at the front and back edges and lift. Alternatively, tilt the chair to one side and lift the seat rail.

WARNINGS

3 Year Limited Warranty

Drive brand products are warranted to be free of defects in materials and workmanship for three years from the date of original consumer purchase. This warranty reflects confidence in the product's quality and assurance of dependable service.

Exclusions: This warranty does not cover failures due to owner misuse, negligence, or normal wear and tear. Non-durable components like rubber accessories, casters, and grips, subject to normal wear, require periodic replacement and are not covered.

For questions about your Drive device or warranty, please contact an authorized Drive dealer.

PDF preview unavailable. Download the PDF instead.

BLS18FBD-ELR-DOC 1 Adobe PDF Library 9.9 Adobe InDesign CS5.5 (7.5)

Related Documents

Preview Drive Medical Adjustable Tension General Use Wheelchair Back Cushion
Detailed information on the Drive Medical Adjustable Tension General Use Wheelchair Back Cushion, including its features, product codes 14300 and 14301, and specifications like fluid-resistant cover and 350 lbs weight capacity.
Preview Drive Scout Mobility Scooter: Owner's Handbook and Operation Guide
Comprehensive owner's handbook for the Drive Scout mobility scooter. Learn about safe operation, maintenance, troubleshooting, specifications, and warranty for your Scout scooter.
Preview Drive Medical Shower Bench Assembly and Usage Guide
Comprehensive guide for assembling and using the Drive Medical Shower Bench (RTL12203KDR), including safety precautions, maintenance, and warranty information.
Preview Drive Delta 1000 Ultra Light Plus Full Electric Bed Assembly and Operating Manual
This manual provides assembly and operating instructions for the Drive Delta 1000 Ultra Light Plus Full Electric Bed, item # 15033. It includes detailed steps for assembly, maintenance, safety checks, and specifications.
Preview Drive Medical Rollator Assembly, Safety, and Maintenance Guide
Comprehensive guide for the Drive Medical Rollator, covering assembly instructions, safety precautions, hand brake adjustment, maintenance, and limited lifetime warranty. Includes information in English, French, and Spanish.
Preview Drive Medical Hospital Beds 15005LP, 15004P, 15003P: Owner's Assembly and Operating Manual
This manual provides comprehensive assembly, operating, and safety instructions for Drive Medical's 15005LP Full Electric, 15004P Semi-Electric, and 15003P Manual Hospital Beds. It includes detailed technical specifications, maintenance procedures, troubleshooting guidance, and important safety warnings.
Preview Drive Nitro Rollator User Manual and Warranty Information
Comprehensive guide to the Drive Nitro Rollator, including setup, usage, safety instructions, maintenance, and warranty details. Learn how to unfold, adjust height, use brakes, and care for your rollator.
Preview Delta 1000 Ultra Light Plus Semi Electric Bed: Owner's Assembly and Operating Manual
Comprehensive guide for assembling, operating, and maintaining the Drive Medical Delta 1000 Ultra Light Plus Semi Electric Bed (Item # 15030), including safety warnings and specifications.