MICROTECH 141088015A Micron Computerized Multi Force Caliper
Product Specifications
- Calibration Date: Rechargeable
- Auto Off: Yes
- Password Memory: Yes
- Statistics Memory: Yes
- Wireless: Yes
- USB: Yes
- HID: Yes
- FW Update: Available
- Formula Timer: Yes
- Temperature Compensation: Yes
- Linear Correction: Yes
- Analog Scale: Yes
Product Usage Instructions
Cleaning and Maintenance
Clean the oil from the measuring surfaces of the caliper using a cloth. Do not leave the caliper uncharged for an extended period as it may result in internal battery failure.
Charging and Power
For charging, connect the USB cable to the device’s built-in rechargeable Li-Pol battery. To switch on the device, push the button for 1 second. To switch off the device, push the button for 3 seconds or allow it to auto switch off. Data transfer can be done through programming via the menu.
Force Control
The electronic multi-force control module allows for measuring with a requested force in the range of 1-20N. Different force modes are available such as Disable, Indicate, Auto, and Force. The force indication on the force bar is color-coded for easy reference.
Data Transfer and Memory
You can save measuring data to the internal device memory by touching the data area on the screen or pushing a button. The saved data can be viewed or sent wirelessly or via USB connection to Windows PC, Android, or iOS devices. Different memory modes like Standard or Folder system can be activated through the memory menu.
Analog Scale and Probe Mode
The analog scale can be activated for probe mode, and folders system memory allows for storing up to 2000 values. The product also supports memory export and statistics tracking.
Frequently Asked Questions (FAQ)
Q: How do I know if the wireless connection is established?
A: The device will indicate “Wireless Connected” on the screen when the wireless connection is established.
Q: How can I set limits using the product?
A: You can set limits using the Limits mode with color indications on the main screen for Go and NoGo zones.
SPECIFICATION
Item No |
Range | Reso- lution | Accu- racy* | eForce | Jaws | Measuring HUB | Analog scale | Graphic | Preset | Go/NoGo | Max/Min | Formula | Timer | Temp comp | Linear corr. | Calibr date | Recharge | Auto off | Password | Memory | Statistics | WIRELESS | USB HID | FW update | |
mm | inch | mm | μm | N | mm | ||||||||||||||||||||
141088015A | 0-150 | 0-6” | 0,001 | ±5 |
1-20 |
40/16 | • | • | • | • | • | • | • | • | • | • | • | • | • | • | 1000 val | • | • | • | • |
141088020A | 0-200 | 0-8” | ±8 | 50/17 | • | • | • | • | • | • | • | • | • | • | • | • | • | • | • | • | • | • | |||
141088030A | 0-300 | 0-12” | ±10 | 60/20 | • | • | • | • | • | • | • | • | • | • | • | • | • | • | • | • | • | • |
* For internal jaw and depth bar accuracy according DIN 862
TECHNICAL DATA
Parameters | |
LED display | color 2,0 inch |
Resolution | 320×240 |
Indication system | MICS 3.0 |
Power supply | Rechargeable Li-Pol battery |
Battery capacity | 400mAh |
Charging port | micro-USB / Magnetic port |
Case material | Aluminum |
Buttons | Switch (Multifunctional) |
Wireless data transfer | Long range/HID |
Clean the oil from the measuring surfaces of caliper by cloth. DO NOT LEAVE UNCHARGED CALIPER FOR A LONG TIME. THIS MAY RESULT TO INTERNAL BATTERY FAILURE!
CORRECT MEASURING
MAIN INFO
FORCE CONTROL
Сaliper with electronic multi-Force control module, gives possibility measure all time with a requested FORCE in range 1-20N
FORCE MODES:
- DISABLE – FORCE mode OFF
- INDICATE – Indicating requested FORCE and status on FORCE bar
- AUTO – Indicating requested FORCE and status on FORCE bar and Saving to memory actual lenght value upon reaching requested FORCE (1-20N). FORCE -Indicating actual FORCE value during pushing
FORCE Indication on FORCE bar:
- Green color – actual FORCE OK
- Yellow color – lower FORCE
- Red color – over FORCE
For ZERO setting also use Force control module (push Force up to Green indication and set ZERO by push for 0 button.
MAIN SCREEN
MAIN SCREEN on 2D mode
FUNCTIONS
LIMITS mode GO/NOGO
COLOR INDICATION LIMITS ON MAIN SCREEN Go NoGo
PEAK mode MAX/MIN/MODULE/RANGE
SAVING DATA TO MEMORY OR SENDING WIRELESS/USB BY TIMER
TIMER mode
FORMULA mode
RESOLUTION selection
DISPLAY settings
Linear correction error’s on device
LINEAR error compensation
TEMP compensation
WIRELESS data transfer
USB OTG data transfer
HUB connection
2D MODE
FORCE
- DISABLE – FORCE mode OFF
- INDICATE – Indicating requested FORCE and status on FORCE bar
- AUTO – Indicating requested FORCE and status on FORCE bar and Saving to memory actual lenght value upon reaching requested FORCE (1-20N). FORCE -Indicating actual FORCE value during pushing
RESET to Factory settings
EXTRA
MEMORY manager setting
CALIBRATION date info
Device INFO
Information about device
- Firmware version
- MAC adress for WIRELESS connection
INDUSTRY 4.0 INSTRUMENTS
MICROTECH
innovative measuring instruments
04080, Kyiv, Ukraine, Nizhneyurkovska str., 45a, 61001, Kharkiv, Ukraine, str. Rustaveli, 39
www.microtech.ua
tool@microtech.ua
Documents / Resources
![]() |