SciCan BRAVO G4 Autoclave Quick Reference Guide
Consult the Operator's Manual (located on the USB key) before using this autoclave.
For support, contact SciCan at +49(0) 7561 98 343 -0.
Basic Operation
- Power unit ON.
- Fill with demineralised/distilled water. NEVER USE TAP WATER ?
- Load configurations.
- Close unit.
- Run cycle.
- Open unit.
- Unload instruments.
- If required, drain manually.
Ventilation Requirements:
- Front & Side: 0.78 in / 20 mm clearance
- Back: 2.75 in / 70 mm clearance
Filling Options:
- Top Filling
- Front Filling
Draining Options:
- Manually (To Release)
- Automatic
Maintenance Schedule
Daily
Clean the door gasket and internal part of the door with a clean wet cloth. Check for wear and replace if required. Completely empty the waste tank before refilling the water reservoir.
Weekly
Clean the chamber and appropriate accessories with a wet cloth. If required, use a neutral detergent (without alkalines and bleaches). Carefully rinse with demineralized/distilled water and remove any residue. Inspect and clean chamber anti-dust filter. Replace if necessary. Ensure area around BRAVO G4 is dust free.
Monthly
Drain & clean the internal water reservoir as per the instructions outlined in the Operator's Manual. If applicable, clean the external water tank used for automatic filling.
Every 3 - 6 Months
Check the bacteriological filter and replace as required. Lubricate the door lock. Using a clean cloth, remove any residues and lubricate the inside of the bushing with a film of the silicone-based grease provided. Inspect and clean the chamber drain filter. Replace if required.
Every Year or 1,000 Cycles
Replace the door gasket.
Spare Parts and Accessories
Part Number | Description |
---|---|
97400145 | Door Gasket 17L/22L |
97467176 | Door Gasket 28L |
97290160 | Bacteriological Filter |
97290210 | Chamber Drain Filter |
99950390 | Chamber Rack 17L |
99950317 | Chamber Rack 22L |
99951041 | Chamber Rack 28L |
99950394 | Pouch Rack |
99951317 | Tray 17L |
99951318 | Tray 22L |
99951042 | Tray 28L |
97467066 | Tray Extractor |
Important!
Users of this device must be sure to select the most appropriate sterilization cycle according to the recommendations of the leading infection control authorities and local regulatory guidelines / recommendations.