SciCan HYDRIM® M2 G4 Quick Reference Guide

Operator's Manual should be consulted before using this washer disinfector.

Basic Operation

  1. 1. Power unit ON.
    Image: Front view of the SciCan HYDRIM M2 G4 unit with the power button illuminated.
  2. 2. Load Instruments.
    Image: The interior of the HYDRIM M2 G4 with instrument trays being loaded.
  3. 3. Close the door.
    Image: The front door of the HYDRIM M2 G4 is shown being closed.
  4. 4. Select cycle and press START.
    Image: The touchscreen interface of the HYDRIM M2 G4 displaying cycle options, with a finger pressing the START button.

Touchscreen

Refilling the Water Softener Salt Reservoir

  1. Unscrew lid.
  2. Fill to the top (max. 1 kg / 2.2 lbs.).
  3. Screw lid on tightly.
  4. Run a complete cycle, with no instruments.

Image: An illustration showing the water softener salt reservoir being filled with tablet salt.

IMPORTANT!

If the salt container is being filled for the first time, it MUST be filled with at least 1 litre of water before adding the salt. Only tablet salt should be used in the softening unit. Fine grained salt such as table salt should NOT be used as it could clog the unit. If salt was spilled while refilling, run a cleaning cycle. The water softener salt is also available from SciCan. Please contact your local dealer.

Cleaning the Chamber

Pour 1.0L of vinegar or citric acid-based instrument neutralizer into the chamber and follow these steps:

Process: Pour neutralizer, select 'Cleaning' cycle on touchscreen, run cycle.

Image: An illustration depicting the chamber cleaning process: pouring a liquid into the chamber, selecting the 'Cleaning' cycle on the touchscreen, and the unit running the cycle.

IMPORTANT!

Do a Chamber Cleaning cycle every 25 cycles. Failure to clean the chamber can promote corrosion.

Replacing the HIP™ Cleaning Solution

Recommended precautionary measures:

Image: An illustration showing a person wearing gloves and eye protection while handling a cleaning solution container.

Troubleshooting

Refer to Operator's Manual for a more comprehensive list. Call for service if suggested recommendations do not resolve the error message.

CF2 Chamber Filling: Water is not filling the chamber
Check that the water hoses are not kinked.
Check that water shut-off valves are opened.
CF15 Water Reservoir Overflow: Overflow switch triggered
Call for Service.
CF20 Condenser System: Unit is not able to cool down
Restart the program.
Call for Service.
CF21 Dosing System: Detergent is not dispensing
Check that the water hoses are not kinked.
Tighten detergent cap.
Check for leaks.
Call for Service.

Spare Parts and Accessories

Part NumberDescription
CS-HIPLHIP Cleaning Solution (2 x 3.8L)
01-112594SWater Softener Salt
01-108305SWater Test Kit
01-114390SHEPA Filter

Important Information

IMPORTANT! Terminal sterilization of instruments MUST follow processing in the HYDRIM M2 G4.

For non-invasive medical devices, SciCan recommends terminal sterilization after processing in the SciCan HYDRIM M2 G4 according to local guidelines. For invasive medical devices, such as dental handpieces, terminal sterilization must be performed after processing in the SciCan HYDRIM M2 G4.

Models: HYDRIM M2 G4, Automated Instrument Washers

File Info : application/pdf, 1 Pages, 1.72MB

PDF preview unavailable. Download the PDF instead.

96-113757 EU EN R2 HYDRIM M2 QRG

References

Adobe PDF Library 15.0 Adobe InDesign 15.1 (Macintosh)

Related Documents

Preview SciCan HYDRIM C61W G4 Quick Reference Guide
Quick reference guide for the SciCan HYDRIM C61W G4 instrument washer, detailing basic operation, touchscreen controls, maintenance procedures like refilling salt and cleaning the chamber, and troubleshooting common errors. Includes information on spare parts and accessories.
Preview SciCan BRAVO G4 Autoclave Quick Reference Guide
A concise guide to the basic operation and maintenance of the SciCan BRAVO G4 Autoclave, including a maintenance schedule and spare parts list.
Preview Hydrim C61W G4 Preventative Maintenance Checklist | SciCan
A comprehensive preventative maintenance checklist for the SciCan Hydrim C61W G4 sterilizer, detailing daily, periodic, and annual service procedures and reference materials.
Preview SciCan BRAVO G4 Autoclave Quick Reference Guide
A concise guide to the basic operation, touchscreen interface, ventilation, filling and draining options, and maintenance schedule for the SciCan BRAVO G4 autoclave.
Preview SciCan OPTIM OS1 Wipes Quick Reference Guide
A quick reference guide for using SciCan OPTIM OS1 Wipes and OPTIM OS1 SOLUTION Liquid Spray for disinfecting PVS elastomeric impression materials and surfaces.
Preview SciCan BRAVO G4 Quick Reference Guide
Essential guide for SciCan BRAVO G4 autoclave operation, maintenance, and troubleshooting. Covers basic steps, touchscreen use, and upkeep schedules for optimal sterilization.
Preview SciCan OPTIM OS1 Surface Cleaner Disinfectant Wipes
SciCan OPTIM OS1 is a one-step cleaner-disinfectant that removes soil and kills bacteria, viruses, and TB bacteria in just 1 minute. Features LAS Active System for effective cleaning and disinfection on hard, non-porous surfaces.