Basic Operation
-
Power On
Power on the unit. The power button is located inside the kick plate.
-
Load Instruments
Load instruments and close the door.
-
Select Cycle
Select the desired cycle from the touchscreen.
-
Start Cycle
Press START to begin the cycle.
Touchscreen Cycles
-
P1 Rinse & Hold Cycle
Prevents soil from drying on instruments. Suitable for incomplete loads that will not be washed within the hour. Note: This is NOT a cleaning cycle.
-
P2 Regular Cycle
For moderately soiled loose instruments.
-
P3 Heavy Duty Cycle
For heavily soiled instruments and cassettes.
Touchscreen Controls
- [Door Lock]
- [User/Technical Menu]
- [Network Connection]
- [USB]
- [Water Softener/Detergent Level/Door Status]
- [Enter]
- [Back]
Refilling the Water Softener Salt Reservoir
- Unscrew the lid.
- Fill to the top (max. 0.5 kg / 1.1 lbs.).
- Screw the lid on tightly.
- Run a complete cycle with no instruments.
IMPORTANT! If the salt container is being filled for the first time, it MUST be filled with at least 1 litre of water before adding the salt. Only tablet salt should be used. Fine-grained salt, such as table salt, should NOT be used as it could clog the unit. Water softener salt is available from SciCan. Please contact your local dealer.
Cleaning the Chamber
Pour 0.5L of vinegar or citric acid-based instrument neutralizer into the chamber and follow these steps:
P0 Cleaning
IMPORTANT! Do a Chamber Cleaning cycle every 25 cycles. Failure to clean the chamber can promote corrosion.
Replacing the HIP™ Cleaning Solution
Recommended precautionary measures for safe handling of HIP solution:
- Wear undamaged surgical gloves and eye protection.
- After handling the product, remove and dispose of gloves, and wash hands prior to removing eye protection.
Troubleshooting
Refer to the Operator's Manual for a more comprehensive list. Call for service if suggested recommendations do not resolve the error message.
Error Code / Symptom | Description / Checks |
---|---|
CF2 Chamber filling failure |
|
CF23 Top wash arm low RPM |
|
CF24 Bottom wash arm low RPM |
|
Spare Parts and Accessories
- CS-HIPC-U: HIP- Ultra Cleaning Solution (8 x 750 mL)
- 01-112594S: Water Softener Salt
- 01-108305S: Water Test Kits
- 01-114390S: HEPA Filters
- WTI-SK: Wash Test Indicator - Starter Kit
- WTI-RK: Wash Test Indicator - Refill Kit
Note: Terminal sterilization of instruments MUST follow processing in the HYDRIM C61W G4.
MSDS available for download at www.scican.com
HYDRIM is a registered trademark and HIP is a trademark of SciCan Ltd. Manufactured by SciCan Ltd. 1440 Don Mills Road, Toronto, Ontario, Canada, M3B 3P9.