WTW 9220 Online TOC-Analyzer

The WTW 9220 Online TOC Analyzer is designed for monitoring organic total carbon (TOC) in water quality, water treatment, and process water applications. More information on application possibilities and areas of use is available on the WTW website.

Product Views and Dimensions

Front view showing the analyzer's front panel, featuring a 7-inch WSVGA display with a capacitive touchscreen. The width is 55.9 cm.

Side view illustrating the analyzer's profile, revealing internal components and connections. The depth is 25.4 cm.

Rack installation view, depicting the analyzer mounted in a rack with a reagent tray. This view shows both side and front aspects. The height for this configuration is 113.7 cm.

Bottom view of the analyzer. The depth is 28 cm.

Overall dimensions for the analyzer are approximately 55.9 cm (width) x 25.4 cm (depth) x 113.7 cm (height) when configured for rack installation.

A visual representation shows the internal setup including reagent bottles, tubing, and electronic components within the analyzer.

Technical Specifications

CategorySpecificationValue
Model9220 Online TOC
PerformanceMeasurement PrincipleHeated Sodium Persulfate Oxidation
Measurement TechnologyNon-dispersive Infrared Detection (NDIR)
Compliance with RegulationsUSEPA 415.3 (Source and Drinking Water); SM 5310C (Water and Wastewater)
Measurement Range0 to 25 ppm (Standard); adjustable up to 100 ppm
Accuracy±5%; Precision: 2% RSD
Detection Limit0.015 ppm
Sample Preparation / Analysis Time5 to 9 minute intervals
Sample Preparation ConnectionsSample Preparation Connections6 (included)
Sample Preparation Valve Controls4 (included)
Sample Preparation ValvesUp to 4 (optional)
User Interface7" WSVGA display with capacitive touchscreen (Windows® CE-based)
GeneralDevice Housing CertificationsIEC 61326-1, IEC 61010-1, cETLus (ETL and CSA standards)
Protection RatingIP66; NEMA 4
WeightAnalyzer: 16.6 kg; Analyzer with reagent tray, PGM, and full reagent containers: 44.4 kg
Device Warranty2 Years
Reagents and Required ReagentsReagent Container10% Sodium Persulfate, 5% Phosphoric Acid, DI Water
Reagent Container Type5 L Polyurethane, high density
Reagent Lifespan (liquid)Nominal 30 days at 25 °C; 90 days at temperatures below 4 °C
Sample and Gas RequirementsFlow Rate to Sample Inlet50 to 1,000 ml/min using the sample inlet device
Inlet Pressure1 to 20 psig with the sample inlet device
Sample Temperature Range5 to 45 °C
Gas Requirement (internally generated)Process gas module (included); consumption ≤ 100 ml/min. CO2-free air
Power and CommunicationPower Supply100 to 240 VAC, 70VA, 50/60 Hz
Input Relays2 (Remote Start, Remote Stop); 5A/30 VDC Max - potential-free contact
Output Relays2 (System Alarm, Sample Alarm); 5A/30 VDC Max - potential-free contact
Analog Outputs4 (4 to 20 mA; user-configurable concentrations)
Digital OutputsRS-485/422 Modbus RTU protocol or ASCII standard
Data ExportTo PC via USB flash drive (Microsoft® Excel®-compatible .csv file format)
EnvironmentOperating Temperature Range5 to 45 °C
HumidityUp to 90% relative humidity (non-condensing)
Storage Temperature Range5 to 45 °C

Product Summary

Model: TOC 9220

Description: Online analyzer for determining total organic carbon (TOC) using sodium persulfate oxidation. Features 4 x 4-20 mA outputs, 2 x relays, and RS 485 Modbus protocol. Measurement range: 0 to 25 mg/L.

Order Number: 860200

PDF preview unavailable. Download the PDF instead.

v1 , WTW, Prozess-Segment Adobe PDF Library 17.0

Related Documents

Preview Alyza IQ PO4 Analyzer for Orthophosphate Measurement by WTW
Discover the Alyza IQ PO4, the most economical analyzer for direct orthophosphate measurement at the edge of basins. Learn about its revolutionary Multiport Valve, low reagent consumption, and precise results for wastewater treatment applications.
Preview WTW DW/P Drinking Water Panel: Operating Manual
Comprehensive guide to the WTW DW/P Drinking Water Panel, covering installation, safety, operation, maintenance, and technical specifications for water analysis.
Preview WTW VARION Plus, AmmoLyt Plus, NitraLyt Plus: Ammonium and Nitrate Measurement Sensors
Discover WTW's advanced sensors for precise ammonium and nitrate measurement in water quality monitoring. Featuring VARION Plus, AmmoLyt Plus, and NitraLyt Plus, these solutions offer easy operation, cost savings, and high performance in challenging media, integrating seamlessly with the IQ Sensor Net system.
Preview WTW Continuous Color Measurement Sensors for Water Analysis
Discover WTW's advanced sensors for continuous, compliant color measurement in water. Featuring ColorVis, CarboVis, and NiCaVis models, these instruments ensure reliable analysis of wastewater and surface water, adhering to ISO standards and offering benefits like turbidity compensation and ultrasonic cleaning.
Preview WTW VARION Plus, AmmoLyt Plus, NitraLyt Plus: Ion-Selective Electrodes for Ammonium and Nitrate Measurement
Detailed information on WTW's VARION Plus, AmmoLyt Plus, and NitraLyt Plus ion-selective electrode sensors for accurate ammonium and nitrate measurement in water analysis, featuring easy operation, cost savings, and advanced capabilities.
Preview IQ Sensor Net System Overview for Wastewater Treatment
Discover the IQ Sensor Net system by WTW (Xylem) for wastewater treatment plants. This overview details its advantages, features, and system comparisons for optimal monitoring.
Preview WTW TCML 412 N/M12-2 Chlorine Sensor Operating Manual
User manual for WTW TCML 412 N and TCML 412 M12-2 amperometric sensors for total chlorine measurement. Covers installation, operation, maintenance, safety, and technical specifications.
Preview WTW FCML 412 N/M12 pH-Independent Free Chlorine Sensor Operating Manual
This manual provides comprehensive instructions for the WTW FCML 412 N and FCML 412 M12 pH-independent amperometric sensors used for free chlorine measurement in water. It covers setup, operation, maintenance, and technical specifications for reliable water quality monitoring.