Alyza IQ PO4: The Most Economical Analyzer for Orthophosphate Measurement
Brand: WTW (a Xylem brand)
Tagline: Let's Solve Water
Your Benefits at a Glance
The revolutionary Alyza IQ PO4 offers precise results with extremely low liquid consumption, thanks to its new wet-chemical analysis system. It provides significant advantages:
- Minimizes reagent consumption and waste volume.
- Requires minimal maintenance.
- A service contract is not mandatory.
- Delivers reliable results, even in low measurement ranges.
Key Features
Multiport Valve
Easy and quick to change without the need for a service technician or maintenance contract.
Reagents
Safely packaged, long-lasting, and easy to replace.
Operation
Clear measurement value display, even for 2-channel devices, with monitoring of reagent levels.
Integration into IQ Sensor NET
The Alyza IQ PO4 integrates seamlessly as a sensor into the IQ Sensor NET system, compatible with System 2020 and 282/284. It powers the IQ Sensor NET with 10W.
For more information on the IQ Sensor NET, visit: www.IQ-Sensor.Net
For the latest information on the Alyza IQ PO4, visit the landing page: www.WTW.com/Alyza-IQ
Orthophosphate Measurement
Method
The Molybdate-Vanadate method (also known as the Yellow method) detects orthophosphate (dissolved PO43-). A color reagent containing molybdate and vanadate is added to the sample. In an acidic environment, the PO43- ions are incorporated into the resulting polyanions. Depending on the concentration of PO43- ions, a yellow color develops, the intensity of which is proportional to the PO43- amount. The colored sample is then analyzed in a photometer at approximately 420 nm.
Application Areas for Orthophosphate Determination
Control of Coagulant Dosing
- Automatic cleaning reduces maintenance effort and increases operational reliability.
- Reduces coagulant costs through reliable and fast measurements, preventing overdosing.
- One device can handle up to two sample streams.
Monitoring of Wastewater Treatment Plant Effluent
- Precise results through automatic 1- and 2-point calibration.
- Simplified daily work through self-diagnostics with reagent level display.
- Declining effluent limits are easily met due to the low measurement range.
Configuration Examples for Alyza IQ PO4 in IQ Sensor NET
Example 1 - System 2020
Configuration: MIQ/TC 2020 3G, MIQ/JB, SACIQ-7.0, IQ Sensors (selected as needed)
Part Number: 470022
Includes: MIQ/TC 2020 3G (Terminal/Controller), MIQ/JB (Junction Box), Alyza IQ PO4 (Analyzer with SNCIQ cable and power supply), 3 SACIQ (Sensor Connection Cables), 3 IQ Sensors (e.g., Turbidity, NH4, Oxygen).
Example 2 - System 282/284
Configuration: DIQ/S 284-CR6, Alyza IQ PO4, SACIQ-7.0, IQ Sensors (selected as needed)
Part Number: 472130
Includes: DIQ/S 284-CR6 (Controller), Alyza IQ PO4 (Analyzer with SNCIQ cable and power supply), 2 SACIQ (Sensor Connection Cables), 2 IQ Sensors (e.g., Oxygen, NH4), 1 IQ Spectral Sensor (e.g., COD).
Note: Neither Example 1 nor Example 2 requires a MIQ/PS power supply module, as the Alyza IQ provides 10W to the systems.
Technical Data
Alyza IQ PO4
- Model: Molybdate-Vanadate Method (Yellow Method)
- Measurement Range: Devices equipped with one measurement range. Available as 1- and 2-channel devices.
- Sample Streams/Channels: 1- and 2-channel devices available.
- Cleaning: Automatic cleaning with cleaning solution.
- Calibration: Automatic 1- or 2-point calibration.
- Operating Temperature: -20 to +40 °C
Measurement Ranges:
- MB1: 0.02 - 15.00 mg/l PO4-P (Display: 0.00 - 15.00 mg/l PO4-P, Resolution: 0.01 mg/l PO4-P, Accuracy: ±2% ±0.02 mg/l)
- MB2: 0.2 - 50.0 mg/l PO4-P (Display: 0.0 - 50.0 mg/l PO4-P, Resolution: 0.05 mg/l PO4-P, Accuracy: ±2% ±0.2 mg/l)
For further technical data, please refer to datasheet D2.25.
Ordering Information
Model | Description | Part Number |
---|---|---|
Alyza IQ PO4-111 | PO4 Analyzer, 1-channel, with MB1, Yellow Method, connectable to IQ SENSOR NET systems 2020 and 282/284, powers the IQ SENSOR NET with 10 W; includes 2 m SNCIQ cable; reagent set ordered separately | 825511 |
Alyza IQ PO4-112 | PO4 Analyzer, 2-channel, with MB1, Yellow Method, connectable to IQ SENSOR NET systems 2020 and 282/284, powers the IQ SENSOR NET with 10 W; includes 2 m SNCIQ cable; reagent set ordered separately | 825512 |
Alyza IQ PO4-121 | PO4 Analyzer, 1-channel, with MB2, Yellow Method, connectable to IQ SENSOR NET systems 2020 and 282/284, powers the IQ SENSOR NET with 10 W; includes 2 m SNCIQ cable; reagent set ordered separately | 825521 |
Alyza IQ PO4-122 | PO4 Analyzer, 2-channel, with MB2, Yellow Method, connectable to IQ SENSOR NET systems 2020 and 282/284, powers the IQ SENSOR NET with 10 W; includes 2 m SNCIQ cable; reagent set ordered separately | 825522 |
Reagent Sets
Model | Description | Part Number |
---|---|---|
R-Set PO4/1-1 | Reagents for Alyza IQ PO4-X1X with MB1 | 827550 |
R-Set PO4/1-2 | Reagents for Alyza IQ PO4-X2X with MB2 | 827551 |
SC-Set PO4/1-1_0/1 | Calibration standards and cleaning solution for Alyza IQ PO4-X1X with MB1; calibration standards with 0 mg/l and 1 mg/l | 827555 |
SC-Set PO4/1-1_0/10 | Calibration standards and cleaning solution for Alyza IQ PO4-X1X with MB1; calibration standards with 0 mg/l and 10 mg/l | 827556 |
SC-Set PO4/1-2_10/40 | Calibration standards and cleaning solution for Alyza IQ PO4-X2X with MB2; calibration standards with 10 mg/l and 40 mg/l | 827557 |
Xylem Analytics Germany Sales GmbH & Co. KG, WTW Dr.-Karl-Slevogt-Straße 18, 2362 Weilheim, Germany
Tel: +49 881 183-0 Fax: +49 881 183-420 Info.WTW@Xyleminc.com www.xylemanalytics.com
All names are registered trademarks or trademarks of Xylem Inc. or one of its subsidiaries. Technical changes reserved. ©2019 Xylem Analytics Germany Sales GmbH & Co. KG.