Xylem
Let's Solve Water

Your Perfect Partner for Ammonium and Nitrate Measurement

WTW - a xylem brand

Very Easy

The new matrix adjustment: Easier and faster than ever! Benefit from intuitive operation for simple adjustments.

Less Costs

Cross compensation: One electrode, several sensors. Compensate multiple sensors simultaneously, saving costs.

More Possible

Ammonium measurement in heavily polluted media: Measure high NH4-N concentrations (up to 2,000 mg/l) without chemicals.

Proven Quality

Leveraging over 15 years of experience in ion-selective on-line measurement, WTW offers sophisticated and reliable measurement techniques without compromises.

Applications include control of intermittent nitrification/denitrification, optimization/monitoring of simultaneous nitrification/denitrification, SBR process control, and measurement of high ammonium load in centrate and process water.

VARION® Plus

Multiparameter in situ probe with short response time, variably applicable. Ideal for fast measurement of Ammonium and Nitrate with dynamic compensation of potassium and/or chloride.

[Diagram: VARION Plus sensor, a multiparameter in situ probe, with labels for NH4 (Ammonium), NO3 (Nitrate), and Reference. Also indicates K+ (Potassium) and Cl- (Chloride) compensation.]

AmmoLyt® Plus

Single parameter in situ probe for the measuring of Ammonium with dynamic compensation of potassium.

[Diagram: AmmoLyt Plus sensor, a single parameter in situ probe for Ammonium, with labels for NH4 (Ammonium) and Reference. Also indicates K+ (Potassium) compensation.]

NitraLyt® Plus

Single parameter in situ probe for the measuring of Nitrate with dynamic compensation of chloride.

[Diagram: NitraLyt Plus sensor, a single parameter in situ probe for Nitrate, with labels for NO3 (Nitrate) and Reference. Also indicates Cl- (Chloride) compensation.]

Calibration and Electrodes

Calibration Free: WTW's ISE sensors measure directly in the process without chemicals or sample taking. Mature technology makes calibration with standard solutions obsolete, ensuring long-term stable measurement values.

Single Exchange of Electrodes: The single electrode principle ensures only the necessary component is replaced.

Robust Electrodes: WTW's robust ISE electrodes simplify handling and cleaning.

Direct Connection to the IQ Sensor Net

Benefit from connection of up to 20 sensors, measurement of all important parameters, and maximum communication possibilities.

For more detailed information, visit: www.xylemanalytics.com/en/landingpages/iq-sensor-net

Technical Data

VARION® Plus • AmmoLyt® Plus • NitraLyt® Plus
VARION® Plus: Common reference electrode, two measuring electrodes, one compensation electrode AmmoLyt® Plus: Common reference electrode, Ammonium measuring electrodes, one compensation electrode NitraLyt® Plus: Common reference electrode, Nitrate measuring electrodes, one compensation electrode
Maximum configuration Ammonium measurement Ammonium measurement Nitrate measurement
Integrable Electrodes:
Reference electrode VARION® Plus Ref VARION® Plus Ref VARION® Plus Ref
Measuring electrode VARION® Plus NH4 VARION® Plus NH4 VARION® Plus NO3
Compensation electrode VARION® Plus K VARION® Plus K VARION® Plus Cl
Measuring Ranges/ Resolution
NH4-N: 1 ... 2,000 mg/l / 1 mg/l;
0.1 ... 100.0 mg/l / 0.1 mg/l
NH4+: 1 ... 2,580 mg/l / 1 mg/l;
0.1 ... 129.0 mg/l / 0.1 mg/l
NO3-N: 1 ... 1,000 mg/l / 1 mg/l;
0.1 ... 100.0 mg/l / 0.1 mg/l
NO3-: 5 ... 4,500 mg/l / 5 mg/l;
0.5 ... 450.0 mg/l / 0.5 mg/l
Compensation Ranges
K+: 0.1 ... 1,000 mg/l / 0.1 mg/l K+: 0.1 ... 1,000 mg/l / 0.1 mg/l Cl-: 0.1 ... 1,000 mg/l / 0.1 mg/l
Temperature Measurement Integrated NTC thermistor, Range 0 °C ... + 40 °C, Accuracy ± 0.5 K, Resolution 0.1 K
Temperature Compensation 32 °F ... 104 °F (0 °C ... + 40 °C)
Matrix adjustment Yes (adjustment against photometric measurement or other measurement technique)
Calibration Procedures Calibration against any reference value, 2-point-calibration possible with multiple standard solutions
Ambient Conditions Operating temperature: 32 °F ... 104 °F (0 °C ... + 40 °C), storing temperature: 32 °F ... 104 °F (0 °C ... + 40 °C)
pH Range pH 4 ... pH 8.5 pH 4 ... pH 8.5 pH 4 ... pH 11
Measuring Accuracy in laboratory standard solutions ± 5% of measured value ± 0.5 mg/l
Working Life (typically) All electrodes: 18 months (in typical application - sewage plants)
Mechanical
Sensor body V4A Stainless Steel 1.4571
Temp. sensor V4A Stainless Steel 1.4571
Electrode connector POM
Protective cap POM
Protection rating IP 68 (0.2 bar, with installed electrodes)
Max. Pressure Maximum 0.2 bar (incl. SACIQ sensor connection cable; with installed electrodes)
Power Consumption 0.2 Watt
Dimensions 15.43 x 1.57 in. (392 x 40 mm, length x diameter), incl. SACIQ sensor connection cable
Weight Approx. 1.48 lb (approx. 670 g, incl. protective cap, without SACIQ sensor connection cable)
Guaranty VARION® Plus 700 IQ, AmmoLyt® Plus 700 IQ, NitraLyt® Plus 700 IQ: 2 years for defects of quality
Electrodes: 12 months for defects of quality

Ordering Information

Sets include Probe, reference electrode, corresponding measuring and compensation electrode.

Model Description Order No.
AmmoLyt® Plus SET/Comp for the measurement of Ammonium with compensation electrode 107072
NitraLyt® Plus SET/Comp for the measurement of Nitrate with compensation electrode 107082
VARION® Plus AN/A comp SET for the measurement of Ammonium with compensation, plus Nitrate measurement 107066
VARION® Plus AN/N comp SET for the measurement of Nitrate with compensation, plus Ammonium measurement 107068

Xylem Analytics Germany Sales GmbH & Co. KG, WTW

Am Achalaich 11, 82362 Weilheim, Germany

Phone: +49 881 1830 | Fax: +49 881 183-420 | Info.WTW@xylem.com | www.xylemanalytics.com

All names are registered tradenames or trademarks of Xylem Inc. or one of its subsidiaries. Technical changes reserved. © 2017 Xylem Analytics Germany Sales GmbH & Co. KG.

PDF preview unavailable. Download the PDF instead.

v1 , WTW, Prozess Segment Adobe PDF Library 17.0

Related Documents

Preview WTW VARION Plus, AmmoLyt Plus, NitraLyt Plus: Ion-Selective Electrodes for Ammonium and Nitrate Measurement
Detailed information on WTW's VARION Plus, AmmoLyt Plus, and NitraLyt Plus ion-selective electrode sensors for accurate ammonium and nitrate measurement in water analysis, featuring easy operation, cost savings, and advanced capabilities.
Preview WTW VARION® Plus, AmmoLyt® Plus, NitraLyt® Plus Sensoren für Ammonium- und Nitratmessung
Entdecken Sie die WTW VARION® Plus, AmmoLyt® Plus und NitraLyt® Plus Sensoren von Xylem für präzise und einfache In-Situ-Messungen von Ammonium und Nitrat in Wasser. Erfahren Sie mehr über Kalibrierfreiheit, Robustheit und Integration ins IQ Sensor Net.
Preview WTW VARION® Plus Digital ISE Combination Sensor for Ammonium and Nitrate
Technical datasheet for the WTW VARION® Plus digital ISE combination sensor, measuring Ammonium and Nitrate ion-selectively with automatic compensation. Details specifications, measurement ranges, and ordering information.
Preview WTW VARION® Plus, AmmoLyt® Plus, NitraLyt® Plus: Soluzioni per Misurazione Ammonio e Nitrato
Scopri le soluzioni WTW VARION® Plus, AmmoLyt® Plus e NitraLyt® Plus per la misurazione accurata e semplice di ammonio e nitrato in acqua. Caratteristiche avanzate, calibrazione facile e robustezza per applicazioni industriali.
Preview Capteurs WTW VARION Plus, AmmoLyt Plus, NitraLyt Plus : Mesure d'Ammonium et de Nitrate
Découvrez les capteurs ion-sélectifs avancés de WTW pour la mesure précise de l'Ammonium et du Nitrate dans l'eau. Apprenez les caractéristiques, spécifications techniques et informations de commande des VARION Plus, AmmoLyt Plus et NitraLyt Plus.
Preview Alyza IQ PO4 Analyzer for Orthophosphate Measurement by WTW
Discover the Alyza IQ PO4, the most economical analyzer for direct orthophosphate measurement at the edge of basins. Learn about its revolutionary Multiport Valve, low reagent consumption, and precise results for wastewater treatment applications.
Preview WTW Continuous Color Measurement Sensors for Water Analysis
Discover WTW's advanced sensors for continuous, compliant color measurement in water. Featuring ColorVis, CarboVis, and NiCaVis models, these instruments ensure reliable analysis of wastewater and surface water, adhering to ISO standards and offering benefits like turbidity compensation and ultrasonic cleaning.
Preview IQ Sensor Net System Overview for Wastewater Treatment
Discover the IQ Sensor Net system by WTW (Xylem) for wastewater treatment plants. This overview details its advantages, features, and system comparisons for optimal monitoring.