ProSense Field Device Configurator Software Help File

Manual Number: Field Device Configurator Help File

Issue: 1st Edition

Issue Date: 04/2023

Publication History:

Issue Date Description of Changes
1st Edition 04/2023 Original

1. General

ProSense Field Device Configurator PC software is a utility program that allows users to easily configure ProSense XTH2-0-UNV and XTD2-0-UNV temperature transmitters. Download your free copy of the ProSense Field Device Configurator at www.AutomationDirect.com and connect your ProSense temperature transmitter through an XT-USB configuration cable. (XT-USB purchased separately).

Connection Examples

XTH2-0-UNV Connection

A diagram shows a laptop connected via USB-Port to an XT-USB Configuration Cable, which connects to the XTH2-0-UNV device via a cable socket.

XTD2-0-UNV Connection

A diagram shows a laptop connected via USB-Port to an XT-USB Configuration Cable, which connects to the XTD2-0-UNV device via a cable socket.

2. Operation

2.1 Home

After ProSense Field Device Configurator has been installed and started, the Home Screen appears. If the Manual Connection icon does not appear, select Settings and enable the Manual Connection Mode (see Section 2.2).

The Home screen displays two options: 'Automatic Connection' and 'Manual Connection'.

Figure 1 shows the Home Screen with options for Automatic Connection and Manual Connection.

On the home screen, you have the option to choose Automatic Connection or Manual Connection. If the Manual Connection icon does not appear, select Settings and enable the Manual Connection Mode (see Section 2.2). The Automatic Connection will try to connect to the device using the available COM port on your PC. If it is unable to connect or you have multiple COM devices available, it will prompt you to use the Manual Connection mode. Selecting Manual Connection will allow you to select the COM Port and Baud rate manually as seen in Fig. 2. Once selected, click the connect button.

Figure 2 shows a connection dialog with fields for Protocol, Com Ports, and Baud rate, and a 'Connect' button.

2.2 Settings

Selecting Settings allows you to change the language, switch between Basic and Expert Setup modes, and enable or disable the Manual Connection mode.

Figure 3 shows the Settings screen with options for Language, Mode (Basic setup/Expert), and Manual Connection Mode (Enable/Disable).

2.3 Setup

After connection has been made with the transmitter, the side menu and either the Basic or Expert Setup screen will be displayed (depending on the Setup Mode selected in Settings, see Section 2.2).

Figure 4 shows the left-hand navigation menu with options like Logout, Setup, Create Report, Save/Restore, and Settings. The main panel shows device information, including Device Tag, Unit, Sensor type, Sensor value, Access status, and configuration parameters like Lower range value output, Upper range value output, and Failure mode.

2.3.1 Logout

The Logout button in Fig. 4 will disconnect you from the transmitter.

2.3.2 Setup

Setup in Fig. 4 will display either the Basic or Expert parameter session. The option for Basic and Expert is found in the Settings.

2.3.3 Create Report

Create Report in Fig. 4 allows you to create a PDF file of the Basic or Expert Setup.

2.3.4 Save/Restore

Save/Restore in Fig. 4 allows you to save the current configuration to a file or restore a previously saved configuration file to the connected device.

2.3.5 Settings

Settings in Fig. 4 allows you to change the language, switch between Basic and Expert setup modes, and enable or disable the manual connection mode.

2.3.6 Info

The Info button on the bottom of the window displays contact information, licensing information, and how to retrieve logging data. See Figure 5.

Figure 5 displays the 'About' section with contact details for Automation Direct, a disclaimer regarding software copyright, and instructions for error reporting (Ctrl+Alt+L). It also lists software licenses used.

2.4 Basic Setup

The availability of some parameters is dependent on the selection of other parameters (for example, the Reference Junction parameter is only available when a Thermocouple input type is previously selected).

Figure 6 shows the Basic Setup screen for an XTD2-0-UNV device, displaying fields for Device Tag, Unit, Sensor type, Sensor value, Access status, Connection type, Lower range value output, Upper range value output, and Failure mode.

Basic Settings

Sensor Types:

Input Type Programmable Measuring Range Limits Min. Span
Inputs
Pt100 RTD
Pt200 RTD
Pt500 RTD
Pt1000 RTD
(to IEC 751)
a=0.003851)
-328 to 1562°F (-200 to 850°C)
-328 to 1562°F (-200 to 850°C)
18°F (10°C)
Pt100 RTD
(to JIS C1604)
(a=0.0039)
-328 to 932°F (-200 to 500°C) 18°F (10°C)
Ni100 RTD
Ni120 RTD
(to DIN 43760)
(a=0.006180)
-76 to 482°F (-60 to 250°C) 18°F (10°C)
Ni100 RTD
Ni120 RTD
(to OIML, GOST)
(a=0.006170)
-6 to 356°F (-60 to 180°C) 18°F (10°C)
Pt50 RTD
Pt100 RTD
(to GOST)
(a=0.00390)
-301 to 2012°F (-185 to 1100°C)
-328 to 1562°F (-200 to 850°C)
18°F (10°C)
Pt100 (Callendar van Dusen)
Nickel polynomial
Copper polynomial
The measuring range limits are specified by entering the limit values that depend on the coefficients A to C and R0. 18°F (10°C)
Cu50 RTD
Cu100 RTD
(to OMIL, GOST)
(a=0.004280)
-292 to 392°F (-180 to 200°C) 18°F (10°C)
Cu50 RTD
(to OMIL, GOST)
(a=0.004260)
-8 to 392°F (-50 to 200°C) 18°F (10°C)

Selection of the unit for all measured values.

Input Type Programmable Measuring Range Limits Min. Span
Inputs
Resistance Ω
RTDs:
• Connection type: 2-, 3-, or 4-wire connection
• Software compensation of cable resistance possible in the 2 wire system (0-300)
• Sensor cable resistance max. 500 Ω per cable in the 3 and 4 wire system
• Sensor current: ≤ 0.3mA
10 to 400 Ω
10 to 2000 Ω
10 Ω
Thermocouples:
Type A
Type B
Type E
Type J
Type K
Type N
Type R
Type S
Type T
(to IEC 60584, Part 1)
32 to 4532°F (0 to +2500°C)
104 to 3308°F (104 to +1820°C)
-482 to 1832°F (-250 to +1000°C)
-346 to 2192°F (-210 to +1200°C)
-454 to 2501°F (-270 to +1372°C)
-454 to 2372°F (-270 to +1300°C)
-58 to 3214°F (-50 to +1768°C)
-58 to 3214°F (-50 to +1768°C)
-328 to 752°F (-200 to +400°C)
90°F (50°C)
Thermocouples:
Type C
Type D
(to ASTM E988)
32 to 4199°F (0 to +2315°C)
32 to 4199°F (0 to +2315°C)
90°F (50°C)
Thermocouples:
Type L
Type U
(to DIN 43710)
-328 to 1652°F (-200 to +900°C)
-328 to 1112° (-200 to +600°C)
90°F (50°C)
Thermocouple:
Type L
(to GOST)
-28 to 1472°F (-200 to +800°C) 90°F (50°C)
Thermocouples:
• Internal cold junction (Pt100) or external programmable fixed value, -40 to 185°F (-40 to 85°C)
• Accuracy of cold junction: ± 1.8°F (1°C)
• Sensor current: 30nA
• Maximum sensor wire resistance 10kΩ
(If the sensor wire resistance is greater than 10 kΩ, an error message is output in accordance with NAMUR NE89.)
Millivolt (mV) -20 to 100 mV 5 mV

Basic Setup Continued

Device Tag: Custom identification name for this transmitter. Up to 32 characters.

Connection type: RTD wiring connection (only available when an RTD Sensor type is selected). Options: 2-wire, 3-wire, 4-wire.

Units: Temperature engineering units. Options: °C, °F, K, Ohm, mV.

Lower Range Value Output: Input the low end of the desired temperature range (4mA value). Range: -50000 to 50000.

Upper Range Value Output: Input the high end of the desired temperature range (20mA value). Range: -50000 to 50000.

Polynomial Coefficient R0, A, and B: Only available when Polynomial RTD Sensor type is selected. Enter coefficients for custom linearization of an RTD sensor.

Calendar/van Dusen coefficient R0, A, B, and C: Only available when RTD Platinum (Callendar/van Dusen) sensor type is selected. Enter coefficients for custom linearization of an RTD sensor.

Sensor lower limit: Only available when Polynomial or Callendar/van Dusen RTD Sensor types are selected. Lower calculation limit for special sensor linearization.

Sensor upper limit: Only available when Polynomial or Callendar/van Dusen RTD Sensor types are selected. Upper calculation limit for special sensor linearization.

2-wire compensation: Input the resistance value for two-wire compensation in RTDs. Only available when 2-wire compensation type is selected.

Reference Junction: Selection of the internal cold junction compensation, fixed value, or external comparison to a temperature value. (only available when a Thermocouple Sensor type is selected).

RJ preset value: Input fixed temperature comparison value with Fixed Value is selected. Range: -15 to 360.

2-wire compensation: Input of cable resistance compensation (Only available when a RTD Sensor type with 2 wire connection is selected). Range: 0.0 to 30.0 ohms.

Failure mode: Input the desired output current value on a sensor break or short circuit. Options: Max 21.5-23mA user selectable, Min. ≤ 3.6 mA.

Expert Setup

Figure 7 shows the Expert Setup screen, which includes all Basic Setup parameters plus additional options. It displays fields for Sensor value, Access status, Lower range value output, Upper range value output, Failure mode, Current trimming (4mA and 20mA), and Damping. It also includes sections for Diagnostics and Simulation.

Expert Setup includes all the Basic Setup parameters as well as the following:

Diagnostics

Simulation

2.6 Device Settings and Password Protection

Device Settings

System

Device Management

Device Tag: Custom identification name for this transmitter. Up to 32 characters.

Mains filter: Select the mains filter for the A/D conversion.

User Management

By default, there is no write-protection on the device. By defining a password, you enable write-protection and the defined password must be entered to switch into maintenance role to change settings.

Define Password: Enter and confirm a new password. Must consist of at least 4 and a maximum of 16 characters and can contain letters and numbers.

Figure 8 shows the User Management screen with options to Define Password, Enter Password, Change Password, Delete Password, and Reset Password.

Enter Password: Select user role and input maintenance password.

Change Password: Enter old password and setup new password.

Delete Password: Select delete password which sets the device to no write-protection.

Reset Password: In the event you have forgotten the password, contact Automation Direct Technical Support for the reset password for your device.

PDF preview unavailable. Download the PDF instead.

fdconfig Adobe PDF Library 17.0 Adobe InDesign 18.2 (Windows)

Related Documents

Preview ProSense Thermocouple Probes with Connection Head - Technical Specifications and Overview
Detailed information on ProSense thermocouple probes with connection heads, including types J and K, specifications, dimensions, wiring, and available accessories from AutomationDirect.
Preview ProSense FMM Series (-1002) Magnetic-Inductive Flow Meters
Explore the ProSense FMM Series (-1002) Magnetic-Inductive Flow Meters, designed for reliable liquid flow rate detection. This document provides an overview, features, specifications, and selection guides for these robust and compact flow meters.
Preview ProSense FMM Series (-1001) Magnetic-Inductive Flow Meters
AutomationDirect's ProSense FMM Series (-1001) Magmeters are designed for reliable flow rate detection of conductive media up to 160 GPM. These stainless steel, mechanically-robust meters offer in-line mounting, a 4-digit numeric display with pushbutton setup, and selectable engineering units. They feature two outputs for remote monitoring of binary or analog status, and are a reliable alternative to traditional flow meters.
Preview ProSense DPM3-P Series 1/8 DIN Digital Panel Meters
The ProSense DPM3-P series digital panel meters offer a simple, feature-packed solution for counter, chronometer, frequency, tachometer, rate, and pulse width modulated (PWM) applications. These 1/8 DIN meters feature a 5-digit tri-color LED display, selectable decimal points, and accept various input types including AC voltage, magnetic sensors, NPN/PNP sensors, NAMUR sensors, TTL/24V encoders, and switched contacts. They are available with optional dual SPDT relay outputs and a 4-20mA analog output. The meters are powered by an external AC or DC supply and feature an IP65 rated front panel. Configuration parameters can be locked to prevent unauthorized changes. The series is backed by a 3-year warranty.
Preview ProSense 10K-3 Thermistor Probes with Lead Wire Transition
ProSense 10K-3 thermistor probes feature a 1/4 inch diameter, 316 stainless steel sheath for harsh environments. Available in 6", 12", or 18" probe lengths with a 4-point crimp and sleeve transition to a 6-foot PVC cable. These probes are made in the USA and offer a temperature sensing range of -40 to 125°C (-40 to 257°F). Technical specifications include a 10kΩ no-load resistance at 25°C and an accuracy of ± 0.2°C from 0°C to 70°C. Accessories such as brass and stainless steel compression fittings are also available.
Preview ProSense RTD Probes, Compression Fittings, Bayonet Adapters, and Connectors
Explore ProSense temperature sensing solutions including RTD probes with connection heads, compression mounting fittings, bayonet adapters, and thermocouple/RTD connectors. Detailed specifications, dimensions, and part numbers for industrial applications.
Preview ProSense DPM4 Advanced High Speed Process Panel Meters User Manual
User manual for the ProSense DPM4 Advanced High Speed Process Panel Meters, covering installation, operation, programming, and technical specifications. Includes details on input configuration, output options, and troubleshooting.
Preview ProSense FMM Series (-1002) Magnetic-Inductive Flow Meters
Automation Direct's ProSense FMM Series (-1002) Magmeters offer reliable detection of conductive liquid flow rates up to 158.5 GPM. These compact, stainless steel meters feature a 4-digit display, selectable engineering units, and two analog outputs for remote monitoring of flow and temperature. They are suitable for various industrial process applications.