Wassermann Wamix-Economy Vacuum Mixer

Model: Wamix-Eco (Item No. 171972)

USER MANUAL

Introduction

Dear customer,

Thank you for choosing a product from the Wassermann range. Wassermann Dental-Maschinen incorporates the highest standards of quality and the latest technology. In order to enjoy maximum performance and years of trouble-free operation, please read this user manual carefully before you connect this device and start work, and operate the device according to the recommended guidelines. The operation safety and the functionality of this device can only be guaranteed if you follow both the general safety guidelines and the applying laws to prevent accidents as well as the precautions given in this user manual. Wassermann Dental-Maschinen GmbH is not liable for any damages which occur due to inappropriate usage or faulty operation of this device.

ℹ️ Make sure that anyone using this device has read and understood this user manual. Keep this user manual in a safe place where it can be referred to as required at any time.

CE The unit complies with the relevant EU guidelines.

The unit is subject to the EU guidelines 2012/19/EU (WEEE Directive).

Company Address:

Wassermann Dental-Maschinen GmbH
Rudorffweg 15-17
21031 Hamburg, Germany
Phone: +49 (0)40 730 926-20/-24
Fax: +49 (0)40 730 37 24
Email: export@wassermann-dental.com
Website: www.wassermann.hamburg
Facebook: f/WassermannDentalmaschinen

Contents

1 Features

The Wamix-Economy units are modern vacuum mixers used not only for mixing all kinds of plasters, embedding media and silicones, but also designed for specialised dental procedures and for vacuum-rinsing of prostheses.

  • Designed for long life
  • Flexible use as well as table unit (table stand not included)
  • Low-maintenance and easy to clean
  • For all types of dental plasters, investments and silicones
  • Digital timer
  • Inclined motor for easy insertion of the mixer beaker
  • Comfortable single-hand operation
  • High-quality workmanship, made of metal
  • Can be mounted in just few steps

2 Safety Symbols used in this Manual

⚠️Warning! This is a warning of risk situations and dangers. Failure to observe this warning could be life-threatening. These warnings must be observed.

ℹ️Information! This symbol draws your attention to specific features that have to be observed.

3 Safety Guidelines

Configuring and operating this equipment requires precise knowledge and observance of the instructions in this user manual. The equipment is designed only for its intended application.

⚠️WARNING: Servicing and repairs should be carried out only by authorised specialists. Disconnect the power plug before starting any maintenance work.

The unit is designed to operate on 100-240 V / 50/60 Hz. Make sure that the equipment is connected to the correct power source.

A safety angle underneath each unit prevents it coming loose once it has been mounted in the bracket.

Allow the condensers to discharge before opening the unit.

4 Responsibility for Operation or Damage

The responsibility for operating the device lies exclusively with the owner or user if said device is incorrectly serviced, maintained or altered by persons not employed by an authorised dealer or if the device is used in a manner contrary to its specified purpose. The unit has to be maintained and operated in accordance with this user manual. Wassermann Dental-Maschinen GmbH is not responsible for damage arising from the nonobservance of these instructions.

Warranty and responsibility provisions contained in the sales and supply conditions of Wassermann Dental-Maschinen GmbH are not extended by these instructions.

4.1 Warranty

ℹ️ The warranty period for our equipment is 12 months. If faults occur within the warranty period, contact your dental depot or get in touch directly with our service department.

Your equipment should only be operated in perfect condition. If faults occur which could harm operators or third parties, the unit should not be used until it has been fixed.

This warranty does not cover damage caused by improper use, external mechanical causes, transport damage or interference with the unit by unauthorized persons.

5 Application

The Wamix units are used for mixing all kinds of plasters, embedding media and silicones under vacuum. These units are also suitable for vacuum rinsing of models for special dental procedures. The insulating layer between the model and the prosthesis cannot separate prematurely because the model remains moister in the pressure-embedding unit than when water is used.

Only use the device for this type of application.

6 Before Starting

6.1 Storage

The unit location is very important when it comes to workplace safety, even if it is only to be set up there for a short period. The room should be dry, well ventilated and vibration-free. An even temperature and wooden supports also help. The unit should not be stored or set up outdoors. If the unit is to be stored for an extended period, protect it from moisture and dust.

6.2 Transport

Before transporting the unit, ensure that it has been unplugged from the power socket. Make sure that it is packed correctly in order to avoid accidental damage.

ℹ️ Be sure to check for any transport damage when unpacking the goods. Note down any damage if found.

6.3 Installation

Open the box, remove the packing materials, and carefully lift out the device and accessories. Check the included accessories. The device with stand has to stand horizontally on a steady and even surface.

6.4 Wall Mounting

If mounting the unit on the wall, make sure that the wall is strong enough to take the weight. Use the supplied bracket to mount it horizontally on the wall. For drawing a drilling scheme to drill the dowel holes please see page 9. Leave at least 30-35 cm free space beneath the unit.

⚠️ Install the device in a place where it will not block the working area and the functionality (take the dimensions into account).

ℹ️ Do not install the unit outdoors or in places without proper ventilation. Before start-up, be sure the device reaches room temperature.

6.4.1 Mounting Instruction Fastening Set for Wall Mounting

Please check the included accessories:

QuantityName
1Safety angle
1Phillips head screw M 4x8
1Washer Ø 4,3
1Serrated washer Ø 4,3
1Slotted screw 4 x 16
2Slotted screw 6 x 50
2Screw anchor S8
1Screw anchor S5

Mounting Instruction: Please drill the holes as indicated in the following drilling template. Please make sure before that the wall is really suitable for mounting a device (aerated concrete walls and drywall are problematic).

Drilling Template:

Bohrung=drilling, tief=deep

A diagram shows two holes to be drilled: the upper hole is 8 mm in diameter and 55 mm deep, located 140 mm from the left edge and 150 mm from the top. The lower hole is 5 mm in diameter and 30 mm deep, located 34 mm from the left edge.

7 Installation/ Start-up

7.1 Before Starting the Unit, Connect Up the Following:

ℹ️ Please check that the mains voltage and the voltage requirements of the mixer are the same, then plug in the power cable.

7.2 Start-up

  • Switch on the green mainswitch (Fig. 4).

Fig. 4: Image of a green, illuminated power switch.

ℹ️ Accidental mixing is not possible because the motor cannot start until the beaker is held to the rim by the vacuum.

⚠️ Before you start your vacuum mixer, make sure that the safety hook has been fastened under the unit. Push the hook to the wall using the slot and fasten the safety hook to the wall. Then retighten the screw on the chassis to prevent it coming loose.

⚠️Warning: Use only beakers with the new curved seal (171708). (For details please see 10.3 Accessories).

ℹ️ If the unit will not be used for an extended period, switch off the vacuum mixer.

8 Operation

The Wamix-Eco concept is characterized by its simplicity. Only the times for the pre-vacuum and the mixing time are set.

A diagram shows the device's digital display. The top line shows '18.4' in red, labeled 'prevacuum (red)'. The bottom line shows '2.2.8' in blue, labeled 'mixing time (bue)'. Buttons labeled '+' and '-' are visible, along with a 'SET' button.

1. Set the mixing timer (blue) (default 35 sec.)

  • Press SET twice (short) - time 35 starts flashing.
  • Press + or - to adjust the time (e.g., 40).
  • Wait 5 seconds until the time (e.g., 40) stops flashing.
  • Continue with step 2, or turn off then turn on.

2. Set the pre-vacuum timer (red) (default 8 sec.)

  • Press SET once (short) - time 8 starts flashing.
  • Press + or - to adjust the time (e.g., 10).
  • Wait 5 seconds until the time (e.g., 10) stops flashing.
  • Turn off then turn on.

ℹ️Important! If "P0" or "P1" appears in the display, please wait 5 seconds.

8.1 General Operating Instructions

All instructions for using the unit, whether in verbal or written form, are based on our own experience and experimentation and can only be regarded as guidelines.

ℹ️ Fill the beaker to the Max marking. Please pay special attention to the lower Max marking for Silicon. Otherwise, the vacuum would draw any excess material into the system. Excess material would be appeared in the "control window" (Fig. 3).

ℹ️ When mixing silicones, make sure that the beaker is only filled to the max. height for silicone. Otherwise, silicone could penetrate the interior of the unit itself.

ℹ️ The beaker will come away automatically once the vacuum is removed. To avoid damaging the seal, never remove the beaker by force.

ℹ️ The sealing surfaces between the beaker and the plate must be clean - likewise the rotor fastenings and the beaker lid - to ensure trouble-free operation.

9 Troubleshooting

FaultCauseRemedy
Mixer does not startBlown fuseReplace fuse and determine cause; otherwise, call Service Center
No mixing time programmedSet mixing time
No rpm programmedAdjust the rpm
Vacuum too lowFilter (vacuum) blockedCheck filter – clean and/or change (see 10.2.1 Fig. 8)
No vacuumDirty or leaking seal on vacuum flange for mixing beakerClean or replace seal
Beaker filled above Max. mark, i.e., mixing ingredients in system/"control window"Clean system, i.e., maintenance by Service Center
Pump brokenMaintenance by Service Center
Beaker falls outWrong couplingChanging the coupling (171708)
Wrong program parameterPlease check the mixing time of the program. The mixing time must be larger than 0s.
Malfunction timerDisplay Reset:
press SET (>3 sec.)
"P0" starts flashing
press + or - until 1 appears
"P0"= 0 (default)
wait 5 sec. until P0 stops flashing
press SET (>3 sec.)
"P0" starts flashing
press SET (short)
"P1" appears
press + or - until 2 appears
"P1" = 2 (default)
wait 5 sec. until P0 stops flashing

ℹ️ If the above recommendations do not solve the problem, contact your dental depot or our service department.

10 Care and Maintenance

10.1 Cleaning

⚠️ Disconnect the power plug before starting any maintenance work. The identification plate must always be kept in easily legible condition and must not be removed.

ℹ️ Remove external dirt from time to time with some form of cold cleaner. Use only cold cleaners to avoid damaging the paintwork or removing the lettering. The equipment should be cleaned at regular intervals to ensure trouble-free operation. It requires only normal cleaning (sponge, damp cloth, mild detergent) and no further chemical additives.

Solvents must not be used to clean the front panel!

Keep the beaker clean. When not in use, fill with water to just below the mixer.

Cleaning mixing beakers in an ultrasonic bath, especially if plaster solvents are also used in the beaker, will inevitably result in changes to the structure of the beaker material and will therefore damage the mixing beaker.

10.2 Maintenance

ℹ️ Disconnect the power plug before you open the unit. Allow the condensers to discharge before opening the unit.

Our vacuum mixers do not require much servicing. Make sure the unit is kept clean and that the rim seal for the beaker (vacuum seal) is not damaged.

10.2.1 Changing the Filter

The protective filter is located in the housing. No special expertise is required to replace it.

  • Pull out the mains plug.
  • Dismantle the cover.
  • Disconnect vacuum meter hose and plug connection.
  • Dismantle the casing and pull it forwards.
  • Disassemble the protective filter, assemble a new one and secure with a cable tie.
  • Note installation position (direction of flow) (see sketch).
  • Assembly in reverse order.

A diagram shows the internal components of the Wamix unit, with the protective filter labeled (Item No. 610021).

Fig. 8: Internal view showing protective filter location.

10.3 Scope of Delivery/ Accessories

Item no. 171970 Wamix-Eco

Included in delivery:

  • Beaker, medium cpl. 350 ml
  • Mounting set for wall fastening

The Wamix units are delivered as wall models but are convertible as counter devices at any time. A table stand is available as extra.

Accessories

AccessoryItem no.:
Table stand (W 208 x H 450 x D 275 mm)171744
Beaker, extra small cpl. 75 ml171295
Beaker, small cpl. 200 ml171230
Beaker, medium cpl. 350 ml171283
Beaker, large cpl. 600 ml171237
Beaker, extra large cpl. 1000 ml171269
Vacuum washing beaker, 600 ml171275

10.4 Repairs

⚠️ Servicing or repairs to the unit must only be carried out by qualified technicians. Only original spare parts are to be used. Responsibility for the product is voided if unauthorised persons alter it or if inappropriate components are installed.

10.5 Spare Parts

If necessary, please contact our service hotline.

Service Hotline: 0049 (0)40 730 926 -20/-24

11 Technical Data

Item no. 171972 Wamix-Eco

Voltage220–240 V / 50/60 Hz
Power consumption max.0,44 A
Output94 W
Speed500 min⁻¹
Vacuum*approx. 130 mbar (absolute)
Mixing time999 sec. adjustable
W x H x D195 x 310 x 180 mm
Weight7,1 kg
Sound level≤ 70 dB (A)
Safety classIP 22

ℹ️ Technical changes reserved.

* „Vacuum" means the state of a gas when in a container the pressure of the gas and thus the particle number density is lower than outside or when the pressure of the gas is lower than 300 mbar (smaller than the lowest atmospheric pressure occurring on the earth's surface).

12 Disposing of the Unit

The unit has to be disposed of by an authorized recycling operation. The selected company has to be informed of all possibly health-hazardous residues in the unit.

12.1 Information on Disposal for Countries within the EU

♻️ To conserve and protect the environment, prevent environmental pollution and improve the recycling of raw materials, the European Commission adopted a directive that requires the manufacturer to accept the return of electrical and electronic units for proper disposal or recycling.

Within the European Union, units with this symbol should not therefore be disposed of in unsorted domestic waste.

For more information regarding proper disposal, please apply at your local authority.

13 EU Declaration of Conformity

In accordance with 2014/35/EU (Low Voltage Directive) and 2014/30/EU (Electromagnetic Compatibility Directive) and 2006/42/EC (Machinery Directive) and 2011/65/EU (RoHS Directive)

Manufacturer:Wassermann Dental-Maschinen GmbH
Rudorffweg 15-17
21031 Hamburg
Germany
Product description:Vacuum Mixers for dental applications
Model:Wamix-EconomyItem no.171972
Applicable standards:DIN EN 61010-1
DIN EN 61000-6-3
DIN EN 61000-6-1
DIN 45635-1
DIN EN 60335-1

This is to confirm that the above mentioned machine complies with the described EU guidelines. This declaration becomes invalid if the machine is modified without our approval.

Place, date: Hamburg, 2023-07-20

Signature: Sven Wassermann (Managing Director)

Company stamp: Wassermann Dental-Maschinen GmbH, Rudorffweg 15-17, 21031 Hamburg

Notes

No specific notes were provided in the document.

PDF preview unavailable. Download the PDF instead.

EN manual Wamix-ECO Writer LibreOffice 7.4

Related Documents

Preview Wassermann Wamix-Comfort IV Vacuum Mixer User Manual
User manual for the Wassermann Wamix-Comfort IV vacuum mixer, detailing features, installation, operation, maintenance, troubleshooting, and technical specifications for dental applications.
Preview Wassermann DTC Economy Dental Cleaner User Manual
User manual for the Wassermann DTC Economy dental cleaning unit. Provides detailed instructions on features, safety guidelines, operation, installation, maintenance, troubleshooting, and technical specifications for efficient dental tray and retention plate cleaning.
Preview Wassermann Sidomix III: Operating Manual for Dental Mixing and Dosing Device
Comprehensive operating manual for the Wassermann Sidomix III, a 1:1 mixing and dosing device for dental applications. Covers features, safety instructions, installation, operation, troubleshooting, maintenance, technical specifications, and disposal.
Preview Wassermann Wasi-Steam Classic Steam Cleaner User Manual
Comprehensive user manual for the Wassermann Wasi-Steam Classic steam cleaner, detailing features, safety instructions, operation procedures, maintenance, troubleshooting, technical specifications, and EU compliance for dental applications.