Midmark PowerVac® Care Guide

To ensure years of reliable service, follow the care instructions outlined in this guide. If service is required, contact your authorized Midmark dealer.

Limited Warranty

Scope of Warranty: Midmark Corporation ("Midmark") warrants to the original purchaser its new PowerVac® products and components (except for components not warranted under "Exclusions") manufactured by Midmark to be free from defects in material and workmanship under normal use and service. Midmark's obligation under this warranty is limited to the repair or replacement, at Midmark's option, of the parts or the products the defects of which are reported to Midmark within the applicable warranty period and which, upon examination by Midmark, prove to be defective.

Applicable Warranty Period: The applicable warranty period, measured from the date of installation for the original user, shall be five (5) years or 10,000 usage hours (whichever comes first) for all warranted products and components and ten (10) years or 20,000 usage hours (whichever comes first) for the pump only.

Exclusions: This warranty does not cover and Midmark shall not be liable for the following: (1) repairs and replacements because of misuse, abuse, negligence, alteration, accident, freight damage, or tampering; (2) products which are not installed, used, and properly cleaned as required in the Midmark "Installation" manual and or PowerVac® Care Guide for this applicable product; (3) products considered to be of a consumable nature; (4) accessories or parts not manufactured by Midmark; (5) charges by anyone for adjustments, repairs, replacement parts, installation, or other work performed upon or in connection with such products which is not expressly authorized in writing in advance by Midmark.

Exclusive Remedy: Midmark's only obligation under this warranty is the repair or replacement of defective parts. Midmark shall not be liable for any direct, special, indirect, incidental, exemplary, or consequential damages or delay, including, but not limited to, damages for loss of profits or loss of use.

No Authorization: No person or firm is authorized to create for Midmark any other obligation or liability in connection with the products.

This warranty is Midmark's only warranty and is in lieu of all other warranties, express or implied. Midmark makes no implied warranties of any kind including any warranties of merchantability or fitness for any particular purpose. This warranty is limited to the repair or replacement of defective parts.

Important Information

Safety Symbols

These symbols may appear on your equipment and/or in the manuals. Warnings and cautions are provided in the manuals where applicable.

Safety Symbol Key

Intended Use

To provide vacuum suction during general examinations and procedures conducted by qualified dental professionals.

Electromagnetic Interference

This Midmark PowerVac® and Evacuation control box are designed and built to minimize electromagnetic interference with other devices. However, if interference is noticed between another device and this unit:

Disposal of Equipment

At the end of product life, the unit(s), accessory, and other consumable goods may become contaminated from normal use. Consult local codes and ordinances for proper disposal of equipment, accessories and other consumable goods.

Transportation & Storage Conditions

Note: The PowerVac® is electrically designed to be only sold in U.S.A. or Canada.

This product has been evaluated with respect to electrical shock, fire and mechanical hazards only, in accordance with UL60601-1 and CAN/CSA C22.2 NO. 601.1.

Classifications: Class 1, Type B Applied Part

MEDICAL EQUIPMENT CLASSIFIED CULUS 94XM UL60601-1 CAN/CSA C22.2 No.601.1 AA2508i

Midmark Corporation
For contact information, go to: www.midmark.com

Daily Maintenance

A. Clean Vacuum Piping System.

Clean vacuum piping system with a non-foaming vacuum system cleaner. Midmark recommends Precision Clense, Order No. 77000570.

Note: By using a non-foaming vacuum system cleaner on a daily basis, the PowerVac® separation tank will stay rinsed and clean. This practice helps protect the internal tank controls and eliminates the need to rinse the separator tank. (See PowerVac® Service and Parts manual located on www.midmark.com regarding rinsing the separation tank.)

Note: For the first two weeks of using Precision Clense, flush lines at the beginning of the day with water and check traps twice a day. After the first two weeks, flush with water and check traps as needed.

  1. Start with the furthest operatory from the vacuum.
  2. Suction cleaning solution. (Follow mixture instructions provided with the vacuum system cleaner.)

Note: Alternate air and vacuum while cleansing; submerge and remove HVE and SE from mixture several times until mixture is consumed.

  1. Keep the cleansed vacuum lines (HVE and SE) open and move to the next operatory.
  2. Repeat steps 1-3 until all vacuum lines are cleansed.

Note: PowerVac® MUST run for 5 minutes after all operatories are cleansed.

Vacuum Level Adjustment Procedure

Adjust the center mounted screw inside the valve.

Use 1/4 inch nutdriver and phillips screwdriver. Each full clockwise turn increases the vacuum level by approximately 2" Hg.

Caution: Never operate the equipment without complete and proper grounding. The On/Off switch controls only the secondary circuit power. The main power source must be turned off to remove all power in the control box.

WARNING ⚠️ Motors are thermally protected with automatic reset. Unit may start without warning.

Quarterly Inspection

A. Check Gearbox Lube Level.

The gear lube level is set at the factory to 1/2 full. A site glass is provided to visually inspect the gear lube level. If the gear lube level is 1/3 full or less, fill as needed and monitor oil level. Replace absorbent pad if required.

B. Inspect Vacuum Relief Valve Filter

If the mechanical room contains excessive contaminated air, the VRV filter may become dirty. If it appears clogged, contact Midmark Technical Service at 1.800.MIDMARK or refer to the PowerVac® Service and Parts manual located on Technical Library at www.midmark.com, Midmark P/N 004-0494-00.

Annual / 2,000 HR Inspection

A. Inspect Belt

If the belt appears loose, makes noise, abnormally worn, contact Midmark Technical Service at 1.800.MIDMARK or refer to the PowerVac® Service and Parts Manual located on Technical Library at www.midmark.com.

B. Use Conditioning Fluid

To maintain proper vacuum performance pour conditioning fluid (part# 002-1910-02) into intake hose. (See maintenance kit instructions (003-10248-00) located on www.midmark.com for proper procedures.)

C. Inspect Tank

If required, use garden hose adapter and flush tank with garden hose. Refer to procedure in the PowerVac® Service and Parts Manual located on Technical Library at www.midmark.com.

Note: Every 10 years and or 20,000 hours of usage, Change Gear Lube. See PowerVac® Service and Parts Manual for procedure.

Dealer Information

Dealer (name / phone):

Model Number:

Serial Number:

Date of Purchase:

Log

Date: Services Performed: Technician:
Models: PowerVac Dry Vacuum, Dry Vacuum, PowerVac Vacuum, Vacuum, PowerVac

File Info : application/pdf, 2 Pages, 414.50KB

PDF preview unavailable. Download the PDF instead.

003-1587-00 ?sfvrsn=efb8d305 4

References

Adobe PDF Library 15.0 Adobe InDesign 16.3 (Windows)

Related Documents

Preview Midmark Manifold Installation Kit Installation Guide
Detailed installation guide for the Midmark Manifold Installation Kit, covering models 002-11029-00, 002-11058-00, 002-11059-00, and 002-11060-00. Includes step-by-step instructions and tubing connection diagrams for dental delivery systems.
Preview Midmark M9/M9D & M11 PC Board Kit Installation Guide
This guide provides step-by-step instructions for installing and configuring the PC Board Kit for Midmark M9, M9D, and M11 sterilizers. It covers board removal, installation, unit configuration, and serial number setup.
Preview Midmark Dental Halogen Light Wiring Diagrams
Detailed wiring diagrams and technical specifications for Midmark Dental Halogen Lights, covering various mounting types and models. Includes part numbers, voltage information, and color codes for installation and maintenance.
Preview Midmark M11 UltraClave Steam Sterilizer: Installation and Operation Manual
Comprehensive guide for installing, operating, and maintaining the Midmark M11 UltraClave Steam Sterilizer. Covers features, safety, troubleshooting, and specifications.
Preview Midmark Din Rail Contactor Replacement Kit Installation Guide
Instructions for replacing a contactor on a Din Rail using the Midmark Din Rail Contactor Replacement Kit (002-10390-00). Includes safety warnings and step-by-step procedures.
Preview Midmark Digital Vital Signs Device: Quick Reference Guide and Procedural Hints
A concise guide to operating the Midmark Digital Vital Signs Device, providing step-by-step instructions and procedural hints for blood pressure, SpO2, and temperature measurements.
Preview Midmark Preva Dental X-Ray System: Installation and Service Manual
Comprehensive guide for installing and servicing the Midmark Preva Dental X-Ray System, covering safety, operation, components, and troubleshooting for dental professionals.
Preview Midmark Progeny Vantage Panoramic X-ray System Installation Guide
This installation guide provides comprehensive instructions for setting up the Midmark Progeny Vantage Panoramic X-ray System. It covers pre-installation planning, physical installation on various mounts, cable management, software client setup, and the installation of the optional Cephalometric Extension.