Midmark Manifold Installation Kit
Model Numbers: 002-11029-00 (Syringe), 002-11058-00 (Low Speed Handpiece), 002-11059-00 (High Speed Handpiece), 002-11060-00 (Scaler)
Applies to Models: All
Special Tools: None
Note: Some components may vary depending on the manifold kit number.
Installation Overview
This document provides instructions for installing the Midmark Manifold Installation Kit. The kit includes components for syringe, low speed handpiece, high speed handpiece, and scaler connections.
Referenced Components: Manifolds (Four Handpieces, Fifth Handpiece Manifold), Mounting Plate Screws, Manifold Screws, Sleeve Clamp, Uni-Clamps.
Installation Steps
Step 1: Power Down
Turn the power switch to the 'Off' position. Unplug the dental delivery system from the power cord.
[Diagram Description: Illustration of a power switch in the 'Off' position and a power cord being unplugged from the unit.]
Step 2: Release Pressure
Release pressure from the pressure tank. Hold a cloth on the tube and open the valve.
[Diagram Description: Diagram demonstrating pressure release from a tank: hold a cloth on the tube and open the valve.]
Step 3: Open Unit Lid
Remove the mat from the lid. Open the top lid by pulling the lever back and lifting the lid completely up.
[Diagram Description: Illustration showing how to open the unit's top lid by pulling a lever back and lifting.]
Step 4: Disconnect Tubing (High Speed Manifold Kit Example)
Remove tubing from the desired manifold block and label each tube as needed for identification.
[Diagram Description: Close-up of colored tubing connected to a manifold block. Arrows indicate removal of tubing, with a note to label each tube.]
Step 5: Remove Old Mounting Plate (High Speed Manifold Kit Example)
Remove the two screws securing the manifold mounting plate.
[Diagram Description: Close-up of a manifold mounting plate with arrows pointing to two screws to be removed.]
Step 6: Replace Manifold (High Speed Manifold Kit Example)
Remove the old manifold by unscrewing the two screws. Replace it with the new manifold and secure it using the provided two screws.
[Diagram Description: Side-by-side comparison of an 'Old Manifold' and a 'New Manifold'. Instructions show removing the old manifold and installing the new one with screws.]
Step 7: Install Mounting Plate
Install the manifold mounting plate using the two screws.
[Diagram Description: Illustration of the manifold mounting plate being installed with screws.]
Step 8a: Four Handpiece Manifold Tubing Connections
Connect tubing with sleeve clamps and uni-clamps to the manifold components in the delivery system as shown in the diagram.
[Diagram Description: Schematic for 4-handpiece manifold tubing connections. Shows Air Pressure Regulator, Water Bottles, Syringe, Low Speed Handpiece, and Scaler. Tubing (YL, BL, CL, RD) connects to the manifold using Uni-clamps and Sleeve Clamps.]
Step 8b: Five Handpiece Manifold Tubing Connections
Connect tubing with sleeve clamps and uni-clamps to the manifold components in the delivery system as shown in the diagram.
[Diagram Description: Schematic for 5-handpiece manifold tubing connections for the High Speed for Fifth Handpiece Model. Shows Handpiece Air Control and Curing Light. Tubing (YL, BL, WH, CL) connects to the manifold using Uni-clamps and Sleeve Clamps.]
Step 9: Close Unit Lid
Close the lid and push down on the middle section until a 'Click' sound is heard.
[Diagram Description: Illustration of the unit lid being closed, with an instruction to push down on the middle until a 'Click' is heard.]
Step 10: Install Lid Mat
Install the mat onto the lid.
[Diagram Description: Illustration of a mat being placed onto the unit lid.]
Step 11: Power On