Extech UV505 UV-AB Light Meter

User Manual

Contact: [phone-number] | [website-address] | [email-address]

Introduction

Thank you for selecting the Extech Model UV505 UV-AB Light Meter. This device measures UV-AB ultraviolet light in the wavelength range of 290~390nm. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Please visit our website for more information.

Features

General Safety

Please read the entire User Manual before operating this device. Use the meter only as specified and do not attempt to service or open the meter housing. Do not allow children to handle the meter or the protective cover. Please dispose of batteries and meter responsibly and in accordance with all applicable laws and regulations.

UV Safety Information

Use extreme caution when using this meter to test an ultraviolet light source. Exposure to UV radiation can be hazardous depending on the exposure time, intensity of radiation, wavelength and an individual's sensitivity to ultraviolet.

Meter Description

The Extech UV505 meter features the following components:

Note: The battery compartment and a tripod mount are located on the back of the meter.

LCD Description

The LCD display shows the following information:

Button Descriptions

Operation

Meter Power

Short press the power button to switch the meter ON/OFF. If the LCD does not switch on, check the batteries located in the rear battery compartment. The meter powers off automatically after 15 minutes of inactivity.

Zero the Sensor

With the sensor's protective cover attached, short press the Zero button to zero the display before taking measurements. The ZERO display alert will appear while the meter is zeroing. Press the Zero button again to exit the mode; the ZERO display alert will switch off.

Taking Measurements

Remove the sensor cover and aim the light sensor at the UV light source. View the measurement on the meter's display.

Selecting Range and Unit of Measure

Short press the Unit of measure toggle button to toggle the range (mW/cm² and μW/cm²); the lower display area will show the selected range.

Data Hold

Short press the Data Hold button to freeze or unfreeze a reading on the display.

Backlight

The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas. Short press the Backlight button to turn on the backlight. The backlight will automatically turn off after 10 seconds.

Measurement Considerations

Maintenance

Battery Replacement

  1. Power OFF the meter.
  2. Press and hold the locking button above the battery compartment (back of the meter) and slide the battery compartment door off in a downward motion.
  3. Replace the two `AAA` 1.5V batteries, observing correct polarity. Re-assemble the meter before use.

Battery Safety: Please dispose of batteries responsibly; never dispose of batteries in a fire, as batteries may explode or leak. If the meter is not to be used for 60 days or more, remove the batteries and store them separately.

As consumers, users are legally required to take used batteries to appropriate collection sites, the retail store where the batteries were purchased, or wherever batteries are sold. Never dispose of used batteries or rechargeable batteries in household waste.

Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end-of-life devices to a designated collection point for the disposal of electrical and electronic equipment.

Cleaning and Storage

Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. Use a dry lint-free dust cloth to clear debris and dust from the sensor. Avoid any materials that could potentially scratch the sensor surface.

Specifications

Display

Backlit 3-3/4 digit (4000 count) LCD. Size: 35 x 30mm (1.4 x 1.2"). Sampling rate: Approx. 3 seconds.

Measurements

UV sensor spectrum: Ultraviolet Light (UV-AB). Peak Sensitivity Wavelength: 365 nm. Input Overload: "OL".

General

Tripod mount: On rear panel. Power: 2 x 1.5V `AAA` batteries. Battery life: Approx. 50 hours. Automatic power off: After approx. 15 minutes of inactivity. Operating Temperature/Humidity: 5 to 40°C (41 to 104°F) / 80% RH max. Operating Altitude: 2000 meters (7000ft) maximum. Storage Temperature/Humidity: -10 to 60°C (14 to 140°F) / 70% RH max. Dimensions/Weight: 133 x 48 x 23mm (5.2 x 1.9 x 0.9") / 90g (3.2oz). Safety: For indoor use only. EMC: EN61326 (1997), A1 (1998), A2 (2001); Pollution degree 2.

Electrical Specifications (23±5°C)

UnitRangeResolutionAccuracy
mW/cm²0 to 39.990.01± 4% full scale
μW/cm²0 to 39991± 4% full scale

PDF preview unavailable. Download the PDF instead.

uv505-manual

Related Documents

Preview Extech UV505 UV-AB Light Meter User Manual
User manual for the Extech UV505 UV-AB Light Meter. Provides instructions on features, operation, safety, maintenance, and technical specifications for measuring UV-AB light.
Preview Extech UV505 UV-AB Light Meter User Manual
This user manual provides comprehensive information for the Extech UV505 UV-AB Light Meter. Learn about its features, specifications, safe operation, maintenance, and customer support for accurate UV-AB light measurement.
Preview Extech UV505 Pocket UV-AB Light Meter Datasheet
Datasheet for the Extech UV505 Pocket UV-AB Light Meter, detailing its features, specifications, and applications for measuring UV-AB light in various environments.
Preview Extech UV505 UV-AB Light Meter User Manual
User manual for the Extech UV505 UV-AB Light Meter. This document provides detailed information on the device's features, operation, specifications, safety precautions, and maintenance for measuring UVA and UVB light.
Preview Manuale Utente Misuratore di Radiazione UV-AB Extech UV505
Manuale utente completo per il Misuratore di Radiazione UV-AB Extech UV505. Include istruzioni operative, specifiche tecniche, informazioni di sicurezza e manutenzione.
Preview Extech UV505 UV-AB Light Meter User Manual - Specifications and Operation
User manual for the Extech UV505 UV-AB Light Meter. Features include measurement of UV-AB light, wavelength range, accuracy, safety guidelines, operation instructions, maintenance, and detailed specifications.
Preview Extech UV505 UVA-UVB Light Meter User Manual
User manual for the Extech UV505 UVA-UVB Light Meter, detailing its features, operation, safety precautions, specifications, maintenance, and warranty information.
Preview EXTECH UV505 Pocket UV-AB Light Meter Datasheet
Product datasheet for the EXTECH UV505 Pocket UV-AB Light Meter, detailing its features, specifications, and ordering information for UV-AB light measurement applications.