Extech UV-AB Light Meter

Model UV505

USER MANUAL

Contact: 1.800.544.2843 | www.calcert.com | sales@calcert.com

Introduction

Thank you for selecting the Extech Model UV505 UV-AB Light Meter. This device measures UV-AB ultraviolet light in the wavelength range of 290~390nm. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Please visit our website.

Features

General Safety

Please read the entire User Manual before operating this device. Use the meter only as specified and do not attempt to service or open the meter housing. Do not allow children to handle the meter or the protective cover. Please dispose of batteries and meter responsibly and in accordance with all applicable laws and regulations.

UV Safety Information

Use extreme caution when using this meter to test an ultraviolet light source. Exposure to UV radiation can be hazardous depending on the exposure time, intensity of radiation, wavelength and an individual's sensitivity to ultraviolet.

Meter Description

The Extech UV-AB Light Meter, Model UV505, is a compact, handheld device. It features a circular UV sensor at the top, a large LCD screen displaying a measurement of '3999 uW/cm²', and several control buttons below: HOLD, ZERO, UNIT, and a power symbol. The device is branded EXTECH INSTRUMENTS.

Device Components:

  1. Light Sensor (protective cover not shown)
  2. LCD Display
  3. Backlight Button
  4. Zero Button
  5. Power ON-OFF Button
  6. Unit of Measure Toggle Button
  7. Display Data Hold Button

Note: Battery compartment and tripod mount are located on the back of the meter.

LCD and Button Details

LCD Description

  1. Measurement Reading
  2. Data Hold Indicator
  3. Zero Mode Indicator
  4. Low Battery Indicator
  5. Unit of Measure (μW/cm² or mW/cm²)

Button Descriptions

Operation

Meter Power

Short press the [power-button] to switch the meter ON/OFF. If the LCD does not switch on, check the batteries located in the rear battery compartment. The meter powers off automatically after 15 minutes of inactivity.

Zero the Sensor

With the sensor's protective cover attached, short press the ZERO button to zero the display before taking measurements. The ZERO display alert will appear while the meter is zeroing. Press the ZERO button again to exit the mode; the ZERO display alert will switch off.

Taking Measurements

Remove the sensor cover and aim the light sensor at the UV light source. View the measurement on the meter's display.

Selecting Range and Unit of Measure

Short press the UNIT button to toggle the range (mW/cm² and μW/cm²); the lower display area will show the selected range.

Data Hold

Short press the H button to freeze or unfreeze a reading on the display.

Backlight

The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas. Short press the [backlight-button] to turn on the backlight. The backlight will automatically turn off after 10 seconds.

Measurement Considerations

Maintenance

Battery Replacement

  1. Power OFF the meter.
  2. Press and hold the locking button above the battery compartment (back of the meter) and slide the battery compartment door off in a downward motion.
  3. Replace the two 'AAA' 1.5V batteries observing correct polarity. Re-assemble the meter before use.

Battery Safety

Please dispose of batteries responsibly; never dispose of batteries in a fire, batteries may explode or leak. If the meter is not to be used for 60 days or more, remove the batteries and store separately.

Never dispose of used batteries or rechargeable batteries in household waste. As consumers, users are legally required to take used batteries to appropriate collection sites, the retail store where the batteries were purchased, or wherever batteries are sold.

Disposal

Do not dispose of this instrument in household waste. The user is obligated to take end-of-life devices to a designated collection point for the disposal of electrical and electronic equipment.

Cleaning and Storage

Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. Use a dry lint free dust cloth to clear debris and dust from the sensor. Avoid any materials that could potentially scratch the sensor surface.

Specifications

FeatureDetails
DisplayBacklit 3-3/4 digit (4000 count) LCD
Size: 35 x 30mm (1.4 x 1.2")
Sampling rateApprox. 3 seconds
MeasurementsUltraviolet Light (UV-AB)
UV sensor spectrum290 to 390 nm
Peak Sensitivity Wavelength365 nm
Input Overload Display"OL"
Tripod mountOn rear panel
Power2 x 1.5V 'AAA' batteries
Battery lifeApprox. 50 hours
Automatic power offAfter approx. 15 minutes of inactivity
Operating Temperature/Humidity5 to 40°C (41 to 104°F) / 80% RH max
Operating Altitude2000 meters (7000ft) maximum
Storage Temperature/Humidity-10 to 60°C (14 to 140°F) / 70% RH max
Dimensions/Weight133 x 48 x 23mm (5.2 x 1.9 x 0.9") / 90g (3.2oz)
SafetyFor indoor use only. EMC: EN61326 (1997), A1 (1998), A2 (2001); Pollution degree 2

Electrical Specifications (23±5°C)

UnitRangeResolutionAccuracy
mW/cm²0 to 39.990.01± 4% full scale
μW/cm²0 to 39991± 4% full scale

PDF preview unavailable. Download the PDF instead.

uv505-manual Adobe PDF library 9.00

Related Documents

Preview Extech UV505 Pocket UV-AB Light Meter Datasheet
Datasheet for the Extech UV505 Pocket UV-AB Light Meter, detailing its features, specifications, and applications for measuring UV-AB light in various environments.
Preview Extech UV505 UV-AB Light Meter User Manual
User manual for the Extech UV505 UV-AB Light Meter. This document provides detailed information on the device's features, operation, specifications, safety precautions, and maintenance for measuring UVA and UVB light.
Preview Extech UV505 UV-AB Light Meter User Manual
User manual for the Extech UV505 UV-AB Light Meter. Provides instructions on features, operation, safety, maintenance, and technical specifications for measuring UV-AB light.
Preview EXTECH UV505 Pocket UV-AB Light Meter Datasheet
Product datasheet for the EXTECH UV505 Pocket UV-AB Light Meter, detailing its features, specifications, and ordering information for UV-AB light measurement applications.
Preview Manuale Utente Misuratore di Radiazione UV-AB Extech UV505
Manuale utente completo per il Misuratore di Radiazione UV-AB Extech UV505. Include istruzioni operative, specifiche tecniche, informazioni di sicurezza e manutenzione.
Preview Extech UV505 UV-AB Light Meter User Manual
User manual for the Extech UV505 UV-AB Light Meter, detailing features, operation, safety, specifications, and maintenance for measuring ultraviolet light.
Preview EXTECH UV505 Pocket UV-AB Light Meter Datasheet | Extech Instruments
Detailed datasheet for the EXTECH UV505 Pocket UV-AB Light Meter, featuring specifications, features, and applications for measuring UV-AB light. Includes product overview, technical details, and ordering information.
Preview Extech UV505 Pocket UV-AB Light Meter Product Datasheet
Product datasheet for the Extech UV505 Pocket UV-AB Light Meter. Details features, specifications, and ordering information for this portable UV irradiance measurement device.