Extech UV-AB Light Meter
Model UV505
USER MANUAL
Contact: 1.800.544.2843 | www.calcert.com | sales@calcert.com
Introduction
Thank you for selecting the Extech Model UV505 UV-AB Light Meter. This device measures UV-AB ultraviolet light in the wavelength range of 290~390nm. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Please visit our website.
Features
- Professional, high quality UV meter
- Measures ultra-violet irradiation
- Wavelength range 290 nm to 390 nm
- Two measurement ranges for greater accuracy (0~3999 μW/cm² / 0~39.99 mW/cm²)
- Auto power off after 15 minutes
- Lightweight, ergonomic design
General Safety
Please read the entire User Manual before operating this device. Use the meter only as specified and do not attempt to service or open the meter housing. Do not allow children to handle the meter or the protective cover. Please dispose of batteries and meter responsibly and in accordance with all applicable laws and regulations.
- To prevent damage to this precision device, do not attempt measurements outside of the ranges specified in this manual.
- Do not use this meter in direct sunlight or any place where it is excessively hot and/or humid.
- Do not submerge the meter in liquid.
UV Safety Information
Use extreme caution when using this meter to test an ultraviolet light source. Exposure to UV radiation can be hazardous depending on the exposure time, intensity of radiation, wavelength and an individual's sensitivity to ultraviolet.
Meter Description
The Extech UV-AB Light Meter, Model UV505, is a compact, handheld device. It features a circular UV sensor at the top, a large LCD screen displaying a measurement of '3999 uW/cm²', and several control buttons below: HOLD, ZERO, UNIT, and a power symbol. The device is branded EXTECH INSTRUMENTS.
Device Components:
- Light Sensor (protective cover not shown)
- LCD Display
- Backlight Button
- Zero Button
- Power ON-OFF Button
- Unit of Measure Toggle Button
- Display Data Hold Button
Note: Battery compartment and tripod mount are located on the back of the meter.
Operation
Meter Power
Short press the [power-button] to switch the meter ON/OFF. If the LCD does not switch on, check the batteries located in the rear battery compartment. The meter powers off automatically after 15 minutes of inactivity.
Zero the Sensor
With the sensor's protective cover attached, short press the ZERO button to zero the display before taking measurements. The ZERO display alert will appear while the meter is zeroing. Press the ZERO button again to exit the mode; the ZERO display alert will switch off.
Taking Measurements
Remove the sensor cover and aim the light sensor at the UV light source. View the measurement on the meter's display.
Selecting Range and Unit of Measure
Short press the UNIT button to toggle the range (mW/cm² and μW/cm²); the lower display area will show the selected range.
Data Hold
Short press the H button to freeze or unfreeze a reading on the display.
Backlight
The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas. Short press the [backlight-button] to turn on the backlight. The backlight will automatically turn off after 10 seconds.
Measurement Considerations
- Keep the light sensor clean and free of scratches. Dust lightly with dry lint free cloth.
- Avoid reflections or shadows on the sensor while taking measurements.
- For best accuracy, repeat the measurement several times to ensure that the light source has remained stable.
Maintenance
Battery Replacement
- Power OFF the meter.
- Press and hold the locking button above the battery compartment (back of the meter) and slide the battery compartment door off in a downward motion.
- Replace the two 'AAA' 1.5V batteries observing correct polarity. Re-assemble the meter before use.
Battery Safety
Please dispose of batteries responsibly; never dispose of batteries in a fire, batteries may explode or leak. If the meter is not to be used for 60 days or more, remove the batteries and store separately.
Never dispose of used batteries or rechargeable batteries in household waste. As consumers, users are legally required to take used batteries to appropriate collection sites, the retail store where the batteries were purchased, or wherever batteries are sold.
Disposal
Do not dispose of this instrument in household waste. The user is obligated to take end-of-life devices to a designated collection point for the disposal of electrical and electronic equipment.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. Use a dry lint free dust cloth to clear debris and dust from the sensor. Avoid any materials that could potentially scratch the sensor surface.
Specifications
Feature | Details |
---|---|
Display | Backlit 3-3/4 digit (4000 count) LCD Size: 35 x 30mm (1.4 x 1.2") |
Sampling rate | Approx. 3 seconds |
Measurements | Ultraviolet Light (UV-AB) |
UV sensor spectrum | 290 to 390 nm |
Peak Sensitivity Wavelength | 365 nm |
Input Overload Display | "OL" |
Tripod mount | On rear panel |
Power | 2 x 1.5V 'AAA' batteries |
Battery life | Approx. 50 hours |
Automatic power off | After approx. 15 minutes of inactivity |
Operating Temperature/Humidity | 5 to 40°C (41 to 104°F) / 80% RH max |
Operating Altitude | 2000 meters (7000ft) maximum |
Storage Temperature/Humidity | -10 to 60°C (14 to 140°F) / 70% RH max |
Dimensions/Weight | 133 x 48 x 23mm (5.2 x 1.9 x 0.9") / 90g (3.2oz) |
Safety | For indoor use only. EMC: EN61326 (1997), A1 (1998), A2 (2001); Pollution degree 2 |
Electrical Specifications (23±5°C)
Unit | Range | Resolution | Accuracy |
---|---|---|---|
mW/cm² | 0 to 39.99 | 0.01 | ± 4% full scale |
μW/cm² | 0 to 3999 | 1 | ± 4% full scale |
Copyright © 2017 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form. ISO-9001 Certified
Related Documents
![]() |
Extech UV505 Pocket UV-AB Light Meter Datasheet Datasheet for the Extech UV505 Pocket UV-AB Light Meter, detailing its features, specifications, and applications for measuring UV-AB light in various environments. |
![]() |
Extech UV505 UV-AB Light Meter User Manual User manual for the Extech UV505 UV-AB Light Meter. This document provides detailed information on the device's features, operation, specifications, safety precautions, and maintenance for measuring UVA and UVB light. |
![]() |
Extech UV505 UV-AB Light Meter User Manual User manual for the Extech UV505 UV-AB Light Meter. Provides instructions on features, operation, safety, maintenance, and technical specifications for measuring UV-AB light. |
![]() |
EXTECH UV505 Pocket UV-AB Light Meter Datasheet Product datasheet for the EXTECH UV505 Pocket UV-AB Light Meter, detailing its features, specifications, and ordering information for UV-AB light measurement applications. |
![]() |
Manuale Utente Misuratore di Radiazione UV-AB Extech UV505 Manuale utente completo per il Misuratore di Radiazione UV-AB Extech UV505. Include istruzioni operative, specifiche tecniche, informazioni di sicurezza e manutenzione. |
![]() |
Extech UV505 UV-AB Light Meter User Manual User manual for the Extech UV505 UV-AB Light Meter, detailing features, operation, safety, specifications, and maintenance for measuring ultraviolet light. |
![]() |
EXTECH UV505 Pocket UV-AB Light Meter Datasheet | Extech Instruments Detailed datasheet for the EXTECH UV505 Pocket UV-AB Light Meter, featuring specifications, features, and applications for measuring UV-AB light. Includes product overview, technical details, and ordering information. |
![]() |
Extech UV505 Pocket UV-AB Light Meter Product Datasheet Product datasheet for the Extech UV505 Pocket UV-AB Light Meter. Details features, specifications, and ordering information for this portable UV irradiance measurement device. |