Elitech RC-5/5+ Series: Multilingual Specification

Overview

The RC-5 series are used to record the temperature/humidity of foods, medicines and other goods during storage, transportation and in each stage of the cold chain including cooler bags, cooling cabinets, medicine cabinets, refrigerators, laboratories, reefer containers and trucks.

RC-5 is a classic USB temperature data logger used in a wide range of applications around the world. RC-5+ is an upgraded version that adds the functions, including automatic PDF reports generation, repeat start without configuration, etc.

Specifications

ParameterRC-5RC-5+/TE
ModelRC-5RC-5+/TE
Temperature Measurement Range-30°C~+70°C (-22°F~158°F)*
Temperature Accuracy±0.5°C/±0.9°F (-20°C~+40°C); ±1°C/±1.8°F (others)
Resolution0.1°C/°F
MemoryMaximum 32,000 points
Logging Interval10 seconds to 24 hours10 seconds to 12 hours
Data InterfaceUSB
Start ModePress button; Use softwarePress button; Auto start; Use software
Stop ModePress button; Auto stop; Use software
SoftwareElitechLog, for macOS & Windows system
Report FormatPDF/EXCEL/TXT** by ElitechLog softwareAuto PDF report; PDF/EXCEL/TXT** by ElitechLog software
Shelf Life1 year
CertificationEN12830, CE, RoHS
Protection LevelIP67
Dimensions80 x 33.5 x 14 mm
Weight20g

* At ultralow temperature, the LCD is slow but does not affect normal logging. It will be back to normal after the temperature rises. ** TXT for Windows ONLY

Operation Instructions

  1. Battery Activation
    1. Turn the battery cover counterclockwise to open it.
    2. Gently press the battery to hold it in position, then pull out the battery insulator strip.
    3. Turn the battery cover clockwise and tighten it.
  2. Install Software

    Please download and install the free ElitechLog software (macOS and Windows) from Elitech US: www.elitechus.com/pages/download or Elitech UK: www.elitecheu.com/pages/software or Elitech BR: www.elitechbrasil.com.br

  3. Configure Parameters

    First, connect the data logger to the computer via USB cable, wait until [USB Port] icon shows on the LCD; then configure via: ElitechLog Software: - If you do not need to change the default parameters (in Appendix); please click Quick Reset under Summary menu to synchronize local time before usage; - If you need to change the parameters, please click the Parameter menu, enter your preferred values, and click the Save Parameter button to complete the configuration.

    Warning! For first time user or after battery replacement: To avoid time or time zone errors, please make sure you click Quick Reset or Save Parameter before usage to sync and configure your local time into the logger.

  4. Start Logging

    Press Button: Press and hold the [Left Button] for 5 seconds until the [Logging] icon shows on the LCD, indicating the logger starts logging.

    Auto Start (RC-5+/TE only):

    • Immediate Start: The logger starts logging after removed from the computer.
    • Timed Start: The logger starts counting after being removed from the computer; it will start logging automatically after the set date/time.

    Note: If the [Logging] icon keeps flashing, it means the logger configured with start delay; it will start logging after the set delay time elapses.

  5. Mark Events (RC-5+/TE only)

    Double click the [Right Button] to mark current temperature and time, up to 10 groups of data. After marked, it will be indicated by Log X on the LCD screen (X means the marked group).

  6. Stop Logging

    Press Button*: Press and hold the [Right Button] for 5 seconds until the [Stopped] icon shows on the LCD, indicating the logger stops logging.

    Auto Stop: When the logging points reach the maximum memory points, the logger will stop automatically.

    Use Software: Open ElitechLog software, click Summary menu, and Stop Logging button.

    Note: *Default stop is via Press Button, if set as disabled, the button stop function will be invalid; please open the ElitechLog software and click Stop Logging button to stop it.

  7. Download Data

    Connect the data logger to your computer via USB cable, wait until the [USB Port] icon shows on the LCD; then download via:

    • ElitechLog Software: The logger will auto-upload data to ElitechLog, then please click Export to select your desired file format to export. If data failed for auto-upload, please manually click Download and then follow the export operation.
    • Without ElitechLog Software (RC-5+/TE only): Simply find and open the removable storage device ElitechLog, save the auto generated PDF report to your computer for viewing.
  8. Reuse the Logger

    To reuse a logger, please stop it first; then connect it to your computer and use ElitechLog software to save or export the data. Next, reconfigure the logger by repeating the operations in 3. Configure Parameters*. After finished, follow 4. Start Logging to restart the logger for new logging.

    Warning! * To make space for new loggings, all previous logging data inside the logger will be deleted after re-configuration. If you forgot to save/export data, please try to locate the logger in the History menu of ElitechLog software.

  9. Repeat Start (RC-5+/TE only)

    To restart the stopped logger, you may press and hold the [Left Button] to start logging quickly without reconfiguration. Please backup data before restart by repeating 7. Download Data - Download via ElitechLog Software.

Status Indication

1. Buttons

OperationsFunction
Press and hold the left button for 5 secondsStart logging
Press and hold the right button for 5 secondsStop logging
Press and release the left buttonCheck/Switch interfaces
Press and release the right buttonBack to the main menu
Double click the right buttonMark events (RC-5+/TE only)

2. LCD Screen

LCD Icons:

  • 1: [Battery Level Indicator]
  • 2: Stopped
  • 3: Logging
  • Not started
  • 4: Connected to PC
  • 5: High Temperature Alarm
  • 6: Low Temperature Alarm
  • 7: Logging Points (current)
  • 8: No Alarm/Mark Success
  • 9: Alarm/Mark Failure
  • 10: Month
  • 11: Day
  • 12: Maximum Value
  • 13: Minimum Value

3. LCD Interface

ParameterDisplay
TemperatureCurrent Value
Logging PointsCurrent Count
Current TimeHH:MM
Current DateM-D
Maximum TemperatureMAX Value
Minimum TemperatureMIN Value

Battery Replacement

  1. Turn the battery cover counterclockwise to open it.
  2. Install a new and wide-temperature CR2032 button battery into the battery compartment, with its + side facing upward.
  3. Turn the battery cover clockwise and tighten it.

What's Included

  • Data Logger x 1
  • User Manual x 1
  • Certificate of Verification x 1
  • CR2032 button battery x 1

Precautions

  • Please store your logger at room temperature.
  • Please pull out the battery insulator strip in the battery compartment before using.
  • For first time user: please use ElitechLog software to synchronize and configure the system time.
  • Do not remove the battery from the logger while it's recording.
  • The LCD will be auto off after 15 seconds of inactivity (by default). Press the button again to turn on the screen.
  • Any parameter configuration on ElitechLog software will delete all logged data inside the logger. Please save data before you apply any new configurations.
  • Do not use the logger for long-distance transport if the battery icon is less than half [Battery Level Indicator].

Appendix: Default Parameters

ParameterRC-5RC-5+RC-5+TE
Logging Interval15 minutes2 minutes2 minutes
Start ModePress ButtonPress ButtonPress Button
Start Delay000
Stop ModeUse SoftwarePress ButtonPress Button
Repeat StartDisableEnableEnable
Circular LoggingDisableDisableDisable
Time ZoneUTC +00:00UTC +00:00
Temperature Unit°C°C°C
High Temperature Limit60°C//
Low Temperature Limit-30°C//
Calibration Temperature0°C0°C0°C
Temporary PDFEnableEnable
PDF LanguageChinese/EnglishChinese/English
Sensor TypeInternalInternalExternal

PDF preview unavailable. Download the PDF instead.

RC-5 instructions Adobe PDF library 17.00

Related Documents

Preview Elitech RC-5+ Temperature Data Logger: User Manual and Specifications
This manual provides detailed instructions for using the Elitech RC-5+ USB temperature data logger, covering setup, operation, data download, and analysis with Elitech software. Features and technical specifications are also included.
Preview Elitech RC-4HC Temperature and Humidity Data Logger Specifications
Comprehensive specifications and features for the Elitech RC-4HC PDF temperature and humidity data logger, detailing its capabilities for cold chain monitoring, industrial applications, and data export.
Preview Elitech IoT & Cold Chain Data Loggers | Comprehensive Product Catalog
Explore Elitech's extensive range of IoT and cold chain data loggers, including temperature, humidity, and multi-use devices. Discover features, technical specifications, and software solutions for reliable monitoring and compliance.
Preview Elitech RC-4 Series Temperature and Humidity Data Loggers - User Manual
Comprehensive user manual for Elitech RC-4 and RC-4HC series temperature and humidity data loggers. Covers specifications, operation, setup, and troubleshooting for cold chain monitoring.
Preview Elitech RC-4 Pro Multilingual Product Manual
Comprehensive guide to the Elitech RC-4 Pro temperature and humidity data logger, covering product overview, technical specifications, operating instructions, and precautions.
Preview Elitech RC-51H User Manual: Temperature & Humidity Data Logger Guide
Comprehensive user manual for the Elitech RC-51H multi-use temperature and humidity data logger. Learn about its features, operation, technical specifications, and how to use it for monitoring and recording environmental data.
Preview Elitech RC-61/GSP-6 User Manual: Temperature and Humidity Data Logger
This user manual provides comprehensive guidance for the Elitech RC-61/GSP-6 temperature and humidity data logger, covering its features, specifications, operation, and important usage notes for cold chain monitoring applications.
Preview Elitech RC-51 Multi-use Temperature Data Logger: Specifications & User Guide
Comprehensive technical specifications, usage instructions, and overview for the Elitech RC-51 multi-use temperature data logger, designed for monitoring food, pharmaceuticals, and chemicals in storage and transportation.