HIOKI GENNECT One SF4000
A rock-solid control tower for all your measurement operations
GENNECT One software empowers users to consolidate the capabilities of multiple HIOKI measuring instruments. It allows for real-time data viewing across various instruments on a single graph, automatic download and organization of measurement files into a unified folder, and remote control of instrument screens. Essentially, GENNECT One transforms the management of multiple instruments into a seamless experience, as if operating a single device, significantly advancing measurement capabilities.
? The name GENNECT is derived from the Japanese word for "field" (genba) and the English word "connect."
Connect to and manage instruments with a computer
A diagram illustrates the system architecture: multiple HIOKI measuring instruments are connected via ? (LAN) cables to a ? (HUB). The HUB then connects to a ? (computer). A ? (USB) connection is also shown for specific instruments.
Instrument Categories Supported:
- Power Analysis: PW6001, PW3390
- Monitoring Power Quality: PQ3100, PQ3198
- Understanding Power Consumption: PW3360, PW3365
- Voltage and temperature management: LR8450, LR8410, LR8400s
- Waveform Analysis: MR6000
- UPS Inspection: BT3554, BT3554-50
Collect and Display measured values by instrument
Collect values in graphs and lists (Logging)
The Logging function acquires measurement data at regular intervals from multiple instruments. This data is then displayed and stored on the PC in real time. This feature requires no custom program development.
Combine images and other elements (Dashboard)
The Dashboard feature allows users to create custom displays by arranging measurements, background images, and other elements on the screen. Measured values can be viewed on the dashboard in real time. This feature also requires no custom program development.
Change instrument settings from your office
Change instrument settings from a computer
GENNECT One enables remote control, allowing users to change instrument settings and start or stop measurements directly from a computer. This function supports 1 unit at a time for simultaneous operation.
Instrument clock synchronization
The clock of connected measuring instruments can be synchronized with the PC clock.
Collect and organize measurement files from scattered locations
Transfer measurement files to a computer (Automatic file transfer)
Measurement data stored in instruments can be automatically transferred to a PC. The data is transferred after it is saved to the instrument.
Review acquired files on a single time axis (Data import & Time-series viewer)
Measurement data stored in instruments can be manually imported to a PC for review in a single time series. The Time-series viewer supports various analysis functions, including statistical analysis, waveform search, event mark search, editing, CSV output, and image output. Up to 512 items can be displayed (or 32 items simultaneously in graphs).
Centralized management of measurement files
Combine measurement files acquired with GENNECT One into a single data list for efficient management by measurement site or time. Users can easily manage files by setting titles, classifying folders, inputting comments, and adding search tags.
A diagram shows a file list interface on a computer screen, displaying files organized by date and folder, with options to manage titles, folders, comments, and search tags. Clicking a measurement file launches the associated browsing software, provided it is installed beforehand.
Listed measurement files and Viewing Software:
Measurement files can be obtained via automatic or manual file transfer, GENNECT Cross/Remote, internal memory, external storage, or image/PDF data.
Supported viewing software includes:
- PQ ONE (for PQ3100, PQ3198)
- POWER LOGGER VIEWER SF1001 (for PW3360, PW3365)
- Logger utility (for LR8400, LR8410, LR8450, LR8450-01)
- Waveform Viewer (Wv) or WAVE PROCESSOR 9335 (for MR8875, MR6000)
- MR6000 Viewer
- Graph viewer for real-time measurement of GENNECT One
Note: Some software is free, while others are paid. Data acquired by logging functions of GENNECT Cross and GENNECT Remote are also supported.
Functionality and Specifications
Functionality Overview
Functionality | Interface | Compatible Instruments |
---|---|---|
Data import | LAN, USB* | PQ3100, PQ3198, PW3360, PW3365, PW3390, PW6001, LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000, BT3554-50, BT3554, 3554 |
Instrument clock synchronization | LAN | PQ3100, PQ3198, PW3360, PW3365, PW3390, PW6001, LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000, BT3554-50, BT3554, 3554 |
Logging | LAN | PW6001, PW3390, PQ3100, PQ3198, PW3365, PW3360, LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000 |
Dashboard | LAN | PW6001, PW3390, PQ3100, PQ3198, PW3365, PW3360, LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000 |
Remote control | LAN | PQ3100, PQ3198, PW3360, PW3365, PW3390, PW6001, LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000 |
Automatic file transfer | LAN | PQ3100, PQ3198, PW3360, PW3365, LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000 |
Time-series viewer | LAN | PW3390, PW6001, PQ3100, PQ3198, PW3365, PW3360, LR8450, LR8450-01, LR8410, LR8400s, MR6000, BT3554-50, BT3554, 3554 |
*USB interface is supported for BT3554, 3554 models only.
Logging
- Interface: ? LAN
- Function: Generates graph and list displays of measured values from LAN-connected instruments in real time. Acquires measured values (current values) displayed on instruments at a set interval (as short as 1 sec.) using the computer's timer.
- Logging intervals: 1, 2, 5, 10, 30 sec., 1, 2, 5, 10, 30 min., 1 hour
- Measurement parameters:
- PW6001, PW3390: Basic Measurement items, Harmonic measurement items (Instantaneous value per interval)
- PQ3100, PQ3198, PW3365, PW3360: Voltage, Current, Power (Instantaneous value per interval, MAX, MIN, AVE values between intervals)
- LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000: Various measurement items such as Temperature, Analog input, etc. (Instantaneous value per interval)
- Number of items able to monitor and log: Max. 512 items (Max. 32 items when simultaneously displaying graphs)
- Recording time: Continuous measurement/Set time, file segmentation: 1 day / 1 hour. Stops logging when PC storage capacity drops below 512MB.
Dashboard
- Interface: ? LAN
- Function: Monitors measured values in real time from LAN-connected instruments. Displays alarms in real time when a measured value exceeds a threshold. Displays measured values on a user-selected image. Acquires measured values (current values) displayed on instruments at a set interval (as short as 1 sec.) according to the computer's timer.
- Monitor interval: 1, 2, 5, 10, 30 sec., 1, 2, 5, 10, 30 min., 1 hour
- Measurement parameters:
- PW6001, PW3390: Basic Measurement items, Harmonic measurement items (Instantaneous value per interval)
- PQ3100, PQ3198, PW3365, PW3360: Voltage, Current, Power (Instantaneous value per interval, MAX, MIN, AVE values between intervals)
- LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000: Various measurement items such as Temperature, Analog input, etc. (Instantaneous value per interval)
- Number of measurement items: Max. 512 items
Remote control
- Interface: ? LAN
- Function: Controls LAN-connected instruments from a computer.
- Compatible instruments: PQ3100, PQ3198, PW3360, PW3365, PW3390, PW6001, LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000
- Number of simultaneous operation: 1 unit at a time
Automatic file transfer
- Interface: ? LAN
- Function: Automatically sends files saved by LAN-connected instruments to a computer.
- Compatible instruments: PQ3100, PQ3198, PW3360, PW3365, LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000
Data import
- Interface: ? LAN, ? USB*
- Function: Acquires files saved by LAN-connected instruments with a computer.
- Compatible instruments: PQ3100, PQ3198, PW3360, PW3365, PW3390, PW6001, LR8400, LR8401, LR8402, LR8410, LR8450, LR8450-01, MR6000, BT3554-50, BT3554, 3554
- *USB is supported for BT3554, 3554 models only.
Other Functions
- Files loading: Data file (.hok) obtained by GENNECT Cross for iOS/Android (Logging, General Measurement, image and battery formats only). Note: No direct Bluetooth® connection; use smartphone app for Bluetooth® data collection.
- Instrument clock synchronization: Supported for listed instruments.
- CSV output: Output 512 items at once (logging function).
- Report creation: PDF, Word and Excel formats.
Operating Environment
- Supported OS: Windows Vista (32bit/64bit SP2 or more), Windows 7 (32bit/64bit), Windows 8.1 (32bit/64bit), Windows 10 (32bit/64bit)
- Software environment: Microsoft .NET Framework 4.0, 4.5, 4.6, 4.7
- CPU: 2 GHz or more operating clock
- Memory: 4 GB or more
- Display: Resolution 1024 x 800 or more
- Hard disk: Free space of 1 GB or more
Company Information
HIOKI E. E. CORPORATION
HEADQUARTERS
81 Koizumi,
Ueda, Nagano 386-1192 Japan
https://www.hioki.com/
Note: Company names and product names appearing in this brochure are trademarks or registered trademarks of various companies. All information correct as of Dec. 23, 2020. All specifications are subject to change without notice.