Apera Instruments 2310-C High-Range Conductivity Electrode

User Manual

Brief Introduction

Apera Instruments 2310-C High-Range Conductivity Electrode is featured with a proprietary platinum black ring sensor for accurate readings in an ultra-wide conductivity range up to 2000 mS/cm.

Technical Specifications

Range 20 to 2000 mS/cm
Housing POM
Constant K=10
Connector BNC
Cable length 3 Ft
Dimension Ø12*145 mm
Temperature sensor N/A
Operating temperature 32 to 176°F (0 - 80°C)

A diagram shows the conductivity electrode with a BNC connector and a protective cap.

How to Install the Electrode

  1. Find the conductivity BNC socket (where it shows Cond) on the conductivity meter; Open the rubber cap; Insert the BNC connector of the electrode to the BNC socket while twisting clockwise until it's locked.
  2. After connecting the new electrode to your meter, adjust the meter's constant K to 10 according to the instruction of your meter's manual. Then perform a 2 to 3-point calibration to ensure the accuracy.

How to Use the Electrode

  1. There is a proper amount of deionized water in the cap on top of the electrode. The conductivity sensor is soaked in it to keep its sensitivity.
  2. Before measuring, loosen the bottle cap, then pull out the electrode while twisting counterclockwise. Place the storage bottle at a safe position.
  3. Rinse the electrode with distilled or deionized water and blot-dry it with clean tissue or filter paper. Rinse the electrode again with your test solution if possible for the best result. Never rub the sensor.
  4. Insert the electrode into your sample solution and stir it for a few seconds in the solution to remove potential air bubbles. Then wait for the stable reading and take the measurement.
  5. After measurement, insert the electrode back into the storage cap and twist it on. If the water in the cap is contaminated, rinse out the storage cap and fill in new deionized or distilled water.

How to Maintain the Electrode

  1. Always rinse the electrode with distilled or deionized water before and after each test and calibration.
  2. Store the electrode in the storage cap filled with distilled or deionized water when not in use.
  3. Keep the electrode connector clean and dry. Use cotton balls with isopropyl alcohol to clean if it gets dirty and then blow-dry it. This is to prevent a potential short circuit, which will undermine the electrode's performance.

Limited Warranty

Apera Instruments warrants this electrode to be free from defects in material and workmanship and agrees to repair or replace free of charge, at the option of APERA INSTRUMENTS, any malfunctioned or damaged product attributable to the responsibility of APERA INSTRUMENTS for a period of SIX MONTHS from the delivery.

This warranty does not cover any damages due to: Accidental damage, transportation, storage, improper use, failure to follow the product instructions, unauthorized repair, normal wear and tear, or any other actions or events beyond Apera Instruments' reasonable control.

Models: 2310-C High-Range Conductivity Electrode, 2310-C, High-Range Conductivity Electrode, Conductivity Electrode, Electrode

File Info : application/pdf, 2 Pages, 47.71KB

PDF preview unavailable. Download the PDF instead.

2310-C-EN

References

GPL Ghostscript 9.50

Related Documents

Preview LabSen 371 Plastic Flat pH Electrode User Manual | APERA INSTRUMENTS
User manual for the APERA INSTRUMENTS LabSen 371 Plastic Flat pH Electrode, detailing technical specifications, usage, maintenance, and warranty information for surface pH measurements.
Preview Apera EC20 Pocket Conductivity Tester: Instruction Manual & Guide
Comprehensive instruction manual for the Apera EC20 Pocket Conductivity Tester. Learn about setup, calibration, conductivity measurement, technical specifications, and warranty information.
Preview LabSen 751 pH Electrode User Manual - Apera Instruments
User manual for the Apera Instruments LabSen 751 Stainless Steel Sheath Spear pH Electrode, detailing features, technical specifications, usage, maintenance, and warranty for food pH testing.
Preview Apera Instruments EC20 Pocket Conductivity Tester Instruction Manual
User manual for the Apera Instruments EC20 Pocket Conductivity Tester, covering battery installation, keypad functions, calibration, conductivity measurement, technical specifications, and warranty information.
Preview Apera LabSen 241-6 Semi-Micro pH Electrode User Manual
User manual for the Apera LabSen 241-6 Semi-Micro pH Electrode, detailing its features, technical specifications, usage, maintenance, and warranty information for micro sample analysis.
Preview Apera LabSen 241-180 pH Electrode User Manual
User manual for the Apera Instruments LabSen 241-180 Micro pH Electrode, detailing its features, technical specifications, usage, maintenance, and limited warranty.
Preview LabSen 373 Flat pH Electrode User Manual | Apera Instruments
User manual for the Apera Instruments LabSen 373 Flat pH Electrode, detailing technical specifications, usage, maintenance, and warranty information for accurate surface pH measurements.
Preview APERA INSTRUMENTS PC950 Benchtop pH/Conductivity Meter Instruction Manual
Comprehensive instruction manual for the APERA INSTRUMENTS PC950 Benchtop pH/Conductivity Meter, covering specifications, operation, calibration, maintenance, and troubleshooting for laboratory use.