Apera Instruments EC20 Pocket Conductivity Tester Instruction Manual

1. Battery Installation

Please install batteries according to the following steps. *Please note the correct direction of battery installation:

The Positive Side ("+") OF EVERY SINGLE Battery MUST FACE UP. (WRONG INSTALLATION OF BATTERIES WILL CAUSE DAMAGE TO THE TESTER AND POTENTIAL HAZARDS!) ⚠️

(batteries are pre-installed)

Illustration showing seven steps for installing AAA batteries into the Apera EC20 tester. Step 1 shows the tester with the battery cap removed. Steps 2-6 show the insertion of four AAA batteries, noting the positive (+) terminal facing up. Step 7 shows the battery compartment closed.

2. Keypad Functions

Short press: < 2 seconds

Long press: > 2 seconds

  • [MEAS] Button: 1. Short press to turn on, long press to turn off; 2. When turned off, long press to enter setup; 3. In mode setting, short press to change parameter;
  • [CAL] Button: 1. When turned on, long press to enter calibration mode. 2. In calibration mode, short press to confirm calibration; 3. In mode setting, short press to confirm parameter selection.

Diagram of the Apera EC20 conductivity tester, labeling the Battery Cap, LCD Screen, Keypad, BPB Probe, and Probe Cap.

3. Complete Kit

Visual representation of the Apera EC20 Pocket Conductivity Tester kit contents, including the tester unit, a lanyard, four AAA batteries, and two bottles of calibration solution.

4. Calibration

4.1 Rinse the probe in distilled water and dry it with a clean tissue. Short press [MEAS] to turn on the tester.

4.2 Long press [CAL] to enter calibration mode; Short press [MEAS] to exit.

4.3 Dip the probe into 1413 µS/cm calibration solution. Stir gently, leave it to stand. LCD displays ? (Diagram 3), short press [CAL] to complete 1-point calibration. The tester returns to measuring mode, and calibration icon [M] appears at the bottom left of LCD.

Close-up of the Apera EC20 tester's LCD screen during calibration. It displays a conductivity reading of '1433' and a calibration solution value of '1413', accompanied by a stable smiley face icon (?) and the calibration indicator 'M'.

4.4 Rinse probe in distilled water and dry it. Follow the steps in 4.2-4.3 to complete 2nd point calibration in 12.88 mS/cm calibration solution. Tester returns to measurement mode, calibration icons [M] [H] display on bottom left of LCD.

5. Conductivity Measurement

5.1 Short press [MEAS] to turn on tester. Rinse probe in distilled water and dry it with a clean tissue to remove excess water.

5.2 Stir the probe in the sample solution gently, leave it to stand. Get readings after the smile icon (?) comes up and stays.

5.3 Rinse off the tester thoroughly in distilled water after each test.

6. Notes

6.1 For first-time use or if the tester hasn't been used for a long time, soak the probe in 12.88 mS/cm solution for 15-30 minutes to restore sensitivity and speed up response time.

6.2 The tester uses 1413 µS and 12.88 mS standard calibration solutions. Users can perform 1- or 2-point calibrations as needed. For most circumstances, calibrating in 1413 µS for the first point meets testing requirements.

6.3 Unit Information: 1 mS/cm = 1000 µS/cm. Readings greater than 2000 µS/cm automatically switch to mS/cm.

6.4 Self-Diagnosis Functions

SymbolSelf-Diagnosis InformationHow to Fix
Er1Wrong calibration solution, which exceeds the recognizable range of the meter.1. Check if calibration solution is correct. 2. Check if probe is damaged.
Er2[CAL] is pushed before measurement is stable (? comes up and stays).Wait for the smile icon to stay, then short press [CAL].

6.5 The tester is pre-calibrated from the factory. Users can typically use it immediately or test its accuracy with calibration solutions. If a large error is detected, calibrate before use.

7. Parameter Setting

7.1 Parameter Setting Reference Chart

SymbolParameter Setting ContentCodeFactory Default
P1Select conductivity rangesAut-µS-mSAut
P2Select Temperature Unit°C - °F°C
P3Restore to factory defaultNo - YesNo

7.2 How to Setup Parameters

When turned off, long press [MEAS] to enter setup. Short press [MEAS] to switch between P1-P2-P3. Short press [CAL], parameter flickers; short press [MEAS] to choose; short press [CAL] to confirm parameter selection. Long press [MEAS] to switch off.

7.3 Parameter Setting Notes

a) Select conductivity ranges (P1): Aut-- All ranges: 0~199.9 µS/cm, 20~1999 µS/cm, 2.0~20.0mS/cm. Factory default is Aut.

b) P2-Select temperature unit (°C-°F): In P1 mode, press [MEAS] to enter P2. Press [CAL], °C or °F flickering. Press [MEAS] to choose °C-°F. Press [CAL] to confirm parameter selection. Long press [MEAS] to return to measurement mode.

c) Restore to factory default: Select Yes to restore calibration to theoretical values and parameter settings to original values. This function can be used if the meter's calibration or measurement performs abnormally, allowing it to revert to factory default settings for subsequent calibration or measurement.

8. Technical Specifications

RangeResolutionAccuracy
Conductivity0~199.9 µS/cm, 20~1999µS/cm, 2.0~20.00mS/cm0.1/1 µS/cm, 0.01 mS/cm±1% F.S±1 digit
Calibration points1-2 points auto calibration
Automatic Temp. Compensation0-50°C
Temp.Range0-50°C
Resolution0.1°C
Accuracy±0.5°C

9. Other Functions & Parameters

FeatureDetails
Indication of calibration points[M] [H]
Indication of stable measurements?
Self-Diagnosis informationEr1, Er2
Low battery reminderFlashes to remind to replace batteries
Auto Power-offPower-off in 8 minutes if no operation
Waterproof levelIP67, floats on water
Power SupplyAAA batteries*4
Battery LifeUp to 1000 hours of continuous operation
Dimensions/WeightInstrument: 40*31*178mm/107g; Carrying case: 190*165*140mm/438g

10. Warranty

Apera Instruments warrants this instrument to be free from defects in material and workmanship. Apera Instruments (Europe) GmbH agrees to repair or replace, free of charge, any malfunctioned or damaged product attributable to its responsibility for a period of two years from delivery (a six-month limited warranty applies to electrodes). The warranty period is the time limit for providing free service for products purchased by customers, not the service life of the tester or electrodes.

This limited warranty does not cover any damages due to:

  • transportation;
  • storage;
  • improper use;
  • failure to follow the product instructions or to perform any preventive maintenance;
  • modifications;
  • combination or use with any products, materials, processes, systems, or other matter not provided or authorized in writing by Apera Instruments;
  • unauthorized repair;
  • normal wear and tear; or
  • external causes such as accidents, abuse, or other actions or events beyond Apera Instruments' reasonable control.

Contact Information

APERA INSTRUMENTS (Europe) GmbH

Wilhelm-Muthmann-Straße 18

42329 Wuppertal, Germany

Contact: info@aperainst.de

Website: www.aperainst.de

Tel: +49 202 51988998

Models: EC20 Pocket Conductivity Tester, EC20, Pocket Conductivity Tester, Conductivity Tester, Tester

File Info : application/pdf, 6 Pages, 459.87KB

PDF preview unavailable. Download the PDF instead.

EC20-EN-3

References

Microsoft Word für Microsoft 365

Related Documents

PreviewApera EC20 Pocket Conductivity Tester: Instruction Manual & Guide
Comprehensive instruction manual for the Apera EC20 Pocket Conductivity Tester. Learn about setup, calibration, conductivity measurement, technical specifications, and warranty information.
PreviewApera Instruments EC400S Portable Conductivity Meter Instruction Manual
Comprehensive instruction manual for the Apera Instruments EC400S Portable Conductivity Meter, covering features, specifications, operation, calibration, sample measurement, maintenance, parameter settings, and USB data communication.
PreviewApera EC1 Pocket Conductivity Tester Instruction Manual
Comprehensive instruction manual for the Apera EC1 Pocket Conductivity Tester, detailing its features, operation, calibration, technical specifications, and warranty information.
PreviewAPERA INSTRUMENTS PC950 Benchtop pH/Conductivity Meter Instruction Manual
Comprehensive instruction manual for the APERA INSTRUMENTS PC950 Benchtop pH/Conductivity Meter, covering specifications, operation, calibration, maintenance, and troubleshooting for laboratory use.
PreviewApera Instruments PC60 Premium 5-in-1 Tester User Manual
Comprehensive user manual for the Apera Instruments PC60 Premium 5-in-1 Tester, covering pH, Conductivity, TDS, Salinity, and Temperature measurements. Includes setup, calibration, operation, maintenance, and troubleshooting.
PreviewApera SX823-B pH/Conductivity Portable Meter Instruction Manual
This instruction manual provides detailed information on the Apera SX823-B portable pH/Conductivity meter, covering its features, specifications, operation, calibration procedures, maintenance, and troubleshooting.
PreviewSALT20 Pocket Salinity Tester User Manual - Apera Instruments
User manual for the Apera Instruments SALT20 Pocket Salinity Tester, covering installation, operation, calibration, technical specifications, and warranty information provided by Labco Scientific.
PreviewApera Instruments SX731 Portable pH/Conductivity/ORP Meter User Manual
User manual for the Apera Instruments SX731 Portable pH/Conductivity/ORP Meter, covering its features, technical specifications, operation, calibration, maintenance, and warranty.