Fisher & Paykel AIRVO™ 2 User Manual

Brand: Fisher & Paykel Healthcare

This manual provides comprehensive instructions for the setup, operation, and maintenance of the AIRVO™ 2 humidifier.

Before You Start

This User Manual is intended for healthcare professionals and applies to AIRVO 2 units with LOT numbers 130621 and above. It is crucial to read this manual, including all warnings, and to watch the AIRVO 2 Video Guide for safe and effective use. Both resources should be kept for future reference. Before initial use, the AIRVO 2 must be set up according to the AIRVO 2 Technical Manual. Cleaning and disinfection between patients must follow the Disinfection Kit Manual (900PT600). For assistance, contact your Fisher & Paykel Healthcare representative.

Table of Contents

  1. Overview (Intended Use, Warnings, AIRVO 2 and Accessories)
  2. Setting up AIRVO 2
  3. Using AIRVO 2 (Target dew-point temperature, Target flow, Oxygen, Alarms)
  4. Reprocessing (Schedule for changing accessories, Filter replacement, Servicing)
  5. Technical Information

1. Overview

The AIRVO 2 is a humidifier with an integrated flow generator that delivers high-flow, warmed, and humidified respiratory gases to spontaneously breathing patients via various patient interfaces. It is designed for use in hospitals and long-term care facilities. The unit is restricted by USA Federal Law for sale by or on the order of a physician.

Intended Use

The AIRVO 2 is intended for the treatment of spontaneously breathing patients who benefit from high-flow, warmed, and humidified respiratory gases, including those with bypassed upper airways. The flow rate can range from 2 to 60 L/min, depending on the patient interface.

Warnings

AIRVO 2 and Accessories

The AIRVO 2 system includes the main unit, heated breathing tube, chamber, and various patient interfaces. Key components include:

Tube & Chamber Kits and Patient Interfaces:

Tube & Chamber KitInterfaces
900PT531 (Heated breathing tube, MR290 auto-fill chamber and adapter)OPT316 (Nasal Cannula - Infant), OPT318 (Nasal Cannula - Pediatric), OPT842 (Nasal Cannula - Small), OPT844 (Nasal Cannula - Medium), OPT846 (Nasal Cannula - Large), OPT870 (Tracheostomy Interface), RT013 (Mask Interface Adapter - 22mm)
900PT501 (Heated breathing tube, MR290 auto-fill chamber and adapter)(For use with OPT316/OPT318 only)

Cleaning and Disinfection: Disinfection Kit (900PT600), Disinfection Filter (900PT601), Cleaning Sponge-Stick (900PT602), Clean Storage Cover (900PT603).

Miscellaneous: Pole mounting tray (900PT405), Hospital stand (900PT421), Oxygen inlet extension kit (900PT422), Filter holder (900PT912), Air filter (900PT913), Wigglepads (OPTO12), Oxygen Tubing (OPT014).

2. Setting Up AIRVO 2

  1. Before You Begin: Mount the AIRVO 2 on a pole mounting tray (900PT405) below patient head height. Unpack the tube & chamber kit.
  2. Install Water Chamber: Remove blue port caps, attach the adapter, and connect the water supply tube. Fit the water chamber to the unit by pressing the finger guard and sliding it until it clicks. Ensure the chamber is securely in place.
    • Warnings: Do not start the unit without the chamber. The water becomes hot during use. Avoid tilting the unit to prevent water entry. Empty the chamber before transport.
    • Cautions: For MR290, do not use if dropped or run dry with the "water out" alarm activated.
  3. Connect Water Bag: Attach the sterile water bag to the hanging bracket above the unit and insert the bag spike. Open the vent cap. The chamber will auto-fill. Ensure the water bag or chamber does not run dry. Check that water flow is maintained below the fill line.

MR290: Flow setting vs usage time (2-litre sterile water bag)

L/min251015202530354045505560
hrs3791527651383025221917151413

3. Using AIRVO 2

  1. Switch On Unit: Plug in the power cord securely. Ensure the unit is dry. Press the On/Off button.
  2. Check Disinfection Status: The unit displays whether it is safe for use on a new patient based on its last disinfection.
  3. Warm-up: The unit warms up, displaying target temperature, flow, and oxygen values. These values pulse until they reach their targets. This is the "Summary screen".
  4. Junior Mode: For Optiflow Junior nasal cannulas (OPT316/OPT318), activate Junior Mode by holding the Mode button for 5 seconds. This limits target settings to 34°C and 2-25 L/min. The screen icons indicate Junior Mode. Deactivate by holding the Mode button for 5 seconds.

4. Configure Target Settings

Target Dew-Point Temperature

The AIRVO 2 can be set to three target dew-point temperatures: 37°C (98.6°F), 34°C (93°F) [for compliance issues], and 31°C (88°F) [for face masks only]. Settings may be limited by Junior Mode or pre-set configurations. The unit defaults to 37°C after disinfection.

To change the target dew-point temperature setting: Hold Up and Down buttons for 3 seconds to unlock. Use Up/Down buttons to select the new setting. Press Mode to lock.

Transport Mode

Activate Transport Mode by holding the "Audio pause" button for 5 seconds. This enables a low-power, low-humidity mode for 20 minutes during patient transport. Deactivate similarly.

Target Flow

Set flows between 10-60 L/min (10-25 L/min in 1 L/min increments, 25-60 L/min in 5 L/min increments). Settings may be limited by Junior Mode or pre-set configurations. The unit remembers the last flow setting.

To change the target flow setting: Follow the same steps as for changing the dew-point temperature.

5. Oxygen

Supplementary oxygen can be connected (up to 60 L/min). The AIRVO 2 has an oxygen analyzer. Use continuous oxygen monitoring for at-risk patients.

Warnings

Connect Oxygen

Connect the oxygen source output to the oxygen inlet port on the back of the unit. Push the tube firmly.

Adjust Oxygen

Adjust the oxygen source to achieve the desired oxygen fraction. The reading may take minutes to stabilize. The display shows the acceptable range. If the fraction exceeds 95%, the reading pulses red and the device beeps.

Warnings: If patient's peak inspiratory demand exceeds the unit's flow, the inspired oxygen fraction may be lower due to ambient air entrainment. Ensure adequate blood saturation levels.

6. Connect Your Patient

Wait for the "Ready for use" symbol on the Summary screen. Connect the patient interface to the heated breathing tube. Monitor flow and oxygen values and adjust oxygen as needed. The initial airflow will feel warm, which is normal. The patient should breathe normally through their interface.

7. During Use

A screensaver activates if no button is pressed for 1 minute after the "Ready for use" symbol appears. If excess condensate accumulates in the heated breathing tube, drain it into the water chamber by lifting the patient end of the tube.

8. After Use

Switch off the unit by pressing the On/Off button.

Alarms

The AIRVO 2 features visual and auditory alarms to alert users to interruptions in patient treatment. An intelligent alarm system processes sensor data and compares it to pre-programmed limits.

Alarm Signals

Alarm Conditions

All alarms are classified as "Medium Priority". The unit prioritizes alarms if multiple conditions occur simultaneously.

MessageMeaningAffects Delivery Of:Delays
Fault (E###)Internal fault, unit shut down. Restart unit. If persists, contact representative.Oxygen, humidity< 5 seconds
Check tubeCannot detect heated breathing tube. Check for damage and correct connection. Replace if persists.Oxygen, humidity< 5 seconds
Check for leaksSystem leak detected. Check water chamber seating and connection. Check tube for damage and connection. Check nasal interface and filter fit.Oxygen, humidity< 5 seconds
Check for blockagesSystem blockage detected. Check tube, interface, air filter, and filter holder for blockage. Activate Junior Mode if applicable.Oxygen, humidity< 10 seconds
O₂ too lowMeasured oxygen level below limit. Check oxygen source connection and adjust level.Oxygen< 20 seconds
O₂ too highMeasured oxygen level exceeds limit. Adjust oxygen source level.Oxygen< 20 seconds
Cannot reach target flowUnit cannot reach target flow. Check tube/interface for blockage. Verify target flow setting is appropriate for the interface. Unit may select new settings.Oxygen10 +/- 1 minutes
Check waterChamber has run out of water. Replace chamber and water bag. Unit enters Warm-up Mode upon replacement. Ensure water bag/chamber does not run dry.HumidityFlows > 20 L/min: < 20 minutes. Flows ≤ 20 L/min: < 40 minutes
Cannot reach target temperatureUnit cannot reach target temperature. Likely cause: high flow rate in low ambient conditions. Consider decreasing target flow.Humidity30 +/- 3 minutes
Check operating conditionsUnit operating in unsuitable ambient conditions (below 10°C or above 30°C). Allow unit to acclimatize or restart.Humidity60 +/- 6 seconds
[Power out]Unit disconnected from mains power. Auditory alarm sounds for 120 seconds.Oxygen, humidity< 5 seconds

Alarm Limits

Most alarm limits are pre-programmed but can be adjusted by authorized personnel. Examples include O₂ too low (factory set at 21%) and O₂ too high (factory set at 95%).

Checking Alarm System Functionality

To check the alarm system, remove the heated breathing tube. The "Check tube" visual and auditory alarms should activate. If not, do not use the unit and contact Fisher & Paykel Healthcare.

Auditory Information Signals

4. Reprocessing

The AIRVO 2 must be cleaned and disinfected between patients per the Disinfection Kit Manual (900PT600). Follow aseptic techniques to minimize contamination. Proper hand hygiene, avoiding contact with ports, and safe disposal of consumables are essential.

Schedule for Changing Accessories

Accessories require frequent replacement to prevent infection risk. Replace immediately if damaged or discolored, or within the following periods:

Maximum Period of UsePart Number and Description
1 week (single-patient use)All patient interfaces: OPT316 (Nasal Cannula - Infant), OPT318 (Nasal Cannula - Pediatric), OPT842 (Nasal Cannula - Small), OPT844 (Nasal Cannula - Medium), OPT846 (Nasal Cannula - Large), OPT870 (Tracheostomy Interface), RT013 (Mask Interface Adapter - 22mm)
2 weeks (single-patient use)All tube & chamber kits: 900PT501, 900PT531 (Heated breathing tube, MR290 auto-fill chamber and adapter)
3 months or 1000 hours900PT913 (Air filter)

Filter Replacement

When prompted for a filter change:

  1. Remove the filter holder from the unit and take out the old filter.
  2. Replace with a new filter.
  3. Reattach the filter holder, ensuring clips engage correctly.
  4. Press the Mode button to proceed.

Servicing

This device contains no serviceable parts.

5. Technical Information

Symbol Definitions

The manual includes definitions for various symbols used on the device and in the documentation, such as Caution, Type BF Applied Part, Hot Surfaces, IPX1 Drip Proof, Do Not Throw Away, Consult Accompanying Documents, Alternating Current, Class II Double Insulated, Power On/Off, and CE 0123 (93/42/EEC Class IIa).

Product Specifications

SpecificationDetails
Dimensions295 mm x 170 mm x 175 mm (11.6" x 6.7" x 6.9")
Weight2.2 kg (4.8 lb) unit only; 3.4 kg (7.5 lb) packaged
Humidity>33 mg/L at 37°C target; >10 mg/L at 34°C target; >10 mg/L at 31°C target
Maximum temperature of delivered gas43 °C (109 °F)
Supply frequency50-60 Hz
Supply voltage/current100-115 V 2.2 A (2.4 A max); 220-240 V 1.8 A (2.0 A max)
Sound pressure levelAlarms exceed 45dbA @ 1m
Auditory alarm pause115 seconds
Serial portUsed for downloading product data with F&P Infosmart™ software.
Maximum flow range (default)10-60 L/min
Maximum flow range (Junior Mode)2-25 L/min
Maximum oxygen input60 L/min
Warm-up time10 minutes to 31°C (88°F); 30 minutes to 37°C (98.6°F) (with MR290 chamber, 35 L/min flow, 23°C starting temp)
Oxygen analyzer accuracy< ± (2.5% + 2.5% of gas level) (within 25-95% O₂)
Operating conditions18-28 °C (64-82 °F), 30-70% RH

Compliance

The unit is designed to conform to IEC 60601-1, UL 60601-1, CSA C22.2/No. 601.1, AS 3200.1.0, and EN 60601-1 standards. It complies with electromagnetic compatibility requirements (IEC 60601-1-2). Ensure system compliance with IEC 60601-1-1 when connecting external equipment.

Storage and Transport Conditions

Store and transport in environmental conditions of -10°C to 60°C (14°F to 140°F), 10 to 95% RH (non-condensing).

Disposal Instructions

Unit Disposal: Contains electronics. Do not discard with regular waste. Return to Fisher & Paykel Healthcare or dispose according to local electronics disposal guidelines and the WEEE directive.

Consumables Disposal: Place used consumables (interface, breathing tube, chamber) in a waste bag. Hospitals should discard according to their standard procedures for contaminated products.

Models: AIRVO 2, Water Spray Humidifier, AIRVO 2 Water Spray Humidifier, Spray Humidifier, Humidifier

File Info : application/pdf, 16 Pages, 2.12MB

PDF preview unavailable. Download the PDF instead.

myAIRVO-2-User-Manual RAD PDF 3.19.2.2 - https://www.radpdf.com PDFescape Online - https://www escape.com

Related Documents

Preview F&P Airvo 3 User Manual: High Flow Therapy Device Guide
This user manual provides comprehensive instructions for the Fisher & Paykel Healthcare Airvo 3, a device designed for high flow respiratory therapy. It covers device setup, operation, safety guidelines, troubleshooting, and technical specifications for healthcare professionals.
Preview F&P Airvo 2: Respiratory Support and Airway Hydration System
Discover the F&P Airvo 2, an integrated system for Optiflow tracheostomy high flow therapy, offering advanced humidification, respiratory support, and airway hydration for patients.
Preview Fisher & Paykel AIRVO 2 Humidifier User Manual
User manual for the Fisher & Paykel AIRVO 2 humidifier, providing instructions on setup, operation, maintenance, and troubleshooting for healthcare professionals.
Preview Fisher & Paykel AIRVO 2 Humidifier User Manual
User manual for the Fisher & Paykel AIRVO 2 humidifier, providing instructions on setup, operation, maintenance, and troubleshooting.
Preview Fisher & Paykel Healthcare Homecare and Long Term Care Facilities Catalogue
This catalogue from Fisher & Paykel Healthcare presents a comprehensive range of respiratory support products for homecare and long-term care facilities. It details various CPAP masks, including nasal, full face, nasal pillows, and oral types, along with CPAP machines, humidifiers, and breathing circuits. Product specifications, features, and spare parts are listed for each item, covering models such as Evora, Eson, Zest, FlexiFit, Vitera, Simplus, Forma, Brevida, Pilairo, Opus, Oracle, SleepStyle, myAirvo, F&P 810 System, and F&P 550 System.
Preview F&P SleepStyle Use and Care Guide for Sleep Apnea Therapy
Comprehensive guide for Fisher & Paykel Healthcare's F&P SleepStyle Auto and CPAP devices. Covers setup, operation, safety, maintenance, and troubleshooting for Obstructive Sleep Apnea treatment.
Preview F&P Brevida Nasal Pillows Mask: Use and Care Guide
A comprehensive guide from Fisher & Paykel Healthcare detailing the intended use, fitting, assembly, cleaning, and technical specifications of the F&P Brevida Nasal Pillows Mask for CPAP and Bi-Level therapy.
Preview F&P ICON+ CPAP Machine: Features, Technology, and Specifications
Discover the F&P ICON+ CPAP machine from Fisher & Paykel Healthcare, designed for enhanced patient comfort and effective sleep therapy. This product overview details its advanced features like ThermoSmart™ and SensAwake™, connectivity options via InfoSmart™ technologies, product specifications, and performance modes.