Fisher & Paykel AIRVO 2 User Manual

Before You Start

This User Manual is intended for healthcare professionals. It applies to AIRVO™ 2 units with a manufacturing date after 2016-08-04.

Read this manual, including all warnings. Failure to do so may result in injury. Keep it in a safe place for future reference.

Before first use, the AIRVO 2 must be set up according to the AIRVO 2 Technical Manual.

Some accessories may not be available in all countries. Contact your local Fisher & Paykel Healthcare representative for more information.

Other References

1. Overview

The AIRVO 2 is a humidifier with an integrated flow generator that delivers high-flow warmed and humidified respiratory gases to spontaneously breathing patients through various patient interfaces.

Intended Use

The AIRVO 2 is for the treatment of spontaneously breathing patients who benefit from high-flow warmed and humidified respiratory gases, including those with bypassed upper airways. Flow can range from 2 - 60 L/min depending on the patient interface. The AIRVO 2 is intended for use in hospitals and long-term care facilities.

The AIRVO 2 provides respiratory support through a nasal cannula interface, promoting slower, deeper breathing and increasing alveolar ventilation. For patients with nasal cannula or tracheostomy interfaces, it provides airway hydration and may be used with supplemental oxygen.

Warnings

To avoid burns:

To avoid electric shock:

To avoid choking, or inhalation of a foreign object:

Miscellaneous:

2. Setting Up AIRVO 2

1. Before You Begin

The AIRVO 2 should be fixed on a pole mounting tray (900PT405) below patient head height. Open the packaging of the tube & chamber kit (heated breathing tube, MR290 auto-fill chamber, and adapter).

2. Install Water Chamber

Remove blue port caps from the chamber. Remove the bracket holding the water supply tube. Fit the supplied adapter over the two vertical ports on the chamber and push on fully. Clip the water supply tube into position.

Fit the water chamber to the unit by pressing the finger guard and sliding the chamber on, aligning with the blue chamber port ends. Push firmly until the finger guard clicks.

Warnings

Cautions

3. Connect Water Bag

Attach the sterile water bag to the hanging bracket 20cm (8") above the unit. Push the bag spike into the fitting at the bottom of the bag. Open the vent cap on the bag spike. The chamber will automatically fill and maintain the level until the bag is empty.

Ensure the water chamber and/or water bag are never allowed to run out of water for continual humidification.

Check that water flows into the chamber and stays below the maximum water level line. If the level rises above, replace the chamber immediately.

MR290: Flow setting vs usage time (2-litre sterile water bag, at 37 °C target temperature)
L/min 2 5 10 15 20 25 30 35 40 45 50 55 60
hrs 378 151 75 50 37 30 25 21 18 16 15 13 12

Cautions

4. Install Heated Breathing Tube

Attach the heated breathing tube to the unit. Lift the blue plastic sleeve, slide the connector onto the unit, and push the sleeve down to lock.

Warnings

Cautions

5. Select Patient Interface

The AIRVO 2 can be used with various patient interfaces. Refer to the separate user instructions for each interface.

Nasal Cannula: Optiflow™™+, OPT942, OPT944, OPT946; Optiflow™™, OPT842, OPT844, OPT846; Optiflow™™ Junior, OPT316, OPT318.

Tracheostomy Interface: Optiflow™™+, OPT970; Optiflow™™, OPT870.

Mask Interface Adapter: OPT980 / RT013 (with mask). Note: The OPT980/RT013 Mask Interface Adapter is for vented masks only. Do not use sealed masks.

The following table shows target dew-point temperature and target flow settings for different interfaces:

Target Dew-Point Temperature (°C) and Target Flow (L/min)
Patient Interface 31 34 37 2 5 10 15 20 25 50 55 60
OPT316 2 20
OPT318 2 25
OPT942 S 10 50
OPT944 M 10 60
OPT946 L 10 60
OPT970 10 60
OPT980 10 60
OPT842 S 10 50
OPT844 M 10 60
OPT846 L 10 60
OPT870 10 60
RT013 10 60

Low temperature ambient conditions may prevent the unit from reaching a 37°C target temperature setting at high target flow settings. In these cases, consider decreasing the target flow setting.

At altitude, maximum achievable flow rates may be lower than those in the table by 5 L/min per 1000 m (3000 ft).

Warnings

3. Using AIRVO 2

1. Switch On Unit

Plug the unit's power cord into a mains/utility power socket. Ensure the connector is well secured to the rear of the unit. Switch on the unit using the On/Off button.

Warnings

2. Check Disinfection Status

The unit indicates if it is safe for use on a new patient:

3. Warm-Up

The unit warms up, displaying current output dew-point temperature, flow, and oxygen values. These numbers pulse until they approach their target settings. This screen is the "Summary screen".

4. Junior Mode

If using an Optiflow Junior nasal cannula (OPT316/OPT318), activate Junior Mode. This limits target settings to 34°C and 2-25 L/min (in 1 L/min increments).

To activate Junior Mode: Hold the Mode button for 5 seconds. The target settings will change automatically. Colorful icons indicate Junior Mode.

To deactivate Junior Mode: Hold the Mode button for 5 seconds.

5. Configure Target Settings

Press the Mode button to view target settings. These are locked by default.

Target Dew-Point Temperature

You can set the AIRVO 2 to three target dew-point temperatures:

Access to all settings may be limited if the unit is in Junior Mode (limited to 34°C) or if initially set up with tighter limits. The AIRVO 2 returns to the default 37°C after every disinfection cycle.

To change the target dew-point temperature setting:

  1. Hold the Up and Down buttons for 3 seconds to "unlock" the setting. A lock symbol disappears and is replaced by an arrow showing minimum and maximum settings.
  2. Press Up and Down buttons to choose the new setting.
  3. Press the Mode button to "lock" the setting again. The lock symbol reappears.

Target Flow

You can set the AIRVO 2 to flows between 10 L/min and 60 L/min (in 1 L/min increments for 10-25 L/min, and 5 L/min increments for 25-60 L/min). Access may be limited in Junior Mode (2-25 L/min) or with tighter initial limits.

The AIRVO 2 remembers its target flow setting when switched off.

To change the target flow setting: Follow the same steps as for changing the target dew-point temperature.

Oxygen

Supplementary oxygen can be connected to the AIRVO 2 (up to 60 L/min). The unit has an oxygen analyzer to help determine the delivered oxygen fraction. Your unit may have tighter initial limits.

Use continuous oxygen monitoring for patients who may desaturate if oxygen supply is interrupted.

Warnings

Connect Oxygen

Connect the oxygen source output to the oxygen inlet port on the side of the unit. Push the oxygen tube firmly onto the port.

Adjust Oxygen

Adjust the oxygen source level until the desired oxygen fraction is displayed. Readings may take several minutes to stabilize. The oxygen fraction can be set between the displayed minimum and maximum values.

Real-time O₂ measurement is displayed when O₂ is between 25% and 95%. Fractions below 25% and above 95% are displayed as 21% and 100% respectively.

If the oxygen fraction exceeds 95%, the reading pulses red, and the device beeps.

Warnings

6. Connect Your Patient

Wait for the "Ready for use" symbol on the Summary screen. Connect the patient interface to the heated breathing tube. Monitor flow and oxygen values on the Summary screen. Adjust oxygen from the source as needed. The air will feel warm initially; this is normal. The patient should breathe normally through the nose/mouth or tracheostomy.

7. During Use

If the "Ready for use" symbol is displayed for 2 minutes with no button presses, a screensaver will launch.

Condensate Management

Place the unit below head height and flat to allow condensate to drain towards the water chamber, away from the patient.

If excess condensate forms in the heated breathing tube, disconnect the patient interface, lift the patient end of the tube to drain condensate into the water chamber.

At higher target flow rates, reduce the target flow to 30 L/min or below to ensure condensate drains into the chamber.

Minimize local cooling sources acting on the heated breathing tube (e.g., fans, air conditioning). If condensate persists, consider lowering the target temperature, which will decrease humidity output and condensation.

8. After Use

Switch off the unit using the On/Off button.

ALARMS

The AIRVO 2 has visual and auditory alarms to warn of treatment interruptions. An intelligent alarm system processes sensor and target setting information against pre-programmed limits.

Alarm Signals

Auditory Alarm Signal

3 beeps in 3 seconds, repeated every 5 seconds. Press the button to mute the auditory alarm for 15 seconds. The alarm can be reactivated by pressing the button again.

Alarm Conditions

All alarms are assessed as "Medium Priority" for operator proximity within 1 meter. The unit uses an internal priority-ranking system. If multiple alarms occur, the highest priority alarm is displayed.

The following table lists alarm conditions from highest to lowest priority, with causes, solutions, and delays. Alarms affecting oxygen delivery require immediate response to assess patient saturation. Alarms affecting humidity delivery require prompt response to assess potential mucus drying and blockages.

Alarm Conditions (assuming 'Ready for use' mode)
Message Meaning Affects delivery of: Delays
Fault (E###) Internal fault detected; unit shut down. Switch off/on. If persists, note fault code and contact Fisher & Paykel Healthcare representative. Oxygen, humidity. < 5 seconds
Check tube Heated breathing tube not detected or damaged. Check connection and damage. If persists, change tube. Oxygen, humidity. < 5 seconds
Check for leaks Leak detected in the system. Check water chamber seating, heated breathing tube connection, patient interface fit, and filter fit. Oxygen, humidity. < 120 seconds
Check for blockages Blockage detected. Check heated breathing tube, patient interface, air filter, and filter holder. Activate Junior Mode if using Optiflow Junior cannula. Oxygen, humidity. < 10 seconds
O₂ too low Measured oxygen level below limit. Check oxygen source operational status and connection. Adjust oxygen level. Oxygen < 20 seconds
O₂ too high Measured oxygen level exceeds limit. Check AIRVO flow rate. Adjust oxygen level. Oxygen < 20 seconds
Cannot reach target flow Unit cannot reach target flow. Check for blockages in tube/interface. Verify target flow is appropriate for patient interface. Acknowledge prompt. Oxygen concentration may be affected; adjust oxygen source as necessary. Oxygen < 120 seconds
Check water Chamber has run out of water. Chamber float may be damaged. Replace chamber and water bag. Ensure water chamber/bag is not empty. Humidity < 30 minutes
Cannot reach target temperature Unit cannot reach target temperature. Acknowledge prompt. Likely cause: high flow rate in low ambient conditions. Consider decreasing target flow. Oxygen concentration may be affected; adjust oxygen source as necessary. Humidity 30 +/- 3 minutes
Check operating conditions Unsuitable ambient conditions detected. May be due to sudden change. Leave unit running for 30 minutes, then switch off/on. Humidity 60 +/- 6 seconds
[Power out] Disconnected from mains power. Auditory alarm sounds for 120 seconds. Unit restarts automatically if power is reconnected unless "Audio Pause" was pressed. Appropriate patient monitoring is essential. Oxygen, humidity. < 5 seconds

Alarm Limits

Most alarm limits are pre-programmed. Exceptions are listed below. Authorized personnel can change these limits.

Alarm Limits
Alarm condition Factory-set alarm limit Possible preset values
O₂ too low 21% or 25% O₂ 21% O₂
O₂ too high 95% O₂ 30% - 100% O₂ in 5% increments

Warnings: Hazard exists if different alarm presets are used in the same area (e.g., intensive care unit). Extreme alarm limits can render the system useless.

Checking Alarm System Functionality

Remove the heated breathing tube. You should see the "Check tube" visual alarm and hear the auditory alarm. If either is absent, do not use the unit. Refer to the AIRVO 2 Technical Manual for troubleshooting. Contact Fisher & Paykel Healthcare representative if problems persist.

Auditory Information Signals

Auditory information signals are provided:

Auditory Information Signals
Melody Meaning
Ascending sequence of 5 tones "Ready for use" symbol has appeared.
Ascending sequence of 3 tones Activation/deactivation of Junior Mode.
Single tone every 5 seconds Measured oxygen level ≥ 33% at turn-off.
Single tone every 30 seconds Measured oxygen level > 95%.

4. Reprocessing

The AIRVO 2 must be cleaned and disinfected between patients per the Disinfection Kit Manual (900PT600). This should be done as soon as possible after use. The unit uses warmed water and can pose a risk of bacterial colonization and infection if procedures are not followed.

Follow standard aseptic techniques to minimize contamination, including proper hand washing, avoiding contact with connection ports, safe disposal of consumables, and suitable storage after cleaning and disinfection.

Schedule for Changing Accessories

Accessories must be changed frequently to avoid infection risk. Replace parts immediately if damaged or discolored; otherwise, replace within the periods shown:

Accessory Replacement Schedule
Maximum period of use Part number and description
1 week (single-patient use) Patient interfaces excluding Optiflow™™+
OPT316 Nasal Cannula - Infant
OPT318 Nasal Cannula - Pediatric
OPT842 Optiflow™™ Nasal Cannula - Small
OPT844 Optiflow™™ Nasal Cannula - Medium
OPT846 Optiflow™™ Nasal Cannula - Large
OPT870 Tracheostomy Interface
RT013 Mask Interface Adapter - 22mm
2 weeks (single-patient use) Optiflow™™+ patient interfaces
OPT942 Optiflow™™+ Nasal Cannula - Small
OPT944 Optiflow™™+ Nasal Cannula - Medium
OPT946 Optiflow™™+ Nasal Cannula - Large
OPT970 Optiflow™™+ Tracheostomy Interface
OPT980 Optiflow™™+ Mask Interface Adapter
3 months or 1000 hours All tube & chamber kits
900PT551/900PT561 Heated breathing tube, MR290 auto-fill chamber and adapter
900PT501/900PT531 Junior heated breathing tube, MR290 auto-fill chamber and adapter (for use with OPT316/318 only)
900PT913 Air filter (or more often if significantly discolored)

Filter Replacement

After 1000 hours of operation, a prompt will appear to change the air filter. Follow these steps:

  1. Take the filter holder from the back of the unit and remove the filter.
  2. Replace the old filter with a new filter (900PT913).
  3. Reattach the filter holder to the unit, clipping the bottom in first, then rotating upwards until the top clips into place.
  4. Press Mode to move to the "Replace now" screen.
  5. Press Up to select "Now".
  6. Press Mode to confirm. The hours counter resets to zero.

If "Later" is chosen, the prompt will reappear at subsequent disinfection cycles.

Servicing

This device has no internal serviceable parts. AIRVO 2 humidifiers do not require routine servicing or calibration. Refer to the AIRVO 2 Technical Manual for external spare parts.

5. Technical Information

Symbol Definitions

Caution: Consult accompanying documents.

Type BF Applied Part.

Hot Surfaces.

IPX1: Drip Proof.

ATTENTION.

Do not throw away.

Alternating Current.

Class II, Double Insulated.

Power On/Off (Standby).

CE 0123

93/42/EEC Class IIa.

Product Specifications

Dimensions 295 mm x 170 mm x 175 mm (11.6" x 6.7" x 6.9") Target temperature settings 37, 34, 31 °C
Weight 2.2 kg (4.8 lb) unit only, 3.4 kg (7.5 lb) packaged Humidity >33 mg/L at 37 °C target
>10 mg/L at 34 °C target
>10 mg/L at 31 °C target
Supply frequency 50-60 Hz Maximum temperature of delivered gas 43 °C (109 °F) (in accordance with ISO 8185:2007)
Supply voltage/current 100-115 V 2.2 A (2.4 A max)
220-240 V 1.8 A (2.0 A max)
Flow range (default) 10-60 L/min
Sound pressure level Alarms exceed 45dbA @ 1m Flow range (Junior Mode) 2-25 L/min
Auditory alarm pause 115 seconds Maximum oxygen input 60 L/min
Serial port Used for downloading product data using F&P Infosmart™™ software. Oxygen analyzer accuracy < ±4% (within 25-95% O₂)
Warm-up time 10 minutes to 31 °C (88 °F), 30 minutes to 37 °C (98.6 °F) using MR290 chamber at 35 L/min and 23 ± 2 °C (73 ± 3 °F) starting temperature. Operating conditions: 18-28 °C (64-82 °F), 30-70% RH

Designed to conform to the requirements of: IEC 60601-1, UL 60601-1, CSA C22.2/No. 601.1, AS 3200.1.0, EN 60601-1, ISO 8185.

The unit complies with electromagnetic compatibility requirements of IEC 60601-1-2. If interference occurs, try moving the unit, the interfering equipment, or consult your healthcare provider.

Accessory equipment connected to the serial port must be certified to IEC 60601-1 or IEC 60950-1. All configurations must comply with IEC 60601-1-1. If unsure, consult the technical services department or your local representative.

Operating Conditions

Storage and Transport Conditions

Store and transport in environmental conditions of -10 °C to 60 °C (14 °F to 140 °F), 10 to 95% RH, non-condensing.

Disposal Instructions

PDF preview unavailable. Download the PDF instead.

900PT600 AIRVO 2 User Manual PDFium PDFium

Related Documents

Preview F&P Airvo 3 User Manual: High Flow Therapy Device Guide
This user manual provides comprehensive instructions for the Fisher & Paykel Healthcare Airvo 3, a device designed for high flow respiratory therapy. It covers device setup, operation, safety guidelines, troubleshooting, and technical specifications for healthcare professionals.
Preview myAIRVO 2 User Manual: Humidified Respiratory Therapy Device
Comprehensive user manual for the Fisher & Paykel Healthcare myAIRVO 2, detailing setup, operation, cleaning, maintenance, and technical specifications for humidified respiratory therapy.
Preview Fisher & Paykel AIRVO 2 Humidifier User Manual
Comprehensive user manual for the Fisher & Paykel AIRVO 2 humidifier, detailing setup, operation, maintenance, and troubleshooting for healthcare professionals.
Preview Fisher & Paykel Healthcare F&P 950 Adult Optiflow™ Heated Circuit Kit User Instructions
User instructions for the Fisher & Paykel Healthcare F&P 950 Adult Optiflow™ Heated Circuit Kit (REF 950A40), detailing setup, usage, warnings, and technical specifications for this single-use medical device accessory.
Preview F&P Airvo 2: Respiratory Support and Airway Hydration System
Discover the F&P Airvo 2, an integrated system for Optiflow tracheostomy high flow therapy, offering advanced humidification, respiratory support, and airway hydration for patients.
Preview Fisher & Paykel AIRVO 2 Humidifier User Manual
User manual for the Fisher & Paykel AIRVO 2 humidifier, providing instructions on setup, operation, maintenance, and troubleshooting.
Preview Fisher & Paykel Healthcare MR850 Respiratory Humidifier User Instructions
Comprehensive user manual for the Fisher & Paykel Healthcare MR850 Respiratory Humidifier, covering setup, operation, alarms, specifications, and safety information for medical professionals.
Preview Fisher & Paykel Optiflow+ Tracheostomy Direct Connection OPT970 User Manual
This document provides instructions for the Fisher & Paykel Optiflow+ Tracheostomy Direct Connection OPT970, a patient interface for delivering humidified respiratory gases to tracheostomy patients. It includes setup, intended use, warnings, cautions, and usage guidelines for both hospital and home environments.