Microlife XL Blood Pressure Cuff
User Manual
1. Introduction
Thank you for purchasing a Microlife XL cuff. This cuff is suitable for arm circumferences of 12.6" to 20.5", and is easy to adapt to most Microlife monitors.
2. Components of your blood pressure cuff
2.1. Cuff
The cuff assembly consists of the cuff and the cuff hose.
2.2. Cuff Assembly & Connector Types
The cuff assembly includes the cuff, hose, and cuff connector.
Connector Types:
- Advanced connector-1 (Blue)
- Advanced connector-2 (Yellow)
- Advanced connector-3 (All Black)
- Standard connector
3. Select your connector
IMPORTANT: This replacement package includes multiple connector types. It is critical that the correct connector be used.
- Look at your old cuff and connector. Identify the cuff connector's shape and color. The connector is used for plugging the cuff into your monitor and is located at the end of the cuff hose.
- If your cuff has a standard connector:
- a) Find the silver regulatory label (usually located on the back of your device).
- b) Review the label for ANY of the following:
- b1. A triangular "IMT" symbol [? IMT Symbol]
- b2. Any of the model numbers listed below:
- BP3AR1-1PASC
- BP3BM1-3ALDR
- BP3MS1-3BLDR
- BP3MV1-3BHM
- BP3NM1-3WHM
- BP3NM1-3WSMARK
- BP3MV1-3BSMARK
- BP3MV1-3XCVS
- BP3MW1-4YCVS
- BP3NQ1-4BHEB
- c) If any of the conditions are met, use the standard connector with your new cuff.
- d) If you do not meet the conditions in b 1-2, use the Advanced connector 3 (All Black). DO NOT USE THE STANDARD CONNECTOR.
- If your cuff has an advanced connector:
- a) Select a new connector matching the tip color of your current connector (yellow or blue).
- Attach the cuff connector to the end of the cuff hose.
- Connect the new cuff assembly (cuff, hose & connector) to your Microlife automatic blood pressure monitor, and ensure the connection fits snugly. It should not be loose.
- Once the cuff is properly connected to the monitor, prepare the cuff for your measurement.
Questions? Contact customer service: 1-800-568-4147.
4. Applying the cuff
- Ensure clothing does not interfere with the cuff. Remove any tight or bulky items which might restrict the arm.
- Slide the cuff onto the arm, positioning it as illustrated, so that the hose is closer to your lower arm. Make certain that the lower edge of the cuff lies approximately 1/2" (1 to 2 cm) above the elbow and that the hose is closer to the inner side of the arm.
- IMPORTANT: The Artery Mark Indicator must lie over the brachial artery. The artery is located closer to the inside of the upper arm (directly under the front of the armpit).
- TIP: Align artery mark to pinky finger (NOT the center of your arm).
- Tighten the cuff by pulling the end, and secure the cuff by affixing the hook & loop material.
- Make sure the cuff is tightened correctly. It should fit the arm snugly, and you should just be able to slide two fingers between the arm and the cuff.
- Once the cuff is in position and adjusted, ensure the "Sizing Seam" marking falls within the "Size Range" marking. If it does not, the cuff may not be the correct size for your arm.
- Caution: Incorrect cuff sizing / application: Incorrect sizing can cause your systolic blood pressure measurement to swing up or down by 10 mmHg. Similarly, a cuff that is applied too loosely or too tightly will result in an inaccurate reading. Please contact customer service, or visit our website, for help determining and obtaining the correct sized cuff.
Diagram Description: Cuff placement on arm for measurement, showing arm circumference range 12.6" - 20.5", and the 1/2" (1-2 cm) distance above the elbow. The Artery Mark is shown aligned with the artery.
Comment: If it is not possible to fit the cuff to your left arm, it can also be placed on your right arm. However, all measurements should be taken on the same arm.
Using the cuff: Follow all instructions included with your original blood pressure monitor when taking a blood pressure measurement.
Instruction manuals for most monitors produced by Microlife are available on our website: www.microlifeusa.com
5. Care and maintenance
- The cuff contains a sensitive airtight bladder. Handle this cuff carefully. Avoid applying stress which can occur if the cuff and/or hose are twisted or folded. Do not pull on the hose. Keep both away from sharp objects.
- Clean the cuff exterior with a soft, dry cloth. Spots on the cuff can be removed carefully with a damp cloth and soapsuds. Do not submerge in water, place in dishwasher, or use chemical cleaning agents/solvents.
- Do not use the device if you think it is damaged or if anything appears unusual.
6. Limited warranty
Your cuff is warrantied for one year by Microlife USA Inc., against manufacturer defects for the original purchaser only, from date of purchase. The warranty does not apply to consequential and incidental damages, or damage caused by improper handling, and accidents. Professional use, not following the operating instructions, and alterations made to the cuff, are also not included in this warranty. Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you.
Microlife USA Inc. will investigate your concern. A cuff as defined by this warranty, determined to be out of specification, will be replaced and shipped to you at no cost. A cuff as defined by this warranty, determined to be within specification, will be returned to you with a report of findings, at no cost.
Please use the customer service contact information below to reach Microlife USA Inc. regarding any warranty concerns. We ask that you please contact us before sending any product back in order to better identify, and more quickly process, your concern. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.
7. Technical specifications
- Cuff type: XL cuff for arms circumference 12.6" - 20.5", 32 - 52 cm
- Weight: 160 g
- Storage temperature: -20 to +55°C (-4° to +131°F)
- Operation temperature: 10 to 40°C (50° to 104°F)
- Operation/storage humidity: 15 to 90% relative humidity maximum
- Expected service life: 2 years or 10,000 uses (whichever comes first)
Technical alterations reserved. Made in China
8. How to contact us
Distributed by:
Microlife USA, Inc.
1617 Gulf to Bay Blvd.
Clearwater, FL 33755
Toll Free Help Line: 1-800-568-4147
Email: custserv@microlifeusa.com
Website: www.microlifeusa.com
Model Number: XLSC